EasyCare Online is a web based solution for healthcare specialists intended to: assist in the diagnosis of sleep disordered breathing in adult patients through analysis of data recorded by an EasyCare Online compatible home sleep test device. transfer and display, usage and therapeutic information that has been transmitted remotely from the patient's therapy device located in the home. It is intended to support the standard follow-up care of patients that have been prescribed a compatible ResMed therapy device for the treatment of obstructive sleep apnea or respiratory insufficiency. EasyCare Online also provides remote settings capabilities
Device Story
EasyCare Online is a web-based application for healthcare specialists to manage patient data from ResMed therapy devices and home sleep test (HST) devices. Inputs include respiratory nasal airflow, respiratory effort, pulse rate, and oxygen saturation from HST devices, plus usage/therapeutic data from flow generators. Data is transmitted wirelessly or via SD card/internet to a central database. The application displays this data to clinicians/DME providers via web browser. It supports diagnosis of sleep-disordered breathing, monitoring of therapy compliance, and remote adjustment of device settings. By enabling timely clinical intervention and remote monitoring, the device aims to optimize therapy for obstructive sleep apnea and respiratory insufficiency.
Clinical Evidence
No clinical testing was required. Evidence consists of non-clinical verification, including regression and end-to-end system testing to confirm data transfer integrity, and side-by-side testing using three patient breathing pattern scripts to verify that the modified software correctly and accurately displays home sleep test data compared to the ApneaLink Plus PC software.
Technological Characteristics
Web-based application; centralized database architecture; supports wireless, SD card, and file upload/internet data transfer. Integrates with ResMed flow generators and home sleep test devices. Software-based data management and reporting system.
Indications for Use
Indicated for adult patients with sleep disordered breathing, obstructive sleep apnea, or respiratory insufficiency requiring home sleep testing or therapy monitoring.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
#### KJSZ371 NOV 2 2 2013
ResMed
ECO Traditional 510(k) Premarket
Notification
Jul 2013
# 4A 510(k) Summary
| 510(k) Summary | | | | | |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Date Prepared | 26 Jul, 2013 | | | | |
| Submitter | Greg Dockar<br>Senior Regulatory Affairs Manager<br>ResMed Ltd, Australia<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW 2153, Australia<br>+ 61 2 8884 2157 (Phone)<br>+ 61 2 8884 2000 (FAX)<br>GregD@resmed.com.au | | | | |
| Official Contact | Jim Cassi<br>Vice President Quality Assurance Americas<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123<br>+1 858 836 6081<br>+1 858 836 5519<br>Jim. Cassi@resmed.com | | | | |
| Device name | EasyCare Online | | | | |
| Product codes | 73 MNS<br>73 MNR | | | | |
| Classification reference | Ventilator, Continuous, Non-Life-Supporting (21 CFR 868.5895,<br>Product Code 73 MNS)<br>Breathing Frequency Monitor (21CFR 868.2375 product code 73<br>MNR) | | | | |
| Predicate device(s) | EasyCare Online (510(k) number (K123557)<br>ApneaLink Plus (510(k) number (K083575) | | | | |
| Reason for submission | Expanded Intended for Use. | | | | |
{1}------------------------------------------------
EasyCare Online is a web based solution for healthcare specialists intended to:
- assist in the diagnosis of sleep disordered breathing in adult patients through analysis of . data recorded by an EasyCare Online compatible home sleep test device.
- transfer and display, usage and therapeutic information that has been transmitted . remotely from the patient's therapy device located in the home. It is intended to support the standard follow-up care of patients that have been prescribed a compatible ResMed therapy device for the treatment of obstructive sleep apnea or respiratory insufficiency. EasyCare Online also provides remote settings capabilities
## Device Description
EasyCare Online is a web application that can transfer, store, manage and display usage and therapy data from (ResMed compatible flow generators) and diagnostic data from (ResMed compatible sleep study devices). The data is transferred from the device either wirelessly through a communications module, or with the aid of an SD card and internet technology, to a central database and then displayed on the clinician's/DME's computer, through a web based application such as internet Explorer.
EasyCare Online is used to monitor and optimize the therapy of patients diagnosed with sleep apnea or respiratory insufficiency, who are using a ResMed therapy device or home sleep test study device. The application enables patient usage data to be shared across several different user groups for the primary purpose of monitoring patient compliance and home sleep studies. Clinical users. Home Medical Equipment (DME) providers and other healthcare specialists can access data with ResMed approved user accounts. Also, clinical users and DMEs are able to address any clinical issues in a timely manner and provide the necessary patient support, including the modification of device settings. EasyCare Online supports physicians in the diagnosis Sleep Disordered Breathing, by reviewing home sleep test results via their web browser or download the clinical data to their PC for further analysis.
