SAGE Vitrification Kit (ART-8025 and ART-8026) / SAGE Vitrification Warming Kit (ART-8030 and ART-8031)

K173731 · CooperSurgical, Inc. · MQL · Mar 2, 2018 · Obstetrics/Gynecology

Device Facts

Record IDK173731
Device NameSAGE Vitrification Kit (ART-8025 and ART-8026) / SAGE Vitrification Warming Kit (ART-8030 and ART-8031)
ApplicantCooperSurgical, Inc.
Product CodeMQL · Obstetrics/Gynecology
Decision DateMar 2, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6180
Device ClassClass 2

Intended Use

SAGE Vitrification Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. SAGE Vitrification Warming Kit is intended for use in the thawing of virtified oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.

Device Story

SAGE Vitrification and Warming Kits are chemical media solutions used in IVF laboratories to cryopreserve and thaw human oocytes and embryos. The vitrification kit uses equilibration and vitrification solutions containing cryoprotectants (DMSO, ethylene glycol, sucrose) to prepare cells for ultra-rapid cooling. The warming kit uses a series of sucrose-based solutions to rehydrate and remove cryoprotectants from vitrified cells. Used by embryologists in clinical settings; the media supports cell survival during the transition between frozen and physiological states. By enabling successful cryopreservation, the device allows for delayed embryo transfer, potentially reducing risks like Ovarian Hyperstimulation Syndrome (OHSS) and improving clinical pregnancy outcomes.

Clinical Evidence

Clinical evidence provided via literature review of 9 studies. Studies assessed survival, fertilization, and pregnancy rates for vitrified/warmed oocytes, PN zygotes, cleavage-stage embryos, and collapsed blastocysts. Oocyte survival rates ranged from 75.0% to 94.1%; fertilization rates 64.2% to 77.7%; clinical pregnancy rates 33.3% to 40.9%. PN zygote survival was 89.0%–98.6% with pregnancy rates 21.6%–28.2%. Cleavage-stage embryo survival was 97.1% with 41.6% clinical pregnancy rate. Results demonstrate performance comparable to established clinical outcomes.

Technological Characteristics

Media solutions containing MOPS-buffered base, essential/non-essential amino acids, gentamicin sulfate, and human serum albumin. Cryoprotectants: DMSO, ethylene glycol, sucrose. Packaging: Type 1 borosilicate glass vials or polypropylene vials. Sterilization: Aseptic filtration. Performance specs: pH 7.20-7.40; osmolality varies by solution (ES: 2331-2849 mOsm/kg; VS: 5603-6849 mOsm/kg; WS: 257-1535 mOsm/kg).

Indications for Use

Indicated for vitrification and thawing of human oocytes (MII), pronuclear (PN) zygotes, day 3 cleavage-stage embryos, and blastocyst-stage embryos for use in assisted reproductive technology procedures.

