Axial3D Insight
K250369 · Axial Medical Printing Limited · QIH · Sep 18, 2025 · Radiology
Device Facts
| Record ID | K250369 |
| Device Name | Axial3D Insight |
| Applicant | Axial Medical Printing Limited |
| Product Code | QIH · Radiology |
| Decision Date | Sep 18, 2025 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.2050 |
| Device Class | Class 2 |
| Attributes | AI/ML, Software as a Medical Device, PCCP |
Intended Use
Axial3D Insight is intended for use as a cloud-based service and image segmentation framework for the transfer of DICOM imaging information from a medical scanner to an output file. The Axial3D Insight output file can be used for the fabrication of physical replicas of the output file using additive manufacturing methods. The output file or physical replica can be used for treatment planning. The output file or the physical replica can be used for diagnostic purposes in the field of trauma, orthopedic, maxillofacial and cardiovascular applications. Axial3D Insight should be used in conjunction with other diagnostic tools and expert clinical judgment.
Device Story
Axial3D Insight is a cloud-based image processing and segmentation framework. It ingests DICOM imaging (CT/CTA) from medical scanners. The device uses 'AxialML' machine learning models to generate initial segmentations, which are then refined and validated by trained staff to produce a final 3D digital model or physical 3D-printed replica. Used in hospital settings by clinicians and medical visualization engineers for preoperative planning, implant selection, and surgical communication. The output serves as an adjunct to standard DICOM review, aiding in complex anatomical visualization to improve surgical treatment planning and patient outcomes.
Clinical Evidence
Bench testing only. Software verification and validation performed per FDA guidance. Clinical segmentation performance validated by 3 radiologists across 12 cases (orthopedic, trauma, maxillofacial, cardiovascular) using RADPEER scoring. Intended use validation conducted with 9 physicians reviewing 12 cases, confirming device outputs meet end-user needs. Machine learning models validated on independent datasets (Cardiac: 4,838; Neuro: 4,041; Ortho: 10,857; Trauma: 19,134 images).
Technological Characteristics
Cloud-based software platform accessed via web browsers (Edge, Safari, Chrome). Supports CT/CTA modalities. Features include image registration, segmentation (automated + manual editing), volume rendering, and quantification. Machine learning models are locked prior to release; no adaptive/continuous learning in the field. Software updates are cloud-hosted.
Indications for Use
Indicated for patients requiring 3D anatomical modeling for treatment planning or diagnostic support in trauma, orthopedic, maxillofacial, and cardiovascular applications. Used in conjunction with DICOM imaging and expert clinical judgment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
- Axial3D Insight (K232841)
Related Devices
- K222745 — Axial3D Insight · Axial Medical Printing Limited · Jul 3, 2023
- K232841 — Axial3D Insight · Axial Medical Printing Limited · Nov 15, 2023
- K221511 — Axial3D Cloud Segmentation Service · Axial Medical Printing Limited · Jun 23, 2022
- K231607 — Axial3D Cloud Segmentation Service · Axial Medical Printing Limited · Jan 24, 2024
- K250237 — InferOperate Suite · Beijing Infervision Healthcare Medical Technology Co., Ltd. · Sep 15, 2025
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Axial Medical Printing Limited
% May Lee
Senior Consultant
CS Lifesciences Ltd
Suite 10, Dunswood House
1 Dunswood Road, Cumbernauld
Glasgow, G67 3EN
United Kingdom
September 18, 2025
Re: K250369
Trade/Device Name: Axial3D Insight
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: August 19, 2025
Received: August 19, 2025
Dear May Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250369 - May Lee
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{2}
K250369 - May Lee
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250369
Device Name
Axial3D Insight
Indications for Use (Describe)
Axial3D Insight is intended for use as a cloud-based service and image segmentation framework for the transfer of DICOM imaging information from a medical scanner to an output file.
The Axial3D Insight output file can be used for the fabrication of physical replicas of the output file using additive manufacturing methods. The output file or physical replica can be used for treatment planning.
