K212466 · Dia Imaging Analysis, Ltd. · QIH · Dec 9, 2021 · Radiology
Device Facts
Record ID
K212466
Device Name
LVivo Seamless
Applicant
Dia Imaging Analysis, Ltd.
Product Code
QIH · Radiology
Decision Date
Dec 9, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Global Longitudinal Strain (GLS)
—
Correlation > 0.80
Correlation 0.88
—
—
100 patient examinations with GLS results by Automated Functional Imaging (AFI).
—
Indications for Use
LVivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease and Age >18.
Device Story
LVivo Seamless v2.0 is a standalone software application operating offline on a server within a healthcare environment. It receives ultrasound examinations in DICOM format from ultrasound devices. The system automatically selects appropriate clips for Ejection Fraction (EF) and Global Longitudinal Strain (GLS) evaluation. It activates the FDA-cleared LVivo EF and LVivo Strain modules to perform automated analysis. Results are transmitted to the PACS for review by a healthcare professional. The device assists clinicians in evaluating cardiac function by providing automated measurements, which are used to support clinical decision-making. The primary benefit is the automated, non-invasive assessment of cardiac parameters, potentially improving workflow efficiency and consistency in cardiac ultrasound analysis.
Clinical Evidence
Validation study used 100 patient examinations (age >18). The system successfully identified 4CH, 2CH, and 3CH clips in 95% of cases. GLS measurements were compared against GE's Automated Functional Imaging (AFI) as the reference standard. Results showed a correlation of 0.88, mean accuracy of -1.4% ± 3.93% (Bland-Altman), sensitivity of 0.90, and specificity of 0.83, using a normal/abnormal threshold of -18%.
Technological Characteristics
Standalone software application; runs on Windows-based servers; processes DICOM ultrasound clips; utilizes automated image selection and analysis modules (LVivo EF and LVivo Strain); supports manual editing of results; integrates with PACS; offline processing architecture.
Indications for Use
Indicated for non-invasive processing of ultrasound images to detect, measure, and calculate medical parameters of cardiac structure and function in patients >18 years old with suspected disease.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
December 9, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
DiA Imaging Analysis Ltd % George Hattub Senior Project Manager Medicsense USA 291 Hillside Avenue SOMERSET, MASSACHUSETTS 02726
Re: K212466
Trade/Device Name: LVivo Seamless v2.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: November 8, 2021 Received: November 10, 2021
Dear George Hattub:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Division Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name DiA LVivo Seamless
Indications for Use (Describe)
LVivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease and Age >18.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
Pursuant to CFR 807.92, the following 510(k) Summary is provided:
| 1. (a) | Submitter<br>Address: | George J. Hattub<br>Medicsense USA LLC<br>291 Hillside Avenue<br>Somerset, MA 02726<br>ghattub@comcast.net<br>https://www.upwork.com/freelancers/~0196e832ca4b82a2f3?viewMode=1 | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. (b) | Manufacturer<br>Address: | DiA Imaging Analysis Ltd<br>HaEnergia Street 77<br>Beer-Sheva, Israel 8470912 | |
| | Mfg. Phone: | Tel.: +972 77 7648318 | |
| | Contact Person: | Mrs. Michal Yaacobi | |
| | Date: | December 7, 2021 | |
| 2. | Device &<br>Classification<br>Name: | Medical Image Management and Processing System -<br>classified as Class 2 QIH, Regulation Number 21 CFR 892.2050<br>LVivo Seamless | |
| 3. | Predicate Device: | K202546 LVivo Seamless (Reference Device K210053) | |
| 4. | Description: | The LVivo Seamless is a standalone application that extends the LVivo<br>Platform and runs offline on a server in a healthcare environment. The<br>system accepts echo examinations in DICOM format that are sent from an<br>Ultrasound device and automatically selects the adequate clips for EF and<br>GLS evaluation. After the clip selection, the LVivo Seamless activates the<br>FDA cleared LVivo EF and LVivo Strain modules which perform automatic<br>evaluation. The results are sent to the PACS and are evaluated by a<br>healthcare professional. | |
| 5. | Indications for<br>Use: | LVivo platform is intended for non-invasive processing of ultrasound images<br>to detect, measure, and calculate relevant medical parameters of structures<br>and function of patients with suspected disease and Age>18 | |
