ULTRASONIC STRAIGHT/ULTRASONIC KNIFE/ULTRASONIC RING CURRETTE
K020220 · Synergetics, Inc. · LFL · Aug 23, 2002 · SU
Device Facts
Record ID
K020220
Device Name
ULTRASONIC STRAIGHT/ULTRASONIC KNIFE/ULTRASONIC RING CURRETTE
Applicant
Synergetics, Inc.
Product Code
LFL · SU
Decision Date
Aug 23, 2002
Decision
SESE
Submission Type
Traditional
Device Class
Class U
Attributes
Therapeutic
Intended Use
Synergetics Sonotome™ Ultrasonic Aspirator Tips are intended for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g.: bone) tissue is desirable, including: Neurosurgery Gastrointestinal and Affiliated Organ Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecological Surgery Thoracic Surgery Laparoscopic Surgery Thoracoscopic Surgery
Device Story
Synergetics Sonotome™ Ultrasonic Aspirator Tips are surgical accessories designed to attach to third-party ultrasonic surgical aspirator handpieces. The tips function as passive transmission elements, receiving ultrasonic energy from the host handpiece to vibrate at the distal end. This vibration, combined with integrated irrigation and aspiration, enables the fragmentation, emulsification, and removal of soft and hard tissues (e.g., bone) while sparing delicate structures like vessels and ducts. Available in various diameters, shapes, and lengths, including curved configurations for improved visualization. Features include pre-aspiration holes to prevent clogging. Used by surgeons in clinical settings (OR) for diverse surgical specialties. The device is supplied sterile for single-use, allowing for tip replacement without requiring handpiece resterilization. Benefits include selective tissue dissection and improved surgical site access.
Clinical Evidence
Bench testing only. Validation and verification were performed through risk analysis and mechanical performance testing on prototype units. Biocompatibility was assured through the use of 6AL-4V titanium (AMS4928), a material with established safety in similar commercially available systems.
Technological Characteristics
Material: 6AL-4V Titanium (AMS4928). Construction: Machined from one solid piece. Sensing/Actuation: Passive ultrasonic energy transmission from host handpiece. Form factor: Various diameters, shapes, and lengths; includes curved options. Features: Pre-aspiration holes. Sterilization: Terminal sterilization (Tyvek peel pouch). Connectivity: Mechanical attachment to third-party ultrasonic surgical aspirator handpieces.
Indications for Use
Indicated for patients undergoing surgical procedures requiring fragmentation, emulsification, and aspiration of soft and hard (e.g., bone) tissue across various specialties including neurosurgery, GI, urology, plastic/reconstructive, general, orthopedic, gynecological, thoracic, laparoscopic, and thoracoscopic surgery.
Predicate Devices
CUSA Excel Ultrasonic Surgical Aspirator
Related Devices
K111741 — INTEGRA(TM) CUSA NXT(TM) BONE TIPS · Integra Burlington Ma, Inc. (Formerly Integra Radi · Aug 23, 2011
K113570 — INTEGRA CUSA NXT INFERIOR FOWARD BONE TIP · Integra Burlington Ma, Inc. · Feb 22, 2012
K051947 — RADIONICS CUSA EXCEL ULTRASONIC SURGICAL ASPIRATOR SYSTEM WITH BONE · Radionics, A Division of Tyco Healthcare Group LP · Aug 22, 2005
K221835 — CUSA Clarity Ultrasonic Surgical Aspirator System · Integra LifeSciences Corporation · Aug 22, 2022
{0}------------------------------------------------
AUG 2 3 2002
Image /page/0/Picture/1 description: The image shows a sequence of characters and numbers written in black ink on a white background. The sequence starts with the letter 'K', followed by the number '0', and then the numbers '20220'. The handwriting appears somewhat rough and uneven.