The output file or the physical replica can be used for diagnostic purposes in the field of trauma, orthopedic, maxillofacial and cardiovascular applications. Axial3D Insight should be used in conjunction with other diagnostic tools and expert clinical judgment.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
axial3D Patient data made real
K250369
Traditional 510(k) Notification
# 1. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CRF 807.92.
Date Prepared: September 17, 2025
# 2. Applicant Information
Axial Medical Printing Limited
17A Ormeau Avenue
Belfast BT2 8HD
United Kingdom
Tel: +44 (0)28 90183590
# 3. Contact Person
May Lee, Senior Consultant
Suite 10, Dunnswood House
1 Dunnswood Road,
Cumbernauld, Glasgow G67 3EN
United Kingdom
Email: may@cslifesciences.com
# 4. Device Information
Trade Name: Axial3D Insight
Common Name: Automated Radiological Image Processing Software
Classification number: 892.2050
Regulatory Class: II
Product Code: QIH
# 5. Predicate Device
Table 5 - Predicate Device
| Name | Manufacturer | 510(k)# |
| --- | --- | --- |
| Axial3D Insight | Axial Medical Printing Limited | K232841 |
{5}
axial3D Patient data made real
K250369 Traditional 510(k) Notification
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
# 6. Device Description
Axial3D Insight is a secure, highly available cloud-based image processing, segmentation and 3D modelling framework for the transfer of imaging information either as a digital file or as a 3D printed physical model.
# a. Indications for Use
Axial3D Insight is intended for use as a cloud-based service and image segmentation framework for the transfer of DICOM imaging information from a medical scanner to an output file.
The Axial3D Insight output file can be used for the fabrication of physical replicas of the output file using additive manufacturing methods.
The output file or physical replica can be used for treatment planning.
The output file or the physical replica can be used for diagnostic purposes in the field of trauma, orthopedic, maxillofacial, and cardiovascular applications.
Axial3D Insight should be used with other diagnostic tools and expert clinical judgment.
# 7. Comparison of Intended Use to Predicate and Reference Devices
Table 2 – Predicate Device Comparison: Intended Use
| Attribute | Axial3D Insight (Proposed Device) | Axial3D Insight (Predicate Device) | Comparison |
| --- | --- | --- | --- |
| Device Manufacturer | Axial Medical Printing Limited | Axial Medical Printing Limited | N/A |
| Device Name | Axial3D Insight | Axial3D Insight | N/A |
| Device Trade or Proprietary Name | Axial3D Insight | Axial3D Insight | N/A |
| 510(k) Number | K250369 | K232841 | N/A |
| Device Regulation Name | Automated Radiological Image Processing Software | Automated Radiological Image Processing Software | Equivalent |
| Device Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Equivalent |
| Device Product Code | QIH | QIH | Equivalent |
{6}
axial³D
Patient data made real
K250369
Traditional 510(k) Notification
| Attribute | Axial3D Insight (Proposed Device) | Axial3D Insight (Predicate Device) | Comparison |
| --- | --- | --- | --- |
| Device Classification FDA | Class II | Class II | Equivalent |
| Indication for Use | Axial3D Insight is intended for use as a cloud-based service and image segmentation framework for the transfer of DICOM imaging information from a medical scanner to an output file.
The Axial3D Insight output file can be used for fabrication of physical replicas of the output file using additive manufacturing methods.
The output file or physical replica can be used for treatment planning.
The output file or physical replica can be used for diagnostic purposes in the field of trauma, orthopedic, maxillofacial, and cardiovascular applications.
Axial3D Insight should be used in conjunction with other diagnostic tools and expert clinical judgement. | Axial3D Insight is intended for use as a cloud-based service and image segmentation framework for the transfer of DICOM imaging information from a medical scanner to an output file.
The Axial3D Insight output file can be used for fabrication of physical replicas of the output file using additive manufacturing methods.
The output file or physical replica can be used for treatment planning.
The output file or physical replica can be used for diagnostic purposes in the field of trauma, orthopedic, maxillofacial, and cardiovascular applications.
Axial3D Insight should be used in conjunction with other diagnostic tools and expert clinical judgement. | Equivalent |
| Intended Use | Axial Medical Printing Limited, Axial3D Insight provides patient-specific 1:1 scale replica models, either as a digital file or as a 3D printed physical model.
The digital file or 3D printed physical model is intended to be used in conjunction with the DICOM images and expert clinical judgement. The applications for using the physical 3D printed model as a presurgical planning tool are as follows:
Preoperative planning of surgical treatment options including planning for surgical instruments, aiding decisions on implants, and aiding the surgical treatment plan., All planning using the 3D replica model should be carried out with the assistance of the DICOM images
Communication with the surgical team to discuss the surgical | Axial Medical Printing Limited, Axial3D Insight provides patient-specific 1:1 scale replica models, either as a digital file or as a 3D printed physical model.
