NEUROCALM MODEL 1 AND 2

K090866 · Neuromed Devices, Inc. · GZJ · Oct 21, 2009 · Neurology

Device Facts

Record IDK090866
Device NameNEUROCALM MODEL 1 AND 2
ApplicantNeuromed Devices, Inc.
Product CodeGZJ · Neurology
Decision DateOct 21, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

Symptomatic relief and management of chronic, intractable pain. Management of post-surgical pain. Management of post-traumatic pain.

Device Story

NeuroCalm is a battery-powered, portable TENS device for pain management. It delivers AC and DC electrical currents through gold electrodes to the site of pain. The device operates in two configurations: Model 1 provides 10-second treatment cycles; Model 2 provides 35-second treatment cycles. A complete course consists of ten treatment cycles, after which the device becomes inoperable and is discarded. Operation is managed by software-controlled printed circuit board assemblies that optimize waveforms based on the selected treatment duration. The device is intended for use by patients for pain relief.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Battery-powered (12V) TENS device. Noble metal (gold) electrodes. Software-controlled printed circuit board assemblies. Two configurations (Model 1: 10-second cycles; Model 2: 35-second cycles). Disposable after 10 cycles. AC/DC current output.

Indications for Use

Indicated for symptomatic relief and management of chronic, intractable pain, and adjunctive treatment of post-surgical and post-traumatic pain.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K090866 # Traditional 510(k) Summary NeuroCalm Transcutaneous Nerve Stimulator (TENS) Device - 1. Sponsor NeuroMed Corporation 29461 Troon St. Laguna Niguel, CA 92677 Contact Person: Robert Seiple Telephone: (940) 390 0961 Date Prepared: 15 February 2009 2. Device Name > Proprietary Name NeuroCalm TENS Device Common/Usual Name Transcutaneous Nerve Stimulator Classification Name Transcutaneous Electrical Nerve Stimulator for Pain Relief #### 3. Predicate Devices: - 1) Biowave Corporation's, "Homewave Neuromodulation Pain Therapy Device" PMN # K072123, - 2) AEMED, Inc. "StimPad™ TENS System" #K071120 #### র্বা Intended Use - Symptomatic relief and management of chronic, intractable pain ● - . Management of post-surgical pain - . Management of post-traumatic pain. #### ട. Device Description The NeuroCalm device is a battery-powered portable TENS device intended for the relief of chronic intractable pain and adjunctive treatment of postsurgical or post-traumatic acute pain. The device is small, portable and lightweight. The device is available in two configurations: Model 1 and Model 2. The Model 1 configuration provides a 10-second treatment using a proprietary wave form while Model 2 provides a 35-second treatment using a slightly different waveform to alleviate pain. The NeuroCalm TENS device is powered by a 12-volt battery. The power source provides ample power for the 10-course treatment. The device has a two-year shelf life. ### NeuroMed Devices Inc. 29461 Troon St, Laguna Niguel, CA 92677 (tel) 949-495-4274 (fax) 949-606-8191 OCT 21 2009 {1}------------------------------------------------ NeuroMed Devices, Inc. - NeuroCalm TENS Traditional 510(k) Treatment is provided by passing AC and DC current through noble metal (gold) electrodes. Treatment cycles are controlled by software-controlled printed circuit board assemblies. Electrical output and waveforms are optimized depending on the treatment duration desired. A complete course of treatment consists of a 10 x 10-second treatment for Model 1 and a 10 x 35second treatment for Model 2. After the conclusion of the tenth treatment, the device is inoperable and may be discarded (discard as one would discard a battery). #### Basis for Substantial Equivalence: 6. NeuroMed Corporation's TENS device, Biowave's neuromodulation device and AEMED, Inc. "StimPad"m TENS System are similar in intended use, design and function. All are TENS devices (of which there are over 500 cleared devices) and provide pain relief by generating small electrical currents of varying waveforms applied directly to the area experiencing pain. The proposed and predicate devices share the same regulatory classification (Class II, classification code GZJ), are battery powered, and are all softwarecontrolled TENS units that provide the user with pain reduction. The NeuroCalm device is intended for use in pain relief. The conclusion of the technical comparison is that the NeuroCalm TENS device is substantially equivalent to the predicate devices for the indication specified. 29461 Troon St, Laguna Niguel, CA 92677 (tel) 949-495-4274 (fax) 949-606-8191 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem of an eagle with three stylized lines representing its wings and body. The eagle faces to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 NeuroMed Corporation c/o Mr. Daniel W. Lehtonen Senior Staff Engineer - Medical Devices Intertek Testing, Services NA, Inc. 23047 E. Aurora Road Unit B7 Twinsburg, OH 44087 OCT 21 2009 Re: K090866 Trade/Device Name: NeuroCalm TENS Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: II Product Code: GZJ Dated: September 9, 2009 Received: October 6, 2009 Dear Mr. Lehtonen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Kesia Alexander for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K090866 NeuroCalm Transcutaneous Electrical Nerve Stimulator Device Name: Indications for Use: The NeuroCalm Transcutaneous Electrical Nerve Stimulator Device is indicated for symptomatic relief and management of chronic, intractable pain, and adjunctive treatment in the management of post-surgical and post-traumatic pain. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Bruce Drum (Division Sign-O Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K090866
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...