Mojo Full Face Vented Mask; Veraseal 2 Full Face Vented Mask; Innova Full Face Vented Mask; Ascend Full Face Vented Mask
Applicant
Sleepnet Corporation
Product Code
BZD · Anesthesiology
Decision Date
Oct 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sleepnet Mojo Full Face Vented Mask, Veraseal 2 Full Face Vented Mask, Innova Full Face Vented Mask and Ascend Full Face Vented Mask are intended to be used with positive airway pressure devices, such as CPAP or bilevel, operating at or above 3 cm H2O. The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. Mojo is intended for Single patient, multi-use in the home or hospital/institutional environment. Veraseal 2 is intended for Short Term Single Patient use (maximum of 7 days) in the hospital institutional environment. Innova is intended for Single patient, multi-use in the home or hospital/institutional environment. Ascend is intended for Single patient, multi-use in the home or hospital/institutional environment.
Device Story
Full face vented masks (Mojo, Veraseal 2, Innova, Ascend) interface between positive airway pressure (PAP) devices and patient face; deliver pressurized air to treat sleep apnea. Masks feature exhalation port and anti-asphyxia valve (AAV) to flush CO2. Operated by patients or clinicians in home or hospital environments. Connects to standard 22mm CPAP circuits. Provides seal over nose and mouth; enables therapy delivery. Benefits include effective pressure delivery for prescribed therapy. Mojo model includes magnets; contraindications exist for patients with specific metallic/medical implants.
Clinical Evidence
Bench testing only. Conformance to ISO 17510 demonstrated for exhaust flow, resistance to flow, AAV pressure, AAV resistance in single-fault conditions, internal volume/dead space, and CO2 rebreathing. Biocompatibility, transportation, shelf-life, and reprocessing data leveraged from predicate devices.
Technological Characteristics
Full face vented masks with mask shell, gel cushion, vented elbow, forehead arm/pad, and headgear. Standard 22mm connector (ISO 5356-1). Materials identical to predicate (K120463/K190254). Mojo includes magnets (380mT). Non-sterile. ISO 10993-1 compliant for permanent skin/tissue contact.
Indications for Use
Indicated for adult patients (>30kg) requiring positive airway pressure therapy (CPAP or bilevel) at or above 3 cm H2O. Mojo, Innova, and Ascend are for single-patient, multi-use in home or hospital settings. Veraseal 2 is for short-term single-patient use (max 7 days) in hospital settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
Predicate Devices
Mojo Full Face Mask and Veraseal Full Face Mask (K120463)
K190254 — Mojo 2 Full Face Vented Mask, Veraseal 3 Full Face Vented Mask, V3 Full Face Vented Mask · Sleepnet Corporation · Jul 5, 2019
K060273 — SLEEPNET MOJO FULL FACE MASK -SMALL, MODEL NO. 53835, MEDIUM ,MODEL NO.53836,LARGE,MODEL NO. 53837 · Sleepnet Corporation · May 2, 2006
K241520 — Mojo 2 Full Face Non-Vented Mask; Mojo 2 Full Face AAV Non-Vented Mask; Veraseal 3 Full Face Non-Vented Mask; Veraseal 3 Full Face AAV Non-Vented Mask · Sleepnet Corporation · Jun 27, 2024
K230476 — Oran Park Mask · Resmed Pty Ltd (Registration Number: 3004604967) · Oct 19, 2023
K150685 — Skynector CPAP Mask · Sky Wise Medical Instrument (Shen Zhen) Co., Ltd. · Aug 24, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
October 18, 2024
Sleepnet Corporation % Paul Dryden President ProMedic. LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K241830
Trade/Device Name: Mojo Full Face Vented Mask; Veraseal 2 Full Face Vented Mask; Innova Full Face Vented Mask; Ascend Full Face Vented Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: September 19, 2024 Received: September 19, 2024
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows a digital signature. The signature is from Binoy J. Mathews. The signature was created on October 18, 2024 at 15:47:34 with a time zone offset of -04'00'.
