EchoGo Core (2.0)
K213275 · Ultromics Limited · QIH · Dec 20, 2021 · Radiology
Device Facts
| Record ID | K213275 |
| Device Name | EchoGo Core (2.0) |
| Applicant | Ultromics Limited |
| Product Code | QIH · Radiology |
| Decision Date | Dec 20, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.2050 |
| Device Class | Class 2 |
| Attributes | AI/ML, Software as a Medical Device |
Intended Use
EchoGo Core is intended to be used for quantification and reporting of results of cardiovascular function to support physician diagnosis. EchoGo Core is indicated for use in adult populations.
Device Story
EchoGo Core 2.0 is a software application for automated quantification of echocardiographic metrics. Input: DICOM ultrasound images of heart. Operation: AI-based analysis performs automatic view classification, LV contouring, and heart cycle identification (ED/ES frame selection). Workflow: Operator confirms/relabels views, approves/rejects contours and frames. Output: Secondary capture DICOM files and/or structured DICOM reports containing LV function metrics (EF, volumes, GLS, SLS, stroke volume, cardiac output). Used in clinical settings by sonographers/echocardiographers to assist physician diagnosis. Benefits: Provides standardized, automated measurements of cardiac function to support clinical decision-making.
Clinical Evidence
Retrospective, non-interventional validation study using 214 unseen TTE studies. Primary endpoint: Root Mean Square (RMS) error against reference values from TomTec Arena TTA2. Results: RMS error for LV length (3.06-4.59%), volume (8.57-16.59%), EF (6.69-8.50%), stroke volume (10.57-13.68%), GLS (3.36-4.79%), and segmental strain (5.51-9.98%). Inter/intra-operator variability showed Deming RMSE of 0. Data representative of adult population across gender, age, weight, and height ranges.
Technological Characteristics
Software-only device; Windows-based. Complies with IEC 62304:2015 (software lifecycle), ISO 14971:2019 (risk management), IEC 62366-1:2020 (usability), and NEMA PS 3.1-3.20 (DICOM). Uses AI/ML for automated view classification and LV contouring. Connectivity: DICOM-based input/output. Developed under ISO 13485:2016 QMS.
Indications for Use
Indicated for quantification and reporting of cardiovascular function to support physician diagnosis in adult populations.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Reference Devices
- TomTec Arena TTA2 (K150122)
Related Devices
- K191171 — EchoGo Core · Ultromics, Ltd. · Nov 13, 2019
- K210791 — Us2.v1 · Eko.Ai Pte Ltd. D/B/A Us2.Ai · Jul 27, 2021
- K220956 — Libby Echo:Prio · Dyad Medical, Inc. · Jul 20, 2022
- K241430 — EchoMeasure · Icardio.Ai · Oct 10, 2024
- K252105 — Ligence Heart · Ligence Uab · Sep 26, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA acronym and full name are written in blue.
Ultromics Limited % Jaco Jacobs Chief Regulatory and Compliance Officer 4630 Kingsgate, Cascade Way Oxford Business Park South Oxford, Oxfordshire OX4 2SU UNITED KINGDOM
December 20, 2021
## Re: K213275
Trade/Device Name: EchoGo Core (2.0) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: September 27, 2021 Received: September 30, 2021
Dear Jaco Jacobs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K213275
Device Name
EchoGo Core (2.0)
Indications for Use (Describe)
EchoGo Core is intended to be used for quantification and reporting of results of cardiovascular function to support physician diagnosis. EchoGo Core is indicated for use in adult populations.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "ULTROMICS" in bold, dark blue letters. To the left of the word is a circular logo with a gradient of blue and green. The logo appears to be a stylized letter "O" or a ring.
# K213275
510(K) SUMMARY
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ultromics. The logo consists of a circular graphic on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular graphic is a gradient of blue and green, creating a swirling effect.
