ReliefBand is indicated for use in the treatment of nausea, vomiting and retching associated with physician-diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy and morning sickness from pregnancy. ReliefBand is also indicated as an adjunct to antiemetics in reducing postoperative nausea. For over the counter use.
Device Story
Battery-powered, wrist-worn digital therapeutic; provides non-invasive transcutaneous electrical nerve stimulation (TENS) to median nerve at P6 wrist location. User selects pulse amplitude via pushbuttons; device delivers asymmetric biphasic pulses (31 Hz, 350 ms pulse width) to alleviate nausea/vomiting. ReliefBand 1.5 uses disposable batteries; ReliefBand 2.0 uses rechargeable Li-ion. LED display indicates signal intensity and battery status. Used by patients in home or clinical settings for self-management of nausea symptoms. Output provides symptomatic relief, reducing need for or supplementing antiemetic medications.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics and intended use to previously cleared predicate devices.
Technological Characteristics
Wrist-worn TENS device. ReliefBand 1.5: gold-plated copper/nickel/zinc electrodes; CR2025 batteries. ReliefBand 2.0: 316L stainless steel electrodes; rechargeable Li-ion (IEC62133). Output: asymmetric biphasic pulses, 31 Hz, 350 ms width, 8-40 mA amplitude. LED intensity/battery indicators. No software-based algorithm; hardware-controlled pulse generation.
Indications for Use
Indicated for treatment of nausea, vomiting, and retching in patients with physician-diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy, or pregnancy-related morning sickness. Also indicated as an adjunct to antiemetics for postoperative nausea. OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
August 30, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
ReliefBand Technologies LLC Barbara Whitman Vice President, Regulatory Affairs and Quality Assurance 220 Gibraltar Rd Ste 270 Horsham, Pennsylvania 19044
Re: K191547
Trade/Device Name: ReliefBand 1.5 and ReliefBand 2.0 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: July 30, 2019 Received: August 1, 2019
### Dear Barbara Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Kang, PharmD Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191547
Device Name ReliefBand 1.5 and ReliefBand 2.0
#### Indications for Use (Describe)
ReliefBand(R) is indicated for use in the treatment of nausea, vomiting associated with physician-diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy and morning sickness from pregnancy. ReliefBand is also indicated as an adjunct to antiemetics in reducing postoperative nausea. For over the counter use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510K Summary K191547
### Submitter Information
Applicant: Reliefband® Technologies LLC 220 Gibralter Road, Suite 270 Horsham, PA 19044
Phone: 877-735-2263 Ext 122
Contact: Barbara Whitman
Vice President of Regulatory Affairs and Quality Assurance
Date Prepared: July 30, 2019
Device Classification Trade name: ReliefBand® 1.5, ReliefBand® 2.0
Common Name: Nerve Stimulation Therapy Device
Classification Name: Transcutaneous electrical nerve stimulator for pain relief (21 CFR 882.5890)
Product Code: GZJ
Device Classification: Class II
Predicate Devices: ReliefBand® 1.5 (K173233) and ReliefBand® 2.0 (K182960)
### Description of the Device
ReliefBand® models 1.5 and 2.0 are battery-powered non-invasive digital therapeutic devices that are worn on the wrist like a watch. They are indicated for the counter use in the relief of mild to moderate nausea and retching associated with physician-diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy and morning sickness associated with pregnancy as an adjunct to antiemetics in reducing mild to moderate postoperative nausea.
ReliefBand 1.5 and 2.0 include two electrical contacts on the back of the device (ReliefBand 1.5) or on the clasp (ReliefBand 2.0), which must be placed firmly against the underside of the wrist at the P6 location (approximately two fingers breadth above the distal skin crease of the wrist joint, between the tendons of the palmaris longus and flexor carpi radialis muscles). The electrodes provide relief of nausea and vomiting through electrical stimulation of the median nerve on the ventral side of the patient's wrist.
The ReliefBand device has a user display that includes a five-level (ReliefBand 1.5) or 10-level (ReliefBand 2.0) LED that indicates the signal intensity. The patient can easily select the pulse amplitude of the electrical impulse and control the intensity of the stimulation via a pushbutton
{4}------------------------------------------------
located on the face of the devices. A low battery indicator LED is also included on the face of the devices. ReliefBand 1.5 uses disposable batteries while ReliefBand 2.0 has a lithium-ion rechargeable battery. A fully charged battery will last for up to 18 hours of continuous use at the mid-power level or below.
## Indications for Use
ReliefBand is indicated for use in the treatment of nausea, vomiting and retching associated with physician-diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy and morning sickness from pregnancy. ReliefBand is also indicated as an adjunct to antiemetics in reducing postoperative nausea. For over the counter use.
