RESCAN

K140054 · Resmed, Ltd. · BZD · May 16, 2014 · Anesthesiology

Device Facts

Record IDK140054
Device NameRESCAN
ApplicantResmed, Ltd.
Product CodeBZD · Anesthesiology
Decision DateMay 16, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and therapeutic information. This includes the ability to remotely change settings in non-life support devices only. It is intended to be used by Clinicians in conjunction with ResMed compatible therapy devices, using ResMed's proprietary communications protocol.

Device Story

ResScan is a software application used by clinicians in hospital settings to manage patient therapy data from ResMed flow generators. It functions as a data interface, allowing clinicians to download, view, and store patient/machine data (e.g., mask leak, AHI, pressure, minute ventilation, respiratory rate) and generate reports. The device facilitates remote settings management for non-life support devices via removable media or direct PC connection using ResMed's proprietary communication protocol. By enabling clinicians to monitor therapy compliance and adjust device parameters, ResScan supports clinical decision-making and optimizes patient therapy management. It does not perform life-support functions but supports the reporting of multiple therapy programs and modes for various ResMed devices.

Clinical Evidence

Bench testing only. No clinical data was required as the device functions to obtain information from therapeutic devices for which clinical trials have already been conducted or compared with previous predicate results. Verification testing confirmed successful data transfer between flow generators and the ResScan software.

Technological Characteristics

Software-based data management system. Interfaces with ResMed flow generators via serial connection or removable media. Supports Windows Remote Desktop environments. Connectivity via proprietary ResMed communication protocol. No specific hardware materials; software-based functionality.