| 6. | Comparison of<br>Technological<br>Characteristics: | With respect to technology and intended use, DiA's LVivo Seamless is<br>substantially equivalent to its predicate device. Based upon the outcomes<br>from the risk analysis and Performance Testing Evaluation, DiA believes<br>that the modification of the predicate device does not raise additional safety<br>of efficacy concerns. The following comparison table depicts the changes. | |
| | Submitted Device | Predicate Device | Reference Device |
| Features/Characteristics | LVivo Seamless | LVivo Seamless | LVivo Software<br>Application |
| Product Code | same | QiH | QiH |
| Indication for Use | DiA's LVivo platform is<br>intended for non-invasive<br>processing of ultrasound<br>images to detect,<br>measure, and<br>calculate relevant<br>medical parameters<br>of structures and<br>function of patients with<br>suspected disease and<br>Age>18 | LVivo platform is<br>intended for non-<br>invasive processing of<br>ultrasound images to<br>detect, measure, and<br>calculate relevant<br>medical parameters of<br>structures and function<br>of patients with<br>suspected disease. | LVivo platform is<br>intended for non-<br>invasive processing of<br>ultrasound images to<br>detect, measure, and<br>calculate relevant<br>medical parameters of<br>structures and function<br>of patients with<br>suspected disease |
| Modules | LVivo EF, LVivo Strain | LVivo EF | LVivo EF, LVivo SG<br>(LVivo SWM & LVivo<br>Strain), LVivo SAX,<br>LVivo RV and LVivo<br>Bladder |
| Automation | same | yes | yes |
| Manual Adjustment | same | yes | yes |
| Bi plane EF evaluation | same | yes | yes |
| Simultaneous 2CH and<br>4CH evaluation | same | yes | yes |
| Off-line EF evaluation<br>using DICOM clips of any<br>vendor | same | yes | yes |
| Automated ED and ES<br>frames selection | same | yes | yes |
| Dynamic left ventricular | same | yes | yes |
| Manual editing by<br>user capability | Same, added editing<br>capabilities to the<br>output results | yes | yes |
| | | | |
| | | | |
| | | | |
| Visually confirm EF | same | yes | yes |
| Automated rejection of<br>false results | same | yes | yes |
| Volume calculation by<br>standard Simpson's<br>method of discs | same | yes | yes |
| Volume curve<br>calculation | same | yes | yes |
| EF results presentation | same | yes | yes |
| Enables calculation EF<br>results for different<br>cycle | same | yes | yes |
| Algorithm | same | same | yes |
| Calculation speed | same | yes | yes |
| Capability or a part of a<br>bigger package (device)<br>for LV function<br>evaluation | same | yes | yes |
| Segmental Longitudinal<br>Strain Measure | yes | no | yes |
| Global Longitudinal<br>Strain (GLS) Measure | yes | Yes (from LVivo<br>EF module) | yes |
| GLS calculation per view | yes | no | no |
| Operating System | Windows | Windows | Windows/Linux<br>(with Android<br>option for LVivo<br>EF) |
| 510(k) # | K212466 | K202546 | K210053 |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
- 7. Performance A summary of the Performance Evaluation, which was based upon well-Evaluation: established test methods, demonstrated conformity to the intended use. Success criteria:
1. Correct identification of 4ch and 2ch and 3ch views in 90% of the examinations
2. Correlation of 80% between GLS by AFI and GLS by LVivo Seamless
100 patient examinations were used for the validation. Inclusion criteria: Age>18, examinations in which GLS results by Automated Functional Imaging (AFI .GE) were available. No exclusion criteria were applied. The system was able to recognize all three 4CH, 2CH and 3CH clips in 95% of the examinations. The measurements by LVivo Strain running offline on the automatically selected clips were compared to the measurements done routinely in the Echo lab by AFI. The GLS by LVivo Strain was compared to GLS by AFI. Statistical analysis was done by Correlation, Bland-Altman:
| Measurement | Accuracy<br>(Mean±1.96STD) | Data Range | Correlation |
|-------------|----------------------------|------------|-------------|
| GLS | -1.4% ± 3.93% | -4% - -24% | 0.88 |
sensitivity and specificity and of 0.90, and 0.83 respectively. The Normal\Abnormal threshold used for GLS was -18%
- 8. Conclusion: The Intended Use and the technological characteristics in the current device are the same as those in the predicate device, including the addition of the GLS Module, do not affect the safety and effectiveness of the device. The performance tests have been completed and successfully support the device performance. Therefore, DiA Imaging Analysis concludes the LVivo Seamless is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.