Q-HTF, CAT# 2001

K993484 · Invitrocare, Inc. · MQL · May 25, 2000 · Obstetrics/Gynecology

Device Facts

Record IDK993484
Device NameQ-HTF, CAT# 2001
ApplicantInvitrocare, Inc.
Product CodeMQL · Obstetrics/Gynecology
Decision DateMay 25, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6180
Device ClassClass 2

Intended Use

Q-HTF is intended for the retrieval, culture, transport, storage and transfer of human gametes and embryos.

Device Story

Q-HTF is a synthetic, defined culture medium formulated to mimic human fallopian tube fluid. It utilizes a sodium bicarbonate buffer system requiring a CO2-enriched incubation environment. Used in assisted reproductive technology (ART) laboratories by embryologists and clinicians, the medium supports the retrieval, fertilization, growth, transport, and storage of human gametes and embryos. The medium is supplemented with protein and replenished as needed during the culture period, typically up to three days post-fertilization. By providing a controlled chemical environment, the device facilitates embryo development in vitro, supporting clinical decision-making regarding embryo viability for transfer.

Clinical Evidence

No clinical data. Performance is supported by bench testing, specifically mouse embryo assays to confirm functionality and absence of toxicity, alongside endotoxin and sterility testing. Historical clinical use of Human Tubal Fluid media is cited as evidence of safety and effectiveness.

Technological Characteristics

Synthetic, defined culture medium; sodium bicarbonate buffer system; requires CO2 atmosphere; protein supplementation required for use. Formulated based on Quinn et al. (1985) composition of human fallopian tube fluid.

Indications for Use

Indicated for use in assisted reproductive technology procedures involving the manipulation of human gametes and embryos, including as a culture medium for embryos post-fertilization and to support in vitro fertilization in a CO2 incubator environment.

Regulatory Classification

Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for InVitroCare INCORPORATED. Below the logo is the address 11408 Sorrento Valley Road, Suite 202. The text is in black and the background is white. Page 1 of 2 ### 510(k) SUMMARY This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 The assigned 510(k) number is: Submitted by: InVitroCare, Inc. 11408 Sorrento Valley Rd. Suite 202 San Diego, CA 92121 Telephone: (858) 452-1986 Facsimile: (858) 452-1828 E-mail: Invitroc@aol.com Contact: Robert E. Lovins, PhD Date Submitted: October 13,1999 Device Identification: Trade Name: Q-HTF Medium Common Name: Gamete and embryo culture, storage and transfer Medium, Human Tubal Fluid medium Classification Name: Reproductive Media (21CFR, 886.6180) ## Predicate Device: Notice of Final Rule, 63 FR 48428, Docket number 97N-0335 and 510(k) Reference Number K983586 ## Description: Q-HTF is a synthetic, defined culture medium intended for use in assisted reproductive technology procedures. It has been formulated to mimic the composition of the fluid found in fallopian tubes as defined by Quinn etal (Quinn P, Kerin JF, Warnes GM: Fertil Steril 1985;44:493-498). Q-HTF uses a sodium bicarbonate buffering system and is appropriate for those procedures requiring the use of a carbon dioxide atmosphere during incubation. {1}------------------------------------------------ ## Intended Use: K 993484 page 2 of 2 Q-HTF is intended for the retrieval, culture, transport, storage and transfer of human gametes and embryos. #### Design Characteristics: Q-HTF is intended for use as a culture medium, with appropriate protein supplementation, for the support of fertilized embryos. Fertilization is also allowed to occur in Q-HTF when in vitro fertilization techniques are used. The fertilized gamete is then allowed to grow in the media and supplement, which are replenished as needed, until the desired state of development, usually up to three days post fertilization. Since Q-HTF utilizes a sodium bicarbonate buffer system, it is intended to be used in those procedures that require a carbon dioxide atmosphere, as is found in the incubators used by assisted reproductive laboratories. Therefore Q-HTF is primarily used as a medium to support embryo growth, development and culture in vitro. ### Performance Data: Q-HTF medium is assayed by a mouse embryo assay prior to its release to market. This assay assures that the product is both functional for its intended use, the support of embryonic growth and that no toxic components are present in the formulation. Human Tubal Fluid medias have been used in a variety of clinical settings, for the intended use for a number of years. In that time the product has become the standard media used for the retrieval, growth, storage and transport of human gametes and embryos. #### Additional Information: Mouse embryo testing will be performed as a condition of release for Q-HTF medium as well as endotoxin and sterility testing. Results of all release assays will be reported on a lot-specific certificate of analysis and will be indicated on the labeling. #### Conclusion: The conclusion from performance testing, as well as a review of published historical information contained in the professional literature shows that Q-HTF is suitable for its intended use and meets the criteria outlined in the Final Rule, 63 FR48428, Docket number 97N-0335. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle's head, represented by three curved lines that suggest the shape of the head and beak. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 2 5 2000 Robert E. Lovins, Ph.D. President InVitroCare, Inc. 11408 Sorrento Valley Road Suite 202 San Diego, CA 92121 Re: K993484 Q-HTF (Assisted Reproductive Medium) Dated: March 9, 2000 Received: March 13, 2000 Requiatory Class: II 21 CFR §884.6180/Procode: 85 MQL Dear Dr. Lovins: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Pracice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yith diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {3}------------------------------------------------ InVitroCare, Inc. October , 1999 # INDICATIONS FOR USE STATEMENT (Page 1 of 1) 510(k) number: K0931484 Device Names: Q-HTF Medium Indications for Use: Q-HTF Medium is intended for use in assisted reproductive technology procedures that involve the manipulation of gametes and embryos. Specifically, Q-HTF is intended for use a a culture medium for the embryo after fertilization, when used with an incubator, and as a nedim to support in vitro fertilization. Q-HTF is intended to simulate the substances fourd in the human to female reproductive system. (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number. Prescription Use (per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...