FDES106 MINI TENS&EMS DEVICE,FDES106 MULTI-FUNCTION MINI TENS&EMS DEVICE,FDES105 PAIN RELIEF PLASTER,FDES107 ABDOMINAL
K130723 · Famidoc Technology Co., Ltd. · NGX · Feb 7, 2014 · Physical Medicine
Device Facts
Record ID
K130723
Device Name
FDES106 MINI TENS&EMS DEVICE,FDES106 MULTI-FUNCTION MINI TENS&EMS DEVICE,FDES105 PAIN RELIEF PLASTER,FDES107 ABDOMINAL
Applicant
Famidoc Technology Co., Ltd.
Product Code
NGX · Physical Medicine
Decision Date
Feb 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FDES106 (ED406) Mini TENS &EMS Device For program N1, B and H of TENS mode To be used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household work activities. For program E1 and E2 of EMS mode Used to stimulate healthy muscles in order to improve and facilitate muscle performance. FDES106A (ED406A) Multi-function Mini TENS &EMS Device For program N1, B and H of TENS mode To be used for temporary relief of pain associated with sore and aching muscles in the lower back, abdomen, thigh, and arm due to strain from exercise or normal household work activities. For program E1 and E2 of EMS mode Used to stimulate healthy muscles in the lower back, abdomen, thigh, and arm in order to improve and facilitate muscle performance. FDES105 (ED405) Pain Relief Plaster To be used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household work activities. FDES107 (ED407) Abdominal Fitness Belt Used to stimulate healthy muscles in order to improve and facilitate muscle performance for abdominal.
Device Story
Battery-powered (3.0V DC) portable electro-stimulator series (FDES106, FDES106A, FDES105, FDES107) providing TENS and/or EMS therapy. Device delivers pre-programmed electrical pulses via electrode pads or electrode belts applied to skin. User controls intensity and program selection via press buttons. TENS mode provides pain relief; EMS mode induces muscle contractions to improve performance. Used in home environment by patients. Output parameters (waveform, frequency, pulse width) are pre-programmed. Healthcare providers/patients use device to manage muscle soreness or facilitate muscle conditioning. Benefits include non-invasive pain management and muscle training.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-4, and IEC 60601-2-10 standards. Biocompatibility testing performed per ISO 10993-5 and ISO 10993-10.
Indicated for temporary relief of pain associated with sore/aching muscles due to exercise or household work, and for stimulation of healthy muscles to improve/facilitate muscle performance. Applicable to adults using OTC. Specific models target lower back, abdomen, thigh, and arm.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
Slenderton FLEX Abdominal Training system type 515 (K030708)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Famido" in a simple, outlined font. Each letter is clearly defined with a black outline, giving it a clean and slightly retro appearance. The word is presented in a straightforward, horizontal arrangement.
## 510(k) SUMMARY [As required by 21CFR 807.92]
FEB - 7 2014
FDES106(ED406) Series Electro-Stimulator, K (130723 )
#### Submitter's Information [21CFR 807.92(a)(1)] 1.
| Company Name: | Famidoc Technology Co., Ltd |
|------------------|-------------------------------------------------------------|
| Street Address: | No. 212 Yilong Road, Hexi Industrial Zone, Jingxia, Changan |
| | Town |
| City: | Dongguan |
| State/ Province: | Guangdong |
| Country: | China |
| Telephone: | +86(769) 89272488-8674 |
| Fax: | +86(769) 89272498 |
| Contact Person: | Reanny Wang |
| Contact Title: | Vice-general Manager |
| Contact Email: | qa@famidoc.com |
#### 2. Trade Name, Common Name, Classification [21CFR 807.92(a)(2)]
| a) Trade Name: | FDES106(ED406) Mini TENS&EMS Device<br>FDES106A(ED406A) Multi-function TENS&EMS Device |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Electro-Stimulator or Electrical Stimulator |
| Classification Name: | Stimulator, Muscle, Powered, for muscle conditioning<br>per 21 CFR § 890.5850;<br>Transcutaneous Electrical Nerve Stimulator for Pain Relief;<br>Stimulator, Nerve, Transcutaneous, Over-the-Counter<br>per 21 CFR § 882.5890 |
| Device Class: | Class II |
| Product Code: | NUH, NGX |
Section 07-Page 1 of 13
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Famidoc" in a simple, outlined font. The letters are all connected and have a uniform thickness. The word is horizontally oriented and appears to be a logo or brand name.
| b) Trade Name: | FDES105(ED405) Pain Relief Plaster |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | TENS or TENS Device |
| Classification Name: | Transcutaneous Electrical Nerve Stimulator for Pain Relief<br>Stimulator, Nerve, Transcutaneous, Over-the-Counter<br>per 21 CFR § 882.5890 |
| Device Class: | Class II |
| Product Code: | NUH |
| c) Trade Name: | FDES107(ED407) Abdominal Fitness Belt |
| Common Name: | Powered Muscle Stimulator, OTC |
| Classification Name: | Stimulator, Muscle, Powered, for muscle conditioning<br>per 21 CFR § 890.5850; |
| Device Class: | Class II |
| Product Code: | NGX |
- 3. Identification of Predicate Device(s)[21 CFR 807.92(a)(3)]
| | PREDICATE DEVICES | |
|-------------------------|------------------------|----------------------------------------------------------|
| Manufacturer | Endurance Therapeutics | Bio-Medical Research<br>Ltd |
| Legally Marketed Device | T1040™ | Slenderton FLEX<br>Abdominal Training<br>system type 515 |
| 510 (K) Number | K060846 | K030708 |
#### Description of Device[21 CFR 807.92(a)(4)] 4.
The FDES106(ED406) Series Stimulator, which includes models FDES106(ED406), FDES106A(ED406A), FDES105(ED405) and FDES107(ED407), are Transcutaneous Electrical Nerve Stimulator and muscle stimulator for pain relief and/or Electrical Muscle Stimulator. The stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. The parameters of units are controlled by the press buttons. Its intensity level is adjustable according to the needs of patients.
