Us2.v1

K210791 · Eko.Ai Pte Ltd. D/B/A Us2.Ai · QIH · Jul 27, 2021 · Radiology

Device Facts

Record IDK210791
Device NameUs2.v1
ApplicantEko.Ai Pte Ltd. D/B/A Us2.Ai
Product CodeQIH · Radiology
Decision DateJul 27, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Intended Use

Us2.v1 is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracy cardiac ultrasound images, automatically producing a full report with measurements of several and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or licensed primary care providers for clinical decision-making. Us2.v1 is in adult patients. Us2.v1 has not been validated for the assessment of congenital heart disease, pericardial disease, and/or intra-cardiac lesions (e.g. tumours, thrombi). Please note the following limitations: - Poor image capture will lead to poor annotations and subsequent measurements. Multiple image quality algorithms are used to filter out images of poor quality. - Our software complements good patient care and does not exempt the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. - If no gender is present, female referenced guideline values will be used for conclusions. - If Body Surface Area (BSA) is not present, indexed values cannot be provided. - During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, use of suboptimal settings (e.g. gain, contrast, depth), or lack of electrocardiogram capture may lead to lower accuracy of the software.

Device Story

Us2.v1 is an image post-processing software platform for DICOM-format cardiovascular ultrasound images; used by sonographers and clinicians in clinical settings. Input: transthoracic echocardiographic images. Operation: machine learning-based view classification and border detection to automate cardiac measurements (e.g., LVEF, volumes, dimensions, flow parameters). Output: automated markup, quantitative analysis, and summary report. Workflow: sonographer performs markup (manual or automated); physician reviews and uses output to support clinical decision-making. Benefits: provides standardized, automated cardiac measurements to assist in patient assessment and management.

Clinical Evidence

Retrospective non-interventional validation study using 600 previously acquired echocardiograms (421 HFrEF, 179 normal). Primary endpoint: individual equivalence coefficient (IEC) compared to triplicate human expert reads. Success criterion: non-inferiority margin Δ=0.25. Results: 95% confidence intervals for all Us2.v1 measurements fell below the 0.25 margin, demonstrating interchangeability with human reference standards.

Technological Characteristics

Software-only device operating on off-the-shelf PC hardware. Processes DICOM-compliant cardiac ultrasound images. Utilizes machine learning algorithms for view classification and border detection. Provides automated measurements of cardiac structure and function. Cybersecurity controls included. No specific hardware materials or sterilization required.

Indications for Use

Indicated for adult patients to process, analyze, and measure transthoracic cardiac ultrasound images to support cardiologists or primary care providers in clinical decision-making. Not validated for congenital heart disease, valve disease, pericardial disease, or intra-cardiac lesions (tumors, thrombi).

