K142666 · Delta International Services & Logistics S.R.L · GZJ · May 22, 2015 · Neurology
Device Facts
Record ID
K142666
Device Name
Scrambler Therapy MC-5 A Device
Applicant
Delta International Services & Logistics S.R.L
Product Code
GZJ · Neurology
Decision Date
May 22, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Scrambler Therapy MC-5A Device is indicated for: - Symptomatic relief of chronic, intractable pain, post-surgical and posttraumatic acute pain. - Symptomatic relief of acute pain. - Symptomatic relief of post-operative pain.
Device Story
Multi-channel electrical nerve stimulator; generates and controls stimulation impulses via stored program to provide pain relief. Used in clinical settings for symptomatic relief of chronic, intractable, post-surgical, post-traumatic, and post-operative pain. Device operates by delivering controlled electrical impulses to pain sites; healthcare providers apply treatment to manage patient pain levels. Benefits include non-invasive symptomatic pain relief.
Clinical Evidence
Bench testing only; device meets design and performance specifications.
Technological Characteristics
Multi-channel electrical nerve stimulator; Class II; Product Code GZJ; Regulation 882.5890. Operates via stored program-controlled stimulation impulses.
Indications for Use
Indicated for symptomatic relief of chronic, intractable pain, post-surgical pain, post-traumatic acute pain, and post-operative pain in patients requiring pain management.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
K071624 — EVERELIFE TRANSCUTANEOUS MUSCLE STIMULATOR · Everlife Medical Equipment Co., Ltd. · Nov 27, 2007
K052836 — BODIHEALTH SYSTEM · Usa Laser Therapeutics, Inc. · Feb 13, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines above and below them. The profiles are stacked, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 22, 2015
Delta international Service & Logistics S.r.l. Ms. Mariella Giorgieri CEO Piazza Adriana 4 00193 Roma Italy
Re: K142666
Trade/Device Name: Scrambler Therapy MC-5A Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ Dated: May 6, 2015 Received: May 11, 2015
Dear Ms. Giorgieri:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Felipe Aquel -S
Carlos Peña, PhD, MS for Director Division and Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| | 510(k) | Scrambler Therapy MC-5A |
|--|--------|-------------------------|
|--|--------|-------------------------|
INDICATION FOR USE STATEMENT
| 510(k) Number (if you know): | K142666 |
|------------------------------|--------------------------------|
| Device Name: | SCRAMBLER THERAPY MC-5A DEVICE |
Indication for Use:
The Scrambler Therapy MC-5A Device is indicated for:
- Symptomatic relief of chronic, intractable pain, post-surgical and posttraumatic acute pain.
- Symptomatic relief of acute pain.
- Symptomatic relief of post-operative pain.
| Prescription Use <u>X</u><br>(Part 21 CFR 801 Subpart D) | AND / OR | Over-The-Counter Use ______<br>(Part 21 CFR 801 Subpart C) |
|----------------------------------------------------------|----------|------------------------------------------------------------|
|----------------------------------------------------------|----------|------------------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1/1
{3}------------------------------------------------
510(k)
# 510(K) SUMMARY
## 1. SUBMITTER/510(K) HOLDER:
| Company Name: | Delta International Services & Logistics S.r.l. |
|------------------------|----------------------------------------------------------------------------------|
| Company Address: | Piazza Adriana 4<br>00193, Roma - Italy |
| Company Phone: | 039- 3921603399 |
| Company Fax: | 039-063215142 |
| Company e-mail: | ceo@st-team.eu |
| Contact person: | Ms. Mariella Giorgieri<br>CEO<br>Delta International Services & Logistics S.r.l. |
| Date Summary Prepared: | May 13, 2015 |
## 2. DEVICE IDENTIFICATION
| Common Usual Name: | Electrical Nerve Stimulator |
|-------------------------|--------------------------------|
| Trade/Proprietary Name: | Scrambler Therapy MC-5A Device |
| Classification: | Class II |
| Product Code: | GZJ |
| Classification Panel: | 882 Neurological Devices |
| Regulation Number: | 882.5890 |
#### 3. PREDICATED DEVICES
| Predicate device | 510 (k) Holder | 510 (k) No |
|--------------------------------|-----------------------------------|------------|
| SCRAMBLER MC-5A TENS<br>DEVICE | COMPETITIVE TECHNOLOGIES,<br>INC. | K081255 |
{4}------------------------------------------------
#### 4. DEVICE DESCRIPTION
The Scrambler Therapy MC-5A Device is a multi-channel device which allows simultaneous treatment of a number of pain sites. Stimulation impulses are generated and controlled according to a stored program to provide pain relief.
#### 5. INTENDED USE
The Scrambler Therapy MC-5A Device is indicated for:
- · Symptomatic relief of chronic, intractable pain, post-surgical and posttraumatic acute pain.
- · Symptomatic relief of acute pain.
- · Symptomatic relief of post-operative pain.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
Delta International Services & Logistics s.r.l. claims substantial equivalence of the Scrambler Therapy MC-5A Device to the predicate devices based on the intended use, fundamental technology, and operation characteristics. A side-by-side comparison of the Scrambler Therapy MC-5A Device and the cited predicate device is included in the 510(k).
#### 7. PERFORMANCE TESTING
Testing of the Scrambler Therapy MC-5A Device demonstrates that the device meets design and performance specifications.
Page 2/2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.