For
Rachana Visaria Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
### 510(k) Number (if known) K241830
#### Device Name
Mojo Full Face Vented Mask, Veraseal 2 Full Face Vented Mask, Innova Full Face Vented Mask, Ascend Full Face Vented Mask
### Indications for Use (Describe)
The Sleepnet Mojo Full Face Vented Mask, Veraseal 2 Full Face Vented Mask, Innova Full Face Vented Mask and Ascend Full Face Vented Mask are intended to be used with positive airway pressure devices, such as CPAP or bilevel, operating at or above 3 cm H2O. The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed.
Mojo is intended for Single patient, multi-use in the home or hospital/institutional environment.
Veraseal 2 is intended for Short Term Single Patient use (maximum of 7 days) in the hospital institutional environment.
Innova is intended for Single patient, multi-use in the home or hospital/institutional environment.
Ascend is intended for Single patient, multi-use in the home or hospital/institutional environment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| | |
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Sponsor: Sleepnet Corporation 5 Merrill Industrial Drive Hampton, NH 03842 | Tel - 603-758-6625 |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Sponsor Contact: | Jennifer Kennedy – Director of Regulatory and Quality |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name: | Mojo Full Face Vented Mask<br>Veraseal 2 Full Face Vented Mask<br>Innova Full Face Vented Mask<br>Ascend Full Face Vented Mask |
| Common/Usual Name: | Patient interface for CPAP |
| Regulation Number:<br>Regulation Code:<br>Product Code: | 21CFR 868.5905<br>Non-continuous ventilator (IPPB)<br>BZD |
| Predicate Device:<br>Regulation Number:<br>Regulation Code:<br>Product Code: | K120463 - Mojo Full Face Mask and Veraseal Full Face Mask<br>21CFR 868.5905<br>Non-continuous ventilator (IPPB)<br>BZD |
### Device Description:
The Sleepnet Mojo Vented Full Face Mask, Veraseal 2 Full Face Vented Mask, Innova Full Face Vented Mask and Ascend Full Face Vented Mask are used with positive pressure devices, such as CPAP and Bi-level.
The modifications to the cleared Mojo Full Face mask (K120463) are:
- Addition of contraindication and updated warning related to presence of magnets ●
- Change to the headgear material ●
- Addition of an XL size ●
The modifications to the cleared Veraseal Full Face mask (K120463) are:
- Modified design, now called the Veraseal 2 Full Face Vented Mask ●
- . Addition of an XL size
Other modifications include:
- Addition of the Innova Full Face Vented Mask version ●
- Addition of the Ascend Full Face Vented Mask version ●
### Indications for Use:
The Sleepnet Mojo Full Face Vented Mask, Veraseal 2 Full Face Vented Mask, Innova Full Face Vented Mask and Ascend Full Face Vented Mask are intended to be used with positive airway pressure devices, such as CPAP or bilevel, operating at or above 3 cm H2O.
{5}------------------------------------------------
### 510(k) Summary Page 2 of 12 18-Oct-24
The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed.
Mojo is intended for Single patient, multi-use in the home or hospital/institutional environment. Veraseal 2 is intended for Short Term Single Patient use (maximum of 7 days) in the hospital/institutional environment.
Innova is intended for Single patient, multi-use in the home or hospital/institutional environment.
Ascend is intended for Single patient, multi-use in the home or hospital/institutional environment.
