Khelix Steerable Electrophysiology Catheters, Khelix Loop Fixed Electrophysiology Catheters, Khelix Fixed Loop Steerable Electrophysiology Catheters, Khelix Variable Loop Steerable Electrophysiology Catheters

K211327 · CathRx, Ltd. · DRF · Feb 10, 2022 · Cardiovascular

Device Facts

Record IDK211327
Device NameKhelix Steerable Electrophysiology Catheters, Khelix Loop Fixed Electrophysiology Catheters, Khelix Fixed Loop Steerable Electrophysiology Catheters, Khelix Variable Loop Steerable Electrophysiology Catheters
ApplicantCathRx, Ltd.
Product CodeDRF · Cardiovascular
Decision DateFeb 10, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Intended Use

The Khelix Diagnostic Electrophysiology Catheters are indicated for the electrical recording or stimulation of endocardial structures. The loop versions of the catheters are intended for obtaining and recording electrograms from the atrial region of the heart.

Device Story

Khelix Diagnostic Electrophysiology Catheters are cardiac catheters used by cardiologists during EP studies to map heart electrical activity, identify arrhythmia origins, and assess ablation therapy effectiveness. The device comprises a handle and shaft with a platinum-iridium electrode array (4-20 electrodes). It connects to standard EP laboratory equipment (ECG monitors, stimulators, impedance-based navigation systems) via a Khelix Diagnostic Extension Cable. The catheter is inserted into the heart to record endocardial signals or provide electrical stimulation. By providing high-fidelity electrograms, the device assists physicians in diagnosing cardiac arrhythmias and guiding therapeutic interventions, potentially improving patient outcomes by enabling precise treatment planning and monitoring.

Clinical Evidence

Bench testing only. Testing included sterilization validation (ISO 11135), biocompatibility (ISO 10993, ASTM 756), packaging/shelf-life validation (ASTM F1980, ISO 11607), electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-27), and functional performance (ISO 10555-1). All tests passed.

Technological Characteristics

Materials: PEBAX insulation, platinum-iridium electrodes. Sensing/Actuation: Electrical recording/stimulation via passive electrodes. Form factor: 6Fr/7Fr steerable/loop catheters, 115cm length. Connectivity: Interfaces with external EP lab equipment via extension cable. Sterilization: Ethylene Oxide (EtO).

Indications for Use

Indicated for electrical recording or stimulation of endocardial structures in patients undergoing electrophysiology (EP) studies to diagnose or treat abnormal heart rhythms. Loop versions are specifically for atrial electrogram recording.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. February 10, 2022 CathRx Ltd % Amy Oakes Quality & Regulatory Consultant Acorn N Oakes LLC 333 N. Dobson Rd. Suite 5 Chandler, Arizona 85224 Re: K211327 Trade/Device Name: Khelix Diagnostic Electrophysiology Catheters Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: DRF Dated: April 6, 2021 Received: May 3, 2021 Dear Amy Oakes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211327 Device Name Khelix Diagnostic Electrophysiology Catheters Indications for Use (Describe) The Khelix Diagnostic Electrophysiology Catheters are indicated for the electrical recording of endocardial structures. The loop versions of the catheters are intended for obtaining and recording electrograms from the atrial region of the heart. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ![]() ## SECTION 5: 510(K) SUMMARY #### Submitter's Name and Address: CathRx Ltd Unit 8, 2-8 South St Rydalmere NSW 2116, Australia #### Owner Name and Information lan Fong CathRx Ltd Chief Executive Officer +61 411 421 354 lan.fong@cathrx.com ## US Agent and Official Correspondent Contact Name and Information: Amy Oakes Acorn N Oakes, LLC Quality & Regulatory Consultant 333 N. Dobson Rd Suite 5 Chandler AZ 85224 USA (480) 304-2165 acornnoakes@cox.net #### Date Prepared: March 24, 2021 #### Device Information: | Trade/Proprietary Name: | Khelix Diagnostic Electrophysiology Catheters | |-------------------------|-----------------------------------------------------------| | Common Name: | Diagnostic Electrophysiology Catheters | | Classification Name: | Electrode Recording Catheter or electrode recording probe | | Classification Number: | 21 CFR 870.1220 | | Product Code: | DRF | #### Predicate Device: | 510(k) Number | 510(k) Title | Manufacturer | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|------------------|--| | K113213 | Lasso NAV 2515 Variable | Biosense Webster | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | | Table 5.1: Predicate Device {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for CathRx. The logo features a stylized heart shape on the left, followed by the text "CATH" in bold, sans-serif font. To the right of "CATH" is "RX" enclosed in a rounded rectangle. #### Manufacturer: CathRx Ltd Unit 8, 2-8 South St Rydalmere NSW 2116, Australia ## Manufacturing Facilities: Manufacturing CathRx Ltd Unit 8, 2-8 South St Rydalmere NSW 2116, Australia Sterilization 5 and 7 Widemere Rd, Wetherill Park NSW 2164, Australia ### Device Description: Khelix Diagnostic Electrophysiology Catheters are a range of electrophysiology diagnostic cardiac catheters designed to facilitate electrophysiological mapping of the heart. The devices are intended for the electrical recording and/or stimulation of endocardial structures. The Khelix Diagnostic Electrophysiology Catheter range is used by cardiologists in electrophysiology (EP) studies to determine the cause of an abnormal heart rhythm, to locate the site of origin of an abnormal heart rhythm, to decide the best treatment for an abnormal heart rhythm, and/or to check the effectiveness of ablation therapy. An EP study is sometimes conducted before placement of an implantable cardioverter/defibrillator (ICD) to determine the best type of device, and to monitor treatment effectiveness. The device consists of a catheter handle and catheter shaft containing an array of platinum-iridium electrodes for stimulation and recording. The subject device comes in a variety of models based on the distal catheter configuration including Steerable and Loop models. The device interfaces with standard electrophysiological recording equipment, such as ECG monitoring equipment, impedance based navigational equipment and cardiac stimulation equipment, via the Khelix Diagnostic Extension Cable (supplied separately). The device is available in different device models with several variants according to electrode configurations, curve types and loop diameters, as applicable. A full list of model variants is detailed at the end of this summary. The subject catheters use 4 to 20 electrodes with variable spacing to provide for variation in physician preference. The catheter is intended to be used with the Khelix Diagnostic Extension Cable. The extension cable connects distally to the catheter (sterile field) and proximally to standard electrophysiology (EP) laboratory equipment such as heart signal monitors/records and stimulators (non-sterile field) via shielded tip pins. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for CathRx. The logo features a stylized heart shape on the left, followed by the text "CATH" in bold, sans-serif font. To the right of "CATH", the letters "RX" are enclosed in a black box with rounded corners. The Khelix Diagnostic Electrophysiology Catheter range is supplied sterile. The devices are sterilized using an ethylene oxide (EtO) sterilization process. ### Indications for Use: The Khelix Diagnostic Electrophysiology Catheters are indicated for the electrical recording or stimulation of endocardial structures. The loop versions of the catheters are intended for obtaining and recording electrograms from the atrial region of the heart. ## Technological Characteristics: A high-level comparison of the proposed and predicate devices is in Table 5.2 below: | Characteristic | Predicate Device | Subject Device | Comparison | |--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Product Code | DRF | DRF | Identical | | Operating Principles / Mechanism of Action | Electrical Stimulation & Recording | Electrical Stimulation & Recording | Identical | | Materials | PEBAX insulation with platinum electrodes | PEBAX insulation with platinum electrodes | Identical | | Key technological characteristics | Equivalent | | | | Fr Size | 7 Fr | 6Fr, 7Fr | Equivalent | | Working Length | 115 cm | 115 cm | Identical | | Deflection Plane | In Plane | In Plane | Identical | | Deflection Type | Unidirectional | Unidirectional | Identical | | # of Electrodes | 10-20 | 4-20 | Equivalent | | Electrode Spacing | 2-6-2mm (20 pole)<br>8-8-8mm (10 pole)<br>In Lasso Nav fixed loop catheters:<br>5-5-5mm (15mm)<br>6-6-6mm (20mm)<br>8-8-8mm (25mm) | 2-5-2mm<br>2-2-2mm<br>5-5-5mm<br>2-8-2mm<br>2-10-2mm<br>2-12-2mm | Equivalent | | Tip Electrode | Passive | Passive | Identical | | Loop Diameter | 25-15 mm | 25-15 mm | Identical | | Curve Shape | D Curve (6.4 cm) | D-curve (40.7mm)<br>Other curve sizes available:<br>F-curve (48.3mm)<br>G- curve (56.6mm)<br>J-curve (64.9mm)<br>P-curve (80.8mm)<br>Q-curve (96.7mm)<br>R-curve (113.3mm) | Equivalent | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for CathRx. The logo consists of a heart shape with a line going through it, followed by the text "CATH" in bold, sans-serif font. To the right of "CATH" is the text "RX" in a similar font, enclosed in a black box with rounded corners. The logo appears to be for a medical or pharmaceutical company, given the "RX" abbreviation often used for prescriptions. | Characteristic | Predicate Device | Subject Device | Comparison | |-------------------------|------------------|-----------------|------------| | Accessory Devices | Extension Cable | Extension Cable | Identical | | Environment of use | Cath Lab | Cath Lab | Identical | | Sterility & Reusability | Single Use | Single Use | Identical | | Sterilization Method | Ethylene Oxide | Ethylene Oxide | Identical | Table 5.2: Comparison Table The purpose, design, materials, function, and intended use of the Khelix Diagnostic Electrophysiology Catheters are substantially equivalent to the predicate devices. ### Function and Safety Testing: Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Khelix Diagnostic Electrophysiology Catheters. Testing includes the following: | Test | Test Methodology | Result | |--------------------------------|------------------------------------------------------------------|--------| | Sterilization Validation | ISO 11135 | Pass | | SAL = 10-6 | ISO 11135 | Pass | | EO Residuals | ISO 10993-7 | Pass | | Biocompatibility | ISO 10993<br>ASTM 756 | Pass | | Packaging Validation | ASTM 4332<br>ASTM 4169 | Pass | | Shelf Life | ASTM F1980<br>ISO 11607 | Pass | | Sterile Barrier Integrity | ASTM F1886/F1886M-16<br>ASTM F2096-11 (2019)<br>ASTM F88/F88M-15 | Pass | | Electrical Safety Testing | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-27 | Pass | | Functional Performance Testing | ISO 10555-1 | Pass | | Attribute Inspection | Visual inspection to<br>workmanship standards | Pass | | Mechanical Characterization | N/A | Pass | | Functional Performance | N/A | Pass | | Simulated Use | N/A | Pass | | Joint Strength (Integrity) | ISO 10555-1 | Pass | #### Conclusion: CathRx concludes that the Khelix Diagnostic Electrophysiology Catheters are substantially equivalent to the predicate device described herein. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the logo for "CATHRX". The logo features a heart shape with a line running through it on the left side. The text "CATH" is in bold, black letters, and "RX" is in a black box with white letters on the right side. ## Khelix Diagnostic Electrophysiology Catheters ## Steerable Models | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS1-DOU-04P6222 | | 6.4 | In Plane | Unidirectional | 4 | Passive | 2-2-2 | | SS1-DOU-04P6252 | | | | | 4 | | 2-5-2 | | SS1-DOU-04P6555 | | | | | 4 | | 5-5-5 | | SS1-DOU-04P6282 | | | | | 4 | | 2-8-2 | | SS1-DOU-04P62102 | | | | | 4 | | 2-10-2 | | SS1-DOU-06P6222 | | 6.4 | In Plane | Unidirectional | 6 | | 2-2-2 | | SS1-DOU-06P6252 | | | | | 6 | | 2-5-2 | | SS1-DOU-06P6555 | | | | | 6 | | 5-5-5 | | SS1-DOU-06P6282 | | | | | 6 | | 2-8-2 | | SS1-DOU-06P62102 | | | | | 6 | | 2-10-2 | | SS1-DOU-08P6222 | | 6.4 | In Plane | Unidirectional | 8 | | 2-2-2 | | SS1-DOU-08P6252 | | | | | 8 | | 2-5-2 | | SS1-DOU-08P6555 | | | | | 8 | | 5-5-5 | | SS1-DOU-08P6282 | | | | | 8 | | 2-8-2 | | SS1-DOU-08P62102 | | | | | 8 | | 2-10-2 | | SS1-DOU-10P6222 | | 6.4 | In Plane | Unidirectional | 10 | Passive | 2-2-2 | | SS1-DOU-10P6252 | | | | | 10 | | 2-5-2 | | SS1-DOU-10P6555 | | | | | 10 | | 5-5-5 | | SS1-DOU-10P6282 | | | | | 10 | | 2-8-2 | | SS1-DOU-10P62102 | | | | | 10 | | 2-10-2 | CathRx Khelix Diagnostic Electrophysiology Catheters Traditional 510(k) {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for CathRx. The logo features a heart shape with a line running through it on the left side. To the right of the heart is the text "CATH" in bold, sans-serif font, followed by "RX" in a rounded rectangle. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS1-FOU-04P6222 | | | | Unidirectional | 4 | Passive | 2-2-2 | | SS1-FOU-04P6252 | | | | | | | 2-5-2 | | SS1-FOU-04P6555 | | | | | | | 5-5-5 | | SS1-FOU-04P6282 | | | | | | | 2-8-2 | | SS1-FOU-04P62102 | | | | | | | 2-10-2 | | SS1-FOU-06P6222 | | | | | 6 | | 2-2-2 | | SS1-FOU-06P6252 | | | | | | | 2-5-2 | | SS1-FOU-06P6555 | | | | | | | 5-5-5 | | SS1-FOU-06P6282 | | | | | | | 2-8-2 | | SS1-FOU-06P62102 | | | | | | | 2-10-2 | | SS1-FOU-08P6222 | | 7.6 | In Plane | | 8 | | 2-2-2 | | SS1-FOU-08P6252 | | | | | | | 2-5-2 | | SS1-FOU-08P6555 | | | | | | | 5-5-5 | | SS1-FOU-08P6282 | | | | | | | | | SS1-FOU-08P62102 | | | | | | | | | SS1-FOU-10P6222 | | | | | 10 | | 2-2-2 | | SS1-FOU-10P6252 | | | | | | | 2-5-2 | | SS1-FOU-10P6555 | | | | | | | 5-5-5 | | SS1-FOU-10P6282 | | | | | | | 2-8-2 | | SS1-FOU-10P62102 | | | | | | | 2-10-2 | {9}------------------------------------------------ Image /page/9/Picture/0 description: The image contains the logo for CATHRX. The logo features a stylized heart shape with a line extending from the bottom, resembling a medical symbol. The text "CATH" is written in bold, sans-serif font, followed by "RX" enclosed in a rounded rectangle. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS1-GOU-04P6222 | | | | | 4 | Passive | 2-2-2 | | SS1-GOU-04P6252 | | | | | | | 2-5-2 | | SS1-GOU-04P6555 | | | | | | | 5-5-5 | | SS1-GOU-04P6282 | | | | | | | 2-8-2 | | SS1-GOU-04P62102 | | | | | | | 2-10-2 | | SS1-GOU-06P6222 | | | | | 6 | | 2-2-2 | | SS1-GOU-06P6252 | | | | | | | 2-5-2 | | SS1-GOU-06P6555 | | | | | | | 5-5-5 | | SS1-GOU-06P6282 | | | | | | | 2-8-2 | | SS1-GOU-06P62102 | | | | | | | 2-10-2 | | SS1-GOU-08P6222 | | 8.9 | In Plane | Unidirectional | 8 | | 2-2-2 | | SS1-GOU-08P6252 | | | | | | | 2-5-2 | | SS1-GOU-08P6555 | | | | | | | 5-5-5 | | SS1-GOU-08P6282 | | | | | | | 2-8-2 | | SS1-GOU-08P62102 | | | | | | | 2-10-2 | | SS1-GOU-10P6222 | | | | | 10 | | 2-2-2 | | SS1-GOU-10P6252 | | | | | | | 2-5-2 | | SS1-GOU-10P6555 | | | | | | | 5-5-5 | | SS1-GOU-10P6282 | | | | | | | 2-8-2 | | SS1-GOU-10P62102 | | | | | | | 2-10-2 | {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for CathRx. The logo consists of a heart shape with a line running through it on the left side, followed by the text "CATH" in bold, sans-serif font. To the right of "CATH" is "RX" in a similar font, enclosed in a rounded rectangle. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|----------------| | SS1-JOU-04P6222 | | | | | 4 | Passive | 2-2-2 | | SS1-JOU-04P6252 | | | | | | | 2-5-2 | | SS1-JOU-04P6555 | | | | | | | 5-5-5 | | SS1-JOU-04P6282 | | | | | | | 2-8-2 | | SS1-JOU-04P62102 | | | | | | | 2-10-2 | | SS1-JOU-06P6222 | | 10.2 | | | In Plane | | Unidirectional | | SS1-JOU-06P6252 | | | | | | | | | SS1-JOU-06P6555 | | | | | | | | | SS1-JOU-06P6282 | | | | | | | | | SS1-JOU-06P62102 | | | | | | | | | SS1-JOU-08P6222 | | | | | 8 | | 2-2-2 | | SS1-JOU-08P6252 | | | | | | | 2-5-2 | | SS1-JOU-08P6555 | | | | | | | 5-5-5 | | SS1-JOU-08P6282 | | | | | | | 2-8-2 | | SS1-JOU-08P62102 | | | | | | | 2-10-2 | | SS1-JOU-10P6222 | | | | | 10 | | 2-2-2 | | SS1-JOU-10P6252 | | | | | | | 2-5-2 | | SS1-JOU-10P6555 | | | | | | | 5-5-5 | | SS1-JOU-10P6282 | | | | | | | 2-8-2 | | SS1-JOU-10P62102 | | | | | | | 2-10-2 | {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for CathRx. The logo consists of a heart shape with a line running through it on the left side, followed by the text "CATH" in bold, sans-serif font. To the right of "CATH" is "RX" in a similar font, enclosed in a rounded rectangle. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS1-POU-04P6222 | | 12.7 | In Plane | Unidirectional | 4 | Passive | 2-2-2 | | SS1-POU-04P6252 | | 12.7 | In Plane | Unidirectional | | | 2-5-2 | | SS1-POU-04P6555 | | 12.7 | In Plane | Unidirectional | | | 5-5-5 | | SS1-POU-04P6282 | | 12.7 | In Plane | Unidirectional | | | 2-8-2 | | SS1-POU-04P62102 | | 12.7 | In Plane | Unidirectional | | | 2-10-2 | | SS1-POU-04P6101010 | | 12.7 | In Plane | Unidirectional | | | 10-10-10 | | SS1-POU-06P6222 | | 12.7 | In Plane | Unidirectional | 6 | | 2-2-2 | | SS1-POU-06P6252 | | 12.7 | In Plane | Unidirectional | | | 2-5-2 | | SS1-POU-06P6555 | | 12.7 | In Plane | Unidirectional | | | 5-5-5 | | SS1-POU-06P6282 | | 12.7 | In Plane | Unidirectional | | | 2-8-2 | | SS1-POU-06P62102 | | 12.7 | In Plane | Unidirectional | | | 2-10-2 | | SS1-POU-06P6101010 | | 12.7 | In Plane | Unidirectional | | | 10-10-10 | | SS1-POU-08P6222 | | 12.7 | In Plane | Unidirectional | 8 | | 2-2-2 | | SS1-POU-08P6252 | | 12.7 | In Plane | Unidirectional | | | 2-5-2 | | SS1-POU-08P6555 | | 12.7 | In Plane | Unidirectional | | | 5-5-5 | | SS1-POU-08P6282 | | 12.7 | In Plane | Unidirectional | | | 2-8-2 | | SS1-POU-08P62102 | | 12.7 | In Plane | Unidirectional | | | 2-10-2 | | SS1-POU-08P6101010 | | 12.7 | In Plane | Unidirectional | | | 10-10-10 | | SS1-POU-10P6222 | | 12.7 | In Plane | Unidirectional | 10 | | 2-2-2 | | SS1-POU-10P6252 | | 12.7 | In Plane | Unidirectional | | | 2-5-2 | | SS1-POU-10P6555 | | 12.7 | In Plane | Unidirectional | | | 5-5-5 | | SS1-POU-10P6282 | | 12.7 | In Plane | Unidirectional | | | 2-8-2 | | SS1-POU-10P62102 | | 12.7 | In Plane | Unidirectional | | | 2-10-2 | | SS1-POU-10P6101010 | | 12.7 | In Plane | Unidirectional | | | 10-10-10 | {12}------------------------------------------------ Image /page/12/Picture/0 description: The image contains the logo for "CATHRx". The logo features a heart shape with a line extending from the bottom, resembling a heartbeat. The text "CATH" is written in bold, sans-serif font, followed by "Rx" enclosed in a rounded rectangle. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS1-Q0U-04P6222 | | 15.2 | In Plane | Unidirectional | 4 | Passive | 2-2-2 | | SS1-Q0U-04P6252 | | | | | | | 2-5-2 | | SS1-Q0U-04P6555 | | | | | | | 5-5-5 | | SS1-Q0U-04P6282 | | | | | | | 2-8-2 | | SS1-Q0U-04P62102 | | | | | | | 2-10-2 | | SS1-Q0U-04P6101010 | | | | | | | 10-10-10 | | SS1-Q0U-06P6222 | | 6 | | | 2-2-2 | | | | SS1-Q0U-06P6252 | | | | | 2-5-2 | | | | SS1-Q0U-06P6555 | | | | | 5-5-5 | | | | SS1-Q0U-06P6282 | | | | | 2-8-2 | | | | SS1-Q0U-06P62102 | | | | | 2-10-2 | | | | SS1-Q0U-06P6101010 | | | | | 10-10-10 | | | | SS1-Q0U-08P6222 | | 8 | | | 2-2-2 | | | | SS1-Q0U-08P6252 | | | | | 2-5-2 | | | | SS1-Q0U-08P6555 | | | | | 5-5-5 | | | | SS1-Q0U-08P6282 | | | | | 2-8-2 | | | | SS1-Q0U-08P62102 | | | | | 2-10-2 | | | | SS1-Q0U-08P6101010 | | 10 | 10-10-10 | | | | | | SS1-Q0U-10P6222 | | | 2-2-2 | | | | | | SS1-Q0U-10P6252 | | | 2-5-2 | | | | | | SS1-Q0U-10P6555 | | | 5-5-5 | | | | | | SS1-Q0U-10P6282 | | | 2-8-2 | | | | | | SS1-Q0U-10P62102 | | | 2-10-2 | | | | | | SS1-Q0U-10P6101010 | | 10-10-10 | | | | | | {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo for "CATHRx". The logo features a heart shape with a line running through it on the left side. The text "CATH" is in bold, sans-serif font, and the "Rx" is in a rounded rectangle on the right side. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS1-ROU-04P6222 | | 17.8 | In Plane | Unidirectional | 4 | Passive | 2-2-2 | | SS1-ROU-04P6252 | | | | | | | 2-5-2 | | SS1-ROU-04P6555 | | | | | | | 5-5-5 | | SS1-ROU-04P6282 | | | | | | | 2-8-2 | | SS1-ROU-04P62102 | | | | | | | 2-10-2 | | SS1-ROU-04P6101010 | | | | | | | 10-10-10 | | SS1-ROU-06P6222 | | 6 | | | 2-2-2 | | | | SS1-ROU-06P6252 | | | | | 2-5-2 | | | | SS1-ROU-06P6555 | | | | | 5-5-5 | | | | SS1-ROU-06P6282 | | | | | 2-8-2 | | | | SS1-ROU-06P62102 | | | | | 2-10-2 | | | | SS1-ROU-06P6101010 | | | | | 10-10-10 | | | | SS1-ROU-08P6222 | | 8 | | | 2-2-2 | | | | SS1-ROU-08P6252 | | | | | 2-5-2 | | | | SS1-ROU-08P6555 | | | | | 5-5-5 | | | | SS1-ROU-08P6282 | | | | | 2-8-2 | | | | SS1-ROU-08P62102 | | | | | 2-10-2 | | | | SS1-ROU-08P6101010 | | | | | 10-10-10 | | | | SS1-ROU-10P6222 | | 10 | 2-2-2 | | | | | | SS1-ROU-10P6252 | | | 2-5-2 | | | | | | SS1-ROU-10P6555 | | | 5-5-5 | | | | | | SS1-ROU-10P6282 | | | 2-8-2 | | | | | | SS1-ROU-10P62102 | | | 2-10-2 | | | | | | SS1-ROU-10P6101010 | | | 10-10-10 | | | | | {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the logo for CATH RX. The logo features a heart shape with a line running through it on the left side. The text "CATH" is in bold, sans-serif font, and "RX" is in a rounded rectangle on the right side. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS1-YOU-04P6222 | | 6.4 HIS | In Plane | Unidirectional | 4 | Passive | 2-2-2 | | SS1-YOU-04P6252 | | | | | | | 2-5-2 | | SS1-YOU-04P6555 | | | | | | | 5-5-5 | | SS1-YOU-04P6282 | | | | | | | 2-8-2 | | SS1-YOU-04P62102 | | | | | | | 2-10-2 | | SS1-YOU-04P6101010 | | | | | | | 10-10-10 | | SS1-YOU-06P6222 | | | | | 6 | | 2-2-2 | | SS1-YOU-06P6252 | | | | | | | 2-5-2 | | SS1-YOU-06P6555 | | | | | | | 5-5-5 | | SS1-YOU-06P6282 | | | | | | | 2-8-2 | | SS1-YOU-06P62102 | | | | | | | 2-10-2 | | SS1-YOU-06P6101010 | | | | | | | 10-10-10 | | SS1-YOU-08P6222 | | | | | 8 | | 2-2-2 | | SS1-YOU-08P6252 | | | | | | | 2-5-2 | | SS1-YOU-08P6555 | | | | | | | 5-5-5 | | SS1-YOU-08P6282 | | | | | | | 2-8-2 | | SS1-YOU-08P62102 | | | | | | | 2-10-2 | | SS1-YOU-08P6101010 | | | | | | 10-10-10 | | | SS1-YOU-10P6222 | | | | | 10 | 2-2-2 | | | SS1-YOU-10P6252 | | | | | | 2-5-2 | | | SS1-YOU-10P6555 | | | | | | 5-5-5 | | | SS1-YOU-10P6282 | | | | | | 2-8-2 | | | SS1-YOU-10P62102 | | | | | | 2-10-2 | | | SS1-YOU-10P6101010 | | | | | | 10-10-10 | | {15}------------------------------------------------ Image /page/15/Picture/0 description: The image contains the logo for "CATH RX". The logo features a stylized heart shape on the left, followed by the word "CATH" in bold, sans-serif font. To the right of "CATH" is "RX" enclosed in a rounded rectangle. | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS2-POU-20P6242 | | | | | | | 2-4-2 | | SS2-POU-20P6282 | CCD0104 | | | | | | 2-8-2 | | SS2-POU-20P62102 | | | | | | Passive | 2-10-2 | | SS2-POU-20P62122 | CCD0042 | | | | | | 2-12-2 | | SS2-POU-20P62132 | | | | | | | 2-13-2 | | SS2-POU-20P620282 | | | | | | | 20-2-8-2 | | SS2-Q0U-20P6242 | | | | | | | 2-4-2 | | SS2-Q0U-20P6282 | | | | | | | 2-8-2 | | SS2-Q0U-20P62102 | | | | | | Passive | 2-10-2 | | SS2-Q0U-20P62122 | CCD0089 | | | | | | 2-12-2 | | SS2-Q0U-20P62132 | | | | | | | 2-13-2 | | SS2-Q0U-20P620282 | | | | | | | 20-2-8-2 | | SS2-ROU-20P6242 | | | | | | | 2-4-2 | | SS2-ROU-20P6282 | | | | | | | 2-8-2 | | SS2-ROU-20P62102 | | | | | | Passive | 2-10-2 | | SS2-ROU-20P62122 | CCD0090 | | | | | | 2-12-2 | | SS2-ROU-20P62132 | | | | | | | 2-13-2 | | SS2-ROU-20P620282 | | | | | | | 20-2-8-2 | | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection<br>Plane | Deflection Type | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------------------|---------------------|-----------------|--------------------|---------------|--------------| | SS2-YOU-20P6242 | | | | | | | 2-4-2 | | SS2-YOU-20P6282 | | | | | | | 2-8-2 | | SS2-YOU-20P62102 | | | | | | | 2-10-2 | | SS2-YOU-20P62122 | | | | | | Passive | 2-12-2 | | SS2-YOU-20P62132 | | | | | | | 2-13-2 | | SS2-YOU-20P620282 | | | | | | | 20-2-8-2 | ### Loop Fixed Models {16}------------------------------------------------ Image /page/16/Picture/0 description: The image contains the