Harmonic Hand Piece, Blue, Harmonic Hand Piece, Gray

K211273 · Ethicon Endo-Surgery, LLC · LFL · May 21, 2021 · SU

Device Facts

Record IDK211273
Device NameHarmonic Hand Piece, Blue, Harmonic Hand Piece, Gray
ApplicantEthicon Endo-Surgery, LLC
Product CodeLFL · SU
Decision DateMay 21, 2021
DecisionSESE
Submission TypeSpecial
Device ClassClass U
AttributesTherapeutic

Intended Use

The HARMONIC® Blue Hand Piece, when used in conjunction with the HARMONIC® Instruments, is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels. The HARMONIC® Gray Hand Piece, when used in conjunction with the HARMONIC® Instruments, is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels.

Device Story

Hand pieces convert electrical energy from compatible Ethicon generators into mechanical motion for surgical instrument blades. Used in clinical settings by surgeons for soft tissue incisions; provides bleeding control and minimal thermal injury. Devices feature integrated procedure counters (100 procedures for Blue; 95 for Gray) to manage service life; generator displays remaining procedure counts. Includes blade wrench and test tip for assembly. Modifications include updated EEPROM for cybersecurity, harmonized cable design, and improved durability. Output is mechanical vibration of the blade; clinical benefit is precise tissue cutting with hemostasis.

Clinical Evidence

No clinical data provided. Substantial equivalence demonstrated via risk analysis, verification testing, biocompatibility evaluation, and electrical testing.

Technological Characteristics

Electromechanical transducer; converts electrical energy to mechanical motion. Tissue-contacting materials comply with ISO 10993-1. Features EEPROM for procedure counting and cybersecurity. Compatible with GEN11 generators (software 2018-1 or later). Non-sterile; requires sterilization per instructions. Form factor includes permanently attached cable and blade interface.