The digital file or 3D printed physical model is intended to be used in conjunction with the DICOM images and expert clinical judgement. The applications for using the physical 3D printed model as a presurgical planning tool are as follows:
Preoperative planning of surgical treatment options including planning for surgical instruments, aiding decisions on implants, and aiding the surgical treatment plan., All planning using the 3D replica model should be carried out with the assistance of the DICOM images | Equivalent |
3
{7}
axial3D Patient data made real
K250369
Traditional 510(k) Notification
| Attribute | Axial3D Insight (Proposed Device) | Axial3D Insight (Predicate Device) | Comparison |
| --- | --- | --- | --- |
| | Communication with the surgical team to discuss the surgical treatment plan in conjunction with DICOM images
Communication with the patient to discuss the surgical treatment plan in conjunction with DICOM images
Education tool for surgical planning.
The 3D printed physical model can be used for surgical planning in the following applications: orthopedics, trauma, maxillofacial, and cardiac surgery. | treatment plan in conjunction with DICOM images
Communication with the patient to discuss the surgical treatment plan in conjunction with DICOM images
Education tool for surgical planning.
The 3D printed physical model can be used for surgical planning in the following applications: orthopedics, trauma, maxillofacial, and cardiac surgery. | |
| Method of Use | Used in conjunction with other diagnostic tools and expert clinical judgment. | Used in conjunction with other diagnostic tools and expert clinical judgment. | Equivalent |
| Use Environment | Hospital | Hospital | Equivalent |
| OTC or Prescription Device | Prescription Use | Prescription Use | Equivalent |
| V&V | Complies with FDA Guidance Requirement | Complies with FDA Guidance Requirement | Equivalent |
| PCCP | Will have access to approved PCCP | Will have access to approved PCCP | Different – the premise of this 510k is to have the PCCP approved |
# 8. Comparison of Technological Characteristics to the Predicate Device and Reference Device
Table 3 Predicate Comparison: Technology
| Attribute | Axial3D Insight (Proposed Device) | Axial3D Insight (Predicate Device) | Comparison |
| --- | --- | --- | --- |
| Method of Use | software interface | software interface | Equivalent |
| Computer Platform and Operating System | Microsoft Edge (v104), Safari(v15), Chrome (v103), or equivalent | Microsoft Edge (v104), Safari(v15), Chrome (v103), or equivalent | Equivalent |
{8}
axial3D Patient data made real
K250369
Traditional 510(k) Notification
| Attribute | Axial3D Insight (Proposed Device) | Axial3D Insight (Predicate Device) | Comparison |
| --- | --- | --- | --- |
| Supported Modalities | CT and CTA | CT and CTA | Equivalent |
| Image registration | Yes | Yes | Equivalent |
| Segmentation Features | A combination of automated tools with smart editing tools | A combination of automated tools with smart editing tools | Equivalent |
| View Manipulation and Volume Rendering | Yes | Yes | Equivalent |
| Regions and Volumes of Interest (ROI) | Orthopedics / Trauma Cardiovascular Cranio- Maxillofacial | Orthopedics / Trauma Cardiovascular Cranio- Maxillofacial | Equivalent |
| Region/volume of interest measurements and size measurements | Yes | Yes | Equivalent |
| Region/Volume Quantification | Yes | Yes | Equivalent |
# 9. Performance Data
## a. Axial3D Insight Device Validation
Axial3D performed software design verification and validation testing on all three software components of the device. Axial3D has conducted software verification and validation, in accordance with the FDA guidance, General Principles of Software Validation; Final Guidance for Industry and FDA Staff, issued on January 11, 2002. All software requirements and risk analysis have been successfully verified and traced.
In addition to the human factors validation of the Axial3D Insight device, Axial3D conducted two validation studies - the Clinical Segmentation performance and the Intended Use of the device output - the 3D patient specific model.
{9}
axial3D Patient data made real
K250369 Traditional 510(k) Notification
The Clinical Segmentation Performance study was conducted with 3 radiologists reviewing the segmentation of 12 cases across the fields of Orthopedics, Trauma, Maxillofacial and Cardiovascular. Axial3D adopted a peer reviewed medical imaging review framework of RADPEER* to capture the assessment and feedback from the radiologists involved - all cases were scored within the acceptance criteria of 1 or 2a.
* "ACR RADPEER committee white paper with 2016 updates: revised scoring system, new classifications, self-review, and subspecialized reports." Journal of the American College of Radiology 14.8 (2017): 1080-1086.
