K221511 · Axial Medical Printing Limited · LLZ · Jun 23, 2022 · Radiology
Device Facts
Record ID
K221511
Device Name
Axial3D Cloud Segmentation Service
Applicant
Axial Medical Printing Limited
Product Code
LLZ · Radiology
Decision Date
Jun 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Axial3D Cloud Segmentation Service is intended for use as a cloud based service and image segmentation framework for the transfer of DICOM imaging information from a medical scanner to an output file. The Axial3D Cloud Segmentation Service output file can be used for the fabrication of physical replicas of the output file using additive manufacturing methods. The output file or physical replica can be used for treatment planning. The physical replica can be used for diagnostic purposes in the field of orthopedic, maxillofacial and cardiovascular applications. Axial3D Cloud Segmentation Service should be used in conjunction with other diagnostic tools and expert clinical judgment.
Device Story
Cloud-based image segmentation service; transforms DICOM imaging data from medical scanners into digital output files (e.g., STL) for additive manufacturing. Used by clinicians for pre-surgical treatment planning and diagnostic support in orthopedic, maxillofacial, and cardiovascular fields. Operates as a cloud platform; processes medical images to generate 3D anatomical models. Healthcare providers use these physical replicas or digital files to visualize patient-specific anatomy, aiding in surgical planning and clinical decision-making. Benefits include improved anatomical understanding and surgical preparation.
Clinical Evidence
Bench testing only. Validation confirmed measurement accuracy and comparison of generated meshes against the predicate device. Printing validation demonstrated accuracy across compatible 3D printers. No clinical data presented.
Technological Characteristics
Cloud-based software platform; processes DICOM imaging data; generates digital output files for additive manufacturing. Employs image segmentation algorithms to produce 1:1 scale patient-specific anatomical models. Verification and validation performed per FDA guidance; minimal mesh smoothing applied to maintain accuracy to original DICOM data.
Indications for Use
Indicated for adult patients requiring medical image segmentation for treatment planning and diagnostic purposes in orthopedic, maxillofacial, and cardiovascular applications. Used in conjunction with other diagnostic tools and expert clinical judgment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
Axial Medical Printing Limited % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
June 23, 2022
Re: K221511
Trade/Device Name: Axial3D Cloud Segmentation Service Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: Mav 23, 2022 Received: May 24, 2022
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221511
Device Name Axial3D Cloud Segmentation Service
#### Indications for Use (Describe)
Axial3D Cloud Segmentation Service is intended for use as a cloud based service and image segmentation system for the transfer of DICOM imaging information from a medical scanner to an output file.
The Axial3D Cloud Segmentation Service output file can be used for the fabrication of physical replicas of the output file using additive manufacturing methods.
The output file or physical replica can be used for treatment planning.
The physical replica can be used for diagnostic purposes in the field of orthopedic, maxillofacial and cardiovascular applications.
Axial3D Cloud Segmentation Service should be used in conjunction with other diagnostic tools and expert clinical judgment.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
K221511
Image /page/3/Picture/3 description: The image shows the logo for Axial3D. The word "axial" is written in a dark blue color, with a teal diamond shape above the "i". To the right of "axial" is the number "3D" in a smaller font size. Below the logo is the tagline "Patient data made real" in teal.
### Submitters Contact Information
Submitter's Name: Axial Medical Printing Limited 17A Ormeau Avenue Belfast BT2 8HD United Kingdom Tel: +44 (0)28 90183590
### Name of Contact Person
Jenna McGarry, QA/RA Manager
#### Date of submission
25th February 2022
#### Subject Device Name
Device Trade Name: Axial3D Cloud Segmentation Service
Device Common Name: Axial3D Cloud Segmentation Service
Classification Name: System, Image Processing, Radiological (21CFR 892.2050, Product Code LLZ)
### Identification of Legally Marketed Predicate Device
The Axial Medical Printing Limited Axial3D Cloud Segmentation Service is substantially equivalent to the following:
### Predicate Device
Manufacturer: Materialse NV
Trade Name: Mimics InPrint
Common Name: Mimics InPrint
{4}------------------------------------------------
# Classification Name: System, Image Processing, Radiological (21CFR 892.2050,
Product Code LLZ)
510(k) Number: K173619
### Device Description
Axial3D Cloud Segmentation Service is a secure, highly available cloud based image processing, segmentation and 3D modeling framework for the transfer of imaging information to either a digital file or as a 3D printed physical model.
Axial3D Cloud Segmentation Service is made up of a number of component parts, which allow the production of patient-specific 1:1 scale replica models, either as a digital file or as a 3D printed physical model.
### Intended Use/Indications for Use
Axial3D Cloud Segmentation Service is intended for use as a cloud based service and image segmentation framework for the transfer of DICOM imaging information from a medical scanner to an output file.
The Axial3D Cloud Segmentation Service output file can be used for the fabrication of physical replicas of the output file using additive manufacturing methods.
The output file or physical replica can be used for treatment planning.
The physical replica can be used for diagnostic purposes in the field of orthopedic, maxillofacial and cardiovascular applications.
Axial3D Cloud Segmentation Service should be used in conjunction with other diagnostic tools and expert clinical judgment.
