FD TENS 2090, FD TENS 2095

K152676 · Fuji Dynamics , Ltd. · GZJ · Jan 22, 2016 · Neurology

Device Facts

Record IDK152676
Device NameFD TENS 2090, FD TENS 2095
ApplicantFuji Dynamics , Ltd.
Product CodeGZJ · Neurology
Decision DateJan 22, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

As a TENS device, FD TENS 2090 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain. As a EMS device, FD TENS 2090 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes on the affected body area. It is intended for: - Prevention or retardation of muscle disuse atrophy. - Relaxation of muscle spasm. - Muscle re-education. - Maintaining or increasing range of motion. - Increasing local blood circulation. - Immediate post-surgical of calf muscle to prevent venous thrombosis. FD TENS 2095 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

Device Story

FD TENS 2090 and 2095 are transcutaneous electrical nerve stimulators (TENS) and powered muscle stimulators (EMS). Devices generate electrical pulses transmitted via skin-attached electrodes. TENS mode blocks pain signals to brain; EMS mode induces muscle contraction for therapeutic purposes (atrophy prevention, spasm relaxation, circulation, etc.). Devices are intended for prescription use. Healthcare providers or patients apply electrodes to affected body areas. Output parameters (pulse width, frequency, intensity) are adjustable or preset. Clinical benefit includes pain management and physical therapy support. No clinical data provided; substantial equivalence based on bench testing and comparison to predicate devices.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2) testing.

Technological Characteristics

Symmetrical bi-phasic rectangular waveform; TENS/EMS stimulation; adjustable pulse width (50-250µs) and frequency (1-150Hz); powered by electrical current; electrodes for skin contact; safety compliance with IEC 60601-1 and IEC 60601-1-2.

