ALDEN CLASSIC TINTED (POLYMACON) SOFT (SPHERICAL & TORIC) DAILY WEAR CONTACT LENS, TINTED (TRANSPARENT/ENHANCING)

K980554 · Alden Optical Labs., Inc. · LPL · Apr 7, 1998 · Ophthalmic

Device Facts

Record IDK980554
Device NameALDEN CLASSIC TINTED (POLYMACON) SOFT (SPHERICAL & TORIC) DAILY WEAR CONTACT LENS, TINTED (TRANSPARENT/ENHANCING)
ApplicantAlden Optical Labs., Inc.
Product CodeLPL · Ophthalmic
Decision DateApr 7, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Intended Use

The ALDEN CLASSIC TINTED (polymacon) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The ALDEN CLASSIC TINTED (polymacon) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 Diopters.

Device Story

Hydrophilic contact lens fabricated from polymacon; lathe-cut and polished in dry state; becomes soft and pliable when hydrated in aqueous solution. Lens conforms to corneal curvature to provide refractive correction for myopia, hyperopia, and astigmatism. Contains blue tint (Vat blue 6) for visibility/enhancement. Requires maintenance in fully hydrated state using compatible storage solution. Used by patients for daily wear to improve visual acuity.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, material chemistry, and manufacturing process comparisons to predicate devices.

Technological Characteristics

Hydrophilic polymer (polymacon); 38% water content; 62% polymer content. Refractive index 1.43 (hydrated); Dk value 9 x 10^-11 Fatt units. Blue tint (Vat blue 6, CI# 69825). Lathe-cut manufacturing process. Daily wear modality.

Indications for Use

Indicated for correction of visual acuity in aphakic and non-aphakic patients with non-diseased eyes, including myopia, hyperopia, and refractive astigmatism (up to 1.50D for spherical lenses; up to 10D for toric lenses).

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. February 9, 1998 13295 Broadway 800.253.3669 623 Glacier 970.243.5490 970.243.5501 APR - 7 1998 The assigned 510(k) number is: K980554 Alden Optical Laboratories Alden, New York 14004 Mr. Charles H. Creighton Med-Vice Consulting, Inc. Grand Junction, CO 81503 Mr. Martin Dalsing #### Applicant information: Date Prepared: Name: Address Contact Person: Phone Number: US Consultant: Phone Number: Fax Number: #### Device Information: | Device Classification: | Class II | |------------------------|-------------------------------------------------------------------------------------------------------------------------| | Classification Number: | LPL | | Trade Name: | ALDEN CLASSIC TINTED (polymacon) Soft (Spherical<br>& Toric) Daily Wear Contact Lens, Tinted<br>(Transparent/Enhancing) | | Classification Name: | Lenses, Soft Contact, Daily Wear | {1}------------------------------------------------ ### Equivalent Devices: The ALDEN CLASSIC TINTED (polymacon) Soft (Spherical & Toric) Daily Wear Contact Lens is substantially equivalent to predicate devices in terms of intended use and design. Predicate devices include Metro Tint manufactured by Metro Optics and the BENZ-38 manufactured by Benz Research and Development. ## Device Description: The ALDEN CLASSIC TINTED (polymacon) Soft (Spherical & Toric) Daily Wear Contact lenses are fabricated from polymacon, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. A Blue tint (transparent/enhancing), 7,16-Dichloro-6, 15-dihydro-5, and 9,14,18anthrazinetetrone is added to the lens." In the hydrated state, the lens conforms to the curvature of the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (polymacon) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution. The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 38% water by weight. The physical properties of the lens are: | Refractive Index | 1.52 (dry) 1.43 (hydrated) | |---------------------|---------------------------------------------------------------------------------| | Light Transmission: | greater than 70% T | | Water Content | 38 % | | Specific Gravity | 1.28 (dry) 1.18 (hydrated) | | Oxygen Permeability | 9 X 10-11 Fatt Units (cm2/sec)(ml O2/ml x mm Hg @ 35° C), (revised Fatt method) | ## Intended Use: The ALDEN CLASSIC TINTED (polymacon) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The ALDEN CLASSIC TINTED (polymacon) Toric Soft Contact Lenses for daily waar are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism. {2}------------------------------------------------ ## Substantial Equivalence: The device will be manufactured according to specified process controls and a quality assurance program. The device will undergo manufacturing, packaging and sterilization procedures similar to devices currently marketed and distributed by Alden Optical Laboratories. The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the device is equivalent to the BENZ-38 (polymacon), 510(k) #K961103. Being similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate devices, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not roise different questions of i safety and effectiveness than the predicate devices identified above. The following matrix illustrates that the production method, lens function and material of the ALDEN CLASSIC TINTED (polymacon) Soft (Spherical & Toric) Daily Wear Contact Lens, Tinted (Transparent/Enhancing) are substantially equivalent to the predicate device. In addition, the water content, polymer, Dk value, refractive index, specific gravity, and light transmission are as well substantially equivalent to the predicate device. | | Characteristic | ALDEN CLASSIC TINTED | PREDICATE<br>DEVICE | |-----|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------| | 1.) | PRODUCTION<br>METHOD | Lathe-Cur | SAME | | 2.) | LENS FUNCTION | Refractive medium that focuses light<br>rays from near and distant objects on<br>the retina, while compensating for<br>refractive error, including<br>(astigmatism) | SAME | | 3.) | MATERIAL | Hydrophilic Polymer | SAME | | a. | Water Content | 38% | SAME | | b. | Polymer Content | 62% | SAME | | c. | Polymer | polymacon | SAME | | d. | DK Value | 9 | SAME | | e. | Refractive Index | 1.43 (hydrated) | SAME | | f. | Specific Gravity | 1.180 (hydrated) | SAME | | g. | Light Transmission | greater than 70% T | SAME | | h. | Dye Color | Blue (21CDR 73,3119)<br>CJ # 69825 | SAME | #### Substantial Equivalence Matrix - The Dye Color in the SE predicate device "Benz-38" is that of phthalocyanato (2) (copper) being a blue visibility tint and not the Vat blue 6 CI# 69825. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR - 7 1998 Alden Optical Laboratories c/o Mr. Martin Dalsing Med-Vice Consulting, Inc. 623 Glacier Drive Grand Junction, CO 81503 Re: K980554 > Trade Name: Alden Classic Tinted (polymacon) Hydrophilic Contact Lens for Daily Wear (Lathe-cut, enhancement tint) Regulatory Class: II Product Code: 86 LPL Dated: February 9, 1998 Received: February 13, 1998 Dear Mr. Dalsing: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Mr. Martin Dalsing This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597. Sincerely yours, A. Ralph Rosenthal, MD A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT #### Device Name: ALDEN CLASSIC TINTED (polymacon) Soft (Spherical & Toric) Daily Wear Contact Lens, Tinted (Transparent/Enhancing) ## INDICATIONS FOR USE: The ALDEN CLASSIC TINTED (polymacon) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The ALDEN CLASSIC TINTED (polymacon) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 Diopters. ## (PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|--------------------------| | (Division Sign-Off) | | | Division of Ophthalmic Devices | | | 510(k) Number | K980554 | | | <i>Karen Warbenton</i> | | | or | | | (Optional Format 1-2-96) | | Prescription Use | <div style="display:inline-block;"> ✓</div> | |----------------------|---------------------------------------------| | (Per 21 CFR 801.109) | | | Over-The-Counter Use | __________ | |----------------------|------------| |----------------------|------------| ALDEN CLASSIC TINTED (Polymacon) Soft (Spherical & Toric) Daily Wear Contact Lens, Tinted (Transparent/Enhancing) ...14
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...