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. eko.ai Pte. Ltd. d/b/a Us2.ai % Jared Seehafer Regulatory Consultant Enzyme Corporation 611 Gateway Blvd., Suite 120 SOUTH SAN FRANCISCO CA 94080 July 27, 2021 Re: K210791 Trade/Device Name: Us2.v1 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: June 17, 2021 Received: June 24, 2021 Dear Jared Seehafer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210791 Device Name Us2.v1 #### Indications for Use (Describe) Us2.v1 is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracy cardiac ultrasound images, automatically producing a full report with measurements of several and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or licensed primary care providers for clinical decision-making. Us2.v1 is in adult patients. Us2.v1 has not been validated for the assessment of congenital heart disease, pericardial disease, and/or intra-cardiac lesions (e.g. tumours, thrombi). Please note the following limitations: - Poor image capture will lead to poor annotations and subsequent measurements. Multiple image quality algorithms are used to filter out images of poor quality. - Our software complements good patient care and does not exempt the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. - If no gender is present, female referenced guideline values will be used for conclusions. - If Body Surface Area (BSA) is not present, indexed values cannot be provided. - During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, use of suboptimal settings (e.g. gain, contrast, depth), or lack of electrocardiogram capture may lead to lower accuracy of the software. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | <div> <span> 区 Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 5 510(k) Summary | Submission Number: | K210791 | | |--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Submitter's Name: | Eko.ai Pte. Ltd. (d/b/a Us2.ai) | | | Address: | 2 College Road, #02-00, Singapore 169850 | | | Contact Person: | Jared Seehafer | | | Title: | Regulatory Consultant | | | Telephone Number: | 415-857-9554 | | | Fax Number: | 415-367-1279 | | | Email: | jared@enzyme.com | | | Date Summary Prepared: | July 15, 2021 | | | Device Proprietary Name: | Us2.v1 | | | Model Number: | V 1.0.0 | | | Common Name: | Us2.v1 | | | Regulation Number: | 21 CFR 892.2050 | | | Regulation Name: | System, Image Processing, Radiological | | | Product Code: | QIH | | | Device Class: | Class II | | | Predicate Device | Trade name: EchoMD Automated Ejection Fraction<br>Software<br>Manufacturer: Bay Labs, Inc.<br>(now known as Caption Health, Inc.)<br>290 King Street<br>San Francisco, CA 94107<br>Regulation Number: 21 CFR 892.2050<br>Regulation Name: System, Image Processing, Radiological<br>Device Class: Class II<br>Product Code: LLZ<br>510(k) Number: K173780<br>510(k) Clearance Date: June 14, 2018 | | Table 5-1. Subject Device Overview. {4}------------------------------------------------ # 5.1 Device Description Us2.v1 is an image post-processing analysis software device used for viewing and quantifying cardiovascular ultrasound images in DICOM format. The device is intended to aid diagnostic review and analysis of echocardiographic data, patient record management and reporting. The software provides an interface for a skilled sonographer to perform the necessary markup on the echocardiographic image prior to review by the prescribing physician. The markup includes: the cardiac segments captured, measurements of distance, time, area and blood flow, quantitative analysis of cardiac function, and a summary report. The software allows the sonographer to enter their markup manually. It also provides automated markup and analysis, which the sonographer may choose to accept outright, to accept partially and modify, or to reject and ignore. Machine learning based view classification and border detection form the basis for this automated analysis. Additionally, the software has features for organizing, displaying and comparing to reference guidelines the quantitative data from cardiovascular images acquired from ultrasound scanners. # 5.2 Indications for Use Us2.v1 is a software platform that automatically processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or licensed primary care providers for clinical decision-making based on DICOM images produced by FDA approved ultrasound devices. Us2.v1 is indicated for use in adult patients. Us2.v1 has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g. tumours, thrombi). Please note the following limitations: - . Poor image capture will lead to poor annotations and subsequent measurements. Multiple image quality algorithms are used to filter out images of poor quality. - . Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. - If no gender is present, female referenced guideline values will be used for conclusions. . - If Body Surface Area (BSA) is not present, indexed values cannot be provided. . - . During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, use of suboptimal settings (e.g. gain, contrast, depth), or lack of electrocardiogram capture may lead to lower accuracy of the software. {5}------------------------------------------------ # 5.3 Summary of Technological Characteristics Comparison Table 5-2 shows the similarities and differences between the technological characteristics of the two products. Testing demonstrates that the differences do not raise new questions of safety or effectiveness. | Topic | Predicate Device | Subject Device | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--| | Physical<br>Characteristics | Software package that operates<br>utilizing off-the-shelf hardware | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | DICOM Standard<br>Compliance | The software processes DICOM<br>compliant image data | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Modalities | Cardiac echocardiogram | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | User Interface | The software is designed for use on<br>a personal computer that has been<br>received images from a compatible<br>PACS | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Automation level | Fully automated, including clip<br>selection | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | User<br>confirmation/<br>rejection of result | Yes | Yes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Manual editing<br>of automated<br>result by