K130037 · Breathe Technologies, Inc. · BZD · May 17, 2013 · Anesthesiology
Device Facts
Record ID
K130037
Device Name
BREATHE CPAP SYSTEM
Applicant
Breathe Technologies, Inc.
Product Code
BZD · Anesthesiology
Decision Date
May 17, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The continuous positive airway pressure (CPAP) System is indicated for the treatment of obstructive sleep apnea (OSA) in patients > 661b (30 kg). The Breathe CPAP System is intended for single-patient re-use in the home environment.
Device Story
Breathe CPAP System is an AC-powered, microprocessor-controlled flow generator (blower) designed to treat obstructive sleep apnea (OSA). Device delivers physician-prescribed continuous positive airway pressure during inspiratory and expiratory phases to maintain airway patency during sleep. System includes flow generator and nasal pillows mask interface secured by headgear. Used in home environment by patients; operated by patient. Device generates CPAP from room air, transmitting it through patient circuit to nasal interface. Provides therapeutic pressure to prevent airway collapse; benefits patient by reducing OSA symptoms. Healthcare provider prescribes device and monitors therapy efficacy.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility evaluation per ISO 10993-1 (2009) and compliance with electrical safety and performance standards (IEC 60601-1, ANSI/AAMI/IEC 60601-1-2, ASTM F1246-91).
Technological Characteristics
AC-powered, microprocessor-controlled flow generator. Includes nasal pillows mask interface. Compliant with ASTM F1246-91 (2005), IEC 60601-1 (1988/1991/1995), and ANSI/AAMI/IEC 60601-1-2 (2007). Biocompatibility per ISO 10993-1 (2009).
Indications for Use
Indicated for treatment of obstructive sleep apnea (OSA) in patients > 30 kg (66 lb). Intended for single-patient home use.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
K042945 — AURA CPAP SYSTEM · Aeiomed, Inc. · Mar 9, 2005
K052597 — BREATHEX OMEGA CPAP DEVICE, MODEL 322 · Hoffman Laboratories, LLC · Dec 16, 2005
K041010 — REMSTAR AUTO WITH C-FLEX CPAP SYSTEM · Respironics, Inc. · May 5, 2004
K032480 — AUTOSET SPIRIT SYSTEM · Resmed, Ltd. · Oct 16, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the word "Breathe" in a simple, sans-serif font. A curved line is drawn beneath the word, resembling a wave or a breath of air. The text and line are black against a white background, creating a clean and minimalist design.
# Section 5: 510(k) Summary
| Category | Comments | |
|--------------------------|-----------------------------------------|--|
| Sponsor: | Breathe Technologies | |
| | 175 Technology Drive, Suite 100 | |
| | Irvine, CA 92618 | |
| | Tel: 949-988-7700 | |
| Correspondent Contact | Craig Coombs | |
| Information: | President | |
| | Coombs Medical Device Consulting, Inc | |
| | 1193 Sherman St. | |
| | Alameda, CA 94501 | |
| | Office: 510.337.0140 | |
| | Fax: 510.337.0416 | |
| Device Common Name: | CPAP System | |
| Device Classification | 21 CFR 868.5905 | |
| Regulation & Name: | Ventilator, Non-Continuous (Respirator) | |
| Device Classification & | Class II, | |
| Product Code: | BZD | |
| Device Proprietary Name: | Breathe CPAP System | |
# Predicate Device Information:
| Predicate Device: | XT1 CPAP Model 9S-005 |
|-------------------------------------------|--------------------------|
| Predicate Device Manufacturer: | Apex Medical Corporation |
| Predicate Device Premarket Notification # | K070609 |
| Predicate Device Common Name: | CPAP |
| Predicate Device Classification & Name: | 21 CFR 868.5905 |
| Predicate Device Classification & | Class II |
| Product Code: | BZD |
# Reference Device Information:
| Predicate Device: | InnoMed IT Nasal Mask |
|----------------------------------------------------|--------------------------------------------------------------|
| Predicate Device Manufacturer: | InnoMed Technologies |
| Predicate Device Premarket Notification #: | K050171 |
| Predicate Device Common Name: | Ventilator, continuous non-life supporting<br>(accessory to) |
| Predicate Device Classification & Name: | 21 CFR 868.5905 |
| Predicate Device Classification &<br>Product Code: | Class II<br>BZD |
·
# b. Date Summary Prepared
4 January 2013
{1}------------------------------------------------
## c. Description of Device
The Breathe Technologies™ CPAP system is an AC powered Continuous Positive Airway Pressure (CPAP) device that delivers a physician prescribed pressure during the inspiratory phase and expiratory phases to help maintain airway patency during sleep for patients with obstructive sleep apnea (OSA).
The Breathe CPAP System consists of a programmable, microprocessor controlled, flow generator (i.e., blower) and a Nasal Pillows Mask. The flow generator generates CPAP from room air to the patient and transmits it via the attached Nasal Pillows Mask. The end of the patient circuit is a nasal pillows interface which is worn on the nose. The nasal interface is held in place with standard headgear.
#### d. Intended Use
The continuous positive airway pressure (CPAP) System is indicated for the treatment of obstructive sleep apnea (OSA) in patients > 661b (30 kg). The Breathe CPAP System is intended for single-patient re-use in the home environment.
### e. Comparison to Predicate Device
The Breathe Technologies CPAP is substantially equivalent in Intended Use, Indications for Use, technology, design, materials, physician or patient use, and energy source to the predicate Apex Medical XTI CPAP Model 9S-005 (K070609) with the InnoMed IT Nasal Mask (K050171).
- The testing described below demonstrates that the differences in the devices do not raise any unresolved issues of safety or efficacy.
Breathe Technologies concludes that the devices are substantially equivalent.
#### f. Summary of Supporting Data
Biocompatibility data demonstrates that the device is in compliance with ISO 10993-1 (2009) Biological evaluation of medical devices -- Part 1: Evaluation and testing
The Breathe CPAP System is compliant with the following Standards and Guidance:
- FDA Draft Reviewer Guide for Ventilators (July 1995) .
- ASTM F1246-91 (2005), Standard Specification for Electrically Powered Home . Care Ventilators
- IEC 60601 1 (1988), Amendment 1 (1991-11), Amendment 2 (1995): Medical . electrical equipment - General Requirements for Safety.
- ANSI/AAMI/IEC 60601-1-2 (2007), Medical electrical equipment Part 1-2: . General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized symbol that resembles an abstract representation of a human figure or a caduceus. The symbol consists of three curved lines that converge at the bottom, creating a sense of movement or flow.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## May 17, 2013
Breathe Technologies C/O Mr. Craig Coombs President Coombs Medical Devices Consulting, Incorporated 1193 Sherman Street ALAMEDA CA 94501
Re: K130037
Trade/Device Name: Breathe CPAP System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: April 15, 2013 Received: April 18, 2013
## Dear Mr. Coombs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Coombs
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Kwame O. Ulmer -S for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and
Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
Breathe
Breathe CPAP System Traditional Premarket Notification
## Section 4: Indications for Use Statement
510(k) Number (if known):
Device Name: Breathe CPAP System
Indications for Use:
The continuous positive airway pressure (CPAP) System is indicated for the treatment of obstructive sleep apnea (OSA) in patients > 661b (30 kg). The Breathe CPAP System is intended for single-patient re-use in the home environment.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
# Paul H. Shin = 2013.05:15 15:59:53 -04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number;
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.