## Basis for Determination of Substantial Equivalence
The modified EasyCare Online has the following similarities to the previously cleared predicated device:
- Similar intended use .
- Same operating principle .
- 0 Same technology
- Same manufacturing (deployment) process 0
EasyCare Online modifications provide additional features for collection, storage and reporting of respiratory nasal airflow, respiratory effort, pulse rate and oxygen saturation during sleep studies which is used to assist the physician in diagnosis sleep disordered breathing. The physician prescribed device will help to recognize sleep-related respiratory disorders and lead to comprehensive clinical diagnosis and therapy.
{2}------------------------------------------------
With the ApneaLink Plus (also referred to as ApneaLink to Cloud - ALC) Software integration with EasyCare Online, a regression analysis showed that minimal affect on the existing system would occur. Confidence system level testing was conducted to ensure integration of ALC was conducted and found not to change the performance of EasyCare Online. Accordingly, 73 MNS and 73 MNR ResMed compatible devices function correctly with EasyCare Online. These changes do not affect the safety and efficacy of the device, as the EasyCare Online uses the existing data types and communications mechanisms between therapy devices as supported by the predicate device (K123557).
| Feature | EasyCare Online [Primary predicate(K123557)] | ApneaLink Plus [Secondary predicate(K083575)] | EasyCare Online (modified) | Comments |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Intended<br>Use | EasyCare Online<br>transfers and displays<br>to physicians, usage<br>and therapeutic<br>information that has<br>been transmitted<br>remotely from the<br>patient's flow generator<br>located in the home.<br>EasyCare Online also<br>provides remote<br>settings capabilities.<br><br>EasyCare Online is<br>intended to support the<br>standard follow-up care<br>of patients that have<br>been prescribed a<br>compatible ResMed<br>flow generator for the<br>treatment of obstructive<br>sleep apnea or<br>respiratory insufficiency | The ApneaLink Plus<br>device is indicated for<br>use by Health Care<br>Professionals(HCP),<br>where it may aid in the<br>diagnosis of sleep<br>disordered breathing for<br>adult patients.<br>ApneaLink Plus records<br>the following data:<br>patient respiratory nasal<br>airflow, snoring, blood<br>oxygen saturation,<br>pulse and respiratory<br>effort during sleep. The<br>device uses these<br>recordings to produce a<br>report for the HCP that<br>may aid in the diagnosis<br>of sleep disordered<br>breathing or for further<br>clinical investigation. | EasyCare Online is a web<br>based solution for healthcare<br>specialists intended to:<br>• assist in the diagnosis of<br>sleep disordered breathing in<br>adult patients through<br>analysis of data recorded by<br>an EasyCare Online<br>compatible home sleep test<br>device.<br>• transfer and display, usage<br>and therapeutic information<br>that has been transmitted<br>remotely from the patient's<br>therapy device located in the<br>home. It is intended to<br>support the standard follow-<br>up care of patients that have<br>been prescribed a<br>compatible ResMed therapy<br>device for the treatment of<br>obstructive sleep apnea or<br>respiratory insufficiency.<br>EasyCare Online also<br>provides remote settings<br>capabilities | Equivalent<br>Combined and<br>simplified-<br>same intent. |
| Functionality | • Centralised<br>database<br>• Compliance reports<br>• Therapy reports.<br>• Settings<br>management<br>• Patient management | • Diagnostic Home<br>Sleep Test Reports<br>(HSTR) | • Centralised database<br>• Compliance reports<br>• Therapy reports.<br>• Settings management<br>• Patient management<br>• Diagnostic Home Sleep<br>Test Reports (HSTR) | Equivalent<br>Combined<br>features |
| Data<br>transfer<br>Technology | • Wireless<br>• SD Card/Internet | USB connection from<br>ApneaLink Plus to the<br>Personal Computer | • Wireless<br>• SD Card/Internet<br>• File upload/Internet | Equivalent |
{3}------------------------------------------------
ResMed
| Feature | EasyCare Online<br>[Primary<br>predicate(K123557)] | ApneaLink Plus<br>[Secondary<br>predicate(K083575)] | EasyCare Online (modified) | Comments |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Therapy<br>settings | • Pressure<br>• Mode<br>• Comfort | — | • Pressure<br>• Mode<br>• Comfort | Equivalent |
| Patient<br>information | • Mask Leak<br>• AHI<br>• Prescription<br>Pressure<br>• Start Pressure<br>• Minute Ventilation<br>• Respiratory rate<br>• Mode<br>• EPR Level<br>• Pressure Support<br>Level | — | • Mask Leak<br>• AHI<br>• Prescription Pressure<br>• Start Pressure<br>• Minute Ventilation<br>• Respiratory rate<br>• Mode<br>• EPR Level<br>• Pressure Support Level | Equivalent |
