CUSA EXCEL ULTRASONIC SURGICAL ASPIRATOR SYSTEM

K981262 · Valleylab, Inc. · LFL · Jul 6, 1998 · SU

Device Facts

Record IDK981262
Device NameCUSA EXCEL ULTRASONIC SURGICAL ASPIRATOR SYSTEM
ApplicantValleylab, Inc.
Product CodeLFL · SU
Decision DateJul 6, 1998
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The CUSA EXcel Ultrasonic Surgical Aspirator System is intended for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable, including: Neurosurgery Gastrointestinal and Affiliated Organ Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecological Surgery Thoracic Surgery Laparoscopic Surgery Thoracoscopic Surgery The system may also be combined with electrosurgery using the optional CUSA Electrosurgical Module (CEM).

Device Story

Ultrasonic surgical aspirator system; fragments, emulsifies, and removes soft tissue via ultrasonic mechanical energy; preserves vessels/ducts. System includes console, handpiece, titanium tip, irrigation/suction manifold, and footswitch. Optional CUSA Electrosurgical Module (CEM) provides electrosurgical cutting/coagulation. Used in hospital operating rooms by surgeons. Console controls power/irrigation; handpiece delivers ultrasonic energy to surgical site. Coaxial fluid delivery cools tip and clears debris. TISSUE Select feature allows control of fragmentation rate and provides tactile feedback. Output is physical tissue removal; surgeon visualizes site during procedure. Benefits include selective tissue dissection and reduced trauma to surrounding structures.

Clinical Evidence

Bench testing only. No clinical data provided. Verification performed via risk analysis and electro/mechanical performance testing on prototype units. Biocompatibility testing conducted per ISO 10993-1 for unique materials.

Technological Characteristics

Ultrasonic surgical aspirator; titanium handpiece tip; console-based power/control; coaxial irrigation/suction. Operates via ultrasonic mechanical energy. Compatible with external electrosurgical module. Materials biocompatible per ISO 10993-1. Complies with IEC 601-1, IEC 601-2-2, EN 60601-1-2, and ANSI/AAMI HF18. Console is non-microprocessor based (no memory).