Regulatory Classification

Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 2, 2018 CooperSurgical, Inc % Tove Kjær Director Corporate Regulatory Affairs ORIGIO a/s Knardrupvej 2 DK-2760 Målov Denmark Re: K173731 Trade/Device Name: SAGE Vitrification Kit (ART-8025 and ART-8026) SAGE Vitrification Warming Kit (ART-8030 and ART-8031) Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: Class II Product Code: MQL Dated: November 23, 2017 Received: December 6, 2017 Dear Tove Kjær: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K173731 Device Name SAGE Vitrification Kit (ART-8025 and ART-8026) SAGE Vitrification Warming Kit (ART-8030 and ART-8031) Indications for Use (Describe) SAGE Vitrification Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. SAGE Vitrification Warming Kit is intended for use in the thawing of virtified oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # a CooperSurgical Company Image /page/3/Picture/1 description: The image shows the word "origio" in a stylized, sans-serif font. The letters are a deep purple color. The letters are connected to each other, and the "g" is formed by a circle with a small line extending from the top. ## 510(k) SUMMARY K173731 #### I. Submitter information Submitted on behalf of: Mr. James Keller Vice President, Regulatory Affairs and Quality Assurance CooperSurgical, Inc. 95 Corporate Drive Trumbull, CT 06611 USA e-mail: James.Keller@coopersurgical.com Phone: +1 (203) - 601 - 5200 Contact person: Ms. Tove Kjær Director Corporate Regulatory Affairs ORIGIO a/s Knardrupvej 2 DK-2760 Måløv Denmark e-mail: tkjaer@origio.com Phone: +45 4679 0220 - II. Date prepared March 2, 2018 #### III. Device Information: | Device name: | SAGE Vitrification Kit (ART-8025 and ART-8026)<br>SAGE Vitrification Warming Kit (ART-8030 and ART-8031) | |------------------------|----------------------------------------------------------------------------------------------------------| | Common name: | Vitrification Kit and Vitrification Warming Kit | | Classification name: | Reproductive media and supplements | | Classification Number: | 21 CFR 884.6180 | | Product Code: | MQL (Media, Reproductive) | | Regulatory Class: | II | - Predicate device: Vit Kit Freeze (Vitrification Freeze Kit) IV. Vit Kit - Thaw (Vitrification Thaw Kit) K160006 The predicate device has not been subject to a design-related recall. #### V. Indication for use SAGE Vitrification Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. SAGE Vitrification Warming Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. {4}------------------------------------------------ #### VI. Device Description: ### SAGE Vitrification Kit (ART-8025 and ART-8026): This kit includes two solutions (Equilibration Solution and Vitrification Solutions in the kit consist of a MOPS-buffered media containing non-essential and essential amino acids. gentamicin sulfate, and human serum albumin. These solutions also contain the cryoprotectants ethylene glycol, dimethyl sulfoxide (DMSO) and sucrose as described below: Equilibration Solution (ES): 7.5% (v/v) each of DMSO and ethylene glycol. Vitrification Solution (VS): 15% (v/v) each of DMSO and ethylene glycol, and 0.6 M sucrose ### SAGE Vitrification Warming Kit (ART-8030 and ART 8031): This kit includes three solutions (1.0M Sucrose Warming Solution [1.0 M WS], 0.5M Sucrose Warming Solution [0.5 M WS], and MOPS Solution [MS]). All solutions in the kit consist of a MOPS-buffered media containing non-essential and essential amino acids, gentamicin sulfate, and human serum albumin. These solutions contain the cryoprotectant sucrose at the level described in their names (i.e., 1.0M, 0.5M, or no sucrose). Solutions in the SAGE Vitrification Kit and SAGE Vitrification Warming Kit are aseptically-filtered and are provided to users in either 5 ml Type 1 borosilicate glass vials with a rubber closure and an aluminum seal (ART-8025 and ART-8030) or 2 ml polypropylene vials with cap (ART-8026 and ART-8031). The solutions in ART-8025 and ART-8030 are single-use only