K152356 · Respironics, Inc. · BZD · Jan 8, 2016 · Anesthesiology
Device Facts
Record ID
K152356
Device Name
Sapphire
Applicant
Respironics, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Jan 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Sapphire is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Sapphire provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. Sapphire allows read-only access to patients. Sapphire is intended to be used in hospital, institutional, provider, and home care settings.
Device Story
Sapphire is a web-based application for managing patients using compatible sleep and respiratory therapy devices. It collects therapy data wirelessly or via SD card into a centralized database. Healthcare providers (physicians, clinicians, DME providers) access this data through a web interface to view therapy reports, manage patient information, and remotely update prescription or performance settings for non-life support devices. The system supports role-based access control and configuration of compliance rules. By enabling remote monitoring and settings adjustment, the device facilitates clinical follow-up and therapy management, potentially improving patient adherence and treatment outcomes in home and clinical settings.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through non-clinical software verification and validation testing, including system testing of data transfer, display, and remote settings adjustment functionality.
Technological Characteristics
Web-based application; centralized database architecture; supports wireless and SD card data transfer; role-based access control; software-based data management and remote configuration; moderate level of concern for software; cybersecurity controls include encryption of data at rest and in transit.
Indications for Use
Indicated for patients prescribed compatible non-life support therapy devices; used by healthcare representatives (physicians, clinicians, DME providers) for remote data collection, viewing, and adjustment of prescription/performance settings in hospital, institutional, provider, and home care settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
K083816 — RES TRAXX ONLINE, MODEL: R14 · Resmed, Ltd. · Mar 20, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 8, 2016
Respironics Inc. Colleen Witt Manager, Regulatory Affairs 1740 Golden Mile Highway Monroeville, Pennsylvania 15146
Re: K152356
Trade/Device Name: Sapphire Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator Regulatory Class: Class II Product Code: BZD, MNS, MNT, CBK, NOU Dated: December 4, 2015 Received: December 5, 2015
Dear Colleen Witt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152356
Device Name Sapphire
### Indications for Use (Describe)
Sapphire is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Sapphire provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. Sapphire allows read-only access to patients. Sapphire is intended to be used in hospital, institutional, provider, and home care settings.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# TAB 5
# 510(K) SUMMARY
#### I. Submitter
| Official Contact | Colleen Witt<br>Manager, Regulatory Affairs<br>colleen.witt@philips.com<br>Respironics Inc.<br>1740 Golden Mile Highway<br>Monroeville, PA 15146 | | |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | Phone: 724-387-4870 | | |
| | Fax: 724-387-7490 | | |
Date of Preparation January 8, 2016
#### II. Device
| Name of Device: | Sapphire |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Data Management System |
| Device Classification: | Class II |
| Classification Name/<br>Product Code: | Non-continuous ventilator (21 CFR 868.5905, Product Code BZD)<br>Continuous ventilator (21 CFR 868.5895, Product Codes MNS, MNT,<br>CBK, NOU) |
#### Legally Marketed Predicate Devices III.
- K140054 ResScan, ResMed LTD.
- . K132371 EasyCare Online, ResMed LTD.
#### IV. Device Description
Sapphire is a web application that is used to manage patients utilizing compatible sleep and respiratory devices, view therapy data, generate reports and update device settings for applicable therapy devices. Sapphire is intended to be used by healthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) in hospital, institutional, provider and home settings. Device data is transferred either wirelessly or from an SD card to a centralized database. A user can view this data through a web based application on their computer.
{4}------------------------------------------------
Specific roles are assigned within Sapphire. Based on these roles and access rights, users can perform the following tasks within Sapphire:
- Add new users with an assigned role to Sapphire ●
- Add new patients, along with applicable device information, to Sapphire
- Therapy Data Management (including viewing and generating therapy data reports) ●
- . Update a patient's prescription/device settings for non-life support devices
- Configure compliance rules associated with a patient .
#### V. Indications for Use
Sapphire is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Sapphire provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. Sapphire allows read-only access to patients. Sapphire is intended to be used in hospital, institutional, provider, and home care settings.
