syngo.via MI Workflows, syngo MBF

K201195 · Siemens Medical Solutions USA, Inc. · QIH · Nov 18, 2020 · Radiology

Device Facts

Record IDK201195
Device Namesyngo.via MI Workflows, syngo MBF
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeQIH · Radiology
Decision DateNov 18, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

syngo.via molecular imaging (MI) workflows comprise medical diagnostic applications for viewing, manipulation, quantification, analysis and comparison of medical images from single or multiple imaging modalities with one or more time-points. These workflows support functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR), syngo.via MI workflows can perform harmonization of SUV (PET) across different PET systems or different PET reconstruction methods. syngo.via MI workflows are intended to be utilized by appropriately trained health care professionals to aid in the management of diseases, including those associated with oncology, neurology, and organ function. The images and results produced by the syngo.via MI workflows can also be used by the physician to aid in radiotherapy treatment planning.

Device Story

Software-only medical device (syngo.via MI Workflows VB50A) installed on syngo.via platform; processes PET, NM, CT, and MR images. Features include Auto Lung 3D, dynamic summing, automatic layouts, auto ranges, cardiac masking, and FDOPA normals database. Used by clinicians in clinical settings for disease management (oncology, cardiology, neurology) and radiotherapy planning. Transforms raw imaging data into quantitative/visual outputs (e.g., SUV harmonization, myocardial blood flow measurements, striatal/cortical analysis). Clinicians use outputs to assess pathologies like dementia, movement disorders, and epilepsy; aids in clinical decision-making by providing standardized, comparable multimodality data.

Clinical Evidence

Bench testing only. Verification and validation activities confirmed software functions as designed, met performance specifications, and implemented all hazard mitigations. No clinical data presented.

Technological Characteristics

Software-only; runs on syngo.via platform. Standards: IEC 62304 (software lifecycle), ISO 14971 (risk management), NEMA PS 3.1-3.20 (DICOM), ISO 15223-1, IEC 62366-1. Features include automated segmentation, quantification, and database comparison algorithms.