devices, whereas solution in ART-8026 and ART-8031 are to be used within 7 days after opening. | | Subject device | | Predicate device<br>(K160006) | | Comparison | |-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Parameter | SAGE Vitrification<br>Kit<br>ART8025/<br>ART8026 | SAGE Vitrification<br>Warming Kit<br>ART8030/<br>ART8031 | Vit-Kit Freeze | Vit-Kit Thaw | N/A | | Indication for<br>Use | SAGE Vitrification<br>Kit is intended for<br>use in the<br>vitrification of<br>oocytes (MII),<br>pronuclear (PN)<br>zygotes through<br>day 3 cleavage<br>stage embryos<br>and blastocyst<br>stage embryos. | SAGE<br>Vitrification<br>Warming Kit is<br>intended for use<br>in the thawing of<br>oocytes (MII),<br>pronuclear (PN)<br>zygotes through<br>day 3 cleavage<br>stage embryos<br>and blastocyst<br>stage embryos. | Vit Kit-Freeze<br>(Vitrification<br>Freeze Kit) is<br>intended for<br>use in the<br>vitrification of<br>oocytes (MII),<br>pronuclear<br>(PN) zygotes<br>through day 3<br>cleavage stage<br>embryos and<br>blastocyst<br>stage embryos. | Vit Kit- Thaw<br>(Vitrification<br>Thaw Kit) is<br>intended for<br>use in the<br>thawing of<br>oocytes (MII),<br>pronuclear<br>(PN) zygotes<br>through day 3<br>cleavage<br>stage<br>embryos and<br>blastocyst<br>stage<br>embryos. | Same | | Product and performance specification | | | | | | | pH | 7.20-7.40 | 7.20-7.40 | 7.05-7.54 | 7.05-7.44 | Different - The<br>pH range is<br>larger for the | | | Subject device | | Predicate device<br>(K160006) | | Comparison | | Parameter | SAGE Vitrification<br>Kit<br>ART8025/<br>ART8026 | SAGE Vitrification<br>Warming Kit<br>ART8030/<br>ART8031 | Vit-Kit Freeze | Vit-Kit Thaw | N/A | | | | | | | predicate<br>devices than the<br>subject devices.<br>Differences in pH<br>do not raise<br>different Safety<br>and<br>effectiveness<br>(S&E) questions. | | Osmolality<br>(mOsm/kg) | ES:<br>2331-2849<br>VS:<br>5603-6849 | 1.0M WS<br>1255-1535<br>0.5 M WS<br>745-911<br>MS:<br>257-273 | ES:<br>1005-1445<br>VS:<br>1100-1588 | TS:<br>1732-1912<br>DS:<br>857-910<br>WS:<br>268-292 | Different -<br>Osmolality is<br>higher in the<br>subject<br>vitrification kit<br>solutions as<br>compared to the<br>predicate, while<br>the warming<br>solutions of the<br>subject and<br>predicate<br>devices are<br>similar.<br>Differences in<br>osmolality do not<br>raise different<br>S&E questions. | | Endotoxin<br>(EU/mL) | <0.5 | <0.5 | ≤0.6 | ≤0.6 | Similar | | MEA 1-cell (%<br>blastocysts at<br>96h) | ≥80% | ≥80% | ≥80% | ≥80% | Same | | Sterilization | Aseptic Filtration | Aseptic Filtration | Aseptic<br>Filtration | Aseptic<br>Filtration | Same | | Components | | | | | | | Cryoprotectants | Ethylene glycol<br>DMSO<br>Sucrose | Sucrose | Ethylene glycol<br>DMSO<br>Sucrose | Sucrose | Same | | Amino Acids | Yes | Yes | Yes | Yes | Same | | Antibiotics | Gentamicin | Gentamicin | Gentamicin | Gentamicin | Same | | Protein | Human Serum<br>Albumin | Human Serum<br>Albumin | Dextran Serum<br>Supplement | Dextran<br>Serum<br>Supplement | Different -<br>Serum used in<br>the subject and | | | Subject device | | Predicate device<br>(K160006) | | Comparison | | Parameter | SAGE Vitrification<br>Kit<br>ART8025/<br>ART8026 | SAGE Vitrification<br>Warming Kit<br>ART8030/<br>ART8031 | Vit-Kit Freeze | Vit-Kit Thaw | N/A | | | | | | | predicate media<br>are not the<br>same.<br>Differences in<br>serum do not<br>raise different<br>questions of<br>S&E. | | Base medium | Modified Human<br>Tubal Fluid<br>(MOPS-Buffered) | Modified Human<br>Tubal Fluid<br>(MOPS-Buffered) | Medium 199<br>(HEPES-<br>Buffered) | Medium 199<br>(HEPES-<br>Buffered) | Different - The<br>base media of<br>the subject and<br>predicate media<br>are not the<br>same.<br>Differences in<br>base media do<br>not raise<br>different<br>questions of<br>S&E. | #### VII. Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices {5}------------------------------------------------ {6}------------------------------------------------ As noted in the table above, the subject and predicate devices have the same intended use and are technologically comparable. Differences in technological characteristics noted above do not raise different questions of safety or effectiveness. #### Summary of Non-Clinical Performance Testing VIII. The following studies have been performed to support substantial equivalence to the predicate devices: - pH (USP<791>)- 7.2-7.4 . - Osmolality (USP<785>) ● - o ES: 2331-2849 - o VS: 5603-6849 - o 1.0 M WS: 1255-1535 - o 0.5 M WS: 745-911 - o MS: 257-273 - Endotoxin testing (USP<85>) <0.5 EU/ml ● - Mouse Embryo Assay (MEA) 1-Cell MEA: ≥80% blastocysts at 96h ● - Aseptic process validation testing meeting the requirements of EN ISO 13408-1:2015 and ● EN ISO 13408-2:2011. - Sterility testing (USP<71>) No microbial growth ● - Shelf-life testing ART-8025 and ART-8030 relied on testing provided in K073522 ● supporting a 52-week shelf-life as the media and media packaging remained unchanged. Real-time shelf-life testing was conducted on ART-8031 to ensure that the following product specifications were met at time zero and at the end of shelf-life (52 weeks) in closed vials and in vials that were opened and underwent simulated use procedures over the seven days prior to testing: {7}------------------------------------------------ - Osmolality o - Endotoxin o - MEA O - Sterility O #### IX. Clinical Performance Data Published journal articles assessing vitrification and warming media that were identical or similar to the subject devices were provided to support extension of the current indication for vitrifying and warming blastocyst stage embryos to also include vitrification and warming of human MII oocytes, pronuclear (PN) zygotes, and day 3 cleavage-stage embryos. In addition, literature was also provided to support use of the subject media for vitrification and warming of collapsed blastocysts in support of methods included in device labeling. The results included in these studies demonstrated that the subject media was capable of vitrifying and warming the oocytes and embryos as shown by their rates of survival, fertilization, and pregnancy/live birth rates. A summary of the supporting journal articles is shown below: ### Oocytes - Literature 1: Reports results after oocvte vitrification/warming using SAGE™ Vitrification Kit . and SAGE™ Vitrification Warming Kit. This study included 11 subjects who had their surplus MII oocytes vitrified. Post-warming results showed a survival rate of 94.1%, fertilization rate of 67.0%, and a clinical pregnancy rate of 36.4%. - Literature 2: Reports results after oocyte vitrification/warming using SAGE™ Vitrification Kit . and SAGE™ Vitrification Warming Kit. The study included six subjects with tubal factor infertility where surplus MII oocytes were vitrified. Post-warming results showed a survival rate of 75.0%, fertilization rate of 77.7%, and a 33.3% clinical pregnancy rate. - Literature 3: The performance of SAGE™ Vitrification Kit and SAGE™ Vitrification Warming . Kit for vitrification and warming of oocytes was shown to be in-line with general results reported after transfer of frozen/thawed oocytes showing a pregnancy rate of 33.6%. - . Literature 4: The control arm in this study used media and vitrification/warming procedures determined to be comparable to the subject devices. Results for subjects in the control arm of this study that assessed outcomes following vitrification and warming of 2353 MII oocytes showed an oocyte survival rate of 92.1%, clinical pregnancy rate per embryo transfer of 56.4%, and a live birth rate per cycle of 45.3%. - Literature 5: This study used media and vitrification procedures determined to be comparable to the subject devices. Clinical oocyte vitrification and warming using a closed vitrification device (54 study subjects and 413 MII oocytes) resulted in a survival rate of 90.5%, fertilization rate of 64.2%, and a clinical pregnancy rate of 40.9%. ### Pronuclear (PN) Zygotes - Literature 6: This study assessed the effectiveness of two vitrification and warming media systems. The primary media was determined to be