### 1 Submitter
| Company | Ultromics Limited<br>4630 Kingsgate Cascade Way, Oxford Business Park South, Oxford,<br>Oxfordshire, United Kingdom, OX4 2SU |
|---------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact | Dr. Jaco Jacobs |
| Date Prepared | 27 November 2021 |
# 2 Subject Device
| Product Trade Name | EchoGo Core |
|--------------------|--------------------------------------------------------|
| Model Number | 2.0 |
| Medical Speciality | Radiology |
| Regulation | 892.2050 – Picture Archiving and Communications System |
| Product Code | QIH - Automated Radiological Image Processing Software |
| Regulatory Class | II |
EchoGo Core is the product trade name and 2.0 is the model number. For the avoidance of doubt, in this submission we combine the product trade name and model number and refer to the subject device as EchoGo Core 2.0.
#### 3 Predicate and Reference Devices
| Predicate Device | EchoGo Core |
|------------------|-------------------|
| 510(k) | K191171 |
| Manufacturer | Ultromics Limited |
| Reference Device | TomTec Arena TTA2 |
|------------------|-----------------------------|
| 510(k) | K150122 |
| Manufacturer | TomTec Imaging Systems GmbH |
#### Device Description বা
EchoGo Core 2.0 is a software application manufactured by Ultromics to provide a report of left ventricular cardiac function, in the form of secondary capture DICOM files and/or as a structured DICOM report, to aid interpreting physicians with diagnostic decision-making process. EchoGo Core 2.0 applies to ultrasound images of the heart (echocardiograms).
EchoGo Core 2.0 utilizes artificial intelligence (AI) for the operator-assisted automatic quantification of commonly measured echocardiographic metrics . Independent training, test and validation datasets were used for training and performance assessment of the device.
## ECHOGO CORE 2.0
## 510(k) SUMMARY
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular, gradient-colored icon on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The icon has a swirling design with shades of green and blue, creating a sense of movement or flow.
EchoGo Core 2.0 requires an operator at key steps to confirm or relabel automatically labeled acquisition views (if required) and approve the left ventricle segmentations (contours) proposed by the Al.
EchoGo Core 2.0 operates in a sequential workflow manner, that includes:
- 1. Automatic view classification and labeling
- 2. Operator confirmation or relabeling of view classifications
- 3. Contouring of the left ventricle in apical chamber views
- 4. Automated heart cycle identification and end-diastolic and end-systolic frame selection
- 5. Operator confirmation or rejection of left ventricle contouring and frame selection
- 6. Calculation of clinical report metrics
To identify a heart cycle, the heart rate must be known. If an ECG was not taken during acquisition of the echocardiogram, the heart rate may be inferred. The operator is selecting a contoured image clip for each view and approving the selected cycle (and the key end-diastolic (ED) and end-systolic (ES) frames). The operator will review the report produced and may be asked to approve cautions that are added to the report. EchoGo Core 2.0 supports the production of reports for any combination of the three processible views (A2C, A3C and A4C), either embedded as image files in one or more multi-frame true color secondary capture DICOM files, and/or as a DICOM structured report.
# 5 Indications for Use
EchoGo Core 2.0 is intended to be used for quantification and reporting of results of cardiovascular function to support physician diagnosis. EchoGo Core 2.0 is indicated for use in adult populations.
Both the intended use and indications for use statement for the subject and predicate devices are identical. Both the subject and predicate devices are indicated for use in quantification and reporting of cardiovascular function to support the physician's diagnosis in an adult population.
#### Comparison of Technological Characteristics o
A comparison of technological differences between the subject and predicate devices follows.
