CUSA Clarity Ultrasonic Surgical Aspirator System

K190180 · Integra LifeSciences Corporation · LFL · Apr 1, 2019 · SU

Device Facts

Record IDK190180
Device NameCUSA Clarity Ultrasonic Surgical Aspirator System
ApplicantIntegra LifeSciences Corporation
Product CodeLFL · SU
Decision DateApr 1, 2019
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive Surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses: Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological Surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

Device Story

CUSA Clarity Ultrasonic Surgical Aspirator System uses ultrasonic energy to fragment, emulsify, and aspirate soft and hard tissue. The CUSA Electrosurgery Module (CEM) accessory, a nosecone/cable/plug assembly, enables immediate electrosurgical coagulation of bleeding tissue at the surgical site using the same handpiece. Operated by surgeons in OR settings for open or laparoscopic procedures. Output allows real-time tissue removal and hemostasis, aiding surgical precision and reducing blood loss. Benefits include efficient tumor removal and tissue management.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including sterilization, shipping, stability, biocompatibility, EMC, electrical safety, and mechanical performance (handpiece/tip life, thermal effects, capacitive coupling).

Technological Characteristics

Ultrasonic surgical aspirator with CEM accessory. CEM consists of a molded nosecone, cable, and plug. Operates at 23 kHz. Features include electrosurgical coagulation capability. Tested for biocompatibility, electrical safety (EMC), and thermal effects per FDA guidance and recognized standards.