The Intended Use validation study of the device was conducted with 9 physicians reviewing 12 cases across the fields of Orthopedics, Trauma, Maxillofacial and Cardiovascular, as defined in the Intended Use statement of the device. This study concluded successful validation of the 3D models produced by Axial3D demonstrating the device outputs satisfied end user needs and indications for use.
# b. Axial $^{\text{ML}}$ Machine Learning Validation
Axial $^{\text{ML}}$ machine learning models are used to generate an initial segmentation of cases, however the output of these models is not used in isolation to produce the final 3D patient specific model. The segmentations produced by the Axial $^{\text{ML}}$ machine learning models are used by Axial3D trained staff who complete the final segmentation and validation of the quality of each 3D patient specific model produced.
Axial $^{\text{ML}}$ machine learning models were independently verified and validated before inclusion in the Axial3D Insight device. Details of the data used in the validation of each machine learning model is provided below.
Table 4: Software Validation Data
| | Cardiac CT/CTa | Neuro CT/CTa | Ortho CT | Trauma CT |
| --- | --- | --- | --- | --- |
| Number of Images Used for Validation | 4,838 | 4,041 | 10,857 | 19,134 |
| Slice Spacing Range (Min, Max in mm) | 0.4 - 0.8 | 0.44 - 1.0 | 0.3 - 2.0 | 0.2 - 2.0 |
| Slice Spacing Average (in mm) | 0.54 | 0.63 | 0.79 | 0.76 |
| Pixel Size Range (Min, Max in mm) | 0.23 - 0.78 | 0.34 - 0.70 | 0.18 - 0.98 | 0.22 - 0.98 |
| Pixel Size Average (mm) | 0.46 | 0.51 | 0.44 | 0.51 |
*NeuroCT/CTa model is used for cardiology cases.
The variety of image scanner manufacturers and models used within the validation dataset are listed below.
{10}
axial3D Patient data made real
K250369 Traditional 510(k) Notification
| Manufacturer | Model |
| --- | --- |
| GE Medical Systems | Lightspeed Pro 16
Lightspeed Pro 32
Revolution CT
Optima CT660
Discovery CT750 HD |
| Siemens | SOMATOM Definition Flash
SOMATOM Definition Edge
SOMATOM Definition AS
SOMATOM Definition AS+
SOMATOM Perspective
SOMATOM Force
Sensation 16
AXIOM-Artis
Emotion 16 |
| Phillips | IQON Spectral CT
iCT 128
iCT 256
Ingenuity Core 128
Brilliance 62 |
| Toshiba | Aquillon PRIME
Aquillon PRIME SP |
The AxialML machine learning model training data used during the algorithm development was explicitly kept separate and independent from the validation data used.
# 10. Predetermined Change Control Plan (PCCP)
Axial3D Insight contains a Predetermined Change Control Plan (PCCP), which complies with Section 3308 of the Food and Drug Omnibus Reform Act (FDORA) of 2022, enacted on December 29, 2022. Modifications to the AxialML models of Axial3D Insight will be made in accordance with its Predetermined Change Control Plan (PCCP). The PCCP provides a description of the device's planned modifications, a modification protocol to test, verify, validate, and implement the modifications in a manner that ensures the continued safety and effectiveness of the device. This protocol also includes control measures that mitigate risks associated with changes to the AxialML model components, including an impact assessment of the planned modifications, to ensure modifications conducted in line with the modification protocol will not adversely impact the device's performance, safety, or effectiveness associated with its indications for use. The modifications outlined in the PCCP are summarized in the table below. In accordance with the PCCP, all algorithm modifications will be trained, tuned, and locked prior to release of the software to the field. The PCCP does not include provisions for implementation of adaptive algorithms that will continuously learn in the field.
Axial3D Insight is a cloud hosted platform, therefore, any software updates resulting from a modification implemented in line with the authorised PCCP will require no action from the end user. Any impact of the
{11}
axial3D Patient data made real
K250369
Traditional 510(k) Notification
implemented modification on the device's performance, inputs or use will be detailed in an updated IFU, which will be made available on the company website in conjunction with the release of a software update with the implemented modification. This information shall be made available to the user so that they may understand the changes, if applicable, to the device and continue to use the device safely and effectively across the relevant intended use populations and within the indications for use. The PCCP specifies a list of potential modifications to the device software as well as specific verification and validation activities in place to implement these modifications in a controlled manner such that the modified device remains safe and effective.