Patient Population: For adults (>30 kg)
Environment of Use: Home or hospital / institutional environments
{6}------------------------------------------------
## 510(k) Summary Page 3 of 12
18-Oct-24
### Table of Comparison to Predicate
| Attributes | Subject - Mojo, Veraseal 2, Innova, Ascend Full<br>Face Vented Mask | Predicate - Mojo, Veraseal Full Face Mask,<br>K120463 | Explanation of Differences |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Classification<br>CFR | BZD<br>CFR 868.5905 | BZD<br>CFR 868.5905 | Identical |
| List of Devices | Mojo Full Face Vented Mask<br>Veraseal 2 Full Face Vented Mask<br>Innova Full Face Vented Mask<br>Ascend Full Face Vented Mask | Mojo Full Face mask<br>Veraseal Full Face mask | Identical |
| Indications for Use | The Sleepnet Mojo Full Face Vented Mask, Veraseal<br>2 Full Face Vented Mask, Innova Full Face Vented<br>Mask and Ascend Full Face Vented Mask are<br>intended to be used with positive airway pressure<br>devices, such as CPAP or bilevel, operating at or<br>above 3 cm H2O.<br>The masks are to be used on adult patients (>30kg)<br>for whom positive airway pressure therapy has been<br>prescribed.<br>Mojo -<br>Single patient, multi-use in the home or<br>hospital/institutional environment.<br>Veraseal 2 -<br>Short Term Single Patient use (maximum of 7 days)<br>in the hospital/institutional environment.<br>Innova -<br>Single patient, multi-use in the home or<br>hospital/institutional environment.<br>Ascend -<br>Single patient, multi-use in the home or<br>hospital/institutional environment. | The Sleepnet Mojo and Veraseal masks are intended<br>to be used with positive airway pressure devices, such<br>as CPAP or bilevel, operating at or above 3 cm H2O.<br>The masks are to be used on adult patients (>30kg) for<br>whom positive airway pressure therapy has been<br>prescribed.<br>Mojo -<br>Single patient, multi-use and Multi-patient, multi-use<br>Veraseal -<br>Disposable single use; Single patient, multi-use up to<br>7 days<br>The masks are intended for use in the home or<br>hospital/institutional environment. | Similar.<br>The multi-patient use is being<br>removed. |
| Patient Population | Adult (>30 kg) | Adult (>30 kg) | Identical |
| Attributes | Subject - Mojo, Veraseal 2, Innova, Ascend Full<br>Face Vented Mask | Predicate - Mojo, Veraseal Full Face Mask,<br>K120463 | Explanation of Differences |
| Patient type | Patients who are appropriate candidates for positive<br>pressure therapy | Patients who are appropriate candidates for positive<br>pressure therapy | Identical |
| Prescriptive | Yes | Yes | Identical |
| Environment of Use | The masks are intended for use in the home or<br>hospital/institutional environment. | The masks are intended for use in the home or<br>hospital/institutional environment. | Identical |
| Duration of Use | Mojo -<br>Single patient, multi-use<br>Veraseal 2 -<br>Short Term Single Patient use (maximum of 7 days)<br>Innova -<br>Single patient, multi-use<br>Ascend -<br>Single patient, multi-use | Mojo -<br>Single patient, multi-use and Multi-patient, multi-use<br>Veraseal -<br>Disposable single use; Single patient, multi-use up to<br>7 days | Similar.<br>Multi-patient use for the Mojo<br>mask is being removed.<br>Single use for the Veraseal<br>mask is being removed. |
| Principle of Operation | Provides a seal over the face (nose and mouth) to<br>allow for delivery of pressurized air from a positive<br>pressure device. Mask includes an exhalation port<br>and AAV for flushing out exhaled CO2. | Provides a seal over the face (nose and mouth) to<br>allow for delivery of pressurized air from a positive<br>pressure device. Mask includes an exhalation port and<br>AAV for flushing out exhaled CO2. | Identical |
| Therapy Pressure | Greater than 3 cm H2O<br>Typically determined by the equipment to which it is<br>attached. | Greater than 3 cm H2O<br>Typically determined by the equipment to which it is<br>attached. | Identical |
| Anatomical site | Face (seals around nose and mouth) | Face (seals around nose and mouth) | Identical |