user | Yes (on PACS workstation) | Yes (in application) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Automated<br>calculations | Ejection Fraction | Region Measurement LV DecT LV MV-A LV MV-Adur LV MV-E LV e' lateral LV e' septal LV a' lateral LV a' septal LV s' lateral LV s' septal LV LVEDV MOD biplane LV LVEF MOD biplane LV LVESV MOD biplane LV LVSV MOD biplane LV IVSd LV LVIDd LV LVIDs LV LVPWd LV E/e' mean RV RVIDd LA LAESV MOD biplane RA Raa | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Table 5-2. Summary of Technological Characteristics Comparison. {6}------------------------------------------------ ### 5.4 Performance Data Us2.v1 was developed and tested in accordance with Us2.ai's Design Control processes. The device has been subject to extensive safety and performance testing. Non-clinical verification and validationtest results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at unit, module, and system integration levels. Risk managementanalysis generated multiple risk mitigation measures and verification activities. A Cybersecurity Analysis and data security testing were conducted to verify that data and patient protected health information security measures are included in the design of the software. A Human Factors/Usability Engineering study performed according to the principles of AAMI/ANSI HE75 was performed to validate device's usability within the intended user population. A formal retrospective non-interventional validation study was conducted using 600 previouslyacquired echocardiograms to evaluate the performance of the device for each claimed measurement. The patient dataset was constructed to provide a balanced range of gender and body mass index levels. A blinded, anonymized series of 600 unique echocardiographic studies was selected for analysis. Of these, 421 are samples of patients with heart failure with reduced ejection fraction. The remaining 179 examinations were from normal subjects. Both cohorts of subjects were selected fromsets of previously annotated and overread studies with robust inclusion criteria (see below). Additional demographic information and vital signs at the time of examination are available in Table 1. Two subject cohorts were used for analysis across 600 studies: - HFrEF Subjects: 421 studies - - -Normal Subjects: 179 studies Inclusion criteria were: #### HFrEF Subjects Subjects eligible for inclusion in this study met all of the following criteria at screening: - 1. Written informed consent for the original study. - 2. Men and women ≥ 18 years of age. - 3. HFrEF subjects (heart failure with LVEF < 40% determined at the site) - 4. NYHA Functional class II-IV. #### Normal Subjects Subjects eligible for inclusion in this study met all of the following criteria at screening: - 1. Written informed consent for obtained. - 2. Males and females (of non-childbearing potential) between 18 and 55 years of age,inclusive {7}------------------------------------------------ - 3. Body weight > 55.0 kg and body mass index within 18.0 to 32.0 kg/m2, inclusive - 4. In good health, in the opinion of the Investigator, as determined by the following: - a. A physical examination at screening and medical history with no clinically significant abnormalities - b. Not taking medications for the treatment of any chronic or episodic medicaldisease or condition - c. Vital signs within the ranges below measured while the subject is supine after 3 minutes rest. May re-check vital signs twice at screening. * heart rate 40-90 beats per minute; systolic blood pressure (BP) 100-140 mmHg: diastolic BP 50-95 mmHg: respiration rate <25 breaths per minute; oxygen saturation: 95-100% - 5. Acoustic windows adequate for accurate transthoracic echocardiograms, confirmed by the echocardiography core laboratory - 6. Normal cardiac structure and function, as determined by the echocardiography corelaboratory, or if abnormalities are present, they are deemed not clinically significant LVEF > 55% - 7. Normal electrocardiogram (ECG) or, if abnormalities are present, they are deemed notclinically significant by the Investigator | | Normal Cohort | HFrEF Cohort | |--------------------------|--------------------|---------------------| | Age | 40.0 (± 7.5) years | 64.3 (± 12.5) years | | % Male | 72.6% | 67.4% | | Height | 168.1 (± 9.4) cm | 172.2 (± 10.3) cm | | Weight | 76.6 (± 9.2) kg | 93.4 (± 24.9) kg | | Heart Rate | 62.7 (± 7.1) bpm | 72.2 (± 11.9) bpm | | Systolic Blood Pressure | 112.0 (± 9.7) mmHg | 124.6 (± 18.1) mmHg | | Diastolic Blood Pressure | 71.2 (± 7.9) mmHg | 74.1 (± 11.5) mmHg | Table 1. Demographics of Cohort Test datasets were strictly segregated from algorithm training datasets. The primary performance metric is the reference-scaled individual equivalence coefficient (IEC, Barnhart, 2007) across three human readers and one automated read performed by Us2.v1. We define our success criterion as a non-inferiority margin, Δ= 0.25, such that Us2.v1 is deemed interchangeable with the reference if IEC + 1.96*SD(IEC) < Δ. Compared to reference standard echocardiographic human measurements made in triplicate by the independent Cardiovascular Imaging Core Laboratory at Brigham and Women's Hospital, the 95% confidence intervals of automated Us2.v1 measurements all fell below the pre-specifiednoninferiority margin of 0.25 for IEC. {8}------------------------------------------------ Based on the performance as documented in the sum of this testing, the Us2.v1 device has a safety and effectiveness profile that is similar to the predicate EchoMD Automated Ejection Fraction Software device. ## 5.5 Substantial Equivalence Conclusion Us2.v1 is an image processing software which has similar intended use and indications for use statement as the predicate device. The two devices have similar technological characteristics: both use machine learning algorithms to automate the measurement of transthoracic cardiac images. Though the subject device provides more measurements than the predicate, this does not in and of itself produce different questions of safety and effectiveness. This 510(k) submission includes information on the Us2.v1 technological characteristics, as well as performance data and verification and validation activities demonstrating that Us2.v1 is as safe and effective as the predicate, and does not raise different questions of safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...