The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens is indicated for the correction of refractive ametropia (myopia) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens helps protect against transmission of harmful UV radiation to the cornea and into the eye. The eye care professionals may prescribe the lens for single use daily disposable or daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disinfected using a chemical or hydrogen peroxide disinfecting systems.
Device Story
Soft hydrophilic contact lens; manufactured via cast molding; material Methafilcon A (HEMA/MAA copolymer crosslinked with EGDMA); contains UV-absorbing monomer and C.I. Reactive Blue No.4 tint. Supplied in blister packs with sterile isotonic phosphate buffered saline containing sodium hyaluronate and trehalose wetting agents. Prescribed by eye care professionals for daily disposable or frequent replacement wear. Corrects myopia and astigmatism; blocks UV radiation. Patient self-administers lens; performs disinfection for planned replacement schedules using chemical or hydrogen peroxide systems.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing including acute systemic injection, white rabbit ocular irritation, cytotoxicity (ISO 10993-5), extractables, finished lens parameters, light transmittance, refractive index, water content, shelf-life, and mechanical properties.
Technological Characteristics
Material: Methafilcon A (HEMA/MAA copolymer crosslinked with EGDMA); UV-absorbing monomer; C.I. Reactive Blue No.4 tint. Water content: 55%. Manufacturing: Cast molding. Dimensions: Base curve 7.85-10.00 mm, Diameter 12.00-15.00 mm. Refractive index: 1.409. Packaging: Phosphate buffered saline with sodium hyaluronate and trehalose. Sterilization: Not specified.
Indications for Use
Indicated for correction of refractive ametropia (myopia) in aphakic and non-aphakic persons with non-diseased eyes; patients with astigmatism of 2.00 diopters or less; provides UV radiation protection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
K161739 — Aquamax (Etafilcon A) Bi-Weekly Soft, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses · Pegavision Corporation · Dec 19, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 10, 2015
Yung Sheng Optical Co., Ltd. Mr. Wen-Han Chen RA Supervisor 3f-1, No. 6, Jhongke Road, Daya District Taichung City, TW 42881
Re: K151586
Trade/Device Name: Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: June 9, 2015 Received: June 11, 2015
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K151586
Device Name
Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens
### Indications for Use (Describe)
The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact for the correction of refractive ametropia (myopia) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens helps protect against transmission of harmful UV radiation to the cornea and into the eye.
The eye care professionals may prescribe the lens for single use daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disintected using a chemical or hydrogen peroxide disinfecting systems.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
- Traditional 510(k) 1. Type of Submission:
| 2. | <b>Submitter:</b> | Yung Sheng Optical Co., Ltd. |
|----|-------------------------------------------|------------------------------------------------------------------------|
| | <b>Address:</b> | 3F-1, No.6, Jhongke Road, Daya District, Taichung<br>City 42881 Taiwan |
| | <b>Manufacturing<br/>facility:</b> | No.8, Keya 2nd Road, Daya District, Taichung City<br>42881, Taiwan |
| | <b>Phone:</b> | (04) 25658384 #156 、 191 |
| | <b>Fax:</b> | (04) 25658387 |
| | <b>Contact:</b> | Wen-Han Chen / Che-Yu Shen |
| | <b>Date prepared:</b> | May 29, 2015 |
| | <b>Establishment Registration Number:</b> | 3004021238 |
#### 3. Identification of the Device
| Proprietary/Trade<br>name: | Eye Secret 55 UV Aspheric (Methafilcon A)<br>Soft (hydrophilic) Contact Lens | |
|----------------------------|------------------------------------------------------------------------------|--|
| Common Name: | Contact Lens | |
| Classification Name: | Lenses, Soft Contact, Daily Wear | |
| Device Classification: | II | |
| Regulation Number: | 886.5925 | |
| Panel: | Ophthalmic | |
| Product Code: | LPL for Lenses, Soft Contact, Daily Wear | |
| | MVN for Lens, Contact, (Disposable) | |
#### 4. Identification of the Predicate Device
| Predicate Device Name: | Eye Secret 55 UV Aspheric (Methafilcon A) Soft<br>(hydrophilic) Contact Lens |
|-----------------------------------------|------------------------------------------------------------------------------|
| Manufacturer: | Yung Sheng Optical Co., Ltd. |
| 510(k) Number or Clearance Information: | K133735 |
{4}------------------------------------------------
Yung Sheng Optical Co., Ltd. 510(k) notification
### 5. Intended Use and Indications for Use of the subject device
The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens is indicated for the correction of refractive ametropia (myopia) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens helps protect against transmission of harmful UV radiation to the cornea and into the eye.
The eye care professionals may prescribe the lens for single use daily disposable or daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disinfected using a chemical or hydrogen peroxide disinfecting systems.
### 6. Device Description
The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens is manufactured by using cast molding method. The lens material, Methafilcon A, is a random copolymer of 2-hydroxyethyl methacrylate (HEMA) and Methacrylic acid (MAA) crosslinked with ethylene glycol dimethacrylate (EGDMA). A UV absorbing monomer is used to block UV radiation. The average transmittance characteristics are less than 5% in the UVB range of 280 to 315 nm and less than 50% in the UVA range of 316 to 380 nm. The lenses are tinted blue for visibility purposes with the color additives, C.I. Reactive Blue No.4.
