K964019 · Nellcor Puritan Bennett, Inc. · BZD · Jun 18, 1997 · Anesthesiology
Device Facts
Record ID
K964019
Device Name
MORPHEE PLUS ALPHA
Applicant
Nellcor Puritan Bennett, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Jun 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Morphee Plus Alpha is intended to manage obstructive sleep apnea by delivering air pressure through a tube and mask. It is intended for use in either the hospital or home environment.
Device Story
Morphee Plus Alpha is an obstructive sleep apnea management system consisting of a line-powered control unit, air tube, and patient mask. The device delivers positive air pressure to the patient's airway to manage apnea. It operates in two modes: continuous (CPAP) and automatic. In automatic mode, the control unit senses patient respiration and adjusts airway pressure accordingly. The device is intended for use in hospital or home environments. By providing regulated air pressure, the system assists in maintaining airway patency during sleep, potentially reducing apnea events and improving patient outcomes.
Clinical Evidence
Bench testing only. The device underwent electrical safety, electromagnetic compatibility, mechanical, environmental, and biocompatibility testing. No clinical trial data provided.
Technological Characteristics
Line-powered control unit with air tube and mask interface. Features continuous (CPAP) and automatic pressure adjustment modes. Respiratory sensing capability integrated into the control unit. Complies with national and international electrical safety and electromagnetic compatibility standards.
Indications for Use
Indicated for the management of obstructive sleep apnea in patients requiring positive air pressure therapy. Intended for use in hospital or home settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
Predicate Devices
Companion 318 Nasal CPAP System - Nellcor Purtitan Bennett
DPAP Interactive Airway Management System - SleepNet Corporation
Related Devices
K974879 — MODEL 7410 VOYAGER · Healthdyne Technologies, Inc. · Jun 29, 1998
K063830 — REMSTAR AFLEX CPAP SYSTEM · Respironics Inc., Sleep & Home Respiratory Group · Mar 9, 2007
K020730 — ORION NASAL CPAP SYSTEM · Bird Products Corp. · May 22, 2002
K970771 — SULLIVAN AUTOSET PORTABLE II NASAL CPAP SYSTEM · Resmed, Ltd. · Jun 20, 1997
K042945 — AURA CPAP SYSTEM · Aeiomed, Inc. · Mar 9, 2005
Submission Summary (Full Text)
{0}
K964019
JUN 18 1997
510(k) Summary
## General Information
Nellcor Puritan Bennett
10200 Valley View Road
Eden Prairie, MN 55344
Submitter's Name: Chris Hadland
Phone: (612) 941-3006
Fax Number: (612) 829-5423
## Proprietary Name of the Device
Morphee Plus Alpha
## Common Name of the Device
Obstructive Sleep Apnea Management System
## Device Classification
Obstructive sleep apnea management systems have been classified as class II by the Anesthesiology and Respiratory Devices Panel. Devices of this type have a classification code of 73FLS, Apnea Detector (21CFR 868.2375) and 73BZD, Non-continuous Ventilator (868.5905).
## Intended Use
Morphee Plus Alpha is intended to manage obstructive sleep apnea through a predetermined program of positive air pressure administered through a tube and mask.
## Predicate Device Equivalence
Companion 318 Nasal CPAP System - Nellcor Purtitan Bennett
DPAP Interactive Airway Management System - SleepNet Corporation
## Device Description
The Morphee Plus Alpha System consists of a line powered control unit, an air tube and a patient mask. The mask is attached to the patient covering the nose and secured with straps. The air tube attaches to the mask and the control unit allowing airflow from the control unit to the mask as well as respiratory sensing by the control unit.
{1}
The control unit offers both continuous (CPAP) and automatic modes. In the continuous mode, the unit functions as a standard CPAP product. The auto mode allows the control unit to adjust the airway pressure based on the patient's respiration.
## Summary of Performance Testing
The Morphee Plus Alpha System was tested to national and international product safety standards for electrical safety and electromagnetic compatibility. The product was also tested for mechanical, environmental and biocompatibility.
The product met all requirements and is suitable for the intended use.
## Conclusions
We conclude that the Morphee Plus Alpha System meets its stated performance specifications and meets international safety standards. The clinical indications for use are substantially equivalent to the predicate products as well as the theory of operation and basic product design.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Chris Hadland
JUN 18 1997
Nellcor Puritan Bennett Inc.
10200 Valley View Road
Eden Prairie, Minnesota 55344
Re: K964019
*Morphee Plus Alpha
Regulatory Class: II (two)
Product Code: 73 BZD
Dated: March 17, 1997
Received: March 20, 1997
Dear Mr. Hadland:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Page 2 - Mr. Chris Hadland
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}
# Statement of Intended Use
Device Name:
Morphee Plus Alpha
## Intended Use:
The Morphee Plus Alpha is intended to manage obstructive sleep apnea by delivering air pressure through a tube and mask. It is intended for use in either the hospital or home environment.
## Concurrence of CDRH, Office of Device Evaluation
510(k) number: K964019
☑ Prescription Use
— Over the Counter

Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.