CANTOR & SILVER 5X TINTED (HIOXIFILCON A) SOFT (SPHERICAL & TORIC) DAILY WEAR CONTACT WEAR CONTACT LENS, TINTED (TRANSPA
Applicant
Cantor & Silver , Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Apr 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Intended Use
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program. The CANTOR & SILVER 5X Tinted (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10.00 diopters. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
Device Story
Soft contact lens fabricated from hioxifilcon A (copolymer of 2-HEMA and GMA, cross-linked with EGDMA); hydrophilic material becomes pliable when hydrated. Lens covers cornea and extends beyond limbus to provide refractive surface. Tinted with reactive color additives on anterior surface corresponding to iris; tinting process does not alter lens performance. Available in spherical and toric designs for daily wear. Requires hydrogen peroxide-based disinfection. Used by patients for vision correction; prescribed by eye care professionals. Benefits include visual acuity correction and cosmetic enhancement via tinting.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on material chemistry, physical properties, and manufacturing process comparisons to predicate devices.
Technological Characteristics
Material: hioxifilcon A (42% polymer, 58% water). Refractive index: 1.515 (dry), 1.404 (hydrated). Oxygen permeability: 20 x 10^-11 (cm²/sec)(ml O₂/ml x mm Hg @ 35°C). Non-ionic copolymer of 2-HEMA and GMA, cross-linked with EGDMA. Lathe-cut manufacturing. Tinted with reactive color additives. Sterilization via standard procedures.
Indications for Use
Indicated for aphakic and non-aphakic persons with non-diseased eyes for correction of myopia, hyperopia, and refractive astigmatism (up to 1.50D for spherical, up to 10.00D for toric).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
K123431 — DREAM COLOR I (POLYMACON) SOFT (HYDROPHILIC) CONTACT LENS, DREAM COLOR II (POLYMACON) SOFT (HYDROPHILIC) CONTACT LENS, D · Dreamcon Co., Ltd. · Jun 17, 2013
{0}------------------------------------------------
4/21/99
# SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
K990757
Applicant information:
Date Prepared:
March 12, 1999
Name: Address Cantor & Silver Limited Manor Road, Brackley Northamptonshire England NN13 6ED
Contact Person:
Phone Number: Fax:
Mr. David Cantor Managing Director/President 011 44 1280 702002 011 44 1280 703003
Official Correspondent: And US Consultant
Med-Vice Consulting, Inc. Mr. Martin Dalsing President/CCO 623 Glacier Grand Junction, CO 81503 970.243.5490 970.243.5501
Phone Number: Fax Number:
Device Information:
Regulatory Classification: Class II
86 LPL Product Code:
Trade Name:
CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact Lens, Tinted.
Classification Name:
Lenses, Soft Contact, Daily Wear
{1}------------------------------------------------
#### Equivalent Device:
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact Lens is substantially equivalent to the predicate device in terms of intended use and design. The predicate device is the Durasoft 2 Colors manufactured by Wesley-Jessen.
### Device Description:
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact lens are fabricated from hioxifilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. The color enhanced soft contact lenses are tinted with "listed" color additives. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the manufacturing process, the lenses containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed color reactive additive on that portion of the anterior (front) surface of the lens that corresponds to the iris. The finted contact lenses are available in a light, medium or dark shade of the following enhance colors; Blue, Green, Aqua, Yellow, Lavender, Brown, Ultra Violet, and Amber.
The color additive (timing) effect is formed by reacting one or more of the "listed" additives with HEMA based (poly hydroxyethy) methacrylate) soft contact lens. The color additives used are not removed by lens handling and cleaning/disinfecting procedures as recommended. Except for affecting the amount of light transmitted through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens.
In the bydrated state, the lens conforms to the curvature of the eye covering the comea and extending slightly beyond the limbus forming a transparent optical surface. The (hioxifilcon A) soft hydrophilic contact lens a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
{2}------------------------------------------------
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 58% water by weight.
