SAUFLON CLARITI WITH TINT, SAUFLON CLARITI TORIC WITH TINT, SAUFLON CLARITI MULTIFOCAL WITH TINT
Applicant
Sauflon Pharmaceuticals, Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Mar 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Intended Use
The SAUFLON CLARITI (somofilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity. The SAUFLON CLARITI TORIC (somofile) Contact Leas with UV blocker and handling tint is indicated for daily woar for the optical corrective anteropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters. The SAUFLON CLARITI MULTIFOCAL (somofileon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive anetropia (myopia and or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +4.00 Diopters or less. The SAUFLON CLARITI MULTIFOCAL TORIC (somofilion A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and that may require a reading addition of +4.00 Diopters or less. The lenses may be prescribed for daily wear for up to one month with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion as recommended by the eye care professional. Saullon Claili (somofilcon A) Soll (hydrophilic) Contact lens with UV blocker and handling tint helps protect against transmission of harmful UV radiation to the cornea and into the eye. The lens has a itint added to make the lens more visible for handling.
Device Story
Soft contact lens (somofilcon A) for daily wear; corrects refractive ametropia (myopia, hyperopia, astigmatism, presbyopia). Lens material is hydrophilic silicone hydrogel copolymer with benzophenone UV blocker and D&C Green 6 handling tint. Available as single vision, toric, and multifocal designs. Prescribed by eye care professionals; patient performs daily removal, cleaning, and chemical disinfection. Lens covers cornea and adjacent sclera. Provides vision correction and UV radiation protection. Substantial equivalence established via physical, chemical, and clinical comparison to predicate lenses.
Clinical Evidence
Clinical study compared subject device to Air Optix Aqua (Ciba Vision Inc.) using OptiFree Replenish solution. Evaluated safety and efficacy for daily wear. Results confirmed safety, acceptability, and substantial equivalence to predicate device.
Technological Characteristics
Material: somofilcon A (silicone hydrogel copolymer cross-linked with Alkyl bis methacrylate). Water content: 56%. Oxygen permeability (Dk): 60 (Fatt method). Refractive index: 1.4008. UV blocker: benzophenone. Tint: D&C Green 6. Dimensions: 13.0-18.0mm diameter, 7.5-9.5mm base curve. Power range: -30.00 to +30.00 DS. Manufacturing: cast molding. Packaging: borate buffered saline with 0.005% poloxamer 407 in hermetically sealed blister packs.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, and/or presbyopia in phakic or aphakic persons with non-diseased eyes. Supports astigmatism correction up to 10.00 Diopters and reading additions up to +4.00 Diopters.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Lens for Daily Wear (K130342)
Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) Visibility Tinted Contact Lens (K033919/K073459)
Related Devices
K130342 — SAUFLON CLARITI (SOMOFILCON A) SOFT (HYDROPHILIC) CONTACT LENS WITH UV BLOCKER · Sauflon Pharmaceuticals, Ltd. · Aug 5, 2013
K130331 — SAUFLON CLARITI 1 DAY (SOMOFILCON A) SOFT (HYDROPHILIC) DAILY DISPOSABLE CONTACT LENS WITH UV BLOCKER · Sauflon Pharmaceuticals, Ltd. · Jul 26, 2013
K181920 — Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker · CooperVision, Inc. · Dec 11, 2018
K210930 — ACUVUE® OASYS MAX 1-DAY Contact Lenses (senofilcon A) · Johnson & Johnson Vision Care, Inc. · Nov 29, 2021
K141670 — ACUVUE (ETAFILCON A) SOFT CONTACT LENS FOR PRESBYOPIA · Johnson & Johnson Vision Care, Inc. · Sep 18, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(K) SUMMARY AS REQUIRED BY SECTION 807.92(c)
### 1.- SUBMITTER INFORMATION:
| Company Name: | Sauflon Pharmaceuticals Ltd. |
|------------------------|-----------------------------------------------------------------------------------------------------|
| Address: | 49 - 53 York Street<br>Twickenham<br>Middlesex<br>TW1 3LP |
| Phone: | 020 8322 4200 |
| Fax: | 0845 051 8701 |
| Contact Person: | Dr Christopher Smejkal |
| DATE SUMMARY PREPARED: | 24th January 2014 |
| DEVICE NAME: | |
| Trade Name: | Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact<br>Lens with UV Blocker and Handling tint |
Common Name: Soft Contact Lens
| Classification: | CLASS II (21 CFR 886.5925) CODE -LPL, |
|-----------------|---------------------------------------|
| | SOFT (HYDROPHILIC) CONTACT LENS |
Duration of Use:
Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact lens with UV Blocker and handling tint is a daily wear, frequent replacement lens for surface contact of the eye, and intended for daily removal and less than 24 hours contact duration.