| Characteristic | Subject Device<br>EchoGo Core 2.0 | Predicate Device<br>EchoGo Core 1.0 |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Identical | Quantification of cardiovascular function from an echocardiogram |
| Indications for use | Identical | Quantification and reporting of results of cardiovascular function<br>to support physician diagnosis.<br>EchoGo Core is indicated for use<br>in adult populations |
| Anatomical site | Identical | Cardiovascular structures |
| Users | Identical | Accredited echocardiographers<br>and sonographers at Ultromics |
| Machine learning-based algorithm | Yes | Yes |
| Algorithms | Auto-Contouring DICOM Handling | Auto-Contouring DICOM Handling |
| Characteristic | Subject Device<br>EchoGo Core 2.0 | Predicate Device<br>EchoGo Core 1.0 |
| | • Technical QC | • Technical QC |
| | • Auto-View Classification | • Manual Labelling UI |
| | • Manual Labelling UI | • Contour Selection and<br>Approval |
| | • Contour Selection and<br>Approval | • Metrics Calculation |
| | • Metrics Calculation | • Auditing |
| | • Auditing | • Reporting |
| | • Service Webhooks | |
| | • Reporting | |
| Operates on DICOM clips | Yes | Yes |
| Average GLS A2C/A4C views | Yes | Yes |
| Average GLS A2C/A4C/A3C views | Yes | No |
| Per view GLS, EF and LV volume | Yes | No |
| Segmental LS | Yes | No |
| Echocardiogram images on device<br>report | Yes | Yes |
| Horseshoe and bullseye<br>representations | Yes | No |
| Mechanical dispersion plot | Yes | No |
| Strain temporal intensity<br>plot/CAMM plot | Yes | No |
| Contours overlayed on<br>echocardiogram images | Yes | No |
| Auto-view classification | Yes | No |
| Automatic ED/ES frame detection | Yes | Yes |
| EF Reported | Yes | Yes |
| Automated EF calculation | Yes | Yes |
| Simpson's biplane LV volume<br>calculation | Yes | Yes |
| Non-Simpson's bi-plane LV<br>volume calculation methods | Yes | No |
| LV ED/ES volume indexed values | Yes | No |
| EF results shown with video clip | Yes | Yes |
| User confirmation / rejection of<br>result | Yes | Yes |
| Manual editing of automated<br>result by user | No | No |
| LV stroke volume | Yes | No |
| LV length | Yes | No |
| Operating system | Windows | Windows |
| Operating platform | Single software application<br>platform | Single software application<br>platform |
| Software | Complies with IEC 62304:2015<br>and developed under an EDA QSR | Complies with IEC 62304:2006<br>and developed under an EDA QSR |
| Characteristic | Subject Device<br>EchoGo Core 2.0 | Predicate Device<br>EchoGo Core 1.0 |
| | compliant QMS incorporating risk<br>management per ISO 14971:2019 | compliant QMS incorporating risk<br>management per ISO<br>14971:2007 |
| Usability | Complies with IEC 62366-1:2020<br>and general use of FDA guidance<br>on usability engineering | General use of FDA guidance on<br>usability engineering |
| Performance testing | Equivalent to reference<br>comparator TomTec Arena TTA2<br>(K150122) | Equivalent to reference<br>comparator TomTec Arena TTA2<br>(K150122) |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular graphic on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular graphic is a gradient of blue and green, creating a swirling effect.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular icon on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular icon is a gradient of blue and green, creating a swirling effect.
Both the subject and predicate devices have similar technological characteristics and principles of operation. Any technological differences between the subject and predicate devices raise no new concerns with regards to safety and efficacy.
#### Consensus Standards 7
The following consensus standards were used in the design and manufacture of EchoGo Core 2.0.
| Standard | Recognition<br>Number |
|-------------------------------------------------------------------------------------------------|-----------------------|
| ISO 14971:2019 – Medical Devices – Application of Risk Management to Medical Devices | 5-125 |
| IEC 62304:2015 – Medical Device Software - Software Life Cycle Processes | 13-79 |
| IEC 62366-1:2020 – Medical Devices – Application of Usability Engineering to Medical<br>Devices | 5-129 |
| NEMA PS 3.1 – PS 3.20 (2016) – Digital Imaging and Communications in Medicine (DICOM)<br>Set | 12-300 |
| IEC ISO 10918-1:1994 – Digital Compression and Coding of Continuous-tone Still Images | 12-261 |
Ultromics Limited cites conformity to the voluntary standards above. In addition, EchoGo Core 2.0 was designed and manufactured under a QMS that fully conforms to ISO 13485:2016.