Indications for Use

Indicated for patients undergoing surgical procedures requiring fragmentation, emulsification, and aspiration of soft and hard tissue (e.g., bone). Applicable across neurosurgery, plastic/reconstructive, orthopedic, gynecological, thoracic, gastrointestinal, urological, and general/laparoscopic surgeries, including tumor resection and organ-specific procedures (e.g., liver, kidney).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 1, 2019 Integra LifeSciences Corporation Alexandra Wells Regulatory Affairs Specialist 311 Enterprise Drive Plainsboro, New Jersey 08536 Re: K190180 Trade/Device Name: CUSA Clarity Ultrasonic Surgical Aspirator System Regulatory Class: Unclassified Product Code: LFL. LBK Dated: January 31, 2019 Received: February 1. 2019 Dear Alexandra Wells: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html: good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/5 description: The image shows a digital signature. The signature is for Long H. Chen -S. The date of the signature is 2019.04.01. The time of the signature is 15:50:48 -04'00'. Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### K190180 Page 1 of 1 | | DEPARTMENT OF HEALTH AND HUMAN SERVICES | | |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | | Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. | | | <b>Indications for Use</b> | | | 510(k) Number (if known) | To Be Determined | | | Device Name | CUSA® Clarity Ultrasonic Surgical Aspirator System | | | Indications for Use (Describe) | The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable. | | | | The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: | | | | Neurosurgery, Plastic and Reconstructive Surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses: | | | | Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy | | | | Urological Surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy | | | | General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures | | | | Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy | | | Type of Use (Select one or both, as applicable) | <span> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </span> <span> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </span> | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (101)-443-6780 | |----------------------|-------------|----------------------------------------| |----------------------|-------------|----------------------------------------| {3}------------------------------------------------ ### 510(k) SUMMARY A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92 | 807.92(a)(1) - Submitter information | | |---------------------------------------------------------------------------|----------------------------------------------------| | Name | Integra LifeSciences Corporation | | Address | 311 Enterprise Drive Plainsboro, NJ 08536 USA | | Phone Number | 1-609-275-0500 | | Establishment Registration Number | 9004007 | | Name of Contact Person | Alexandra Wells | | Date Prepared | January 31, 2019 | | 807.92(a)(2) – Name of device | | | Trade or Propriety Name | CUSA® Clarity Ultrasonic Surgical Aspirator System | | Common or Usual Name | Ultrasonic Surgical Aspirator | | Classification Name | Instrument, Ultrasonic Surgical | | Classification Panel | General and Plastic Surgery | | Regulation | Unclassified | | Product Code(s) | LFL, LBK | | 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | | | CUSA® Excel+ Ultrasonic Surgical Aspirator System K141668 | | | CUSA® Clarity Ultrasonic Surgical Aspirator System K182809 | | {4}------------------------------------------------ ### 807.92(a)(4) - Device description The device within the scope of this premarket notification is the optional CUSA® Electrosurgery Module (CEM) accessory that is intended to be used with the 23 kHz components of the CUSA® Clarity Ultrasonic Surgical Aspirator System. The CUSA® Clarity Ultrasonic Surgical Aspirator System is the newest device added to the Integra Lifesciences Corporation family of tissue ablation products. There are three (3) systems currently marketed in the United States: CUSA® Clarity Ultrasonic Surgical Aspirator System, CUSA® Excel+ Ultrasonic Surgical Aspirator System and CUSA® NXT Ultrasonic Tissue Ablation System. Of these currently marketed products, a CEM accessory is offered with the CUSA® Excel+ Ultrasonic Surgical Aspirator System that allows for coagulation capabilities at the surgical site. The CEM accessory that is offered with the CUSA® Clarity is nearly identical to its predecessor, modified for use with the CUSA® Clarity system. The CEM accessory will allow the surgeon to apply immediate electrosurgical coagulation to bleeding tissue at the surgical site, with the same handpiece assembly that is removing unwanted tissue. The CEM accessory consists of a nosecone, cable and plug which are molded together and packaged as one unit. {5}------------------------------------------------ | 807.92(a)(5) - Intended use of the device | | |-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated | | | for use in surgical procedures where fragmentation, emulsification and<br>aspiration of soft and hard (e.g. bone) tissue is desirable. | | | The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:<br>Neurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery,<br>Gynecological Surgery and Thoracic Surgery and the following specific<br>uses: | | | Gastrointestinal and Affiliated Organ Surgery – including removal of<br>benign or malignant tumors or other unwanted tissue, including hepatic<br>parenchyma, in open or laparoscopic procedures, hepatic resection,<br>tumor resection, lobectomy or trisegmentectomy, or removal of tissue<br>during liver allotransplantation and donor hepatectomy | | Indications for | Urological surgery- including removal of renal parenchyma during<br>nephrectomy or partial nephrectomy | | Use | General Surgery - including removal of benign or malignant tumors or<br>other unwanted soft tissue in open or minimally invasive general<br>surgical procedures | | | Laparoscopic Surgery - including removal of hepatic parenchyma in<br>laparoscopic hepatic resection, lobectomy or trisegmentectomy, in<br>laparoscopic donor hepatectomy or laparoscopic cholecystectomy or<br>laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic<br>appendectomy, laparoscopic colon resection or laparoscopic partial<br>gastrectomy | | | | {6}------------------------------------------------ 807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate The CEM accessory offered with the CUSA Clarity has the same technological characteristics compared to the predicate device. The main purpose of the CEM accessory is to align CUSA Clarity with CUSA Excel in its ability to provide electrosurgical coagulation to bleeding tissue at the surgical site. The device is maintaining the same underlying technology and intended use of previous CEM accessory used with CUSA Excel. Thus, the majority of the features and technology of the CEM accessory are not new for a CUSA device and benefit longstanding safety and/or efficacy. # 807.92(b)(1-2) – Nonclinical and clinical tests submitted Non-clinical testing was performed to ensure the safety and efficacy of the CUSA Clarity 23 kHz CEM Nosecone accessory. Testing included, but was not limited to: - Sterilization, Shipping and Stability testing per FDA guidance documents and ● recognized standards - Biocompatibility testing per FDA guidance documents and recognized standards ● - EMC and Electrical Safety testing per FDA guidance documents and recognized . standards - Bench testing to verify device requirements, including those listed below ● - o Handpiece and Tip Life with CEM - Functionality within specification during environmental variations - Mechanical and Performance Testing ● - Thermal Effects and Capacitive Coupling per FDA guidance document ● Testing was determined successful for all 13 protocols within the 4 testing categories above and supports the conclusion that all product specifications and design inputs have been met. Integra LifeSciences therefore believes verification testing results for the CUSA Clarity 23 kHz CEM Nosecone accessory support a determination of substantial equivalence when compared with the predicate devices. No clinical studies were performed or required as all conducted performance tests appropriately support a determination of substantial equivalence compared with the predicate devices. # 807.92(b)(3) - Conclusions drawn from non-clinical and clinical data The results of the non-clinical testing indicate that the intended use of the device, fundamental scientific technology, and performance of the CUSA Clarity 23 kHz CEM Nosecone accessory are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...