Bausch + Lomb (Kalifilcon A) Soft Contact Lens, Bausch + Lomb (Kalifilcon A) Soft Contact Lens for astigmatism

K200528 · Bausch + Lomb, Incorporated · LPL · Jun 2, 2020 · Ophthalmic

Device Facts

Record IDK200528
Device NameBausch + Lomb (Kalifilcon A) Soft Contact Lens, Bausch + Lomb (Kalifilcon A) Soft Contact Lens for astigmatism
ApplicantBausch + Lomb, Incorporated
Product CodeLPL · Ophthalmic
Decision DateJun 2, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Intended Use

Bausch + Lomb kalifilcon A Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D. Bausch + Lomb kalifilcon A Contact Lens for Astigmatism is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters. The kalifilcon A contact lens is to be prescribed for single-use disposable wear and is to be discarded after each removal.

Device Story

Hydrophilic silicone hydrogel contact lens (kalifilcon A); 55% water content; UV-blocking monomer; tinted with Reactive Blue Dye 246. Acts as refractive medium to focus light on retina. Prescribed by eye care professionals for daily wear; single-use disposable. Packaged in phosphate buffered saline. Provides vision correction for ametropia and astigmatism; improves patient visual acuity.

Clinical Evidence

3-month controlled clinical study with ~247 patients comparing kalifilcon A to Biotrue ONEday (nesofilcon A). Primary safety endpoint: slit lamp findings > Grade 2. Primary efficacy endpoint: contact lens corrected distance high-contrast visual acuity. All primary endpoints achieved; results demonstrate safety and effectiveness equivalent to predicate.

Technological Characteristics

Material: Hydrophilic siloxane copolymer (kalifilcon A); 55% water content. Oxygen permeability: 107 x 10-11 [cm3O2(STP) x cm]/(sec x cm2 x mmHg) @ 35°C. UV-blocking: <5% UVB (280-315nm), <50% UVA (316-380nm). Refractive index: 1.4011. Manufacturing: Cast molded. Single-use disposable. Sterilized.

