K071624 · Everlife Medical Equipment Co., Ltd. · GZJ · Nov 27, 2007 · Neurology
Device Facts
Record ID
K071624
Device Name
EVERELIFE TRANSCUTANEOUS MUSCLE STIMULATOR
Applicant
Everlife Medical Equipment Co., Ltd.
Product Code
GZJ · Neurology
Decision Date
Nov 27, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To be used in symptomatic relief of chronic intractable pain, post-traumatic and post surgical pain
Device Story
Everlife Transcutaneous Nerve Stimulator is a sequenced system for transcutaneous muscle/nerve stimulation. Device consists of a stimulator unit, a sequencer for channel selection, patient cables, and skin-applied electrodes. It outputs various electrical waveforms to generate desired effects on muscles. Patients control signal intensity for safety and comfort. The system may feature multiple output channels for bilateral treatment or simultaneous/serial treatment of multiple body regions. Used in clinical or home settings to provide symptomatic relief of chronic intractable, post-traumatic, and post-surgical pain. Healthcare providers or patients operate the device to manage pain through electrical stimulation.
Clinical Evidence
No clinical data. Bench testing only. Testing included electrical safety, electromagnetic compatibility, and waveform verification to ensure compliance with standards for nerve and muscle stimulators.
Technological Characteristics
Transcutaneous electrical nerve stimulator; multi-channel output; includes stimulator, sequencer, patient cables, and electrodes. Uses shrouded connectors for lead wire safety. Electrical energy source. System-level waveform testing performed. No specific materials or software architecture details provided.
Indications for Use
Indicated for symptomatic relief of chronic intractable pain, post-traumatic pain, and post-surgical pain in patients requiring transcutaneous electrical nerve stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS for Transcutaneous Nerve Stimulator
### 510(k) No: K071624
| DATE OF<br>SUBMISSION: | June 1, 2007 | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| SUBMITTER: | EVERLIFE MEDICAL EQUIPMENT CO., LTD.<br>NO 58, FU-CHIUN ST.<br>HSIN-CHU CITY, CHINA (TAIWAN) 30067<br>TEL: 886-3-5208829<br>FAX:886-3-5209783 | |
| ESTABLISHMENT<br>REGISTRATION NO: | 3004753827 | NOV 27 2007 |
| OFFICIAL<br>CONTACT: | Dr. JEN, KE-MIN<br>NO 58, FU-CHIUN ST.<br>HSIN-CHU CITY, CHINA (TAIWAN) 30067<br>TEL: 886-3-5208829<br>FAX:886-3-5209783 | |
| TRADE NAME: | EVERLIFE Transcutaneous Nerve Stimulator,<br>T-100201, T-100202, T-100203, T-100801,<br>T-100802, T-100803, T-100501, T100502,<br>SD-100506,T-201110, T-201210, T-201310, T-300708 | |
| COMMON/USUAL<br>NAME: | Transcutaneous Nerve Stimulator | |
| CLASSIFICATION<br>NAME: | Transcutaneous Nerve Stimulator | |
| REGULATION<br>NUMBER: | 882.5890, GZJ, Class II | |
| PREDICATED<br>DEVICE: | APEX Transcutaneous Nerve Stimulator,<br>K902102, K931570, K012643, K021755, K970429,<br>( Please refer to the following comparison page.) | |
| INTENDED USE: | To be used in symptomatic relief of chronic intractable<br>pain, post-traumatic and post surgical pain | |
{1}------------------------------------------------
# Comparison for Predicate Device & Subject Device
| 510K<br>Number | APEX Model<br>(predicate device) | Everlife Model<br>(new device) | Note |
|----------------|----------------------------------|--------------------------------|-------------------------|
| K902102 | COM-TENS | T-100201 | Identical |
| | | T-100202 | Substantial Equivalence |
| | | T-100501 | Substantial Equivalence |
| | | T-100502 | Substantial Equivalence |
| | | T-100801 | Substantial Equivalence |
| | | T100802 | Substantial Equivalence |
| K931570 | COM-TENS<br>(Single Mode) | T-100203 | Identical |
| | | T-100803 | Substantial Equivalence |
| K012643 | Apex Medical<br>LCD TENS-VII | T-300708 | Identical |
| K021755 | TS-1211 | T-201110 | Identical |
| | TS-1211 | T-201310 | Substantial Equivalence |
| K021755 | TS-1212 | T-201210 | Identical |
| K970429 | BIOSCOPE<br>SD TENS | SD-100506 | Identical |
、
{2}------------------------------------------------
| Description of Device: | A sequenced system for transcutaneous muscle stimulation consists of a stimulator, a sequencer for channel selection, patient cable, and electrodes applied to the skin. Various types of waveforms may be output to generate the desired effect on the muscle(s) to be treated, and the patient is given control of the signal intensity for personal safety and comfort. Sequenced system may have more than on output channel in order to operate bilaterally on the body or to treat multiple regions simultaneously or serially in a prescribed sequence. |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Non-Clinical Tests Submitted: | The EVERLIFE Transcutaneous Nerve Stimulator has been tested in accordance with applicable standards for medical device electrical safety, electromagnetic compatibility, and the particular requirements for safety of nerve and muscle stimulators. Accessories also meet safety requirements: 510(k) electrodes are specified, and the patient cable utilizes shrouded connectors to meet lead wire safety requirements. System level testing including waveform testing was performed in combination the EVERLIFE Transcutaneous Nerve Stimulator. |
| Clinical Tests Submitted: | None |
| Conclusion: | As the product description and tests as above, the new device: EVERLIFE Transcutaneous Nerve Stimulators are as safe and effective as, and the function in a manner equivalent to the predicate devices. Thus the new device is substantially equivalent to the predicate devices in this aspect. |
.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the top half of the circle. Inside the circle is an abstract image of an eagle.
NOV 2 7 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Everlife Medical Equipment Co., Ltd. % Dr. Jen Ke-Min No. 58, Fu Chiun Street I Isin Chu City, 30067 Taiwan, ROC
Re: K071624
Trade/Device Name: Everlife Transcutaneous Nerve Stimulator models T-100201, T-100202, T-100501, T-100502, T-100801, T-100802, T-100203, T-100803, T-300708, T-201110, T-201310, T-201210, SD-100506 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ Dated: October 29, 2007 Received: November 6, 2007
Dear Dr. Jen Ke-Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 – Dr. Jen Ke-Min
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter will anow you to ought mating of substantial equivalence of your device to a legally prematics notification: "The PDF Interessification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you active epper for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to (240) 276 0126. Fire Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems (Nedical Donoc Reporting (112)); per general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director
Mark N. Melkerson Director Division of General Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
| 510(k) Number: | K071624 |
|----------------|---------|
|----------------|---------|
Device Name:
EVERLIFE Transcutaneous Nerve Stimulator, T-100201, T-100202, T-100203, J T-100801, T-100802, T-100803, T-100501, T100502, SD-100506, T-201110, T-201210, T-201310, T-300708
#### Indications for Use :
To be used in symptomatic relief of chronic intractable pain, post-traumatic and post-surgical pain.
Prescription Use V
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General, Restorative, and Neurological Device
510(k) Num
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.