| User Interface to<br>administer therapy | Masks have a standard 22mm connector (compliant<br>with ISO 5356-1) that connects to 22mm CPAP<br>circuits. | Mask has a standard 22mm connector (compliant with<br>ISO 5356-1) that connects to 22mm CPAP circuits. | Identical |
| Contraindications | Mojo has an contraindication that is identical to the<br>predicate K120463. Veraseal 2, Innova and Ascend<br>do not have contraindications. | Mojo mask has a contraindication. | Similar. The first<br>contraindication in the Mojo<br>mask is identical to the predicate.<br>New contraindication related to<br>presence of magnets added as<br>per recall event #94169. |
| Attributes | Subject - Mojo, Veraseal 2, Innova, Ascend Full<br>Face Vented Mask | Predicate - Mojo, Veraseal Full Face Mask,<br>K120463 | Explanation of Differences |
| | New Contraindication (Mojo only)<br>Do not use this mask if you or anyone (example:<br>household members, bed partners, caregivers, etc.) in<br>close physical contact with your mask has an active<br>medical implant or metallic implant that will interact<br>with magnets. Implant examples include, but are not<br>limited to, pacemakers, implantable cardioverter<br>defibrillators (ICD), neurostimulators, aneurysm clips,<br>metallic stents, ocular implants, insulin/infusion<br>pumps, cerebral spinal fluid (CSF) shunts, embolic<br>coils, metallic splinter, implants to restore hearing or<br>balance with implanted magnets (such as cochlear<br>implants), flow disruption devices, contact lenses with<br>metal, dental implants, metallic cranial plates, screws,<br>burr hole covers, bone substitute device, magnetic<br>metallic implants/electrodes/valves placed in upper<br>limbs, torso, or higher, etc. If you have any questions<br>regarding the implant, consult your physician or the<br>manufacturer of your implant. | | |
| Warnings | Updated Warning<br>Magnets are used in the mask and headgear clips with<br>a field strength of 380mT. With the exception of the<br>devices identified in the contraindication, ensure that<br>the mask is kept at least 6 inches (approx.16 cm)<br>away from any other medical implants or medical<br>devices that can be impacted by the magnetic fields to<br>avoid possible effects from localized magnetic fields.<br>For instance, the functioning of implants may change,<br>or implants may move within the body due to<br>magnetic attraction/repulsion. This applies to you or<br>anyone in close physical contact with your mask | Unchanged from K120463 | Updated warning for magnets to<br>labeling |
| Attributes | Subject - Mojo, Veraseal 2, Innova, Ascend Full<br>Face Vented Mask | Predicate - Mojo, Veraseal Full Face Mask,<br>K120463 | Explanation of Differences |
| Useful life | Mojo -<br>Single patient use - 6 months<br>Veraseal 2 -<br>Short Term Single Patient use (maximum of 7 days)<br>Innova -<br>Single patient use - 6 months<br>Ascend -<br>Single patient use - 6 months | Mojo -<br>Single patient, multi-use and Multi-patient, multi-use<br>– 6 months<br>Veraseal -<br>Disposable single use; Single patient, multi-use up to<br>7 days | Similar.<br>Multi-patient use for Mojo<br>mask has been removed.<br>Single use for Veraseal mask<br>has been removed. |
| Non-sterile | Yes | Yes | Identical |
| Cleaning methods | Mojo -<br>Mild Soap and water<br>Veraseal 2 -<br>Mild Soap and water<br>Isopropyl alcohol<br>Innova -<br>Mild Soap and water<br>Ascend -<br>Mild Soap and water | Mojo -<br>Mild Soap and water<br>Veraseal -<br>Mild Soap and water<br>Isopropyl alcohol | Identical |
| Features | | | |
| Available sizes | Mojo - 4<br>Veraseal 2 - 4<br>Innova - 3<br>Ascend - 3 | Mojo - 3<br>Veraseal - 3 | Similar.<br>XL size has been added to the<br>Mojo and Veraseal 2 masks.<br>These sizes utilize all the same<br>components as the<br>corresponding large size,<br>except for the gel cushion<br>which is slightly larger in<br>length and width. |
| Shape | Similar | Similar | Similar |
| Incorporates an Exhaust<br>port | Yes | Yes | Identical |