comparable to the subject devices, while the second media assessed was a commercially available media (RapidVit-Cleave) that included different cryoprotectants than the subject devices. Post-warming in the primary media group including 37 subjects and 74 embryos showed a survival rate of 98.6%, with 97% of surviving embryos reaching the cleavage stage. The clinical pregnancy rate was 21.6%. The clinical pregnancy rate for the primary media system was comparable to that observed for the commercially-available comparator media (23.1%). {8}------------------------------------------------ - Literature 7: This study used media and vitrification procedures determined to be ● comparable to the subject devices. A total of 849 pronuclear-stage (PN) zygotes were vitrified during the study period. During this period. 103 cycles of cryopreserved embryo transfer were completed. In total, 339 PN zygotes were thawed resulting in an 89.0% survival rate. The clinical pregnancy rate obtained was 28.2%, resulting in 9 live birth events (18 infants) with 15 ongoing pregnancies. ### Cleavage -Stage Embryos - Literature 8: Reported prospective outcomes following vitrification and warming of cleavage ● stage embryos using the SAGE™ Vitrification Kit and SAGE™ Vitrification Warming Kit. The study included 24 warming cycles after fresh cycle cancellation due to risk of Ovarian Hyperstimulation Syndrome (OHSS). Results showed a post-warming survival rate of 97.1%, a clinical pregnancy rate of 41.6% and an implantation rate of 18.0%. ### Vitrification of Collapsed Blastocysts - Literature 9: Reports data after vitrification/warming of collapsed blastocysts generated ● from lower grade cleavage-stage embryos using the SAGE™ Vitrification Kit and SAGE™ Vitrification Warming Kit. Results from the control group (102 warming cycles) in this randomized controlled trial where blastocyst collapse was performed without additional zonal thinning showed a post-warming survival rate of 98.0% and a clinical pregnancy rate of 35.3%. #### X. Conclusion The results of the performance information summarized above demonstrated that the subject devices are as safe and effective as the predicate devices and supports substantial equivalence. #### References XI. - 1. Pregnancies and deliveries after injection of vitrified-warded oocytes with cryopreserved testicular sperm. (Selman et al, (2010) Fertility and Sterility Vol. 94(7):2927-2929). - 2. Ongoing pregnancies after vitrification of human oocytes using a combined solution of ethylene glycol and dimethyl sulfoxide. (Selman et al. (2006) Fertility 86(4):997-1000). - 3. Assisted reproductive technology in Europe, 2011: results generated from European registers by ESHRE. (Kupka et al (2016) Human Reproduction 31(2): 233-248). - 4. A combination of hydroxypropyl cellulose and trehalose as supplementation for vitrification of human oocytes: a retrospective cohort study. (Coello et al (2016) Journal of Assisted Reproduction Genetics 33(3): 413-421). - 5. Closed vitrification of human oocytes and blastocysts: outcomes from a series of clinical cases. (Gook et al, (2016) J Assist Reprod Genet 33: 1247-1252). - 6. Comparison of two different media for vitrification and rewarming of human zygotes: Prospective randomized study. (Alcolak et al, (2011) Middle East Fertility Society Journal 16:189-193). - 7. Three years of routine vitrification of human zygotes: is it still fair to advocate slow-rate freezing? (Al-Hasani et al. (2016) Reproductive BioMedicine Online 14(3):288-293). - 8. Vitrification is a highly efficient method to cryopreserve human embryos in in-vitro fertilization patients at high risk of developing ovarian hyperstimulation syndrome (Selman et al, (2009) Fertility and Sterility 91(4):1611-1613). - 9. Laser-assisted hatching improves clinical outcomes of vitrified-warmed blastocysts developed from low-grade cleavage-stage embryos: a prospective randomized study. (Wan et al, (2014) Reproductive BioMedicine Online 28:582-589).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...