### Technical Characteristics
The subject and the predicate ReliefBand devices are identical. Only the wording in the indications for use has changed. The are no changes in technical characteristic between the predicate and the subject devices. The differences between the subject ReliefBand and the predicate device do not raise new questions about the safety or effectiveness of the device.
| Item | ReliefBand® 1.5 and 2.0 (Predicate Devices) | ReliefBand® 1.5 and 2.0 (Subject Devices) |
|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Reliefband® is a digital therapeutic device indicated for use in the treatment of nausea, retching and vomiting due to motion sickness, morning sickness associated with pregnancy, and chemotherapy. Reliefband® is also indicated as an adjunct to antiemetic therapy in reducing postoperative nausea.<br>Relief Band 2.0 is indicated for use in the treatment of nausea, retching and vomiting due to motion sickness, chemotherapy and morning sickness associated with pregnancy. ReliefBand 2.0 is also indicated as an adjunct to antiemetics in reducing postoperative nausea. | Reliefband® is indicated for use in the treatment of nausea, vomiting and retching associated with physician diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy and morning sickness from pregnancy. ReliefBand is also indicated as an adjunct to antiemetics in reducing postoperative nausea. For over the counter use. |
| Size | | |
| ReliefBand 1.5 | Teardrop shape, 3.81Wx5.08Lx1.14H cm | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Rectangular shape, 1.45Wx2.35Lx0.41H cm | Identical for ReliefBand 2.0 |
| Weight | | |
| ReliefBand 1.5 | Approximately 34 grams | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Approximately 50 grams | Identical for ReliefBand 2.0 |
| Operating<br>Temperature | | |
| ReliefBand 1.5 | 10°C to 45°C (50°F to 113°F) ±18% output tolerance at operating extremes | Identical |
| ReliefBand 2.0 | | |
| User Controls | | |
| ReliefBand 1.5 | One Push Button: On/Off/Intensity | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Two Push Buttons: On/Off (1), Intensity (2) | Identical for ReliefBand 2.0 |
| Operating Modes | ON/OFF/Set intensity level | Identical |
| ReliefBand 1.5 | Five flashing LED setting indicators (green), 2<br>second flash period during normal operation. One<br>flashing low battery LED indicator (red), 4<br>second flash period alternating with setting<br>LED when in Low Battery operation. | Identical for Relief Band 1.5 |
| ReliefBand 2.0 | Ten diagonal blue LED bars indicate therapy<br>power level. Four circular blue LED segments<br>indicate battery charge level. | Identical for ReliefBand 2.0 |
| Display Face<br>Location | | |
| ReliefBand 1.5 | Ventral side | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Dorsal Side | Identical for ReliefBand 2.0 |
| Output Waveform | Asymmetric biphasic pulses.<br>Ramp up, constant and ramp down phases. | Identical |
| Peak pulse<br>Amplitude | | |
| ReliefBand 1.5 | Five discrete levels between 10-40 mA (nominal),<br>set by intensity selection; nominally constant<br>current, +/-10% at nominal conditions | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Ten discrete levels between 8-40 mA (nominal),<br>set by intensity selection; nominally constant<br>current, +/-10% at nominal conditions | Identical for ReliefBand 2.0 |
| Pulse frequency | Fixed at 31 Hz (32 millisecond period). | Identical |
| Pulse width | Fixed at 350 milliseconds. | Identical |
| Amplitude<br>modulation cycle | Single cycle | Identical |
| Maximum Charge<br>Delivered (500 Ω<br>load) | Approximately 3.1 microcoulombs (max. 3.56<br>microcoulombs including 15% output tolerance) | Identical |
| Electrodes | | |
| ReliefBand 1.5 | Two gold-plated copper, nickel or zinc | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Two hypoallergenic surgical grade (316L)<br>stainless steel contacts | Identical for ReliefBand 2.0 |
| Device Status<br>Indicators | | |
| ReliefBand 1.5 | Five separate green flashing LEDs indicate 5<br>intensity levels. Selection of the intensity level is<br>performed via a pushbutton located on the user<br>display. A sixth blinking LED is used to display<br>the low battery indicator. | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Ten Intensity level LEDS are arranged in a single<br>line. Selection of the intensity level is performed<br>via a pushbutton on the edge of the face. Battery<br>charge indicator provides 4 charge levels and a<br>red blinking low power warning LED. | Identical for ReliefBand 2.0 |
| Power | | |
| ReliefBand 1.5 | 3V, CR2025 Lithium-ion coin cells only, replaceable. | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Rechargeable 40 mA Li-Ion Polymer - IEC62133 standard. Not replaceable. | Identical for ReliefBand 2.0 |
| Regulation of<br>Stimulus Output<br>with Decreasing<br>Battery Voltage<br>Battery Life | Regulation is ± 5% of nominal output. Device stops stimulation when regulation cannot be maintained (device turns itself off). | Identical |
| ReliefBand 1.5 | 18 hours continuous or intermittent use with two 3V CR2025 | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | Approximately three years/ 500 charge cycles. | Identical for ReliefBand 2.0 |
| Low Battery<br>Warning | | |
| ReliefBand 1.5 | Low battery indicator flashes when approximately 10% of battery life remains. | Identical for ReliefBand 1.5 |
| ReliefBand 2.0 | When battery level registers 15% or less, a single LED segment will switch from static blue to flashing red. | Identical for ReliefBand 2.0 |
| Wrist Band | | |
| ReliefBand 1.5<br>ReliefBand 2.0 | Contains latex<br>Latex-Free | Identical for ReliefBand 1.5<br>Identical for ReliefBand 2.0 |
{5}------------------------------------------------
{6}------------------------------------------------
## Performance Specifications
The performance specifications for the applicant ReliefBand devices are identical, and thus substantially equivalent, to the predicate devices.
## Basis for Determination of Substantial Equivalence
Substantial equivalence was determined on the basis that there have been no changes in the physical device, technology, mechanism of action or intended use. Device modifications include only the clarified intended use, which includes an additional anticipated source of nausea (physician-diagnosed migraine) that Reliefband is intended to treat. These changes do not impact the safety of efficacy of the ReliefBand 1.5 or ReliefBand 2.0 devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.