Indications for Use

Indicated for trained healthcare professionals to manage diseases (oncology, cardiology, neurology, organ function) via viewing, manipulation, quantification, and analysis of PET, NM, CT, and MR images. Supports radiotherapy treatment planning.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 18, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The seal is a stylized depiction of an eagle and a caduceus, while the FDA acronym and name are written in blue, with the word "Administration" appearing below the acronym and name. Siemens Medical Solutions USA, Inc. % Mr. Clayton Ginn Regulatory Technical Specialist 2501 N. Barrington Road HOFFMAN ESTATES IL 60192 Re: K201195 Trade/Device Name: syngo.via MI Workflows VB50A Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QIH, LLZ Dated: October 8, 2020 Received: October 9, 2020 Dear Mr. Ginn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K201195 Device Name syngo.via MI Workflows VB50A #### Indications for Use (Describe) syngo.via molecular imaging (MI) workflows comprise medical diagnostic applications for viewing, manipulation, quantification, analysis and comparison of medical images from single or multiple imaging modalities with one or more time-points. These workflows support functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR), syngo.via MI workflows can perform harmonization of SUV (PET) across different PET systems or different PET reconstruction methods. syngo.via MI workflows are intended to be utilized by appropriately trained health care professionals to aid in the management of diseases, including those associated with oncology, neurology, and organ function. The images and results produced by the syngo.via MI workflows can also be used by the physician to aid in radiotherapy treatment planning. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 23 {3}------------------------------------------------ # 510(k) Summary as required by 21 CFR Part 807.87(h) K201195 ## Identification of the Submitter | | Primary Contact: | Alternate Contact: | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Clayton Ginn<br>Regulatory Technical<br>Specialist<br>Siemens Medical Solutions<br>USA, Inc.<br>810 Innovation Drive<br>Knoxville, TN 37932 | Veronica Padharia<br>Regulatory Technical<br>Specialist<br>Siemens Medical Solutions<br>USA, Inc.<br>2501 N. Barrington Road<br>Hoffman Estates, IL 60192 | | Telephone Number: | (865) 898-2692 | (630) 877-5761 | | Fax Number: | (865) 218-3227 | (847) 304-6023 | | Name / Address of<br>Manufacturer | Siemens Medical Solutions USA, Inc<br>Molecular Imaging<br>2501 N. Barrington Road<br>Hoffman Estates, IL 60192<br>USA | | | Date of Submission: | May 1st, 2020 | | | Identification of the product | | | | Device Proprietary Name: | syngo.via MI Workflows VB50A | | | Common Name: | Image Processing Software | | | Code of Federal<br>Regulations | 21 CFR 892.2050 | | | Classification Name: | Picture Archiving and Communication System | | | Product Code: | QIH, LLZ | | | Classification Panel: | Radiology | | | Device Class: | Class II | | | Device Proprietary Name: | syngo.via MI Workflows VB40A | | | Common Name: | Image Processing Software | | | Code of Federal Regulations | 21 CFR 892.2050 | | | Classification Name: | Picture Archiving and Communication System | | | Product Code: | LLZ | | | Classification Panel: | Radiology | | | Device Class: | Class II | | | Manufacturer: | Siemens Medical Solutions USA, inc. | | | 510(k) Number: | K191309 (July 2019) | | | Secondary Predicate Device | | | | Device Proprietary Name: | syngo MBF 2.0 | | | Common Name: | Image Processing Software | | | Code of Federal Regulations | 21 CFR 892.2050 | | | Classification Name: | Picture Archiving and Communication System | | | Product Code: | LLZ | | | Classification Panel: | Radiology | | | Device Class: | Class II | | {4}------------------------------------------------ #### Primary Predicate Device 510(k) Number: synqo.via MI Workflows VB40A is deemed the primary predicate device due to it being the most similar to the device under review of this submission with respect to indications for use and technical characteristics. syngo MBF 2.0 is considered a secondary predicate device because it is a subset of the primary predicate device's Indications for Use and technical characteristics included within this submission. K110494 (April 2011) {5}------------------------------------------------ #### Device Description syngo.via MI Workflows is a software-only medical device which will be delivered on CD-ROM / DVD to be installed onto the commercially available Siemens syngo.via software platform (K191040) by trained service personnel. syngo.via molecular imaging (MI) workflows comprise medical diagnostic applications for viewing, manipulation, quantification, analysis and comparison of medical images from single or multiple imaging modalities with one or more time-points. These workflows support functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR). synqo.via MI Workflows enable visualization of information that would otherwise have to be visually compared disjointedly. syngo.via MI workflows are intended to be utilized by appropriately trained health care professionals to aid in the management of diseases, including those associated with oncology, cardiology, neurology, and organ function. The images and results produced by the syngo.via MI workflows can also be used by the physician to aid in radiotherapy treatment planning. Scenium is a previously cleared software device (K191309) that assists in the display and analysis of images within the MI Neurology workflow of syngo.via MI Workflows. This software enables visualization and appropriate rendering of multimodality data, providing a number of features which enable the user to process acquired image data. Scenium consists of four workflows: - Database Comparison - - -Striatal Analysis - -Cortical Analysis - -Subtraction The Scenium workflows are used to assist the clinician with the visual evaluation, assessment and quantification of pathologies, such as dementia (i.e., Alzheimer's), movement disorders (i.e., Parkinson's) and seizure analysis (i.e., Epilepsy). syngo MBF is a software only product intended for visualization, assessment and quantification of