### 2 .- SUBSTANTIAL EQUIVALENCE:
The sponsor considers the Sauflon Clariti (somofilcon A) Soft (hydrophilic) with UV Blocker and handling tint to be substantially equivalent to the Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Lens for Daily Wear which has been approved pursuant to K130342, and Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) Visibility
{1}------------------------------------------------
Tinted Contact Lens for Daily Wear which has been approved pursuant to K033919/K073459.
### 3.- DESCRIPTION of the DEVICE:
The Sauflon Clariti (somofilcon A) Soft (Silicone Hydrogel) Contact Lens with UV Blocker and handling tint is available as a single vision, toric and multifocal lens. The lens material (somofilcon A) is a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with Alkyl bis methacrylate. When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used to block UV radiation. A D&C Green 6 tint is also included in the composition of the lens for ease of handling and visibility.
The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm
The Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and Handling Tint is spherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:
| • Chord Diameter: | 13.0 to 18.0mm |
|---------------------------|-------------------------|
| • Centre Thickness: | 0.03 to 0.70mm |
| • Base Curve: | 7.5 to 9.5mm |
| • Powers: | -30.00 DS to +30.00 DS |
| • Toric cylinder options: | Between -0.75 and -9.75 |
| • Toric Axis options: | 10° to 180° (5° steps) |
| • Multifocal ADD: | |
Add power of up to +4.00, labelled with indicative add strength to be read in conjunction with the fitting guide.
The physical/optical properties of the lenses are:
| · Refractive Index: | 1.4008 |
|---------------------------|--------|
| · %Transmittance @ 590nm: | 98.30% |
- . %Transmittance @ 590nm:
- %Transmittance @ 280-315nm: 0.71 .
- . %Transmittance @ 316-380nm: 20.62
- . Surface Character: Hydrophilic
- Water Content: . 56%
{2}------------------------------------------------
- 60 x 10-11 (cm2/sec) (ml O2/ml x mmHg) Oxygen Permeability (DK): . at 35°C (Fatt Method for determination of oxygen permeability).
- 1.17 Specific Gravity:
## 4.- INDICATIONS FOR USE
.
The SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.
The SAUFLON CLARITI TORIC (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes that may exhibit astigmatism up to 10.00 Diopters.
The SAUFLON CLARITI MULTIFOCAL (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +4.00 Diopters or less.
The SAUFLON CLARITI MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and that may require a reading addition of +4.00 Diopters or less.
The lenses may be prescribed for daily wear with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion as recommended by the eye care professional.
Sauflon Clariti (somofilcon A) Soft (hydrophilic) Contact lens with UV blocker and handling tint helps protect against transmission of harmful UV radiation to the cornea and into the eve.
{3}------------------------------------------------
The sponsor considers the SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint to be substantially equivalent to the Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear which has been approved pursuant to K130342, and Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear which has been approved pursuant to K033919/K073459.
The following table summarises the primary features for this comparison, illustrating the similarities and differences.
{4}------------------------------------------------
Table 6.1. Comparison of Physical Properties for the SAUFLON CLARITI (Somofilion A) Soft (Hydrophility) Contact Lens with
| | | | Blocker and Handling Tint versus Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) and Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) Visibility<br>Tinted Contact Lenses for Daily Wear |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | PREDICATE DEVICE - SAUFLON<br>CLARITI WITH UV BLOCKER<br>(K130342) | PREDICATE DEVICE – AIR OPTIX<br>(K033919/K073459) | SUBJECT DEVICE - SAUFLON CLARITI WITH<br>UV BLOCKER AND HANDLING TINT |
| LENS MATERIAL | Somofilcon A Silicone Hydrogel | Lotrafilcon B Silicone Hydrogel | Somofilcon A Silicone Hydrogel |
| INDICATIONS FOR<br>USE | Daily wear monthly replacement | Daily wear monthly replacement | Daily wear monthly replacement |
| MANUFACTURING<br>PROCESS | Cast Moulding | Cast Moulding | Cast Moulding |
| WATER CONTENT | 56% | 33% | 56% |
| REFRACTIVE<br>INDEX | 1.40 | 1.42 | 1.40 |
| LIGHT<br>TRANSMITTANCE | ≥96% | ≥96% | ≥96% |
| DK @35°C (EDGE<br>CORRECTED) | 60 (polarographic method) | 110 (Coulometric method) | 60 (polarographic method) |
| COLOUR | No Visibility Tint | Blue Visibility Tint | Blue Visibility Tint |
| TINT | | | |
| TINT | No Visibility Tint | Copper phthalocyanine | D&C Green 6 |
| UV BLOCKER | Benzophenone | None | Benzophenone |
| MODULUS (MPa) | 0.55 | 0.92 | 0.55 |
| TENSILE STRENGTH<br>(MPa) | 1.05 | 0.9 | 1.05 |
| ELONGATION AT<br>BREAK % | 163 | 205 | 163 |
| PACKAGING<br>MATERIALS | Injected molded polypropylene blisters or<br>Sabic PP PCGH19 covered by aluminium<br>foil laminate and blister strips are packed<br>into printed cartons | Injected molded polypropylene blisters covered by<br>aluminium foil laminate and blister strips are<br>packed into printed cartons | Injected molded polypropylene blisters or Sabic PP<br>PCGH19 covered by aluminium foil laminate and blister<br>strips are packed into printed cartons |
| PACKAGING<br>SOLUTION | Borate buffered saline solution containing<br>0.005% poloxamer 407. | Phosphate buffered saline solution | Borate buffered saline solution containing 0.005%<br>poloxamer 407. |
| PACKAGING<br>METHOD | Hermetically sealed blister pack | Hermetically sealed blister pack | Hermetically sealed blister pack |
{5}------------------------------------------------
SAUFLON CLARITI (SOMOFILCON A) SOFT (HYDROPHILIC) CONTACT LENS WITH UV BLOCKER AND HANDLING
TINT
{6}------------------------------------------------
### 6.- PHYSICOCHEMICAL STUDIES
The physical, optical and chemical properties of the lenses as assessed by various test methods show substantial equivalence with the predicate devices as illustrated in the preceding table. Studies were also conducted to verify that leachable substances were either low or below measurable levels to assuage any concerns for its intended use.