Indications for Use

Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic/non-aphakic persons with non-diseased eyes. Spherical powers +20.00D to -20.00D; astigmatism correction up to 5.00 diopters. Single-use disposable.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue text. June 02, 2020 Bausch + Lomb, Incorporated Barbara Klube-Falso Director, Regulatory Affairs 1400 North Goodman Street Rochester, NY 14609 Re: K200528 Trade/Device Name: Bausch + Lomb (kalifilcon A) Soft (hydrophilic) Contact Lens Bausch + Lomb (kalifilcon A) Soft (hydrophilic) Contact Lens for Astigmatism Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: Mav 4, 2020 Received: May 6, 2020 Dear Barbara Klube-Falso: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling {1}------------------------------------------------ (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200528 #### Device Name Bausch + Lomb (kalifilcon A) Soft (hydrophilic) Contact Lens Bausch + Lomb (kalifilcon A) Soft (hydrophilic) Contact Lens for Astigmatism ## Indications for Use (Describe) Kalifilcon A Contact Lens Bausch + Lomb kalifilcon A Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D. Kalifilcon A Contact Lens for Astigmatism Bausch + Lomb kalifilcon A Contact Lens for Astigmatism is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic persons with non-diseased eves, exhibiting astigmatism of up to 5.00 diopters. The lens is to be prescribed for single-use disposable wear and is to be discarded after each removal. Type of Use (Select one or both, as applicable) | <div> <span> <svg height="16" width="16"> <path d="M13.172 4.828l-7.5 7.5-3.536-3.536 1.414-1.414 2.122 2.121 6.086-6.085 1.414 1.414z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY K200528 ## Submitter Information: | Date Prepared: | May 4, 2020 | |-----------------|-----------------------------------------------------| | Name: | Bausch & Lomb Incorporated | | Address: | 1400 North Goodman Street<br>Rochester, NY 14609 | | Contact Person: | Barbara Klube-Falso<br>Director, Regulatory Affairs | | Phone Number: | (585) 338-8503 | | Email: | Barbara.Klube-Falso@bausch.com | ### Device Information: | Trade Name: | Bausch + Lomb (kalifilcon A) Soft (hydrophilic) Contact Lens | |-------------|------------------------------------------------------------------| | | Bausch + Lomb (kalifilcon A) Soft (hydrophilic) Contact Lens for | | | Astigmatism | Regulation Name: Soft (hydrophilic) Contact Lens Device Classification: Class II (21 CFR 886.5925) #### Predicate Device: Bausch + Lomb kalifilcon A Contact Lens is substantially equivalent to the following predicate device: Bausch + Lomb Biotrue ONEday (nesofilcon A) Contact Lens cleared under K113703. #### Device Description : The Bausch + Lomb kalifilcon A material is made from a hydrophilic siloxane copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone and is 55% water by weight when immersed in a sterile phosphate buffered saline with 0.5% poloxamine solution. A UVabsorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280nm to 315nm and less than 50% in the UVA range of 316nm to 380nm. This lens is tinted blue with Reactive Blue Dye 246. The Bausch + Lomb kalifilcon A Contact Lens is to be prescribed for single-use disposable wear. {4}------------------------------------------------ The physical properties of the lens are: | Refractive index | 1.4011 | |---------------------|--------------------------------------------------------------------------------| | Light transmission | 97% | | Water Content | 55% | | Specific Gravity | 1.029 | | Oxygen Permeability | 107 x 10-11[cm³O₂(STP) x cm]/(sec x cm² x<br>mmHg)@35°C (polarographic method) | The lenses will be manufactured with the following parameters: | Diameter | 13.5mm to 15.0mm | |------------------------|--------------------| | Center Thickness | 0.05mm to 0.75mm | | Base Curve | 7.8mm to 9.5mm | | Power Range | +20.00D to -20.00D | | Cylinder Power (Toric) | -0.75D to -5.00D | | Cylinder Axis | 0° to 180º | The lenses are packaged in disposable blister packages containing phosphate buffered saline solution. Blister packages are labeled with lot number, expiration date and applicable lens parameters. Expiration dating is supported by product stability, package integrity, and validation of the sterilization process. #### Intended Use: ## Kalifilcon A Contact Lens Bausch + Lomb kalifilcon A Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D. ## Kalifilcon A Contact Lens for Astigmatism Bausch + Lomb kalifilcon A Contact Lens for Astigmatism is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters. The kalifilcon A contact lens is to be prescribed for single-use disposable wear and is to be discarded after each removal. {5}------------------------------------------------ ## Technological Characteristics (comparison to predicate devices) A side-by-side comparison of the predicate devices to the new device: | Property | Predicate Device<br>Bausch + Lomb<br>Biotrue ONEday (nesofilcon A)<br>Contact Lens | Subject Device<br>Bausch + Lomb<br>kalifilcon A Contact Lens | |---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------| | Intended Use | Indicated for the daily wear<br>correction of refractive<br>ametropia (myopia, hyperopia<br>and astigmatism) in aphakic<br>and/or non-aphakic persons<br>with non-diseased eyes that<br>exhibit refractive astigmatism up<br>to 2.00 diopters or less, that<br>does not interfere with visual<br>acuity. The lens is to be<br>prescribed for single-use<br>disposable wear and is to be<br>discarded after each removal. | Same as predicate | | Functionality | The contact lenses act as a<br>refractive medium that focus<br>light rays from near and distant<br>objects on the retina. | Same as predicate | | Modality | Daily wear contact lens | Same as predicate | | Manufacturing<br>Method | Cast Molded | Same as predicate | | Material Group | Group II<br>Hydrogel<br>(high water, non-ionic) | Group 5-B<br>Silicone Hydrogel<br>(high water, non-ionic) | | USAN Name | nesofilcon A | kalifilcon A | | Water Content | 78% | 55% | | Oxygen<br>Permeability1<br>(edge corrected) | 42 | 107 | | Specific Gravity | 1.039 | 1.029 | | UV Blocker | Yes | Yes | 1- Oxygen Permeability shown was determined by the polarographic method: x 10-11[cm³O₂(STP) x cm]/(sec x cm² x mmHg)@35°C {6}------------------------------------------------ ### Summary of Non-Clinical Testing: The testing performed on the Bausch + Lomb kalifilcon A Contact Lens demonstrated that the device functions in a safe and effective manner. Performance testing included conformance to predetermined specifications, functional test results verify that the device performs as expected and is equivalent to the predicate without creating additional risk to the user. In addition, Bausch + Lomb followed the Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses, May 1994, the following tests were conducted: Toxicology / Biocompatibility In-Vitro Cytotoxicity Ocular Irritation Study Systemic Toxicity Chemistry / Leachables Physical, Chemical and Spectral Properties Leachable Monomer and Additives ### Summary of Clinical Performance Data Bausch + Lomb conducted a 3 month, controlled clinical study with approximately 247 patients, comparing the safety and efficacy of the Bausch + Lomb kalifilcon A Contact Lens to the Bausch + Lomb Biotrue ONEday (nesofilcon A) Contact Lens. The primary safety endpoint was any slit lamp finding greater than Grade 2 over the course of the study. the primary efficacy endpoint was contact lens corrected distance high-contrast visual acuity averaged over all scheduled visits. All primary endpoints were achieved, and the results of the study indicate the test lens is safe and effective, #### Substantial Equivalence Conclusion: The cumulative results of laboratory, in vitro, in vivo testing as well as the clinical study sponsored by Bausch + Lomb demonstrate that the safety. effectiveness and performance of the Bausch + Lomb kalifilcon A Contact Lens are substantially equivalent to the Bausch + Lomb Biotrue ONEday (nesofilcon A) Contact Lens.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...