#### Comparison of Technological Characteristics with the Predicate VI.
## Device
Sapphire has the following similarities to the legally marketed predicate devices:
- Similar intended use
- . Same operating principle
- Similar technologies ●
| Feature/Function | Predicate Device #1 | Predicate Device #2 | Subject Device | Comments |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Device Name:<br>ResScan | Device Name:<br>EasyCare Online | Device Name:<br>Sapphire | Similar to K140054 and K132371.<br>Intended uses have similar intent with respect to transferring device and therapeutic data and changing device settings in compatible non-life support therapy devices. |
| | 510(k) Number:<br>K140054 | 510(k) Number:<br>K132371 | 510(k) Number:<br>K152356 | |
| | Manufacturer:<br>ResMed LTD | Manufacturer:<br>ResMed LTD | Manufacturer:<br>Respironics | |
| | ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and therapeutic information. This includes the ability to remotely change settings in non-life support devices only. | EasyCare Online is a web based solution for healthcare specialists intended to:<br>- Assist in the diagnosis of sleep disordered breathing in adult patients through analysis of data recorded | Sapphire is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Sapphire provides remote patient data collection & | |
| | It is intended to be<br>used by Clinicians<br>in conjunction with<br>ResMed<br>compatible<br>therapy devices,<br>using ResMed's<br>proprietary<br>communications<br>protocol. | by an<br>EasyCare<br>Online<br>compatible<br>home sleep<br>test device.<br>Transfer and<br>display, usage<br>and<br>therapeutic<br>information<br>that has been<br>transmitted<br>remotely from<br>the patient's<br>therapy device<br>located in the<br>home. It is<br>intended to<br>support the<br>standard<br>follow-up care<br>of patients<br>that have<br>been<br>prescribed a<br>compatible<br>ResMed<br>therapy device<br>for the<br>treatment of<br>obstructive<br>sleep apnea<br>or respiratory<br>insufficiency.<br>EasyCare<br>Online also<br>provides<br>remote<br>settings<br>capabilities. | viewing and is<br>intended to be<br>used by<br>healthcare<br>representatives<br>(e.g., Physicians,<br>Clinicians,<br>Durable Medical<br>Equipment<br>providers) in<br>conjunction with<br>compatible non-<br>life support<br>therapy devices to<br>adjust prescription<br>and/or<br>performance<br>settings. Sapphire<br>allows read-only<br>access to patients.<br>Sapphire is<br>intended to be<br>used in hospital,<br>institutional,<br>provider, and<br>home care<br>settings. | |
| Application | PC based<br>application | Web based<br>application | Web based<br>application | Equivalent to<br>K132371 |
| Data storage | Data stored on<br>Clinician's<br>computer | Data stored on<br>centralized<br>database | Data stored on<br>centralized<br>database | Equivalent to<br>K132371. |
| Functionality | Patient<br>management<br>• Display<br>therapy data<br>• Generate<br>reports<br>• Settings<br>management<br>(remote | Patient<br>• management<br>• Display<br>therapy data<br>• Compliance<br>reports<br>• Therapy<br>reports | Patient<br>• management<br>• Display<br>therapy data<br>• Generate<br>reports<br>• Settings<br>management<br>for non-life | Equivalent to<br>K140051<br>Similar to<br>K132371.<br>Sapphire has the<br>same<br>functionality as<br>K132371 with<br>respect to patient<br>management, |
| | setting change<br>functionality<br>only applies to<br>non-life<br>support<br>devices) | Settings management Diagnostic Home Sleep Test Reports | supporting<br>devices | data display,<br>report generating and settings<br>management.<br>Sapphire does not create<br>diagnostic home sleep test reports<br>as was cleared in<br>K132371. |
| Settings<br>management | Ability to remotely<br>change device<br>settings in non-life<br>support devices<br>only. | Ability to remotely<br>change device<br>settings. | Ability to remotely<br>change device<br>settings in non-life<br>support devices<br>only. | Equivalent to<br>K140051 and<br>K132371. |
| Reports | Summary report Detailed report Compliance report | Detailed report Compliance report Therapy report | Detailed Report<br>(includes compliance<br>information) | Similar to<br>K140051 and<br>K132371. |
{5}------------------------------------------------
{6}------------------------------------------------
#### Performance Data: VII.
### Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." This software was considered to have a "moderate" level of concern, since a failure or latent flaw in the software could result in minor harm to the patient.
### Non-Clinical Tests
Software verification and validation testing was performed on Sapphire based on the product requirements. This testing included complete system testing to verify data transfer from therapy devices to Sapphire, through both wireless data transfer. Once data transfer. Once data was transferred, all tests confirmed that Sapphire can display patient and device information, display therapy data including compliance and therapy reports and allow for a user to create and/or edit a patients prescription for applicable therapy device. Verification of cybersecurity requiremented within the system architecture were also confirmed for encryption of data at rest and in transit.
{7}------------------------------------------------
The testing of Sapphire verified that all product requirements have been met with passing test results. The verification and validation testing demonstrated comparable safety and effectiveness of the Sapphire system in comparison to the predicate.
### Clinical Tests
Clinical tests were not required to demonstrate the safety and effectiveness of Sapphire. Product functionality has been adequately assessed by non-clinical tests.
# VIII. Conclusion
Sapphire has passed all of the aforementioned non-clinical tests and required no clinical tests in order to demonstrate safety or effectiveness. It is therefore concluded that Sapphire is substantially equivalent to the predicate devices ResScan, ResMed LTD. (K140054) and EasyCare Online, ResMed LTD. (K132371) in terms of safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.