### 7.- TOXICOLOGY STUDIES
Toxicology testing demonstrates that Clariti (somofilcon A) Soft (Hydrophilic) with UV blocker and handling tint is non-toxic in all cytotoxicity, ocular irritation and acute systemic toxicity tests. This lens is therefore equivalent in terms of biocompatibility to the predicate devices specified.
### 8.- EVALUATION OF DYE LEACHABILITY
Dye leachability was evaluated and no dye was observed to leach from Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Contact Lenses with UV Blocker and handling tint after 14 days soaking in saline at 37°C.
### 9.- HUMAN CLINICAL STUDIES
A clinical study was conducted to evaluate the safety and efficacy of SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) with UV Blocker and handling tint by comparison with Air Optix Aqua hydrophilic contact lenses (Ciba Vision Inc.). Subjects used OptiFree Replenish solution (Alcon Laboratories Inc.) for daily lens maintenance, care and storage. The results of this study showed the safety, acceptability and substantial equivalence of the Sauflon CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint to the predicate device for its intended use.
### 10.- CONCLUSIONS
Based on the above evaluations the SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint is substantially equivalent to the predicate, marketed lenses. Based on these evaluations the SAUFLON CLARITI (somofilcon A) Soft (hydrophilic) Contact Lens with UV Blocker and handling tint has been shown to be safe and effective for its intended use.
{7}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
March 5, 2014
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Sauflon Pharmaceuticals Ltd. Dr. Christopher Smeikal Head of Regulatory and Technical Affairs 49-53 York Street Twickenham. Middlesex TW1 3LP, United Kingdom
Re: K133740
Trade Name: Sauflon Clariti (somofilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker and Handling Tint Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: II Product Code: LPL Dated: January 24, 2014 Received: January 29, 2014
Dear Dr. Smejkal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{8}------------------------------------------------
Page 2 - Dr. Christopher Smejkal
You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
and and the comments of the comments of the
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D.
- Director
..............................................................................................................................................................................
and the commended to the comments of 1000 - 1000
Division of Ophthalmic and Ear,
Nose and Throat Devices . . . . . . .
. . . . . . . .
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
,
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K133740
### Device Name
Sauflon Clariti (somofileon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint.
Indications for Use (Describe)
The SAUFLON CLARITI (somofilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.
The SAUFLON CLARITI TORIC (somofile) Contact Leas with UV blocker and handling tint is indicated for daily woar for the optical corrective anteropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.
The SAUFLON CLARITI MULTIFOCAL (somofileon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive anetropia (myopia and or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +4.00 Diopters or less.
The SAUFLON CLARITI MULTIFOCAL TORIC (somofilion A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and that may require a reading addition of +4.00 Diopters or less.
The lenses may be prescribed for daily wear for up to one month with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion as recommended by the eye care professional.
Saullon Claili (somofilcon A) Soll (hydrophilic) Contact lens with UV blocker and handling tint helps protect against transmission of harmful UV radiation to the cornea and into the eye. The lens has a itint added to make the lens more visible for handling.
Type of Use (Select one or both, as applicable)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ADDITIONAL SHEET
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Kealth (CDRH) (Signature)
> Prescription Use (Part 21 CFR 801 Subpart D)
# Marc W. Robboy -S 2014.02.18 09:13:27 -05'00'
FORM FDA 3881 (1/14)
::
Page 1 of 2
{10}------------------------------------------------
### This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the deta needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.