Table 5: Potential Modifications List
| ID | Summary | Trigger for Retrain | Timeframe |
| --- | --- | --- | --- |
| MOD_ML_001 | Increasing Volume of Training, Tuning and Testing data | Newly available or identified source(s) of data representing the original intended use population | Reviewed periodically every 6 months |
| MOD_ML_002 | Semantic Sub-Labeling | Internal R&D activities demonstrating benefits from introducing semantic sub-labelling that increase the effectiveness of AxialML output as utilized by Axial3D internal engineers | Within 18 months |
| MOD_ML_003 | Introduction of New Assistive Labeling | Internal R&D activities demonstrating performance benefits from introducing assistive labels that increases the effectiveness of AxialML output as utilized by Axial3D internal engineers | Reviewed periodically every 6 months |
| MOD_ML_004 | Model Parameter and Hyperparameter Tuning | Internal R&D activities demonstrating performance benefits from parameter and hyperparameter tuning that increases the effectiveness of AxialML output as utilized by Axial3D internal engineers | Reviewed periodically every 12 months |
| MOD_ML_005 | Performance Increasing Library Updates | Newly available library or dependency updates, of component utilized in the original device that increases the effectiveness of AxialML output as utilized by Axial3D internal engineers | Reviewed periodically every 6 months |
{12}
axial3D Patient data made real
K250369 Traditional 510(k) Notification
## 11. PCCP Performance Verification & Validation
These activities focus on the specific performance and safety of the modified AxialML machine learning models themselves, independent of their integration into the broader system. The purpose of these activities is to verify and validate that the modified ML model meets its design inputs and acceptance criteria before being considered for integration into the Axial3D Insight device.
### a. Verification Activities:
Peer Code Review: Ensures the quality and correctness of the underlying code for the AxialML models.
Unit Testing: Verifies individual components or units of the AxialML code function as intended.
Quantitative 3D Medical Image Segmentation Metric Analysis: This is a crucial step where the performance of the modified ML model is quantitatively assessed using a broad range of independent test datasets that are representative of the intended use population. We utilize a suite of metrics including Dice Coefficient, Pixel Accuracy, Area Under the Curve (AUC), Precision, and Recall to provide a robust indicator of segmentation performance, comparing the candidate modified AxialML model against the model version from the original submission. These metrics are calculated for each individual label and/or as an averaged performance. The performance must demonstrate equivalence or improvement to be considered for the next phase, Validation. This activity serves as a core component of our internal model development and verification, ensuring that any model modifications conducted in line with the modification protocol must consistently meet performance standards before further integration and validation.
### b. Validation Activities:
Qualitative Assessment using Independent Test Datasets: Once quantitative metrics show promise, candidate models undergo a qualitative assessment. This involves a pool, minimum of 3, of Axial3D Medical Visualization Engineers reviewing the segmentations generated by the modified AxialML model, against segmentations generated from the unmodified model, for a consistent set of independent test datasets from the original submission. This assessment utilizes a fixed evaluation methodology to define improved, equivalent or reduced performance based on the AxialML Model Design Input Specifications. This activity confirms validation by producing objective evidence, that demonstrates each AxialML Model Design Input Specification has been met and the model output supports Axial Staff in completing anatomical segmentation for patient-specific 3D models, in line with the device intended use.
Quantitative Assessment using Expert Reference Standard: The validation activities also include the use of a fixed number of expert reference standards, independently sourced datasets commissioned from US
{13}
axial3D Patient data made real
K250369 Traditional 510(k) Notification
only sites, representative of the intended use population and independently segmented and reviewed by expert radiologists. These expert reference standards remain unchanged from the original submission and shall be used as a consistent baseline to determine any potential modification has not adversely affected the device performance, as determined by comparing the original model baseline vs modified model performance using a combination of quantitative metrics including DICE, AUC, Precision, Accuracy and Recall. The mean of the identified quantitative metrics must demonstrate equivalence or an improvement for the proposed modified AxialML model to be successfully validated in line with the modification protocol.
Hazard Analysis Review: As part of the Verification & Validation activities, the device hazard analysis is reviewed to identify and address any potential biases and limitations introduced by the model modifications, ensuring that any relevant risk control measures, new or modified, are captured and verified.
## 12. Conclusion:
A review of the indication for use, operating environment, device class, method of use, software level of concern, technical characteristics, and technical specifications indicate the Axial3D Insight is substantially equivalent to the FDA cleared and marketed device, Axial3D Insight (K232841). Both devices are to be used as an image segmentation tool to create 3D replicas from 2D medical images. The addition of a PCCP is the only difference between the proposed device and the predicate and reference devices. The differences between the proposed device and the predicate device do not raise any new concerns of safety or efficacy.
10