| Attributes | Subject – Mojo, Veraseal 2, Innova, Ascend Full Face Vented Mask | Predicate – Mojo, Veraseal Full Face Mask, K120463 | Explanation of Differences |
| Components of the mask | • Mask shell<br>• Gel bladder (cushion)<br>• Vented elbow assembly<br>• Forehead arm/forehead pad<br>• Headgear | • Mask shell<br>• Gel bladder (cushion)<br>• Vented elbow assembly<br>• Forehead arm/forehead pad<br>• Headgear | Similar |
| Shell design | Mojo – soft shell<br>Veraseal 2 – hard shell<br>Innova – hard shell<br>Ascend – hard shell | Mojo – soft shell<br>Veraseal 2 – hard shell | Identical |
| Patient Contact per<br>ISO 10993-1 | Skin contact<br>Externally Communicating with tissue (air pathway)<br>Permanent contact | Skin contact<br>Externally Communicating with tissue (air pathway)<br>Permanent contact | Identical |
| Performance Characteristics | | | |
| Deadspace | Mojo<br>Small 160 ml<br>Medium 185 ml<br>Large 205 ml<br>XL 210 ml<br>Veraseal 2<br>Small 155.3 ml<br>Medium 181.7 ml<br>Large 196.6 ml<br>XL 200.6 ml<br>Innova<br>Small 155.3 ml<br>Medium 181.7 ml<br>Large 196.6 ml<br>Ascend<br>Small 165 ml<br>Medium 212 ml<br>Large 221 ml | Mojo<br>Small 160 ml<br>Medium 185 ml<br>Large 205 ml<br>Veraseal<br>Small 138 ml<br>Medium 165 ml<br>Large 181 ml | Similar.<br>XL size has been added to the<br>Mojo and Veraseal 2 masks.<br>The deadspace for all subject<br>devices is within 5% of the<br>predicate Mojo mask. The CO2<br>washout profile of all the<br>subject devices has been<br>verified to be adequate. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
## 510(k) Summary Page 6 of 12
18-Oct-24
{10}------------------------------------------------
## 510(k) Summary Page 7 of 12
18-Oct-24
{11}------------------------------------------------
## 510(k) Summary Page 8 of 12
18-Oct-24
| Attributes | Subject – Mojo, Veraseal 2, Innova, Ascend Full<br>Face Vented Mask | | Predicate – Mojo, Veraseal Full Face Mask,<br>K120463 | Explanation of Differences | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Exhaust flow<br>characteristics | Mojo<br>Pressure (cmH2O)<br>3<br>10<br>20<br>Veraseal 2<br>Pressure (cmH2O)<br>3<br>10<br>20<br>Innova<br>Pressure (cmH2O)<br>3<br>10<br>20<br>Ascend<br>Pressure (cmH2O)<br>3<br>10<br>20 | Flow (lpm)<br>17.88<br>32.57<br>48.90<br>Flow (lpm)<br>22.08<br>38.50<br>54.30<br>Flow (lpm)<br>22.08<br>38.50<br>54.30<br>Flow (lpm)<br>26.80<br>41.50<br>58.00 | Mojo<br>Pressure (cmH2O)<br>3<br>10<br>20<br>Veraseal<br>Pressure (cmH2O)<br>3<br>10<br>20 | Flow (lpm)<br>21.20<br>37.20<br>53.10<br>Flow (lpm)<br>21.10<br>36.40<br>53.00 | Similar.<br>The exhaust flow<br>characteristics of the subject<br>devices are similar to the<br>predicate. |
| AAV Opening/Closing<br>Pressure | Mojo<br>Opening - 1.10 cm H2O<br>Closing - 1.60 cm H2O<br>Veraseal<br>Opening - 2.1 cm H2O<br>Closing - 2.28 cm H2O<br>Innova<br>Opening - 2.1 cm H2O<br>Closing - 2.28 cm H2O<br>Ascend<br>Opening - 1.92 cm H2O<br>Closing - 1.85 cm H2O | | Mojo<br>Opening - 1.10 cm H2O<br>Closing - 1.60 cm H2O<br>Veraseal<br>Opening - 1.29 cm H2O<br>Closing - 1.45 cm H2O | | Similar.<br>Pass/Fail criteria as per ISO<br>17510 is < 3 cm H2O. All<br>subject devices meet the<br>criteria. |
{12}------------------------------------------------
| Resistance to Flow<br>(Pressure drop) | Mojo<br>50 lpm - 0.40 cm H2O<br>100 lpm - 0.65 cm H2O<br>Veraseal 2<br>50 lpm - 0.07 cm H2O<br>100 lpm - 0.39 cm H2O<br>Innova<br>50 lpm - 0.07 cm H2O<br>100 lpm - 0.39 cm H2O<br>Ascend<br>50 lpm - 0.11 cm H2O<br>100 lpm - 0.59 cm H2O | Mojo<br>50 lpm - 0.40 cm H2O<br>100 lpm - 0.65 cm H2O<br>Veraseal<br>50 lpm - 0.49 cm H2O<br>100 lpm - 0.89 cm H2O | Similar.<br>There are no acceptance<br>criteria in ISO 17510 for<br>resistance to flow (pressure<br>drop). The resistance values<br>are disclosed in the IFU's. |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Magnets | Mojo only<br>4 Magnets are used in the mask (2) and headgear<br>clips (2) with a field strength of 380mT. These are<br>identical to K190254. | No magnets | Reference K190254 |
| CO2 washout profile<br>ISO 17510:2015…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.