K211155 · BMC Medical Co., Ltd. · BZD · Mar 9, 2022 · Anesthesiology
Device Facts
Record ID
K211155
Device Name
Auto CPAP System
Applicant
BMC Medical Co., Ltd.
Product Code
BZD · Anesthesiology
Decision Date
Mar 9, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Auto CPAP System is a CPAP (Continuous Positive Airway Pressure) device designed for the treatment of adult Obstructive Sleep Apnea (OSA) only, either in the hospital or at home. Auto CPAP System is for prescription use only. It is a travel CPAP device intended for single-patient use. LightTrip App is a mobile application for patients to remotely operate BMC Mini serial devices, and transmit, store and display usage and treatment information. It can also receive parameters and device firmware upgrade data from the cloud, and then transmit these data to the device. The LightTrip app also allows healthcare professionals to remotely configure compatible OSA therapy devices. LightTrip App is for prescription use only.
Device Story
Microprocessor-controlled, blower-based system; generates continuous positive airway pressure (4-20 cmH2O) to treat adult OSA. Inputs: patient respiratory signals; outputs: pressurized air to oral/nasal airway. Used in home or hospital; operated by patient or healthcare professional. Includes Bluetooth module for connectivity with LightTrip mobile app. App enables remote device operation, parameter configuration by clinicians, and data transmission/storage/cloud synchronization. Benefits: portable travel-sized CPAP therapy; remote monitoring and configuration capabilities.
Clinical Evidence
Bench testing only. Performance verified per ISO 80601-2-70:2015 (flow rate, static/dynamic pressure, sound level, display accuracy). Biocompatibility evaluated per ISO 10993 and ISO 18562. Electrical safety and EMC tested per AAMI/ANSI ES 60601-1 and IEC 60601-1-2/11. Mechanical testing (shock/vibration) performed.
Technological Characteristics
Blower-based CPAP; 4-20 cmH2O pressure range; 0.5 cmH2O increments. Bluetooth connectivity for mobile app. Non-sterile. Class II, Type BF. IP22 ingress protection. Software-controlled respiratory event detection (OSA, hypopnea, snore).
Indications for Use
Indicated for treatment of adult Obstructive Sleep Apnea (OSA) in hospital or home settings. Prescription use only.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
K961626 — SOLO CPAP SYSTEM · Respironics, Inc. · Jul 18, 1996
K200480 — DreamStation 2 System, DreamStation 2 Advanced System · Respironics, Inc. · Jul 10, 2020
K032480 — AUTOSET SPIRIT SYSTEM · Resmed, Ltd. · Oct 16, 2003
K201620 — Luna® G3 BPAP 25A · 3B Medical, Inc. · Jan 29, 2021
K180388 — Transcend 365 miniCPAP System · Somnetics International, Inc. · Nov 30, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 9, 2022
BMC Medical Co., Ltd. Amy Wang RA Room 901, Building 1, No. 28 Pingguoyuan Road Beijing, Shijingshan 100041 China
Re: K211155
Trade/Device Name: Auto CPAP System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: February 7, 2022 Received: February 7, 2022
Dear Amy Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K21155
Device Name Auto CPAP System
Indications for Use (Describe)
Auto CPAP System is a CPAP (Continuous Positive Airway Pressure) device designed for the treatment of adult Obstructive Sleep Apnea (OSA) only, either in the hospital or at home.
Auto CPAP System is for prescription use only. It is a travel CPAP device intended for single-patient use.
