Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.
Device Story
Vitrification Freeze and Thaw Kits are media solutions for cryopreservation of human embryos and oocytes. Freeze kit contains Equilibration and Vitrification solutions (ethylene glycol, DMSO, human serum albumin, Quinn's Advantage Medium with HEPES, sucrose). Thaw kit contains four thawing solutions (human serum albumin, Quinn's Advantage Medium with HEPES, sucrose in TS1/TS2). Used in clinical embryology laboratories by trained personnel. Solutions facilitate cryoprotection during vitrification and rehydration during thawing. Output is preserved or thawed biological material for subsequent transfer. Benefits include successful storage and recovery of reproductive cells/embryos for assisted reproduction.
Clinical Evidence
Bench testing only. Performance validated via Mouse Embryo Assay (MEA) (≥80% development to expanded blastocyst at 96 hours), sterility testing (USP <71>), endotoxin testing (USP <85> <0.5 EU/mL), and shelf-life stability testing. Aseptic processing validated per ISO 13408-1/2. Transportation testing performed per ASTM D4169-22.
Technological Characteristics
Media solutions containing ethylene glycol, DMSO, sucrose, human serum albumin, and Quinn's Advantage Medium with HEPES. Packaged in polypropylene vials. Sterilization via aseptic filtration. pH 7.30-7.70. Osmolality varies by solution (e.g., ES 700±70 mOsm/kg). Storage 2-8°C. Shelf life 7 weeks.
Indications for Use
Indicated for vitrification and thawing of human oocytes (MII), pronuclear (PN) zygotes, day 3 cleavage stage embryos, and blastocyst stage embryos for use in in vitro assisted reproduction.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
SAGE Vitrification Kit (ART-8025 and ART-8026) and SAGE Vitrification Warming Kit (ART-8030 and ART-8031) (K173731)
Related Devices
K223265 — CryoX Vitrification Freeze Kit / Thaw Kit · Zhejiang Horizon Medical Technology Co., Ltd. · May 11, 2023
K171748 — Vitrification Kit and Thawing Kit · Kitazato Corporation · Dec 14, 2017
K160006 — Vit Kit-Freeze Oocytes, Embryos and PN zygotes Vitrification Freeze Kit, Vit Kit-Thaw Oocytes, Embryos and PN zygotes Vitrification Thaw Kit · Irvine Scientific · May 4, 2016
K233846 — Vitrification Solution Set and Warming Solution Set · Ecmpc, LLC · Aug 27, 2024
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2025
Jiangsu Ruifuda Medical Device Co., Ltd
% Olivia Meng
RA Director
Guangzhou Osmunda Medical Device Technical Service Co., Ltd.
14F, Bldg C, Ancillary Project of Phase IV Standard
Industrial Park, Guangzhou International Bio-Island
Guangdong, 510320
CHINA
Re: K242089
Trade/Device Name: Vitrification Freeze Kit (RFD-0101);
Vitrification Thaw Kit (RFD-0201, RFD-0202)
Regulation Number: 21 CFR 884.6180
Regulation Name: Reproductive Media And Supplements
Regulatory Class: II
Product Code: MQL
Dated: March 4, 2025
Received: March 7, 2025
Dear Olivia Meng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242089 - Olivia Meng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K242089 - Olivia Meng
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K242089
Device Name
Vitrification Freeze Kit (RFD-0101); Vitrification Thaw Kit (RFD-0201, RFD-0202)
Indications for Use (Describe)
Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
K242089
Page 1 of 5
# 510(k) Summary
K242089
## 1. Contact Details
510(k) Owner
Jiangsu Ruifuda Medical Device Co., LTD
Address
1st Floor, No. 567-12, Huaguoshan Avenue, Economic and Technological Development Area, Lianyungang, Jiangsu, China
Contact
Mr. Wang Jun
Phone
008613911776721
Email
wang.jun@jsrfd.cn
Submission Correspondent
Guangzhou Osmunda Medical Device Technical Service Co., Ltd.
Address
14F, Building C, Ancillary Project of Phase IV Standard Industrial Park Guangzhou International Bio-Island Guangzhou Guangdong 510320 China
Contact
Ms. Olivia Meng
Phone
008618825133860
Email
hui.meng@osmundacn.com
Date Prepared
April 10, 2025
## 2. Device Name
Trade Name
Vitrification Freeze Kit (RFD-0101);
Vitrification Thaw Kit (RFD-0201, RFD-0202)
Common Name
Assisted Reproduction Media
Classification Name
Reproductive media and supplements
Regulation
884.6180
Class
Class II
Product Code(s)
MQL (Media, Reproductive)
## 3. Predicate Device
510(k) #
K173731
Sponsor
Origio A/S
Trade Name
SAGE Vitrification Kit (ART-8025 and ART-8026)
SAGE Vitrification Warming Kit (ART-8030 and ART-8031)
The predicate device has not been subject to a design-related recall.
{5}
K242089
Page 2 of 5
# 4. Device Description
The Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos during the in vitro assisted reproduction. The Vitrification Freeze Kit includes two solutions: Equilibration Solution (ES) and Vitrification Solution (VS). Both solutions contain ethylene glycol, dimethyl sulfoxide (DMSO), human serum albumin, and Quinn's Advantage® Medium with HEPES. The VS also contains additional sucrose to confer intracellular and extracellular cryoprotective effects. The solutions are available in three specifications: 2 mL, 4.5 mL, and 5 mL, all packaged in polypropylene (pp) vials. The kit includes one vial of each solution.
The Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos during in vitro assisted reproduction. The Vitrification Thaw Kit includes 4 solutions: Thawing Solution 1 (TS1), Thawing Solution 2 (TS2), Thawing Solution 3 (TS3), and Thawing Solution 4 (TS4). All four solutions contain human serum albumin and Quinn's Advantage® Medium with HEPES. TS1 and TS2 also contain sucrose. The solutions are available in three specifications: 2 mL, 4.5 mL, and 5 mL, all packaged in polypropylene (pp) vials. Model RFD-0201 includes one vial of each solution in the kit, and Model RFD-0202 contains two vials of TS1 and one vial of TS2, TS3, and TS4 each in the kit.
# 5. Indications for Use
Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.
# 6. Comparison of the Subject and Predicate Device Intended Use and Technological Characteristics
A comparison of the intended use and technological features of the subject and predicate devices are described in the table below:
| | K242089
Vitrification Freeze Kit
(RFD-0101); Vitrification
Thaw Kit (RFD-0201,
RFD-0202) | K173731
SAGE Vitrification Kit
(ART-8025 and ART-8026); SAGE
Vitrification Warming Kit
(ART-8030 and ART-8031) | Comparison |
| --- | --- | --- | --- |
{6}
K242089
Page 3 of 5
| Indications for Use | Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos. | SAGE Vitrification Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
SAGE Vitrification Warming Kit is intended for use in the thawing of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. | Same |
| --- | --- | --- | --- |
| Conditions for Use | Prescription Use Only | Prescription Use Only | Same |
| Solutions | Vitrification Freeze Kit: Equilibration Solution; Vitrification Solution
Vitrification Thaw Kit: Thawing Solution 1; Thawing Solution 2; Thawing Solution 3; Thawing Solution 4 | SAGE Vitrification Kit: Equilibration Solution; Vitrification Solution
SAGE Vitrification Warming Kit: 1.0 M Sucrose Warming Solution; 0.5 M Sucrose Warming Solution; MOPS Solution | Different: There are differences in the medium solutions for the subject and predicate devices; however, these differences do not raise different questions of S&E. |
| Formulation | Ethylene glycol; DMSO; Sucrose; Gentamicin; Human Serum Albumin; Quinn’s Advantage® Medium with HEPES | Ethylene glycol; DMSO; Sucrose; Gentamicin; Human Serum Albumin; Modified Human Tubal Fluid (MOPS-Buffered); Amino Acids | Different: The formulation of the predicate device is not known; however, the differences in media product formulations do not raise different questions of S&E. |
| Sterilization | Aseptic filtration | Aseptic filtration | Same |
| Sterility | No growth | No growth | Same |
{7}
K242089
Page 4 of 5
| pH | 7.30-7.70 | 7.20-7.40 | Different: The subject device has a higher pH range than the predicate device. This difference in pH range does not raise different questions of S&E. |
| --- | --- | --- | --- |
| Endotoxin (EU/ml) | <0.5 | <0.5 | Same |
| Osmolality (mOsm/kg) | ES: 700±70 (1:3 dilution)
VS: 1,000±80 (1:5 dilution)
TS1: 1500-1900
TS2: 840-920
TS3: 240-300
TS4: 240-300 | ES: 2331-2,849
VS: 5,603-6,849
1.0 M WS: 1,255-1,535
0.5 M WS: 745-911
MS: 257-273 | Different: The subject device has a lower osmolality level than the predicate device. This difference in osmolality specifications does not raise different questions of S&E. |
| MEA | One-cell system: ≥ 80% embryos developed to expanded blastocyst at 96 hours | One-cell system: ≥ 80% blastocyst at 96 hours | Same |
| Storage | 2-8°C | 2-8°C | Same |
| Shelf Life | 7 weeks | 52 weeks | Different: The subject device has a shorter shelf life than the predicate device. Difference does not raise different questions of S&E. |
As shown in the table above, there are differences in the indications for use statements and technological features of the subject and predicate devices. However, the subject and predicate devices have the same intended use and the differences in technological features do not raise different questions of safety and effectiveness.
{8}
K242089
Page 5 of 5
# 7. Summary of Non-Clinical Performance Testing
The following studies have been performed to support substantial equivalence to the predicate devices:
- Aseptic filtration and aseptic filling validation, per ISO 13408-1:2008 – Aseptic Processing of Health Care Products – Part 1 General Requirements (including Amendment 1 (2013)) and ISO 13408-2:2018 – Aseptic Processing of Health Care Products – Part 2 Sterilizing Filtration.
- Shelf-life testing was conducted to support the 7-week shelf-life for the subject device through demonstration that the product specifications (shown below) were met at time 0 and after real-time aging. Testing conducted is shown below:
Appearance: pH (USP <791>): 7.30-7.70
Osmolality (USP <785>) (mOsm/kg):
Equilibration Solution (ES): 700±70 (ES:purified water=1:3)
Vitrification Solution (VS): 1000±80 (VS:purified water=1:5)
Thawing Solution 1 (TS1): 1500-1900
Thawing Solution 2(TS2): 840-920
Thawing Solution 3 (TS3): 240-300
Thawing Solution 4 (TS4): 240-300
Endotoxin testing (USP <85>): <0.5 EU/mL
Mouse Embryo Assay (MEA): One-Cell System: ≥80% embryos developed to expanded blastocyst at 96 hours
Sterility testing (USP <71>): No microbial growth
- Transportation testing per ASTM D4169-22 and USP <1207.2>.
# 8. Conclusions
The results of the performance testing described above demonstrated that the Vitrification Freeze Kit and Vitrification Thaw Kit are as safe and effective as the predicate devices and support a determination of substantial equivalence.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.