logo for "CATHRx". The logo features a stylized heart shape with a line extending from it on the left side. To the right of the heart is the text "CATH" in bold, followed by "Rx" in a black box. | Product Reference<br>Number | Allocated Item<br>Number | Curve | Loop Type | Loop Size (mm) | # of<br>Electrodes | Tip Electrode | Spacing (mm) | |-----------------------------|--------------------------|----------|-----------|----------------|--------------------|---------------|--------------| | SCF-S15-10P7555 | | Straight | Fixed | 15 | 10 | Passive | 5-5-5 | | SCF-S20-10P7666 | | | | 20 | | | 6-6-6 | | SCF-S25-10P7888 | | | | 25 | | | 8-8-8 | | SCF-S30-10P7999 | | | | 30 | | | 9-9-9 | Fixed Loop Steerable Models | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection Type | Loop Type | Loop Size<br>(mm) | # of<br>Electrodes | Tip Electrode | Spacing<br>(mm) | |-----------------------------|--------------------------|----------------------|-----------------|-----------|-------------------|--------------------|---------------|-----------------| | SC1-B15U-10P7555 | CCD0051 | | Unidirectional | Fixed | 15 | 10 | Passive | 5-5-5 | | SC1-B20U-10P7666 | CCD0052 | 5.1 | | | 20 | | | 6-6-6 | | SC1-B25U-10P7888 | CCD0053 | | | | 25 | | | 8-8-8 | | SC1-D15U-10P7555 | | | | | 15 | | | 5-5-5 | | SC1-D20U-10P7666 | | 6.4 | | | 20 | | | 6-6-6 | | SC1-D25U-10P7888 | | | | | 25 | | | 8-8-8 | | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection Type | Loop Type | Loop Size<br>(mm) | # of<br>Electrodes | Tip Electrode | Spacing<br>(mm) | |-----------------------------|--------------------------|----------------------|-----------------|-----------|-------------------|--------------------|---------------|-----------------| | SC2-B15U-14P7333 | | 5.1 | Unidirectional | Fixed | 15 | 14 | Passive | 3-3-3 | | SC2-B20U-14P7444 | | | | | 20 | | | 4-4-4 | | SC2-B30U-14P7777 | | | | | 30 | | | 7-7-7 | | SC2-B15U-20P7232 | | | | | 15 | 20 | | 2-3-2 | | SC2-B20U-20P7242 | | | | | 20 | | | 2-4-2 | | SC2-B25U-20P7262 | | | | | 25 | | | 2-6-2 | | SC2-B30U-20P7272 | | | | | 30 | | | 2-7-2 | | SC2-D15U-14P7333 | | 6.4 | | | 15 | 14 | | 3-3-3 | | SC2-D20U-14P7444 | | | | 20 | 4-4-4 | | | | | SC2-D30U-14P7777 | | | | 30 | 7-7-7 | | | | | SC2-D15U-20P7232 | | | | 15 | 20 | 2-3-2 | | | | SC2-D20U-20P7242 | | | | 20 | | 2-4-2 | | | | SC2-D25U-20P7262 | | | | 25 | | 2-6-2 | | | | SC2-D30U-20P7272 | | | | | 30 | | 2-7-2 | | CathRx Traditional 510(k) Khelix Diagnostic Electrophysiology Catheters Page 14/15 K211327 {17}------------------------------------------------ Image /page/17/Picture/0 description: The image contains the logo for "CATHRX". The logo features a stylized heart shape on the left, followed by the text "CATH" in bold, sans-serif font. To the right of "CATH", the letters "RX" are enclosed in a rounded rectangle. ## Variable Loop Steerable Models | Product Reference<br>Number | Allocated Item<br>Number | Curve Length<br>(cm) | Deflection Type | Loop Type | Loop Size<br>(mm) | # of<br>Electrodes | Tip Electrode | Spacing<br>(mm) | |-----------------------------|--------------------------|----------------------|-----------------|-----------|-------------------|--------------------|---------------|-----------------| | SC2-B2515U-10P7888 | CCD0069 | 5.1 | Unidirectional | Variable | 25-15 | 10 | Passive | 8-8-8 | | SC2-B2515U-20P7262 | CCD0040 | 5.1 | Unidirectional | Variable | 25-15 | 20 | Passive | 2-6-2 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...