## synergetics, inc.
## PREMARKET NOTIFICATION 510(k) SUMMARY [As required by 21 CFR 807.92(c)]
| Prepared by: | Sue Oster<br>Director of Quality Assurance and Regulatory Affairs |
|-------------------------|-------------------------------------------------------------------|
| Contact Person: | Same |
| Preparation Date: | January 2, 2002 |
| Device Name: | Aspirator tips for ultrasonic surgical instrument systems |
| Proprietary/Trade Name: | Synergetics, Inc. Sonotome™ Ultrasonic Aspirator Tips |
| Common/Usual Name: | Ultrasonic aspirator tips |
| Classification Name: | Instrument, Ultrasonic Surgical |
The following table provides a summary of the safety and effectiveness comparison between the predicate device and the proposed device:
| Name: | Predicate Device | Proposed Device | Safety & Effectiveness Comparison |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CUSA Excel Ultrasonic Surgical<br>Aspirator | Synergetics Sonotome Ultrasonic<br>Aspirator Tips | |
| Intended Use | Indicated for use in surgical<br>procedures where fragmentation,<br>emulsification and aspiration of<br>soft tissue is desirable. | Indicated for use in surgical<br>procedures where fragmentation,<br>emulsification and aspiration of<br>soft and hard (e.g.: bone) tissue<br>is desirable. | No difference in intended<br>use. No safety or<br>effectiveness issues<br>identified. |
| Materials and<br>Assembly<br>Methods | Tips made of 6AL-4V Titanium.<br>Tips are drawn and machined<br>from one solid piece of titanium. | Tips made of 6AL-4V Titanium.<br>Tips are drawn and machined<br>from one solid piece of titanium. | Identical in material and<br>construction. No safety or<br>effectiveness issues<br>identified. |
| Packaging | Supplied non-sterile in packages<br>of four individual tips. Also<br>supplied in a procedural kit. | Supplied sterile in packages of<br>three. Individual tips are<br>supplied in a Tyvek peel pouch. | Terminal sterilization of the<br>proposed device prior to<br>distribution should provide<br>greater flexibility for the<br>surgeon to change tips<br>during the procedure if<br>necessary |
Synergetics Sonotome™ Ultrasonic Aspirator Tips are accessories that are attached to the handpiece of an ultrasonic surgical aspirator system manufactured by another company. The tip, which simply transmits power from the system handpiece, is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. It allows the selective dissection of target
{1}------------------------------------------------
tissues while preserving vessels, ducts and other delicate structures. The main features on the Synergetics Sonotome™ Ultrasonic Aspirator Tips are:
- . A variety of tip diameters, shapes and lengths are available for specific surgical applications.
- Curved tips are bent so as to provide clear visualization of the surgical site, . unobstructed by the handpiece.
- Individually packaged, sterile tips can be replaced without the need for ● resterilization of the handpiece.
- Preaspiration holes in the tips minimize clogging and keep the tip clear of ● debris.
Validation and verification of the Ultrasonic Aspirator Tips will be accomplished through a combination of analysis and testing. This process will include a Risk Analysis and a mechanical performance test on prototype units.
The biological safety of the Ultrasonic Aspirating Tips has been assured through the selection of materials that demonstrate appropriate levels of biocompatibility. The material used (6AL-4V titanium, AMS4928) is the same as the material that is used in the predicate device and in other similar existing systems that are commercially available in the United States.
## Intended Use:
Synergetics Sonotome™ Ultrasonic Aspirator Tips are intended for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g.: bone) tissue is desirable, including:
- . Neurosurgery
- Gastrointestinal and Affiliated Organ Surgery .
- Urological Surgery ●
- . Plastic and Reconstructive Surgery
- General Surgery ●
- . Orthopedic Surgery
- . Gynecological Surgery
- . Thoracic Surgery
- . Laparoscopic Surgery
- . Thoracoscopic Surgery
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Synergetics, Inc. Sue Oster Director, Quality/Regulatory Affairs 88 Hubble Drive St. Charles, Missouri 63304
Re: K020220
Trade/Device Name: Synergetics Sonotome™ Ultrasonic Aspirator Tips Regulation Number: 878.4400 Regulation Name: Ultrasonic surgical instrument Regulatory Class: Class II Product Code: LFL Dated: April 26, 2002 Received: August 1, 2002
Dear Ms. Oster:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Sue Oster
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Colin M. Wittwer, Ph.D., M.D.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K 020220/A'
page *1* of *1*
510(k) NUMBER (IF KNOWN): K020220
DEVICE NAME: Ultrasonic/Stright, knife, ring curvette
INDICATIONS FOR USE:
Indications for Use:
Synergetics Sonotome™ Ultrasonic Aspirator Tips are intended for use in surgical procedures where fragmentation, emulsification and aspiration of so and Hard (e.g .: bone) tissue is desirable, including:
- Neurosurgery
- Gastrointestinal and Affiliated Organ Surgery
- Urological Surgery
- Plastic and Reconstructive Surgery
- General Surgery
- Orthopedic Surgery
- Gynecological Surgery
- Thoracic Surgery
- Laparoscopic Surgery
- Thoracoscopic Surgery
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter-Use
(Optional Format 1-2-9)
for
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
510(k) Number _
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.