RESOLUTION 45 (SPHERICAL) AND ULTRA GEL (TORIC) (HIOXIFILCON B) SOFT DAILY WEAR CONTACT LENS, )CLEAR & BLUE VISIBILITY T

K983496 · Opti-Centre Laboratories · LPL · Mar 29, 1999 · Ophthalmic

Device Facts

Record IDK983496
Device NameRESOLUTION 45 (SPHERICAL) AND ULTRA GEL (TORIC) (HIOXIFILCON B) SOFT DAILY WEAR CONTACT LENS, )CLEAR & BLUE VISIBILITY T
ApplicantOpti-Centre Laboratories
Product CodeLPL · Ophthalmic
Decision DateMar 29, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Intended Use

The Resolution 45 (Spherical) (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The Ultra Gel (Toric) (hioxifileon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 diopters.

Device Story

Resolution 45 (Spherical) and Ultra Gel (Toric) are daily wear soft contact lenses fabricated from hioxifilcon B polymer; lathe-cut from lens blanks. Lenses are hydrophilic; 48% water content when hydrated; conform to corneal curvature to provide refractive correction. Used by patients with myopia, hyperopia, or astigmatism; prescribed by eye care professionals. Lenses function as refractive media to focus light on retina. Benefits include visual acuity correction for aphakic and non-aphakic patients.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, material characterization, and comparison of physical properties (water content, refractive index, oxygen permeability, light transmission) against predicate devices.

Technological Characteristics

Material: hioxifilcon B (hydrophilic polymer). Water content: 48% ±2%. Oxygen permeability (Dk): 15 x 10^-11 (cm2/sec)(ml O2/ml x mm Hg @ 35°C). Refractive index: 1.404 (hydrated). Specific gravity: 1.136 (hydrated). Light transmission: >95%. Production: Lathe-cut. Form factor: Spherical and Toric soft contact lenses.