LightTrip App is a mobile application for patients to remotely operate BMC Mini serial devices, and transmit, store and display usage and treatment information. It can also receive parameters and device firmware upgrade data from the cloud, and then transmit these data to the device. The LightTrip app also allows healthcare professionals to remotely configure compatible OSA therapy devices. LightTrip App is for prescription use only.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">□</span> Over The Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name | Auto CPAP System |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Model | M1 Mini |
| Common/Usual Name | Auto CPAP System |
| Date Prepared | March 8, 2022 |
| Sponsor Identification | BMC Medical Co., Ltd.<br>Room 901, Building 1, No.28 Pingguoyuan Road,<br>Shijingshan, Beijing 100041, CHINA |
| Submission Correspondent | Amy Wang<br>BMC Medical Co., Ltd.<br>Room 901, Building 1, No.28 Pingguoyuan Road,<br>Shijingshan, Beijing 100041, CHINA |
| Phone | 86-18701556244 |
| Fax | 86-22-82939881 |
| Email | wangliping@bmc-medical.com |
| Establishment Registration # | 3008566132 |
| Classification | BMC Medical Co., Ltd.<br>Room 901, Building 1, No.28 Pingguoyuan Road,<br>Shijingshan, Beijing 100041, CHINA |
| Classification | Class II Device (21 CFR 868.5905) |
| Classification Name | Noncontinuous ventilator |
| Classification Panel | Medical Device |
| Products Code | BZD |
| Medical Specialties | Anesthesiology |
| Predicate Device(s) | Luna CPAP and Auto CPAP System (K153387) |
| | Menai System (K160836) |
{4}------------------------------------------------
| Reason for Submission: | New Device |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Auto CPAP System is a CPAP (Continuous<br>Positive Airway Pressure) device designed for the treatment<br>of adult Obstructive Sleep Apnea (OSA) only, either in the<br>hospital or at home.<br>Auto CPAP System is for prescription use only. It is a travel<br>CPAP device intended for single-patient use.<br><br>LightTrip App is a mobile application for patients to<br>remotely operate BMC Mini serial devices, and transmit,<br>store and display usage and treatment information. It can also<br>receive parameters and device firmware upgrade data from<br>the cloud, and then transmit these data to the device. The<br>LightTrip app also allows healthcare professionals to<br>remotely configure compatible OSA therapy devices.<br>LightTrip App is for prescription use only. |
| Device Description | Auto CPAP System is a microprocessor controlled,<br>blower-based system that generates continuous positive<br>airway pressure (CPAP) to support treatment of obstructive<br>sleep apnea. The system provides fixed or auto-adjust<br>pressure from 4 to 20 cmH2O above the ambient atmospheric<br>pressure to a patient's oral/nasal airway.<br>The Auto CPAP System is designed and developed on the<br>basis of the predicate device Luna CPAP and Auto CPAP<br>System (K153387). The parameters and algorithms of the<br>two devices, such as the treatment mode (CPAP and<br>AutoCPAP), the pressure range (4 to 20 cm H2O), and the<br>Respiratory Event Detection, are consistent. Compared with<br>the predicate device, the subject device does not include a<br>humidifier and an LCD screen. The parameters of subject<br>device are displayed and set via the LightTrip App. |
#### Substantial Equivalence:
The subject and predicate devices have the following similarities:
- Same Indications for Use ●
- Similar operating principle ●
- Similar technologies
Auto CPAP System retains similar functionality and performance features of the primary predicate device Luna CPAP and Auto CPAP System (K153387). It utilizes the same algorithm as the primary predicate device Luna CPAP and Auto CPAP System (K153387) for respiratory event detection and therapy for sleep disordered breathing events. The subject device and the primary
{5}------------------------------------------------
predicate device share the similar intended use, similar operating principal, and are manufactured and packaged with similar processes.
Compared with the primary predicate device, the subject device makes the following main modifications:
- By removing the humidifier, the subject device is provided in a portable smaller footprint than the predicate for user convenience, especially for using on an aircraft or in travel. Choosing a mask with a humidifier component can improve the moisture level in the breathing air and offer relief from possible dryness.
- The subject device includes a Bluetooth module, which enables the device to be connected with the LightTrip App on the smartphone.
- LightTrip App is a mobile application of Auto CPAP System. The patient and the health care professional can set parameters of the device via this application. Patients can set the comfort settings, turn the device on and off using the LightTrip App. However, only health care professional is able to remotely change the prescription settings. The device can upload the running data to LightTrip App where they can be displayed, stored and uploaded to the cloud platform. LightTrip App can also obtain the remote parameter setting of the cloud platform and realize the on-line firmware upgrade of the device.
The substantial equivalence comparison is provided below.