The lenses are available as aspheric lenses. Each finished lens is supplied in a plastic blister container with a sterile isotonic phosphate buffered saline solution containing sodium hyaluronate and trehalose wetting agents.
### 7. Summary of Clinical Study
Methafilcon A lenses have been used widely. Their safety and effectiveness have been well documented and cleared by FDA "Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens (K133735)" submitted by Yung Sheng Optical Co., Ltd. is an example.
{5}------------------------------------------------
Yung Sheng Optical Co., Ltd. 510(k) notification
Clinical study for the subject device is not required for the premarket notification as the USAN name and process are the same as the above mentioned predicate device.
### 8. Non-Clinical Testing
A series of preclinical testing were performed to demonstrate the safety and effectiveness of the Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens. The results of all testing demonstrated that the safety and effectiveness of the subject device are equivalent to the predicate device, Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens (K133735). The following tests were conducted as recommended by the FDA Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, 1994:
- · Toxicity
- 1. Acute Systemic Injection Study: The lens material meets the requirements of the systemic injection test and is considered non-toxic.
- 2. White Rabbit Ocular Irritation Test: Ocular irritation test was performed and produced no ocular irritation.
- 3. Cytotoxicity Test: The test article meets the requirements of ISO 10993-5.
- Extractables (Leachability)
- Finished Lens Parameters .
- · Light Transmittance
- Refractive Index .
- . Water Content
- . Shelf-life
- Mechanical Properties Comparative Testing .
- . Physical Compatibility Test with Contact Lens Care Solution and Packaging Solution
The results of the non-clinical testing demonstrated that the Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens is substantially equivalent to the predicate device.
{6}------------------------------------------------
Yung Sheng Optical Co., Ltd. 510(k) notification
## 9. Substantial Equivalence Determination
The information presented in this submission establishes the substantial equivalence of the subject device "Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens" to the predicate device. The following table summarized the substantial equivalence comparison information.
| | Subject Device | K133735 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Eye Secret 55 UV Aspheric<br>(Methafilcon A) Soft (hydrophilic)<br>Contact Lens | Eye Secret 55 UV Aspheric<br>(Methafilcon A) Soft (hydrophilic)<br>Contact Lens |
| Intended Use | The Eye Secret 55 UV Aspheric<br>(Methafilcon A) Soft (hydrophilic)<br>Contact Lens is indicated for the<br>correction of refractive ametropia<br>(myopia) in aphakic and<br>not-aphakic persons with<br>non-diseased eyes. The lenses may<br>be worn by person who exhibit<br>astigmatism of 2.00 diopters or less<br>that does not interfere with visual<br>acuity. The Eye Secret 55 UV<br>Aspheric (Methafilcon A) Soft<br>(hydrophilic) Contact Lens helps<br>protect against transmission of<br>harmful UV radiation to the cornea<br>and into the eye. The eye care<br>professionals may prescribe the<br>lens for single use daily disposable<br>or daily wear in a Frequent<br>Replacement Program. As<br>prescribed for planned<br>replacement, the lens should be<br>disinfected using a chemical or<br>hydrogen peroxide disinfecting<br>systems. | The Eye Secret 55 UV Aspheric<br>(Methafilcon A) Soft (Hydrophilic)<br>Contact lenses are indicated for the<br>correction of ametropia (myopia) in<br>aphakic and non-aphakic persons with<br>non-diseased eyes in powers from 0.00<br>to -12.00 diopters. The lenses may be<br>worn by persons who exhibit<br>astigmatism of 2.00 diopters or less<br>that does not interfere with visual<br>acuity. Eye Care professionals may<br>prescribe the lens for daily disposable<br>or daily wear in a Frequent<br>Replacement Program. As prescribed<br>for planned replacement, the lens<br>should be disinfected using a chemical<br>or hydrogen peroxide disinfecting<br>systems. |
| Material | Methafilcon A | Methafilcon A |
| Manufacturing<br>Method | Cast Molded | Cast Molded |
| Water Content | 55 % | 55 % |
| | Subject Device | K133735 |
| Device Name | Eye Secret 55 UV Aspheric<br>(Methafilcon A) Soft (hydrophilic)<br>Contact Lens | Eye Secret 55 UV Aspheric<br>(Methafilcon A) Soft (hydrophilic)<br>Contact Lens |
| Powers | -0.50 D ~ -20.00 D | -0.00 ~ -12.00 D |
| Light<br>Transmittance | 95 % (± 5 %) | 95 % (± 5 %) |
| UV-A | < 50 % | < 50 % |
| UV-B | < 5 % | < 5 % |
| Refractive<br>Index | 1.409 (± 0.005 nd) | 1.409 (± 0.005 nd) |
| Base Curve | 7.85 ~ 10.00 (± 0.20 mm) | 8.60 (± 0.20 mm) |
| Diameter | 12.00 ~ 15.00 (± 0.20 mm) | 14.20 (± 0.20 mm) |
| Tint | C.I. Reactive Blue #4 for handling purpose | C.I. Reactive Blue #4 for handling purpose |
| Packaging<br>solution | Phosphate buffered saline solution containing sodium hyaluronate and trehalose. | Phosphate buffered saline solution. |
{7}------------------------------------------------
Yung Sheng Optical Co., Ltd. 510(k) notification
Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens
# 10. Conclusion
After analyzing the bench tests and safety testing data, it can be concluded that the Eye Secret 55 UV Aspheric (Methafilcon A) Soft (hydrophilic) Contact Lens is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.