### Comparison between Proposed Device and Predicate Device
| Product Details | New Device: | Proposed<br>Predicate<br>Device: | Comment: |
|----------------------------------------|------------------------------------------|----------------------------------|----------|
| Device<br>Manufacturer | Axial Medical<br>Printing Limited | Materialise N.V. | N/A |
| Device Name | Axial3D Cloud<br>Segmentation<br>Service | Mimics inPrint | N/A |
| Device Trade or<br>Proprietary<br>Name | Axial3D Cloud<br>Segmentation<br>Service | Mimics inPrint | N/A |
| 510(k) Number | TBC | K173619 | N/A |
{5}------------------------------------------------
# Axial Medical Printing Limited
| Device<br>Regulation<br>Name: | Medical Image<br>Management and<br>Processing<br>System | Picture archiving<br>and<br>communications<br>system | The proposed device and predicate devices are<br>identical | |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Device<br>Regulation<br>Number: | 21 CFR 892.2050 | 21 CFR<br>892.2050 | The proposed device and predicate devices are<br>identical | |
| Device Product<br>Code: | LLZ | LLZ | The proposed device and predicate devices are<br>identical | |
| Device<br>Classification<br>FDA: | Class II | Class II | The proposed device and predicate devices are<br>identical | |
| Intended Use<br>/Indication for<br>Use | Axial3D Cloud<br>Segmentation<br>Service is<br>intended for use<br>as a cloud based<br>service and<br>image<br>segmentation<br>framework for the<br>transfer of<br>DICOM imaging<br>information from<br>a medical<br>scanner to an<br>output file.<br>The Axial3D<br>Cloud<br>Segmentation<br>Service output file<br>can be used for<br>the fabrication of<br>physical replicas<br>of the output file<br>using additive<br>manufacturing<br>methods.<br>The output file or<br>physical replica<br>can be used for<br>treatment<br>planning.<br>The physical<br>replica can be<br>used for<br>diagnostic<br>purposes in the<br>field of<br>orthopedic,<br>maxillofacial and | Mimics inPrint is<br>intended for use<br>as a software<br>interface and<br>image<br>segmentation<br>system for the<br>transfer of<br>DICOM imaging<br>information from<br>a medical<br>scanner to an<br>output file. It is<br>also used as<br>pre-operative<br>software for<br>treatment<br>planning. For<br>this purpose, the<br>Mimics inPrint<br>output file can<br>be used for the<br>fabrication of<br>physical replicas<br>of the output file<br>using traditional<br>or additive<br>manufacturing<br>methods. The<br>physical replica<br>can be used for<br>diagnostic<br>purposes in the<br>field of<br>orthopedic,<br>maxillofacial and<br>cardiovascular<br>applications.<br>Mimics inPrint<br>should be used<br>in conjunction<br>with other<br>diagnostic tools | The proposed device and primary predicate have<br>equivalent indications. | |
| | | cardiovascular<br>applications.<br>Axial3D Cloud<br>Segmentation<br>Service should be<br>used in<br>conjunction with<br>other diagnostic<br>tools and expert<br>clinical judgment. | and expert<br>clinical<br>judgment. | |
| Target<br>Population | | Adult | Adult | Same |
| Method of Use | | Used in<br>conjunction with<br>other diagnostic<br>tools and expert<br>clinical judgment. | Used in<br>conjunction with<br>other diagnostic<br>tools and expert<br>clinical<br>judgment. | The proposed device has an identical method of use<br>to the predicate. |
| Imaging<br>Modality | | Computed<br>tomograghy (CT),<br>CT Angiography<br>(CTA) | DICOM<br>compliant types<br>of imaging<br>information | DICOM compliant types of imaging information |
| Environment | | Hospital | Hospital | The proposed device and predicates have identical<br>target environments |
| OTC or<br>Prescription<br>Device | | Prescription Use | Prescription Use | The proposed device and predicate devices are<br>identical |
| | Level of<br>Concer<br>n | Moderate | Moderate | |
| Software | Verifica<br>tion &<br>Validati<br>on | Complies with<br>FDA Guidance<br>Requirement | Complies with<br>FDA Guidance<br>Requirement | The proposed device and predicate devices are<br>identical |
{6}------------------------------------------------
### Comparison of Technological Characteristics
Comparison shows the Axial3D Cloud Segmentation Service is substantially equivalent in intended use, design, functionality, operating principles and performance characteristics of the predicate device.
Both the predicate and subject device are intended for use as a software interface and image segmentation process to facilitate the transfer of imaging information from a medical scanner to an output file. Both devices use the same segmentation functionality and generate the same output files. Both devices have functionalities to assist pre-surgical planning.
Verification and validation of both the subject and predicate devices have been performed in the same way.
It was found that minimal variances were visible between the Mesh generated from subject device and the predicate device, these variances are a result of mesh smoothing, Axial3D apply minimal smoothing to the STL file generated from the labeled images to retain a higher level of accuracy to
{7}------------------------------------------------
the original DICOM images. The validation highlighted that the subject device performed to a higher standard, than the predicate device.
The printer selection is also based on 3D printer manufacturers' guidance on the most commonly used printing technologies for the production of medical 3D printed anatomical models.
The following technological differences exist between the subject and predicate device:
- . The subject device is a cloud based software platform rather than a standalone software package like the predicate
### Performance Data
Non-clinical Testing
The Axial3D Cloud Segmentation Service device has been validated for its intended use to determine substantial equivalence to the predicate device. Measurement accuracy and comparisons were performed and confirmed to be within specification.
Validation of printing of physical replica models was performed and demonstrated to be accurate when using any of the compatible 3D printers.
#### Conclusion
The characteristics that determine functionality and performance of the subject device, Axial3D Cloud Segmentation Service are similar to the device cleared under K173619.
The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.