Indications for Use

Indicated for soft tissue incisions requiring bleeding control and minimal thermal injury; adjunct or substitute for electrosurgery, lasers, and steel scalpels. No specific age or gender restrictions provided.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 21, 2021 Ethicon Endo-Surgery, LLC % Ms. Ruth James, MSc, RAC Senior Regulatory Affairs Program Lead Ethicon Endo-Surgery, LLC. 4545 Creek Road Cincinnati, OHIO 45242 Re: K211273 Trade/Device Name: Harmonic Hand Piece, Blue, Harmonic Hand Piece, Gray Regulatory Class: Unclassified Product Code: LFL Dated: April 26, 2021 Received: April 27, 2021 Dear Ms. James: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211273 Device Name HARMONIC® Blue Hand Piece HARMONIC® Gray Hand Piece #### Indications for Use (Describe) The HARMONIC® Blue Hand Piece, when used in conjunction with the HARMONIC® Instruments, is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels. The HARMONIC® Gray Hand Piece, when used in conjunction with the HARMONIC® Instruments, is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------| | <span style="text-decoration: underline;"><b></b></span> <span style="font-family: Arial;"><b> X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"><b></b></span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the Ethicon Endo-Surgery logo. The logo consists of a grid of red circles on the left and the text "Ethicon Endo-Surgery" in gray on the right. Below the text is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" in a smaller font. Ethicon Endo-Surgery, LLC Guaynabo, Puerto Rico 00969 # K211273 ## 510(k) Summary | Submitter Information: | Ethicon Endo-Surgery, LLC<br>475 Calle Street<br>Guaynabo, PR 00969 | |------------------------|---------------------------------------------------------------------| |------------------------|---------------------------------------------------------------------| #### Application Correspondent Ms. Ruth James MSc, RAC. Sr. Regulatory Affairs Program Lead Ethicon Endo-Surgery, LLC. 4545 Creek Road Cincinnati, OH 45242 Telephone: (513) 337-3118 Fax: (513) 337-1122 Email: rjames15@its.jnj.com ## Date Prepared May 20, 2021 | Device Trade Name: | HARMONIC® Blue Hand Piece<br>HARMONIC® Gray Hand Piece | |--------------------------------|-----------------------------------------------------------| | Models: | HARHPBL, HARHPBLCN (Blue)<br>HARHPGR, HARHPGRCN<br>(Gray) | | Device Class: | Unclassified | | Panel: | II | | Classification (Product) Code: | 79, General & Plastic Surgery<br>LFL | Legally Marketed Predicate Devices: - . HARMONIC® Hand Piece, Blue (K063192) - . HARMONIC® Hand Piece, Gray (K002906) This predicate has not been subject to a design-related recall. {4}------------------------------------------------ ## Device Description #### HARMONIC® Blue Hand Piece The HARMONIC® Blue Hand Piece is designed to convert electrical energy from a compatible Ethicon Generator to mechanical motion for the instrument blade. The Blue Hand Piece is intended for use exclusively with a compatible HARMONIC Generator. The Blue Hand Piece is permanently attached to a blue cord which connects to the front of the Generator. The Blue Hand Piece is a re-useable instrument with a limited service life. The Blue Hand Piece is programmed with a counter to limit the service life to 100 procedures. The generator will give a Hand Piece error after 100 procedures are completed. The number of activations during a procedure is not limited, and the counter will not log a procedure until the Hand Piece is unplugged from the generator or the generator is powered down. The number of remaining procedures is indicated on the display of the generator in the Systems Settings screen. The Hand Piece is compatible with generator software 2018-1 or later. #### HARMONIC® Gray Hand Piece The HARMONIC® Gray Hand Piece is designed to convert electrical energy from a compatible Ethicon Generator to mechanical motion for the instrument blades. This Hand Piece is intended for use with a compatible Ethicon Generator. The Hand Piece is permanently attached to a cord which connects to the front of the Generator. The Hand Piece is a re-useable instrument with a limited service life. The Hand Piece is programmed with a counter to limit the service life to 95 procedures. The generator will give a Hand Piece error after 95 procedures are completed. The number of activations during a procedure is not limited, and the counter will not log a procedure until the Hand Piece is unplugged from the generator or the generator is powered down. The number of remaining procedures is indicated on the display of the generator in the Systems Settings screen. The Hand Piece is compatible with generator software 2018-1 or later. The Hand Piece is packaged with a Blade Wrench and Test Tip. The Blade Wrench is used to secure instrument blades to the Hand Piece. The Hand Piece. Blade Wrench, and Test Tip are packaged non-sterile and must be sterilized per their insert instructions prior to use. {5}------------------------------------------------ ## Indications for Use for new devices (identical to predicates) The HARMONIC® Blue Hand Piece, when used in conjunction with the HARMONIC® Instruments, is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels. The HARMONIC® Gray Hand Piece, when used in conjunction with the HARMONIC® Instruments, is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels. #### Comparison of technological characteristics with the predicate device At a high level, the new and predicate devices are based on the following same technological elements: | Characteristic | New Devices | Legally marketed Predicate<br>Devices<br>K063192, K002906 | |------------------------------------------------------------------------------------------------|-------------|-----------------------------------------------------------| | Indications for Use | Identical | Identical | | Contraindications | Identical | Identical | | Non-Sterile, Single Patient Use | Yes | Identical | | Bio-Compatibility of Materials:<br>All tissue-contacting materials comply<br>with ISO 10993-1. | Yes | Yes | | Packaging Method | Identical | Identical | | Device Operation | Identical | Identical | The following technological differences exist between the new and predicate devices: - . The new Hand Pieces have a standard connector (tear-drop) interface to GEN11 generators. - . The new Hand Pieces are only compatible with GEN11 operating 2018-1 software and later generators. - . The Electrically erasable programmable read-only memory (EEPROM) in the Hand Pieces has been upgraded with cybersecurity improvements. - . The cable design has been harmonized for both Blue and Gray Hand Pieces. - . The cable components have been modified to improve hand activation durability - . There is a new protective end cap, used in packaging, for the Gray Hand Piece. {6}------------------------------------------------ ## Performance Data Risk analyses for each device modification are provided. Verification testing, Biocompatibility evaluation and Electrical testing were conducted to confirm device modifications do not raise new issues of safety or effectiveness. Clinical studies were not required to demonstrate substantial equivalence. #### Conclusions Risk analyses show that the risk profile of the new devices has not changed as a result of the described changes while having the same intended use as the predicate devices. Verification testing, Biocompatibility evaluation and Electrical testing conducted, confirm the performance of the new devices is consistent with the predicate devices and does not raise any new questions of safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...