The non-ionic lens material, hioxifilcon A is an ultra high molecular weight copolymer of 2-bydroxyethyl methacrylate (2-HEMA) and 2,3-Dibydroxypropyl Methacrylate (Glycerol Methacrylate, GMA) and cross-linked with ethylene glycol dimethacrylate (EGDMA). It consists of 42% hioxifiloon A and 58% water by weight when immersed in normal saline solution buffered with sodium bicarbonate. The physical properties of the lens are:
| Refractive Index | 1.515 (dry) 1.404 (hydrated) |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Light Transmission (tinted) | greater than 70% |
| Color Additives | reactive black 5, reactive blue 21, reactive blue 19, reactive blue 4,<br>reactive blue 163, reactive red 11, reactive red 180, reactive yellow 15,<br>reactive yellow 86, or reactive orange 78 |
| Water Content | 58 % ±2% |
| Specific Gravity | 1.308 (dry) 1.136 (hydrated) |
| Oxygen Permeability | $20 X 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt method) |
#### INDICATED USE:
The CANTOR & SILVER 5X Tinted (hioxifileon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the assigmatism does not interfere with visual acuity. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10.00 diopters. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
{3}------------------------------------------------
#### Substantial Equivalence:
The device will be manufactured according to specified process controls and a Quality Management System certified to ISO 9002 and EN46002 by the National Accreditation of Certification Bodies. The device will undergo manufacturing, packaging and sterilization procedures similar to devices currently marketed and distributed by Cantor & Silver Limited in the United States. The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the device is equivalent to the BENZ-G 5X (hioxifilcon A), 510(k) #K952620. The tinting characteristics and parameters of the lens are substantially equivalent to the Durasoft 2 Colors tinted soft contact lens, manufactured Wesley-Jessen. Being similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate device, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate devices identified above.
The following matrix illustrates that the production method, lens function, and the "listed" color additives and timing characteristics of the CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact Lens are substantially equivalent to the predicate device.
| | Characteristic | CANTOR & SILVER 5X<br>TINTED | DURASOFT 2 COLORS |
|-----|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | PRODUCTION<br>METHOD | Lathe-Cut | Lathe-Cut |
| 2.) | LENS FUNCTION | Refractive medium that focuses<br>light rays from near and distant<br>objects on the retina, while<br>compensating for refractive error,<br>including (astigmatism) | Refractive medium that focuses<br>light rays from near and distant<br>objects on the retina, while<br>compensating for refractive error,<br>including (astigmatism) |
| 3.) | FDA "listed"<br>colored additives | The reactive colored additives<br>consist of reactive black 5,<br>reactive blue 4, reactive blue 19,<br>reactive 21, reactive blue 163,<br>reactive yellow 15, reactive yellow<br>86, reactive orange 78, reactive<br>red 11 and reactive red 180 | The colored pigments consist of<br>iron oxides, chromium oxide<br>greens, titanium dioxide, [phthalo-<br>cyaninato (2-)] copper, carbazole<br>violet and phthalocyanine green. |
| a. | Uses<br>and<br>restrictions | The color additives listed above<br>may be used to color contact<br>lenses in amounts not to exceed<br>the minimum reasonably required<br>to accomplish the intended<br>coloring effect. | The color additives listed above<br>may be used to color contact<br>lenses in amounts not to exceed<br>the minimum reasonably required<br>to accomplish the intended<br>coloring effect. |
| 4.) | Color Additive<br>Characteristics | The color additives used are not<br>removed by lens handling and<br>cleansing/disinfecting procedures.<br>The optical and performance<br>characteristics are not altered by<br>the lens coloring process. | The color additives used are not<br>removed by lens handling and<br>cleansing/disinfecting procedures.<br>The optical and performance<br>characteristics are not altered by<br>the lens coloring process. |
| 5.) | Colors Offered | Blue, Green, Aqua, Yellow,<br>Lavender, Brown, Ultra Violet<br>and Amber | Sky blue, Jade green, Aquamarine<br>and Violet blue |
#### Substantial Equivalence Matrix
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures or profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 1 1933
Mr. Martin Dalsing Med. Vice Consulting, Inc. 623 Glacier Drive Grand Junction, CO
Re: K990757 Trade Name: Cantor & Silver 5X Tinted (hioxifilcon A) Soft Contact Lens Regulatory Class: II Product Code: 86 LPL Dated: April 7, 1999 Received: April 12, 1999
Dear Mr. Dalsing:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{5}------------------------------------------------
Page 2 - Mr. Martin Dalsing
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on. the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
#### CANTOR & SILVER 5X TINTED (hioxifilcon A) Soft (Spherical & Toric) Device Name: Daily Wear Contact Lens, Tinted.
# INDICATIONS FOR USE:
The CANTOR & SILVER SX Tinted (hioxifilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia. The leases may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-disessed eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10.00 diopters. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109)
or
Over-The-Counter Use
(Optional Format 1-2-96)
SRRell
(Division Sign-Off) Division of Ophthalmic Devices
510(k) Number K990757
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.