BLASTFREEZE AND BLASTTHAW

K031486 · Medicult A/S · MQL · Jun 27, 2003 · Obstetrics/Gynecology

Device Facts

Record IDK031486
Device NameBLASTFREEZE AND BLASTTHAW
ApplicantMedicult A/S
Product CodeMQL · Obstetrics/Gynecology
Decision DateJun 27, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6180
Device ClassClass 2

Intended Use

BlastFreeze™: For freezing of human blastocysts. BlastThaw™: For thawing of human blastocysts.

Device Story

BlastFreeze and BlastThaw are reproductive media used for cryopreservation and thawing of human blastocysts. The solutions are based on Earle's Balanced Salts Solution with varying sucrose concentrations; BlastFreeze includes glycerol as a cryoprotectant. The products are used by clinicians in IVF laboratories. The process involves preparing blastocysts using these media to facilitate freezing and subsequent thawing. Clinical outcomes include survival and implantation rates comparable to existing cryopreservation products, with potential for reduced abortion rates. The device is not patient-contacting.

Clinical Evidence

Clinical documentation reports an 80% survival rate and 15% implantation rate for human blastocysts following freezing and thawing. Abortion rates were 14% for IVF patients and 0% for ICSI patients. Results showed no difference between IVF and ICSI procedures. Performance is comparable to existing market products (Ménézo and Veiga, 1997). Bench testing includes sterility, pH, endotoxin levels (≤ 0.1 EU/ml), and mouse embryo assay (blastocyst rate ≥ 80%).

Technological Characteristics

Media based on Earle's Balanced Salts Solution with sucrose; BlastFreeze contains glycerol. Non-patient contacting. Quality controls: sterility, pH, endotoxin testing (USP), and mouse embryo assay (two-cell assay).

Indications for Use

Indicated for the freezing and thawing of human blastocysts in a clinical setting.

Regulatory Classification

Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # KO31486 # 510(k) Summary Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination ### The Products BlastFreeze™ and BlastThawTM, Cat.No. 10532010 and 10542010 ## Indications for use: For freezing and thawing human blastocysts. ### Composition: BlastFreeze and BlastThaw are based on Earle's Balanced Salts Solution and differing concentrations of sucrose. BlastFreeze contains glycerol as the cryprotectant. ### Stability and biocompatibility testing BlastFreeze and BlastThaw have been stability tested. When following the procedure described in our package inserts BlastFreeze and BlastThaw will not be patient contacting that is why biocompatibility testing has not been performed. #### Product testing controls - 1. Sterility - 2. pH - 3. Endotoxin tested ≤ 0.1 EU/ml ( USP) - 4. Mouse Embryo Assay, (two-cell assay, blastocyst rate µ 80 %) For each batch a Certificate of Analysis with the results of the above tests is available. #### Clinical Documentation: The clinical documention show a survival rate of around 80% and an implantation rate of 15% after freezing and thawing of human blastocysts. No difference in results between IVF and ICSI. The abortion rate is low, being 14 % in IVF patients and 0 in ICSI patients. BlastFreeze and BlastThaw products have demonstrated their suitability in daily clinical practice for freezing and thawing of blastocysts giving survival and implantation rates similar to the results obtained by using other products available for freezing and thawing blastocysts ( Ménézo. Y and Veiga.A. 1997 ). In addition, by using BlastFreeze and BlastThaw the abortion rate can be lowered. {1}------------------------------------------------ BlastFreeze and BlastThaw were marketed in Europe in April 2002. There has been no registered complaints and no evidence that the product has been the cause of any serious adverse events in connection with its intended use. Based on the clinical data presented and our experience with the BlastFreeze and BlastThaw products we feel that the safety and effectiveness of the product for its intended use is shown in the present submission and the products are substantially equivalent to the predicated devices, our Blastocyst Freezing Pack and Blastocyst Thawing Pack ( K 991471) Prepared and Submitted by: Date Ronald G. Leonardi, Ph.D. President R & R REGISTRATIONS P.O. Box 262069 San Diego, Ca 92131 858-586-0751 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its head, representing health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the eagle symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 7 2003 Medi-Cult a/s % Ronald G. Leonardi, Ph.D. President R & R Registrations P.O. Box 262069 SAN DIEGO CA 92196-2069 Re: K031486 Trade/Device Name: BlastFreeze™ and BlastThawTM Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: 85 MQL Dated: May 9, 2003 Received: May 14, 2003 Dear Dr. Leonardi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {4}------------------------------------------------ 510(k) Number (if known) k 031486 Device Name: BlastFreeze™ and BlastThaw™ INDICATIONS FOR USE: BlastFreeze™: For freezing of human blastocysts. BlastThaw™: For thawing of human blastocysts. (PLEASE DO NO WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use 1 (Per 21 CFR 801.109) OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) David A. Ayres productive, Abdominal, a De . 10(K) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...