| Home Sleep<br>Test<br>Information<br>(HSTI) | — | • Recording times<br>• Event statistics<br>(including AHI, apnea<br>index, hypopnea<br>index, apnea<br>classification)<br>• Oxy statistics (ODI,<br>oxygen saturation)<br>• Pulse statistics<br>• Breath statistics | • Recording times<br>• Event statistics (including<br>AHI, apnea index,<br>hypopnea index, apnea<br>classification)<br>• Oxy statistics (ODI, oxygen<br>saturation)<br>• Pulse statistics<br>• Breath statistics | Equivalent<br>Side by side<br>testing of ECO<br>(modified) and<br>ApneaLink<br>Plus PC<br>software<br>demonstrated<br>that ECO<br>correctly and<br>accurately<br>displayed<br>home sleep<br>test data<br>belonging to<br>ApneaLink<br>Plus devices. |
## Non Clinical Testing
Design and non-clinical verification activities were performed on EasyCare Online as a result of the updated risk analysis and design requirements. Verification testing included end-to-end testing to verify data transfer integrity between end user browser PC and EasyCare Online was successfully implemented. All tests confirmed the product met the predetermined acceptance criteria. ResMed has determined that new version of EasyCare Online is substantially equivalent to the predicate device, EasyCare Online (K123557).
Side-by-Side testing using the same data showed that both the predicate device ApneaLink Plus (K083575) and EasyCare Online produced the same home sleep test data. EasyCare Online met the predetermined pass/fail criteria.
{4}------------------------------------------------
| Testing | Scope of testing | Result |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Software<br>verification | Regression and End-to-End system testing was conducted to confirm that<br>changes introduced in the code did not impact on previous ECO functionality<br>(K123557) and verify the addition of ApneaLink Plus (K083575) functionality into<br>ECO. | PASS |
| Side-by-side<br>testing | Three patient breathing pattern scripts were used for side by side testing of ECO<br>(modified) and ApneaLink Plus PC software ((K083575) to verify that ECO<br>correctly and accurately displayed home sleep test data belonging to ApneaLink<br>Plus devices. | PASS |
## Clinical Testing
No clinical testing is required for EasyCare Online.
## Conclusion
The modified version of EasyCare Online is substantially equivalent to the predicate device, EasyCare Online (K123557)
The modified version of EasyCare Online is as safe and as effective, and performs substantially equivalent to the predicate devices, EasyCare Online (K123557) and ApneaLink Plus (K083575).
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the emblem.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2013
Resmed Corporation Jim Cassi Vice President, Quality Assurance Americas 9001 Spectrum Center Blvd. SAN DIEGO, CA 92123
Re: K132371 Trade Name: EasyCare Online Regulatory Class: III Product Code: MNS Dated: October 16, 2013 Received: October 21, 2013
Dear Mr. Cassi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Mr. Jim Cassi
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/7 description: The image contains a logo with the text "Tejashri Purohit-Sheth, M.D." and "Clinical Deputy Director." The logo also includes the acronym "BAGRID." The word "FOR" is present at the bottom right of the image.
Erin Keith Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
510(k) Number (if known) K132371
Device Name EasyCare Online
Indications for Use (Describe)
EasyCare Online is a web based solution for healthcare specialists intended to:
• assist in the diagnosis of sleep disordered breathing in adult patients through analysis of data recorded by an EasyCare Online compatible home sleep test device.
· transfer and display, usage and therapeutic information that has been transmitted remotely from the patient's therapy device located in the home. It is intended to support the standard follow-up care of patients that have been prescribed a compatible ResMed therapy device for the treatment of obstructive sleep apnea or respiratory insufficiency. EasyCare Online also provides remote settings capabilities
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
・・・・・・・ FOR FOR FDA USE ONLY . . . . 1-2-2-2 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/13 description: The image shows a digital signature of Anya C. Harry, along with the FDA logo. The signature includes a DN (Distinguished Name) that specifies the U.S. Government, HHS (Health and Human Services), and People. The date and time of the signature are also included, which is November 20, 2013, at 15:25:31 with a time zone offset of -05'00'.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.