Indications for Use

Indicated for patients requiring follow-up care for therapy managed by ResMed compatible flow generators. Used by clinicians to transfer machine/therapeutic data and remotely adjust settings on non-life support devices.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### KI40.054 # RESMED # MAY 1 6 2014 ResScan Traditional 510(k) #### 510(k) SUMMARY [As required by 21 CFR 807.92(c)] | Date Prepared | 15 April 2014 | |---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Name | Nicole Gaddi | | Official Contact | Mr. Jim Cassi<br>Vice President - Quality Assurance Americas<br>9001 Spectrum Center Blvd<br>San Diego CA 92123 USA<br>Tel: (858) 836 6081 | | Device Trade Name | ResScan | | Device Common Name | Continuous Ventilator | | Classification/s &<br>Classification Name/s | 21 CFR 868.5905, 73 BZD (Class II)<br>Non continuous ventilator (IPPB)<br>21 CFR 868.5895, 73 MNS (Class II)<br>Continuous Ventilator<br>21 CFR 868.5895, 73 MNT (Class II)<br>Continuous Ventilator<br>21 CFR 868.5895, 73 CBK (Class II)<br>Ventilator, continuous, facility use | | Legally Marketed<br>Predicate Device<br>Reason for Submission | ResScan (K113815)<br>New Device | #### Indication for Use ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and therapeutic information. This includes the ability to remotely change settings in non-life support devices only. It is intended to be used by Clinicians in conjunction with ResMed compatible therapy devices, using ResMed's proprietary communications protocol. #### Device Description The performance and functional characteristics of ResScan includes all the user friendly features of the predicate device. ResScan allows the clinician to: - Download and view patient and machine data from ResMed flow generators A - Store patient details A - Set machine parameters (Using Removal Media or PC direct connection), non-life support A devices only. - メ Create and print reports - > . Uses Removal Media or PC direct connection as the interface between the flow generator and ResScan - 大 Support for Data Card Reader {1}------------------------------------------------ ### Substantial Equivalence The new device has the following similarities to the previously cleared predicate device. - 入 Similar intended use - > Same operating principle - Same technologies ★ - > Same manufacturing process Design and Verification activities were performed on ResScan as a result of the risk analysis and design requirements. Verification testing included end-to-end testing to confirm that settings were successfully transferred between the flow generator and ResScan, and data captured by the flow generator was sent to ResScan. All tests confirmed the product met the predetermined acceptance criteria. ResMed has determined that the new device is substantially equivalent to the predicate device and any changes do not raise new questions of safety and effectiveness when used for patient compliance management as an adjunct with ResMed flow generators. The new device complies with the applicable requirements referenced in the FDA guidance documents: - FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical > Devices (May 11, 2005) - > FDA Off-the-Shelf Software Use in Medical Devices (September 9, 1999) | Feature | ResScan predicate(K113815) | ResScan (modified) | Comments | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | Intended Use | ResScan™ is intended to<br>augment the standard follow-up care of patients by<br>providing transfer of machine and therapeutic information,<br>including the ability to remotely change settings.<br><br>It is intended to be used by<br>Clinicians in conjunction with<br>ResMed compatible flow<br>generators, using ResMed's<br>proprietary communications protocol. | ResScan™ is intended to<br>augment the standard<br>follow-up care of patients by<br>providing transfer of<br>machine and therapeutic<br>information. This includes<br>the ability to remotely<br>change settings in non-life<br>support devices only.<br><br>It is intended to be used by<br>Clinicians in conjunction<br>with ResMed compatible<br>therapy devices, using<br>ResMed's proprietary<br>communications protocol. | Equivalent:<br>Remote setting change<br>functionality is not<br>applicable to ResMed<br>CBK therapy devices. | | Location of use | • Hospital | • Hospital | Equivalent | | Functionality | • Display of therapy data<br>• Generate reports<br>• Settings management<br>• Patient management | • Display of therapy data<br>• Generate reports<br>• Settings management<br>• Patient management | Equivalent:<br>Remote setting change<br>functionality only<br>applies to non-life<br>support devices | | Compatible<br>flow generators | • Resmed compatible<br>therapy devices (73 BZD,<br>73 MNS and 73 MNT ) | • Resmed compatible<br>therapy devices (73 BZD,<br>73 MNS, 73 MNT and 73<br>CBK ) | Equivalent:<br>Remote setting change<br>functionality only<br>applies to non-life<br>support devices | {2}------------------------------------------------ | Feature | ResScan predicate(K113815) | ResScan (modified) | Comments | |-------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------| | Communication<br>medium | • Serial connection<br>• Removable medium | • Serial connection<br>• Removable medium | Equivalent:<br>Serial connection only<br>applies to non-life<br>support devices | | Patient<br>information | • Mask Leak<br>• AHI<br>• Pressure<br>• Minute Ventilation<br>• Respiratory rate | • Mask Leak<br>• AHI<br>• Pressure<br>• Minute Ventilation<br>• Respiratory rate | Equivalent | | Changeable<br>settings | • Start Pressure<br>• Set Pressure<br>• Ramp time<br>• Setting time | • Start Pressure<br>• Set Pressure<br>• Ramp time<br>• Setting time | Equivalent:<br>Remote setting change<br>functionality only<br>applies to non-life<br>support devices | ### Non-Clinical Testing Design and non-clinical verification activities were performed on ResScan as a result of the updated risk analysis and design requirements. Verification testing included End-to-End bench testing to verify that ResScan, using only Removable Media, can receive patient usage, device settings, therapy summary data, detailed signal data and device log information from the flow generator (data). All tests confirmed the product met the predetermined acceptance criteria. ResMed has determined that the new version of ResScan is substantially equivalent to the predicate device, ResScan (K113815). ResScan met the predetermined pass/fail criteria as defined in the ResScan System Verification Report. #### Clinical Testing Clinical testing was not deemed necessary as identified in the Risk Analysis, as ResScan only obtains patient and machine information from therapeutic devices for which clinical trials have already been conducted, or compared with previous predicate comparison test results. Accordingly no clinical testing is required. #### Summary of additional features from the ResScan (K113815) - > ResMed compatible therapy devices include, e.g. VPAP Bilevel devices (73 MNS). - Life support therapy device (73 CBK). A - Display/reporting of multiple therapy programs for life support therapy devices. メ - Display/reporting of additional modes such as ASVAuto, ACV, PACV, V-SIMV, P-SIMV, & PS. A - A Support for Windows Remote Desktop Environments. The inclusion of these features has been assessed within the risk analysis and no additional safety risks have been found as a result of the inclusion of these features. #### Conclusion The modified ResScan is as safe and as effective as the predicate device, ResScan (K113815) and is deemed substantially equivalent to the predicate device, ResScan (K113815). {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and three horizontal lines below the eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle in a circular fashion. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 16, 2014 Resmed Corporation Mr. Jim Cassi Vice President, Quality Assurance Americas 9001 Spectrum Center Blvd. San Diego, CA 92123 Re: K140054 Trade/Device Name: ResScan Regulation Number: 21 CFR 868.5905 Regulation Name: Non-continuous Ventilator Regulatory Class: II Product Code: BZD, MNS, MNT, CBK Dated: April 16, 2014 Received: April 18, 2014 #### Dear Mr. Cassi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/McdicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin I. Keith, M.S. Acting Director · Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. 510(k) Number (if known) K140054 Device Name ResScan Indications for Use (Describe) ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and thermation. This includes the ability to remotely change settings in non-life support devices only. It is intended to be used by Clinicians in conjunction with ResMed compatible therapy devices, using ResMed's proprietary communications protocol. Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## Please do not write below this line - continue on a separate page if needed. ール 対応 パパーズ (1) 2017 - 1 (FOR FOR FOR FOR FOR FOR (1) (4 ) 2 ) 2 (1) 2 ) 2 (1) 2 ) 2 (1) 2 ) 2 (1) 2 ) 2 (1 ) 2 ) 2 (1 ) 2 ) 2 (1 ) 2 ) 2 (1 ) 2 ) 2 (1 ) 2 ) 2 (1 ) 2 ) 2 (1 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/5/Picture/15 description: The image shows a logo on the left and some text on the right. The text on the right says "Anya C." followed by "2014.05" and "15:10:04". The logo on the left is a stylized design, but it is difficult to make out the details. Anya C. Harry -S 2014.05.15 15:10:04 -04'00'
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...