{6}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Device | |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Auto CPAP System<br>(K211155) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Menai System<br>(K160836) | Comparison |
| Classification | | | | |
| Device<br>Classification | Class II Device | Class II Device | Class II Device | Identical to primary<br>predicate |
| Product Code | BZD | BZD | BZD, MNR | Identical to primary<br>predicate |
| Classification<br>Panel | Anesthesiology | Anesthesiology | Anesthesiology | Identical to primary<br>predicate |
| Regulation<br>Number | 21 CFR 868.5905 | 21 CFR 868.5905 | 21 CFR 868.5905<br>21 CFR 868.2375 | Identical to primary<br>predicate |
| Regulation<br>Name | Noncontinuous ventilator | Noncontinuous ventilator | Noncontinuous ventilator | Identical to primary<br>predicate |
| Sterile | Non-sterile | Non-sterile | Non-sterile | Identical to primary<br>predicate |
| Intended Use and Indications for Use | | | | |
| | Subject Device | Primary Predicate Device | Reference Device | |
| | Auto CPAP System<br>(K211155) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Menai System<br>(K160836) | Comparison |
| Intended Use of<br>the Device | Auto CPAP System is a<br>CPAP<br>(Continuous<br>Positive Airway Pressure)<br>device designed for the<br>treatment<br>of<br>adult<br>Obstructive Sleep Apnea<br>(OSA) only, either in the<br>hospital or at home.<br>Auto CPAP System is for<br>prescription use only. It is a<br>travel<br>CPAP<br>device<br>intended for single-patient<br>use. | The Luna CPAP and Auto<br>CPAP System are intended to<br>deliver positive pressure for the<br>treatment of Obstructive Sleep<br>Apnea. The optional integrated<br>heated humidifier is indicated<br>for<br>the<br>humidification and warming of<br>air from the flow generator.<br>These devices are intended for<br>single<br>patient<br>use<br>by<br>prescription in the home or<br>hospital / institutional<br>environment on adult patients. | The Menai self-adjusting<br>system is indicated for the<br>treatment of Obstructive<br>Sleep Apnea (OSA) in<br>patients (female patients with<br>mild to moderate OSA when<br>using AfH treatment mode)<br>weighing more than 66 lb (30<br>kg). It is intended for home<br>and hospital use. | Similar to primary<br>predicate. |
| Intended Use of<br>the Software<br>Application | LightTrip App is a mobile<br>application for patients to<br>remotely operate BMC<br>Mini serial devices, and<br>transmit, store and display<br>usage and treatment<br>information. It can also<br>receive parameters and<br>device firmware upgrade<br>data from the cloud, and<br>then transmit these data to<br>the device. The LightTrip<br>app also allows healthcare<br>professionals to remotely<br>configure compatible OSA<br>therapy devices.<br>LightTrip App is for<br>prescription use only. | N/A | Monte Carlo is a mobile<br>application for patients to<br>remotely operate a prescribed<br>compatible ResMed machine<br>and transfer, analyze and<br>display usage and therapeutic<br>information. Monte Carlo also<br>allows healthcare<br>professionals to remotely<br>configure compatible OSA<br>therapy devices. | Similar to the<br>secondary predicate. |
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | |
| | Auto CPAP System<br>(K211155) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Menai System<br>(K160836) | Comparison |
| Pressure Delivery | | | | |
| Therapy<br>Delivered | CPAP,<br>AutoCPAP | CPAP,<br>AutoCPAP | CPAP,<br>AutoSet,<br>AutoSet for Her (AfH) | Identical to primary<br>predicate |
| Pressure Range | 4 to 20 cm H2O (in 0.5 cm<br>H2O increments), ≤30 cm<br>H2O under single fault<br>conditions | 4 to 20 cm H2O (in 0.5 cm<br>H2O increments), ≤ 30 cm<br>H2O under single fault<br>conditions | • 4-20 cm H2O (CPAP Mode)<br>• 4-20 cm H2O (AutoSet<br>Mode)<br>• 4-20 cm H2O (AutoSet AfH<br>Mode)<br>• EPR +3 cm H2O (all modes) | Identical to primary<br>predicate |
| Pressure<br>Display<br>Accuracy (hPa) | ±(0.5 hPa+4%) | ±(0.5 hPa+4%) | ±0.5 hPa | Identical to primary<br>predicate |
| Algorithm | | | | |
| Automatic<br>adjusting<br>CPAP<br>algorithm | Yes | Yes | Yes | Identical to primary<br>predicate |