Indications for Use

Indicated for patients requiring symptomatic relief/management of chronic intractable pain or adjunctive treatment for post-surgical/post-traumatic acute pain. EMS function indicated for muscle disuse atrophy prevention/retardation, muscle spasm relaxation, muscle re-education, range of motion maintenance/increase, local blood circulation increase, and post-surgical calf muscle stimulation to prevent venous thrombosis.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 22, 2016 Fuji Dynamics Ltd. Man Man Chung Product Development Manager 1-3. 23/F.. Laws Commerical Plaza 788 Cheung Sha Wan Road Hong Kong, China Re: K152676 Trade/Device Name: FD TENS 2090 and FD TENS 2095 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZJ, IPF Dated: December 16, 2015 Received: December 21, 2015 Dear Man Man Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Michael J. Hoffmann -A for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K152676 Device Name FD TENS 2090 and FD TENS 2095 #### Indications for Use (Describe) As a TENS device, FD TENS 2090 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain. As a EMS device. FD TENS 2090 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes on the affected body area. It is intended for: - Prevention or retardation of muscle disuse atrophy. - · Relaxation of muscle spasm. - · Muscle re-education. - · Maintaining or increasing range of motion. - · Increasing local blood circulation. - · Immediate post-surgical of calf muscle to prevent venous thrombosis. FD TENS 2095 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------|--| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | | | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary [As required by 21 CFR 807.92] #### Device 510(k) number: K152676 #### 1. Applicant Information | Date prepared: | Aug 3, 2015 | | |-----------------|----------------------------------------|--| | Submitter: | Fuji Dynamics Ltd. | | | | Unit 1-3, 23/F., Laws Commercial Plaza | | | | 788 Cheung Sha Wan Road, Kowloon | | | | Hong Kong | | | Contact Person: | Man Man Chung | | | Tel: | (852) 2786 4218 | | | Fax: | (852) 2744 6775 | | #### 2. General Device Information/Trade name | Model No: | FD TENS 2090<br>FD TENS 2095 | |--------------|------------------------------| | Common Name: | Transcutaneous Elec | trical Nerve Stimulator and Powered Muscle Stimulator GZJ, IPF Product code: Classification: Class II #### 3. Predicate Device Information | 510k number | Device Name | Manufacturer | Date cleared | |-------------|--------------|---------------|--------------| | K052813 | FD TENS 2030 | Fuji Dynamics | 2006 | | K063642 | FD EMS | Fuji Dynamics | 2007 | #### 4. Device Description #### FD TENS 2090 FD TENS 2090 can provide both TENS treatments and EMS treatments. During TENS treatments, the FD TENS 2090 generates electrical pulses and transmit it to the electrodes which are attached to the patient's skin. Consequently, the electrical pulses would then pass through the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain. During EMS treatments, FD TENS 2090 generates electrical pulses and transmit it to the electrodes attached to the patient skin, causing the muscle to expand and contract. It is used to relax muscle spasms. prevent or retard atrophy, maintain or increase range of motion, increase local blood circulation, reeducate muscle and provide immediate post-surgical stimulation of calf muscle to prevent venous thrombosis. #### FD TENS 2095 FD TENS 2095 can provide TENS treatment, During treatment, the FD TENS 2095 generates electrical pulses and transmit it to the electrodes, which are attached to the patient's skin. Consequently, the electrical pulses would then pass through the skin to the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain. {4}------------------------------------------------ ## 5. Intended Use: #### FD TENS 2090 As a TENS device, FD TENS 2090 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain. As a EMS device, FD TENS 2090 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes on the affected body area. It is intended for: - · Prevention or retardation of muscle disuse atrophy. - Relaxation of muscle spasm. - · Muscle re-education. - · Maintaining or increasing range of motion. - · Increasing local blood circulation. - · Immediate post-surgical of calf muscle to prevent venous thrombosis. ### FD TENS 2095 FD TENS 2095 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain. #### 6. Comparison to Predicate Device: TENS Program comparison between FD TENS 2090, FD TENS 2095 and FD TENS 2030 | Characteristic | Subject device K152676 | Predicate device K052813 | |---------------------------------------------------------|-------------------------------|---------------------------| | Device | FD TENS 2090 and FD TENS 2095 | FD TENS 2030 | | Waveform | Symmetrical Bi-Phasic | Symmetrical Bi-Phasic | | Shape | Rectangular | Rectangular | | Maximum Voltage<br>Load: 0.50 kohm | (0-peak voltage)<br>48.8V | (0-peak voltage)<br>31.3V | | 02 kohm | 70.4V | 50.7V | | 10 kohm | 88.8V | 85.0V | | Maximum Current<br>Load: 0.50 kohm | 97.6 mA | 62.6 mA | | 02 kohm | 35.2 mA | 25.35 mA | | 10 kohm | 8.88 mA | 8.5 mA | | Maximum pulse width | 250μs | 250μs | | Maximum frequency | 150 Hz | 200 Hz | | Maximum Output<br>Net charge per phase<br>Load: 500 ohm | 10.41 μC | 5.83μC | | Maximum output<br>Charge per phase<br>Load: 500 ohm | 22.13μC | 15.77μC | | 1kohm | 16.33μC | 13.60μC | | Maximum output<br>RMS Current<br>Load: 500 ohm | 11.55 mA rms | 10.99 mA rms | | 1kohm | 8.49 mA rms | 8.95 mA rms | | Max Current Density | 0.15 | 0.1 | {5}------------------------------------------------ | (mA/cm²) | Max Power Density (W/cm²) | | |------------------------|----------------------------------|-----------------------------| | | .00525 | .00182 | | Burst Mode | Burst 1 | Burst II | | a. Pulses per burst | 7 | 7 | | b. Burst per second | 2 bursts per second | 2 | | c. Burst duration | 250 ms | 250ms | | d. Duty cycle | 50% | 50% | | | Burst 2 | Burst I | | a. Pulses per burst | 80 | 80 | | b. Burst per second | 1 burst per 2 seconds | 1 burst per 2 seconds | | c. Burst duration | 1 second | 1 second | | d. Duty cycle | 50% | 50% | | Continuous Mode | P1 CONST | CONSTANT | | a. Pulse width | 59µs default | 50µs default | | | Adjustable from 50µs to<br>250µs | Adjustable from 25 to 250µs | | b. Pulse frequency | 120 Hz default | 120 Hz default | | | Adjustable from 1 to 150 Hz | Adjustable from 1 to 200 Hz | | Pulse width modulation | P4 MODUL 1 | W. MODUL | | a. Pulse width | 50µs to 250µs in 5s | 25µs to 250µs in 8s | | b. Pulse frequency | 120Hz default | 120 Hz default | | | Adjustable from 1 to 150 Hz | Adjustable from 1 to 200 Hz | | Frequency modulation | P5 MODUL 2 | F. MODUL | | a. Pulse frequency | 20 to 100 Hz in 5s | 20 to 100Hz in 8s | | b. Pulse width | 150µs default | 50µs default | | | Adjustable from 50 to 250µs | Adjustable from 25 to 250µs | ## EMS Program comparison between FD TENS 2090 and FD EMS | Characteristic | Subject device K152676 | Predicate device K063642 | |--------------------|------------------------|--------------------------| | Trade name | FD TENS 2090 | FD EMS | | Waveform | Symmetrical Bi-Phasic | Symmetrical Bi-Phasic | | Shape | Rectangular | Rectangular | | Maximum Voltage | (0 to peak voltage) | (0 to peak voltage) | | Load: 500 ohm | 48.4 V | 48.7 V | | 2 kohm | 67.2 V | 60.0 V | | 10 kohm | 87.2 V | 64.7 V | | Max Output current | | | | Load: 500 ohm | 96.8 mA | 97.40 mA | {6}------------------------------------------------ | 2 kohm | 33.6 mA | 30.0 mA | |---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 10 kohm | 8.72 mA | 6.47 mA | | Max pulse width | 300μs | 300μs | | Max frequency | 40 Hz | 60 Hz | | Biphasic waveform | Symmetrical Bi-Phasic<br>Rectangular Waveform | Symmetrical Bi-Phasic<br>Rectangular Waveform | | - Symmetrical phases<br>- Phase Duration | Fixed Pulse width, max 300 μs | Fixed Pulse width, max 300 μs | | Maximum Output Net<br>Charge Per Phase<br>Load: 500 ohm | 13.05μC | 3.57 μC | | Maximum Output Charge<br>Per Phase<br>Load: 500 ohm<br>1 kohm | 24.11 μC<br>17.77 μC | 48.61 μC<br>28.77 μC | | Maximum Output RMS<br>Current<br>Load: 500 ohm<br>1 kohm | 6.62 mA rms<br>4.91 mA rms | 12.5 mA rms<br>7.40 mA rms | | Max Current Density<br>(mA/cm2) | 0.046 | 0.1966 | | Max Power Density (W/cm2) | 0.00152 | 0.00614 | | On Time | 5 or 10 sec fixed | 1 to 9 sec adjustable | | Off Time | 10 or 30 sec fixed | 1 to 9 sec adjustable | | Output pattern -<br>EMS Cycle Mode | Fixed pulse frequency, ramp<br>up/down and work/rest time.<br>Only pulse width is user<br>selectable.<br>(P6,P8,P9) | Fixed pulse widthandwork/rest time.<br>Selectable frequency and ramp<br>up/down time. | | EMS Reciprocal Mode | Fixed pulse frequency, ramp<br>up/down and work/rest time.<br>Only pulse width is user<br>selectable<br>(P7). | Fixed pulse width andwork/rest time.<br>Selectable pulse frequency and ramp<br>up/down time. | | Preset Program,P6 of subject<br>device | Program Name<br>P6<br>EMS Mode<br>Cycle mode<br>Pules Freq: 35Hz<br>Pulse width:300μs<br>Ramp up : 2sec<br>On time : 10sec<br>Ramp down : 2sec<br>Off time : 30sec | Program Name<br>CYCLE 2<br>EMS Mode<br>Cycle mode<br>Pules Freq: 50Hz*<br>Pulse width:300μs<br>Ramp up : 3sec<br>On time : 9sec<br>Ramp down : 3sec<br>Off time : 15sec | | Preset Program,P7of subject<br>device | Program Name<br>P7<br>EMS Mode<br>Reciprocal mode<br>Pules Freq: 40Hz<br>Pulse width:300μs | Program Name<br>RECIPROCAL 3<br>EMS Mode<br>Reciprocalmode<br>Pules Freq: 50Hz*<br>Pulse width:300μs | {7}------------------------------------------------ | | Ramp up : 4sec | Ramp up : 4sec | |--------------------|----------------------------|----------------------| | | On time : 10sec | On time : 8sec | | | Ramp down : 4sec | Ramp down :4sec | | | Off time : 10sec | Off time : 14sec | | Preset Program, P8 | Program Name | Program Name | | | P8 | CYCLE 2 | | | EMS Mode | EMS Mode | | | Cyclemode | Cycle mode | | | Pules Freq: 40Hz | Pules Freq: 50Hz* | | | Pulse width:300μs | Pulse width:300μs | | | Ramp up : 4sec | Ramp up : 3sec | | | On time : 10sec | On time : 9sec | | | Ramp down : 4sec | Ramp down : 3sec | | | Off time : 10sec | Off time : 15sec | | Preset program P9 | Program Name | Program Name | | | P9 | CYCLE 1 | | | EMS Mode | EMS Mode | | | Cyclemode | Cycle mode | | | Pules Freq: 20Hz | Pules Freq: 50Hz* | | | Pulse width:200μs | Pulse width:300μs | | | Ramp up : 2sec | Ramp up : 2sec | | | On time : 5sec | On time : 6sec | | | Ramp down : 2sec | Ramp down : 2sec | | | Off time : 10sec | Off time : 10sec | | Preset Program, P1 | Program Name | Program Name | | | P1 | CONSTANT 1 | | | Pulse width: | Pulse width: | | | 120µs default | 50 µs fixed | | | Adjustable from | Pulse freq: | | | 50µs to 250µs | 30Hz fixed | | | Pulse freq: | -------------------- | | | 120Hz default | Program Name | | | adjustable from 1 to 150Hz | CONSTANT 2 | | | | Pulse width | | | | 200µs fixed | | | | Pulse freq: | | | | 50Hz | #### 7. Non clinical Testing IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) IEC 60601-1-2:2007 (Edition 3) Safety requirement EMC requirement #### 8. Clinical testing None #### 9. Conclusion FD TENS 2090 and FD TENS 2095 have the same intended use and the same technical characteristics as the predicate device(s), FD TENS 2030 (K052813) and FD EMS (K063642). FD TENS 2090, FD TENS 2095 are as safe and as effective as the predicate device. Therefore, FD TENS 2090 and FD TENS 2095 are substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...