# 8 Performance Data
#### 8.1 Software Verification and Validation
EchoGo Core software was developed and tested in accordance with Ultromics' Design Control processes and has been subjected to extensive safety and performance testing. Non-clinical verification and validation test results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at unit, module, and system integration levels. Risk management analysis generated multiple risk mitigation measures and verification activities. Formative usability assessments were conducted during formal development. Cybersecurity and data security testing were conducted to verify that data and patient protected health information security measures are included in the design of the software.
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device is considered as a moderate level of concern, since a failure or latent design flaw could indirectly result in minor injury to the patient through incorrect or delayed information or through the action of a care provider.
## 510(k) SUMMARY
{8}------------------------------------------------
EchoGo Core 2.0 passed all software verification and validation tests.
#### Essential Performance 8.2
Performance testing was conducted to demonstrate substantial equivalence to the predicate device, EchoGo Core 1.0, via a reference comparator device, TomTec Arena TTA2. Pre-determined acceptance criteria were formalised in terms of Root Mean Square (RMS) error against reference values generated using the comparator device, for each reported metric. The acceptance criteria were formalised such that EchoGo Core 2.0 would produce measures of left ventricular (LV) length, volume at end diastole (ED), and end systole (ES), ejection fraction (EF), stroke volume, cardiac output, global longitudinal strain (GLS) and segmental longitudinal strain (SLS) with an RMS error below a set, pre-determined threshold.
A formal retrospective, non-interventional validation study was conducted using 214 previously unseen studies. Left ventricle volume and metrics subsequently derived, such as ejection fraction, were calculated. For global longitudinal strain and segmental strain at peak and end-systole, LV contour Euclidean distances were calculated. The performance dataset used was representative of the intended patient population over a clinically representative range of measures. The population was gender-balanced over clinically typical age, weight, and height ranges. Performance testing included acquisitions from multiple ultrasound system manufacturers to validate essential performance across multiple scanner platforms and settings. Test datasets were strictly segregated from algorithm training datasets. Root mean square error was calculated as the primary endpoint. The primary endpoint was met. Performance against the comparator device is summarised as follows.
| Left Ventricular Metric | % Root Mean Square Error |
|----------------------------------------|--------------------------|
| Length | 3.06 - 4.59 |
| Volume at End Diastole and End Systole | 8.57 - 16.59 |
| Ejection Fraction | 6.69 - 8.50 |
| Stroke Volume | 10.57 - 13.68 |
| Global Longitudinal Strain | 3.36 - 4.79 |
| Systolic Segmental Longitudinal Strain | 5.51 - 9.98 |
Three operators processed a total of 214 TTE studies, respectively. Inter-operator variability demonstrated acceptable performance with a Deming RMSE of 0 for all report metrics. Similarly, intra-operator variability demonstrated acceptable performance with a Deming RMSE of 0 for all report metrics. The repeatability results for EchoGo Core 2.0 demonstrated acceptable performance.
All measurements produced by EchoGo Core 2.0 were deemed to be substantively equivalent to the predicate device and met acceptable levels of performance. We therefore consider EchoGo Core 2.0 to be substantively equivalent to the predicate device and is therefore deemed to be safe and effective.
# 9 Conclusions
The subject device, EchoGo Core 2.0 is as safe and as effective as the predicate device, EchoGo Core 1.0, previously cleared under K191171. The subject and predicate devices have identical intended uses and indications for use, and similar technological characteristics and principles of operation. Any minor differences raise no new concerns with regards to safety and effectiveness. Essential performance of both the subject and predicate devices were evaluated against a comparator reference device, TomTec Arena TTA2, previously
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular graphic on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular graphic is a gradient of blue and green, creating a sense of depth and movement.
cleared under K150122. Performance data demonstrate that EchoGo Core 2.0 and EchoGo Core 1.0 have similar essential performance.
It follows that EchoGo Core 2.0 is substantially equivalent to EchoGo Core 1.0.