RELIEFBAND

K980333 · Woodside Biomedical, Inc. · GZJ · Feb 18, 1998 · Neurology

Device Facts

Record IDK980333
Device NameRELIEFBAND
ApplicantWoodside Biomedical, Inc.
Product CodeGZJ · Neurology
Decision DateFeb 18, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ReliefBand is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy.

Device Story

ReliefBand is a non-invasive nerve stimulation device worn on the wrist; treats nausea and vomiting (NV) via electrical stimulation of the P6 acupuncture point. Device consists of a wristband worn on the ventral or palmar surface of the wrist between the palmaris longus and flexor carpi radialis tendons. Operation involves three pulse amplitude modulation modes (burst/off sequences) and six intensity levels selected by the patient. Powered by silver oxide or zinc-air batteries. User interface displays mode and intensity. Provides symptomatic relief for chemotherapy-induced, motion-related, and pregnancy-related NV. Intended for patient self-use.

Clinical Evidence

Bench testing only. Performance data indicates the device meets functional requirements and specifications for acustimulation of the P6 point.

Technological Characteristics

Non-invasive nerve stimulation device; wrist-worn form factor. Powered by three silver oxide (type 392/IEC SR41) or zinc-air (type 312/IEC PR41) batteries. Features three pulse amplitude modulation modes and six intensity levels. User display for mode/intensity selection. Not waterproof.

Indications for Use

Indicated for treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy in patients requiring acustimulation therapy.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 980333 ## SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 19: FEB 1 8 1998 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. #### SUBMITTER INFORMATION 19.1 | a. | Company Name: | Woodside Biomedical, Inc. | |----|------------------------|---------------------------| | b. | Company Address: | 1132 San Marino Dr. | | | | San Marcos, CA 92069 | | c. | Company Phone: | (760) 761-0907 | | | Company Fax: | (760) 761-0535 | | d. | Contact Person: | Tom Grey | | | | Vice-President of Product | | | | Development | | | | Woodside Biomedical, Inc. | | e. | Date Summary Prepared: | January 26, 1998 | # 19.2. DEVICE IDENTIFICATION | a. | Trade/Proprietary Name: | ReliefBand | |----|-------------------------|------------| | b. | Classification: | Class II | - Product Code: 73BWK C. #### IDENTIFICATION OF PREDICATE DEVICE 19.3 | Device | Manufacturer | 510(k) No. | Date Cleared | |------------|--------------------|------------|----------------| | ReliefBand | Maven Laboratories | K961050 | August 1, 1997 | {1}------------------------------------------------ #### 19.4 DEVICE DESCRIPTION The ReliefBand® Model SW-111 is a non-invasive nerve stimulation therapy device, and is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy. The device is contained within a wristband, and provides relief of NV through electrical stimulation of the P6 acupuncture area on the patient's wrist. This acustimulation wrist band can be worn on either hand on the ventral or palmar (i.e., inside) surface of the wrist, approximately 2-3 fingers breadth above the distal skin crease of the wrist joint between the tendons of the palmaris longus and flexor carpi radialis muscles. There are three operating modes based on pulse amplitude modulation scheme: - 2 second burst, 6 seconds off. This is the first setting in the 3 mode Mode A: sequence. - Mode B: 4 second burst. continuous. This is the second setting in the 3 mode sequence. - 2 second burst, continuous. This is the third setting in the 3 mode Mode C: sequence. There are six intensity levels present for each of the three operating modes, and the intensity setting determines the strength of the stimulation. The user display on the wristband identifies both the operating mode and intensity level, so the patient can select the desired stimulation. The wristband is powered by either three silver oxide batteries (type 392/IEC SR41, 1.5 V), or three zinc-air batteries (type 312/IEC PR41, 1.4V), and should not be immersed in water. {2}------------------------------------------------ #### 19:5 SUBSTANTIAL EQUIVALENCE The Woodside Biomedical, Inc. ReliefBand is of comparable type and is equivalent to the predicate Maven ReliefBand (K961050). #### INTENDED USE 19.6 The ReliefBand is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy. #### 19.7 TECHNOLOGICAL CHARACTERISTICS A comparison of the technological characteristics of the predicate and legally marketed devices has been performed. The results of this comparison demonstrate that the ReliefBand is equivalent to the marketed predicate device. #### 19.8 PERFORMANCE DATA The performance data indicate that the ReliefBand meets the functional requirements and specifications of devices used for acustimulation of the P6 point. #### 19.9 510(K) CHECKLIST This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist has been provided in this submission. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an abstract bird or wave-like form. FEB 1 8 1998 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Carol L. Patterson Consultant for Woodside Biomedical Inc. Patterson Consulting Group, Inc. 18140 Smokesignal Drive San Diego, California 92127 Re: K980333 ReliefBand® Trade Name: Regulatory Class: II Product Codes: GZJ and BWK Dated: January 27, 1998 January 28, 1998 Received: Dear Ms. Patterson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಗಿ substantially equivalent determination assumes compliance with ----the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Failure to Administration (FDA) will verify such assumptions. comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ ### Page 2 - Ms. Carol L. Patterson This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE | 510(k) Number: | To Be Assigned By FDA | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | ReliefBand | | Indications For Use: | The ReliefBand is indicated for use in the treatment of<br>nausea and vomiting (NV) due to chemotherapy, motion<br>sickness, and pregnancy. | PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Adee Division (Sign-Off) Division of General Restorative Devices 510(k) Number L480333 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ JANUARY 26, 1998 . . .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...