medical images: specifically providing quantitative blood flow measurements of PET images. The software sites within the MI Cardiology workflow within syngo.Via MI Workflows. The application supports dynamic Rubidium - PET and dynamic Ammonia – PET images. The application provides visualization and measurement tools, for qualitative and quantitative visualization and assessment of the input data. It provides automatic and manual tools to orient and segment the myocardium. The software calculates measurements of myocardial blood flow, and provides tools, such as a database comparison workflow, for the Clinician to assess these results. {6}------------------------------------------------ The modifications to synqo.via MI Workflows and synqo MBF (within MI Cardiology) software (K191309 and K110494) include the following new features: | Workflow | Workflow-specific Features | |--------------------------------------------|----------------------------------------------------| | MM Oncology | No New Features | | MI General (MI Reading / SPECT Processing) | Auto Lung 3D | | | Dynamic Summing | | | Automatic Layouts | | | Auto Ranges | | | Scanner Data Support for Organ Processing | | | Usability Improvements | | MI Cardiology | Normalization by Rate-Pressure Product (syngo MBF) | | | Cardiac Masking | | | Cardiac Auto Ranges | | | Usability Improvements | | | Updates / redeployment to third party software | | MI Neurology | Normals Database for FDOPA (Scenium) | ### Technological Characteristics The syngo.via MI Workflows VB50A software modifications are based on the commercially available syngo.via MI Workflows VB40A (K191309) and syngo MBF 2.0 (K110494) software. The features introduced into these Clinical Applications do not alter the already existent technological characteristics within the commercially available predicate system. syngo.via MI Workflows is intended to be run on the Siemens syngo.via software platform (K191040) either alone or with other advanced commercially cleared applications. ### Intended Use An individual software program, or group of programs, routines, or algorithms that add specific image processing and/or analysis capabilities to a positron emission tomography (PET) and Single Photon Emission Computed Tomography (SPECT) imaging system configuration. A basic set of application programs and routines is included with such computer controlled imaging systems and they can be upgraded to correct programming errors or to add new system capabilities. Some application software routines or groups of routines (packages) must be combined with specific hardware or firmware accessories or configurations in order to function as intended. Application program packages are typically identified by a proprietary name and "version" or "upgrade" number. The Intended Use for syngo.via MI Workflows is the same and, compared to the primary and reference devices, has not changed. {7}------------------------------------------------ #### Indications for Use syngo.via molecular imaging (MI) workflows comprise medical diagnostic applications for viewing, manipulation, quantification, analysis and comparison of medical images from single or multiple imaging modalities with one or more time-points. These workflows support functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR). syngo.via MI workflows can perform harmonization of SUV (PET) across different PET systems or different PET reconstruction methods. syngo.via MI workflows are intended to be utilized by appropriately trained health care professionals to aid in the management of diseases, including those associated with oncology, cardiology, neurology, and organ function. The images and results produced by the syngo.via MI workflows can also be used by the physician to aid in radiotherapy treatment planning. The syngo.via MI Workflows VB50A Indications for Use has been re-worded for clarification and does not expand the Indications for Use or the Intended Use of the commercially available predicate devices. The updated Indications for Use encompasses the Scenium and syngo MBF devices as used within the MI Neurology and MI Cardiology workflows. ### Performance Testing / Safety and Effectiveness The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk Management has been ensured via risk analyses in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards for development including EN ISO 13485 and IEC 62304. Cybersecurity information in accordance with FDA Guidance documents issued October 2. 2014 has been provided. The Clinical Applications software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the Clinical Applications software and external devices. Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing. The device is designed and manufactured in accordance with Quality System Regulations as outlined in 21 CFR 820. The FDA recognized standards are listed as follows: - Recognition Number 13-79: IEC 62304 Edition 1.1 2015-06 - Recognition Number 12-300: NEMA PS 3.1 3.20 (2016) . {8}------------------------------------------------ - Recognition Number 5-40: ISO 14971:2007 Second Edition . - Recognition Number 5-114: IEC 62366-1 Edition 1.0 2015 . - Recognition Number 5-117: ISO 15223-1 Third Edition 2016 ● #### Statement Regarding Substantial Equivalence: There are no differences in the Indications for Use, Intended Use or Fundamental Technological Characteristics of the syngo.via MI Workflows VB50A software as compared to the currently commercially available syngo.via MI Workflows software (K191309). The Indications for Use has been reworded and does not expand the Indications for Use or Intended Use of the commercially available predicate devices. Both the current and predicate devices are used for viewing, manipulation, quantification, analysis and comparison of medical images from single or multiple imaging modalities with one or more time-points. Additionally, the new features implemented within this release do not raise any new issues of safety and effectiveness as compared to the predicate device. Based on this information, as well as the documentation in support of the modifications, it is Siemens' opinion that the syngo.via MI Workflows software-with the modifications outlined in this application-is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...