Indications for Use

Indicated for correction of visual acuity in aphakic and non-aphakic persons with non-diseased eyes with myopia or hyperopia. Resolution 45 (Spherical) indicated for refractive astigmatism ≤ 0.75 diopters. Ultra Gel (Toric) indicated for refractive astigmatism ≤ 10 diopters.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 3/29/99 # 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### Applicant information: Date Prepared: October 1, 1998 | Name: | Opti-Center Laboratories Inc. | |-----------------|------------------------------------| | Address | 4375 Ouimet Street | | | Sherbrooke (Quebec) Canada JIL 1X5 | | Contact Person: | Robert Mercure | | Phone number: | (819) 564-8114 | | USA Consultant: | Med-Vice Consulting, Inc. | |-----------------|---------------------------| | | Martin Dalsing | | Phone number | (970) 243-5490 | #### Device Information: | Device Classification: | Class II | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Classification Number: | LPL | | Classification Name: | Lenses, Soft Contact, Daily Wear | | Trade Name: | Resolution 45 (Spherical) and Ultra Gel (Toric)<br>(hioxifilcon B) Soft Daily Wear contact Lens, (Clear & Blue<br>Visibility Tint, Lathe-cut from Lens Blank) | {1}------------------------------------------------ #### Equivalent Devices: The Resolution 45 (Spherical) and Ultra Gel (Toric) (hioxifilcon B) Soft Daily Wear Contact Lenses are substantially equivalent to the following predicate devices in terms of intended use and design. Predicate devices include: "Horizon 55 Spherical & Westhin Toric" manufactured by Westcon Contact Lens, and the "BENZ-G 3X" manufactured by Benz Research and Development. #### Device Description: The Resolution 45 (Spherical) and Ultra Gel (Toric) (hioxifileon B) Soft Daily Wear Contact Lenses are fabricated from hioxifilcon B, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (hioxifilcon B) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution. The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 48% water by weight. The physical properties of the lens are: | Refractive Index | 1.515 (dry) 1.404 (hydrated) | |-----------------------------|---------------------------------------------------------------------------| | Color Pigment Name | Phthalocyanine Blue | | Light Transmission (clear) | greater than 95% T | | Light Transmission (tinted) | greater than 95% T | | Water Content | 48 % ±2% | | Specific Gravity | 1.308 (dry) 1.136 (hydrated) | | Oxygen Permeability | 15 X 10-11 (cm2/sec) (ml O2/ml x mm Hg @ 35°C), (revised Fatt<br>method). | #### Intended Use: The Resolution 45 (Spherical) (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The Ultra Gel (Toric) (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 diopters. {2}------------------------------------------------ #### Substantial Equivalence: The device will be manufactured according to specified process controls and a quality assurance program. The device will undergo manufacturing, packaging and sterilization procedures similar to devices currently marketed and distributed by Opti-Center Laboratories, Inc. The established safety profile (preclinical toxicology and manufacturing/chemistry data) of the device is equivalent to the BENZ-G 3X, 510(k) #K964528. Being similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate devices, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate devices identified above. The following matrix illustrates the production method, lens function and material characteristics of the Resolution 45 (Spherical) and Ultra Gel (Toric) (hioxifilcon B) Soft Daily Wear Contact Lens (Clear & Blue Visibility Tint, Lathe-cut from Lens Blank), as well as the predicate devices. | | Characteristic | Resolution 45<br>(Spherical) | Ultra Gel (Toric) | PREDICATE<br>DEVICES | |-----|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1.) | PRODUCTION<br>METHOD | Lathe-Cut | Lathe-Cut | Lathe-Cut | | 2.) | LENS FUNCTION | Refractive medium that<br>focuses light rays from<br>distant, intermediate and<br>near objects on the<br>retina, while<br>compensating for<br>refractive error. | Refractive medium that<br>focuses light rays from<br>near, intermediate and<br>distant objects on the<br>retina, while<br>compensating for<br>refractive error. | Refractive medium that<br>focuses light rays from<br>near, intermediate and<br>distant objects on the<br>retina, while<br>compensating for<br>refractive error. | | 3.) | MATERIAL | Hydrophilic Polymer | Hydrophilic Polymer | Hydrophilic Polymer | | a. | Water Content | 48% | 48% | 55%, 48% | | b. | Polymer Content | 52% | 52% | 45%, 52% | | c. | Polymer | hioxifilcon B | hioxifilcon B | methafilcon A,<br>hioxifilcon B | | d. | DK Value | 15 | 15 | 18.2, 15 | | e. | Refractive Index | 1.404 (hydrated) | 1.404 (hydrated) | 1.408, 1.404 | | f. | Specific Gravity | 1.136 (hydrated) | 1.136 (hydrated) | 1.090, 1.136 | | g. | Light Transmission | greater than 95% T | greater than 95% T | greater than 95% T | ### Substantial Equivalence Matrix {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with three horizontal lines above them, resembling a bird in flight. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 9 1999 Mr. Martin Dalsing Official FDA Correspondent Opti-Center Laboratories 623 Glacier Drive Grand Junction, CO 81503 Re: K983496 Trade Name: Resolution 45 (Spherical) and Ultra Gel (Toric) (hioxifilcon B) Soft Daily Wear Contact Lens (Clear & Blue Visibility Tint, Lathe-cut from Lens Blank) Regulatory Class: II Product Code: 86 LPL Dated: January 20, 1999 Received: January 26, 1999 Dear Mr. Dalsing: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations . {4}------------------------------------------------ Page 2 - Mr. Martin Dalsing · This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, A. Roeyl. Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health ...... . Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT #### Resolution 45 (Spherical) and Ultra Gel (Toric) (hioxifilcon B) Soft Daily Wear Device Name: Contact Lens (Clear & Blue Visibility Tint, Lathe-cut from Lens Blank) #### INDICATIONS FOR USE: The Resolution 45 (Spherical) (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The Ultra Gel (Toric) (hioxifileon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 diopters. #### (PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Ophthalmic Devices 510(k) Number K983496 Prescription Use (Per 21 CFR 801.109) or Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...