RELIEFBAND NST DEVICE, MODEL #'S WB-2, WB-6 AND WB-R
K982436 · Woodside Biomedical, Inc. · GZJ · Oct 8, 1998 · Neurology
Device Facts
Record ID
K982436
Device Name
RELIEFBAND NST DEVICE, MODEL #'S WB-2, WB-6 AND WB-R
Applicant
Woodside Biomedical, Inc.
Product Code
GZJ · Neurology
Decision Date
Oct 8, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ReliefBand® NST™ Device is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, pregnancy, and therapy related to acquired immune deficiency syndrome (AIDS).
Device Story
ReliefBand® NST™ (Models WB-2, WB-6, WB-R) is a non-invasive nerve stimulation therapy device worn on the wrist. It delivers electrical stimulation to nerves located between the palmaris longus and flexor carpi radialis tendons. The device features a single dial for user-controlled peak pulse amplitude adjustment, allowing selection of stimulation intensity. An LED indicator displays operating status (green flash) and low battery (red flash). Powered by two 3V lithium batteries; WB-2 and WB-6 are disposable with non-replaceable batteries, while WB-R is reusable with replaceable batteries. Patients operate the device by selecting intensity levels to alleviate nausea and vomiting symptoms. The device provides a non-pharmacological treatment option for patients suffering from NV associated with chemotherapy, motion sickness, pregnancy, or AIDS therapy.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on performance data indicating equivalence to the predicate device (K980333).
Technological Characteristics
Non-invasive nerve stimulation device. Wrist-worn form factor. Powered by two 3V lithium batteries. Single-dial control for peak pulse amplitude adjustment. LED status indicator. Models vary by intensity settings (3 to 5 levels), reusability, and battery life. No specific materials or software algorithms described.
Indications for Use
Indicated for treatment of nausea and vomiting due to chemotherapy, motion sickness, pregnancy, and AIDS-related therapy. For use by patients experiencing these conditions.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS Woodside Biomedical Inc. ReliefBand® NST™ Device Models WB-2, WB-6, and WB-R
K982436
## SUBMITTER INFORMATION
| A. | Company Name: | Woodside Biomedical, Inc. |
|----|------------------------|-----------------------------------------------------------------------------------|
| B. | Company Address: | 1915 Aston Avenue, Suite 102<br>Carlsbad, CA 92008 |
| C. | Company Phone: | (760) 804-6900<br>Company Fax: (760) 804-6925 |
| D. | Contact Person: | Tom Grey<br>Vice-President of Product<br>Development<br>Woodside Biomedical, Inc. |
| E. | Date Summary Prepared: | July 7, 1998 |
### DEVICE IDENTIFICATION
| A. | Generic Device Name: | Nerve Stimulation Therapy Device |
|----|-------------------------|--------------------------------------------------------|
| B. | Trade/Proprietary Name: | ReliefBand® NST™ (Nerve Stimulation<br>Therapy) Device |
| C. | Classification: | Class II |
| D. | Product Code: | GZJ and BWK |
### IDENTIFICATION OF PREDICATE DEVICE
| Device | Manufacturer | 510(k) No. | Date Cleared |
|-----------------------------|---------------------------|------------|-------------------|
| ReliefBand,<br>Model SW-111 | Woodside Biomedical, Inc. | K980333 | February 18, 1998 |
172
{1}------------------------------------------------
#### DEVICE DESCRIPTION
The ReliefBand® NST™ Device Models WB-2, WB-6, and WB-R are non-invasive nerve stimulation therapy devices, and are indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, pregnancy, and therapy related to acquired immune deficiency syndrome (AIDS). The devices are contained within a wristband, and provide relief of nausea and vomiting through electrical stimulation of the nerves in the patient's wrist.
The devices can be worn on the ventral or palmar (i.e., inside) surface of either or both wrists. approximately 2 fingers breadth above the distal skin crease of the wrist joint, between the tendons of the palmaris longus and flexor carpi radialis muscles.
The ReliefBand® NST™ Device Models WB-2, WB-6, and WB-R have a single dial for setting the stimulation , and incorporate an LED which is used to display a normal operating condition (green color at a slow flash rate) and a low battery condition (red color at a faster flash rate).
All three models have a single operating mode, controlled by a single dial, which controls the peak pulse amplitude of the electrical impulse and thereby determines the intensity of the stimulation. The three models vary by (1) the number of intensity settings available to the user, (2) whether or not the device is reusable, and (3) the length of time that the batteries will last. These features are summarized in Table 1:
| | WB-2<br>(2-Day, Disposable) | WB-6<br>(6-Day, Disposable) | WB-R<br>(Reusable) |
|-----------------------------------------------------------|-----------------------------|-----------------------------|--------------------|
| Identification of<br>discrete impulse<br>intensity levels | Level 1 | Level 1 | Level 1 |
| | | Level 1.5 | Level 1.5 |
| | Level 2 | Level 2 | Level 2 |
| | | Level 2.5 | Level 2.5 |
| | Level 3 | Level 3 | Level 3 |
TABLE 1
![]()
{2}------------------------------------------------
The user dial on the WB-2, WB-6, and WB-R devices identifies the intensity level, so the patient can select the desired stimulation easily. All models are powered by two 3V lithium batteries. The batteries in Models WB-2 and WB-6 are not user replaceable. The batteries in Model WB-R are user replaceable and are commercially available. Battery life for the three models is as follows:
- Model WB-2: approximately 2 days (48 hours) on level 2; batteries not replaceable. ●
- Model WB-6: approximately 6 days (144 hours) on level 2; batteries not replaceable. .
- Model WB-R: approximately 6 days (144 hours) on level 2 before the batteries need . to be replaced.
#### SUBSTANTIAL EQUIVALENCE
The Woodside Biomedical, Inc. ReliefBand® NST™ devices, Models WB-2, WB-6, and WB-R devices are of comparable type and are substantially equivalent to the predicate ReliefBand®, Model SW-111 (K980333).
#### INDICATION FOR USE
The ReliefBand NST™ Device is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, pregnancy, and therapy related to acquired immune deficiency syndrome (AIDS).
## TECHNOLOGICAL CHARACTERISTICS
A comparison of the technological characteristics of the ReliefBand® NST™ Device and the predicate device has been performed. The results of this comparison demonstrate that the ReliefBand® NST™ Device is equivalent to the marketed predicate device.
174
{3}------------------------------------------------
## PERFORMANCE DATA
The performance data indicate that the ReliefBand® NST™ Device Models WB-2, WB-6, and WB-R are substantially equivalent to the ReliefBand® Devices distributed under K980333.
175
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular fashion around the eagle.
8 1998 OCT
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Carol L. Patterson Patterson Consulting Group Consultant for Woodside Biomedical, Inc. 18140 Smokesignal Drive San Diego, California 92127
Re: K982436 Trade Name: ReliefBand® NST™ Device Models WB-2, WB-6, WB-R Regulatory Class: II Product Codes: GZJ and BWK Dated: July 13, 1998 Received: July 14, 1998
Dear Ms. Patterson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2 - Ms. Carol L. Patterson
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number: K982436
Devicc Name:
1. P
ReliefBand® NST™ Device Models WB-2. WB-6, and WB-R
Indications For Use:
The ReliefBand® NST™ Device is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
1000 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 - 100 -
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off)
Division of General Restorative Devices K982436
510(k) Number
Prescription Use (Pcr 21 CFR 801.109)
ાર
Over The-Counter Use .........................................................................................................................................................
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.