| Ramp | Yes, the ramp time ranges<br>from 0 to 60 minutes | Yes, the ramp time ranges<br>from 0 to 60 minutes | Yes, user selected as "Off" to<br>45 minutes in 5 minute<br>increments | Identical to primary<br>predicate |
| Expiratory<br>Pressure Relief | Yes,<br>Reslex® function<br>Level 1-3 | Yes,<br>Reslex® function<br>Level 1-3 | Yes,<br>EPR function | Identical to primary<br>predicate |
| Respiratory<br>Event<br>Detection | OSA detection<br>Hypopnea detection<br>Snore detection | OSA detection<br>Hypopnea detection<br>Snore detection | OSA detection<br>Hypopnea detection<br>Snore detection | Identical to primary<br>predicate |
| Physical Characteristics | | | | |
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Comparison |
| | Auto CPAP System<br>(K211155) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Menai System<br>(K160836) | |
| IEC 60601<br>Classification | Class II, Type BF | Class II, Type BF | Class II, Type BF | Identical to primary<br>predicate |
| Degree of<br>Protection<br>Against Water<br>Ingress | IP22 | IP22 | IP22 | Identical to primary<br>predicate |
| Sound Pressure<br>Level | < 30 dB, when the device is<br>working at the pressure of 10<br>hPa. | < 30 dB, when the device is<br>working at the pressure of 10<br>hPa. | Air tubing connected to<br>AirMini F20 connector with<br>mask: 19 dBA with uncertainty<br>of 3 dBA;<br>Air tubing connected to<br>AirMini N20 or P10 connectors<br>with mask: 15 dBA<br>with uncertainty of 3 dBA | Identical to primary<br>predicate |
| Air Filter | Yes | Yes | Yes | Identical to primary<br>predicate |
| Non-heated<br>Tubing | Yes | Yes | Yes | Identical to primary<br>predicate |
| Humidifier | No | Yes | No | Identical<br>to<br>secondary predicate<br>device. Choosing the<br>appropriate<br>mask<br>with<br>humidification<br>component<br>can<br>improve the moisture<br>level in the breathing<br>air and offer relief<br>from<br>possible<br>dryness |
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | |
| | Auto CPAP System<br>(K211155) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Menai System<br>(K160836) | Comparison |
| Data Reporting | | | | |
| Wireless<br>Function | Bluetooth | N/A | Bluetooth | Identical to<br>secondary predicate<br>device |
| Parameters<br>Transferred by<br>the Mobile App | Usage and treatment data | N/A | Usage and therapeutic data | Identical to<br>secondary predicate<br>device |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
## Performance Data:
The following performance data were used in support of the substantial equivalence determination.
## Bench Testing:
To demonstrate that the performance of the substantially equivalent to the predicate device, the following bench testing were conducted according to ISO 80601-2-70:2015:
- Maximum flow rate testing
- Static pressure testing .
- Dynamic pressure testing ●
- Sound pressure level testing .
- Display accuracy testing .
### Biocompatibility Assessment:
Evaluation and testing of biocompatibility were conducted in accordance with the following standards and guidance documents:
- · ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
{11}------------------------------------------------
• FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" June 16, 2016
• ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
- · ISO 18562-1:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications Part 2: Tests for emissions of particulate matter
• ISO 18562-3:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic compounds (VOCs)
• ISO 18562-4:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 4: Tests for leachables in condensate
#### Electrical Safety and Electromagnetic Compatibility:
Electrical safety and EMC testing were conducted on the subject device Auto CPAP System. The following standards for electrical safety and EMC:
• AAMI/ANSI ES 60601-1:2005/(R) 2012 And A1:2012. Medical Electrical Equipment - Part 1: General Requirements for Basic Safety And Essential Performance
· IEC 60601-1-2:2014 Medical electrical equipments for basic safety and essential performance -
- Collateral standard: Electromagnetic disturbances Requirements and tests
• IEC 60601-1-…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.