Indications for Use

Indicated for soft tissue fragmentation, emulsification, and aspiration in neurosurgery, GI, urological, plastic/reconstructive, general, orthopedic, gynecological, thoracic, laparoscopic, and thoracoscopic surgical procedures.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ l. # SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION CUSA EXCEL ULTRASONIC SURGICAL ASPIRATOR SYSTEM ## Product Description: The CUSA EXcel Ultrasonic Surgical Aspirator System is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. It allows the selective dissection of target tissues while preserving vessels, ducts and other delicate structures. The CUSA EXcel System consists of a console which provides control and power functions, a surgical handpiece which provides ultrasonic mechanical energy to the surgical site, a titanium handpiece tip and flexible irrigation flue, and a suction/irrigation system (manifold tubing and cooling water canister). The CUSA EXcel System accommodates most commercially available suction canisters, and specimen traps (optional). A two-pedal footswitch is provided with the console. The CUSA EXcel System may be used with an external CUSA Electrosurgical Module (CEM), which will provide optional electrosurgical capability. The CEM module is a Force FX-8 Generator (FDA clearance was received for the Force FX Generator under K#944602). This provides the surgeon with the option of delivering cutting or coagulating current through the CUSA handpiece tip to the surgical site. The CEM function is controlled either with the Force FX-8 footswitch or by hand, using the CEM nosecone control button(s). The main features of the CUSA EXcel Ultrasonic Surgical Aspirator System are as follows: - Console is small and easy to move � - Small, lightweight, ergonomic handpiece (CUSA EXcel 36kHz Handpiece) � - Variety of tip diameters and lengths for specific surgical applications � - � External CUSA Electrosurgical Module (CEM) provides optional electrosurgical capability and hemostasis - � Protective, flexible irrigation flue to shield the surgical tip and prevent trauma to surrounding tissue - � Simple to set-up and use - � CUSA EXcel Console is compatible with the currently marketed Valleylab CUSA PFT Ultrasonic 23kHz Straight Handpiece (C2500) - � Ability to change tips in a sterile surgical environment, such as a hospital operating room - � Preaspiration holes in the surgical tip minimize clogging and keep tip clear of debris - � Two-mode suction capability for open and laparoscopic procedures {1}------------------------------------------------ - Fast Flush mode flushes the operative site with irrigation fluid � - Coaxial fluid delivery to maintain cooling and irrigation at the surgical tip, � and to efficiently remove debris for added visualization - Includes TISSUE Select feature for control of fragmentation rate, and � enhanced tactile feedback for added selectivity #### II. Intended Use: The CUSA EXcel Ultrasonic Surgical Aspirator System is intended for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable, including: Neurosurgery Gastrointestinal and Affiliated Organ Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecological Surgery Thoracic Surgery Laparoscopic Surgery Thoracoscopic Surgery The system may also be combined with electrosurgery using the optional CUSA Electrosurgical Module (CEM). - Safety and Performance Data: III. The CUSA EXcel Ultrasonic Surgical Aspirator System has been designed to conform with applicable sections of the following standards: - IEC 601-1 (1988), Medical Electrical Equipment Part 1: General Requirements for Safety - IEC 601-2-2 (1991), Medical Electrical Equipment Part 2: Particular Requirements for the Safety of High Frequency Surgical Equipment - EN 60601-1-2 (1993) Medical Electrical Equipment, Part 1 General Requirements for Safety, 2 Collateral Standard: Electromagnetic Compatibility-requirements and Tests IEC 801-3 (1984), Radiated Electromagnetic Field Requirements ANSI/AAMI HF18 (1993), Electrosurgical Devices - CISPR11, Electromagnetic Compatibility for Industrial-process Measurement and Control Equipment: Emissions Requirements {2}------------------------------------------------ Validation and verification of the CUSA EXcel System will be accomplished through a combination of analysis and testing. This process will include a Risk Analysis and an electro/mechanical performance test on prototype units. The safe performance and proper function of the hardware and external controls of the CUSA EXcel Console will be tested (Note: the EXcel Console contains no microprocessor and no memory). The CUSA EXcel Handpiece and accessories will also be tested, both alone and in conjunction with the Console. The following is a summary of the testing to date, including test criteria, performed on the CUSA EXcel Console and the CUSA EXcel Handpiece. Testing has verified the safe performance and proper function of the product The biological safety of the CUSA EXcel Ultrasonic Surgical Aspirator System devices has been assured through the selection of materials which demonstrate appropriate levels of biocompatibility. Many of the materials used in the CUSA EXcel System are the same as currently used in Valleylab CUSA System 200 devices, which are legally marketed device(s); refer to Attachment 3 for a complete listing and 510(k) references. Those materials which are unique to the CUSA EXcel System are currently undergoing testing in accordance with ISO Standard 10993-1, Biological Evaluation of Medical Devices, Part 1 - Guidance on Selection of Tests, Device Category "External Communicating Device, Blood Path Indirect, contact duration A". - Summary of Substantial Equivalence: IV. The CUSA EXcel Ultrasonic Surgical Aspirator System (Console, Handpiece, and Accessories) is substantially equivalent in function and intended use to the following legally marketed devices: - Valleylab CUSA System 200 (510(k)'s K#921251, K#931902, K#934628) ◆ This console system operates at a frequency of 23kHz. The above submissions included 23kHz handpieces (CUSA PFT Ultrasonic Handpiece) and system accessories. - . Valleylab CUSA System 200H with 37kHz Handpiece (510(k) K#884413). This smaller handpiece operates at a higher frequency (37kHz) for applications which demand greater precision and user visibility. A comparison chart of product features is located in Attachment 6. Included in Attachment 7 is a copy of the advertising literature for the CUSA System 200 Console and CUSA PFT Ultrasonic Handpiece. The surgical indications outlined for the CUSA EXcel Ultrasonic Surgical Aspirator System (refer to section G. Intended Use) are identical to those indications which have been cleared by FDA in previous 510(k) submissions for the CUSA System 200 Ultrasonic Surgical Aspirators. Refer Attachment 8 for a {3}------------------------------------------------ listing of previous CUSA System 200 510(k) submissions cleared by FDA, and the indications for use associated with each submission. With all of these indications, the devices are intended to be used in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. No additional surgical applications are requested for the CUSA EXcel System. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three faces in profile, arranged in a row. The faces are black and are connected to a flowing, ribbon-like shape below. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circle around the image. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 6 1998 .JUL Ms. Julie Ross ·Senior Regulatory Affairs Associate Valleylab, Inc. 5920 Longbow Drive Boulder, Colorado 80301 K981262 Re: Trade Name: CUSA Excel Ultrasonic Surgical Aspirator System Regulatory Class: II Product Code: LFL, LBK Dated: April 6, 1998 Received: April 7, 1998 ### Dear Ms. Ross: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {5}------------------------------------------------ Page 2 - Ms. Julie Ross This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health {6}------------------------------------------------ 11981262 ### Indications for Use 510(k) Number (if known): Not known at the time of submission Device Name: CUSA EXcel Ultrasonic Surgical Aspirator System Indications For Use: The CUSA EXcel Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable, including Neurosurgery, Gastrointestinal and Affiliated Organ Surgery, Urological Surgery, Plastic and Reconstructive Surgery, General Surgery, Orthopedic Surgery, Gynecological Surgery, Thoracic Surgery, Laparoscopic Surgery, and Thoracoscopic Surgery. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Image /page/6/Figure/7 description: The image shows a form titled "Concurrence of CDRH, Office of Device Evaluation (ODE)". The form includes a section for "Prescription Use" and "Over-The-Counter Use", with the "Prescription Use" option checked. There is also a handwritten number "K981262" on the form.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...