{2}------------------------------------------------
The device unit of FDES106(ED406) and FDES106A(ED406A) are same, only the applied part are different: the applied part of FDES106(ED406) is electrode pad; the applied part of FDES106A(ED406A) is electrode belt. They are portable device, battery powered (3.0V DC) multi-function device offering both Transcutaneous Electrical Nerve Stimulator(TENS) and Powered Muscle Stimulator (EMS) qualities in one device.
The FDES106 (ED406) includes FDE\$105 (ED405). FDES106 (ED406) has TENS and EMS two treatment mode 5 programs (N, B, H, E1, E2), FDES105 (ED405) only has TENS mode 5 treatment programs (N1, N2, B, H, F).Their appearance, structure, circuit, software operation is exactly the same, only different output treatment procedure.
FDES107 (ED407) has 10 EMS treatment process, whose intended use, waveform characteristics and is basically the same as FDES106 (ED406) EMS model, The difference is from only operation mode and structure appearance.
Independent channel (by electrode pad or electrode belt) that effectively transfers your desired choice of pre-programmed electrical pulses directly through electrode pad or electrode belt) to suggested area of the body where the electrode are placed, causing minimal muscle contractions. The FDES106 (ED406 ) and FDES106A (ED406A ) have 3 TENS programs and 2 EMS programs; FDES105 (ED405) have 5 TENS programs; FDES107 (ED107) have 10 EMS programs.
### 5. Intended Use[21 CFR 807.92(a)(5)] FDES106(ED406) Mini TENS &EMS Device
#### For program N, B and H of TENS mode
To be used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household work activities.
#### For program E1 and E2 of EMS mode
Used to stimulate healthy muscles in order to improve and facilitate muscle performance.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Famidoe" in a simple, outlined font. The letters are all capitalized except for the "i" and "o", which are lowercase. The word is horizontally oriented and appears to be a logo or brand name.
#### FDES106A(ED406A) Multi-function Mini TENS &EMS Device
#### For program N, B and H of TENS mode
To be used for temporary relief of pain associated with sore and aching muscles in the lower back, abdomen, thigh and arm due to strain from exercise or normal household work activities.
#### For program E1 and E2 of EMS mode
Used to stimulate healthy muscles in the lower back, abdomen, thigh and arm in order to improve and facilitate muscle performance.
#### FDES105(ED405) Pain Relief Plaster
To be used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household work activities.
#### FDES107(ED407) Abdominal Fitness Belt
Used to stimulate healthy muscles in order to improve and facilitate muscle performance for abdominal.
#### 6. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: [21 CFR 807.92(a)(6)]
The FDES106(ED406) Series Electro-Stimulator did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Nonclinical testing was performed in order to validate the design according with the company's specified design requirements, and to assure conformance with the following voluntary design standards:
- > IEC 60601-1, Medical electrical equipment Part 1: General requirements for safety".
- > IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility- Requirements and tests, Interpretation Sheet."
- > IEC 60601-1-11, Medical electrical equipment Part 1-11: General Requirements for basic safety and essential performance - Collateral Standard: Requirements for
Section 07-Page 4 of 13
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Famiddoc" in a simple, outlined font. The letters are evenly spaced and of similar size, creating a balanced appearance. The outline style gives the word a clean and modern look.
medical electrical equipment and medical electrical systems used in the home healthcare environment.
- > IEC 60601-1-4, Medical electrical equipment Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems
- > IEC 60601-2-10, Medical electrical equipment Part 2: Particular requirements for the safety of nerve and muscle stimulators
- > In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
#### 7. Biocompatibility Certification for accessories [21 CFR 807.92(a)(7)]
The materials of applied part are Electrode pad and Electrode belt; the material of enclose is ABS. They are both meet the biocompatibility testing of ISO 10993-5 and ISO 10993-10 standards.
#### 8. Comparison for Predicate Device & Subject Device [21 CFR 807.92(a)(8)}
We present the relevant information for the predicate device here for demonstrating the characteristics of the predicate device.
#### 8.1 Comparison of significant device features
{5}------------------------------------------------
Sponsor: Famidoc Technology Co., Ltd.
File No: XVV-Stimulator A-FDA-07
Version: 1.2
Data: Feb.7, 2014
| Comparison item | Mini TENS&<br>EMS Device<br>FDES106 | Multi-function<br>Mini TENS<br>&EMS Device<br>FDES106A | Pain<br>Relief<br>Plaster<br>FDES105 | New device<br>Abdominal Fitness<br>Belt FDES107 | Predicate device<br>T1040™ | Predicate device<br>Slenderton FLEX<br>Abdominal Training<br>system type 515 |
|-----------------------------------------------|-------------------------------------|--------------------------------------------------------|--------------------------------------|-------------------------------------------------|-----------------------------------------------|------------------------------------------------------------------------------|
| 510K# | | | Pending | | K060846 | K030708 |
| Manufacturer | | | Famidoc Technology Co., Ltd. | | Endurance<br>Therapeutics | Bio-Medical<br>Research Ltd |
| Prescription or OTC | | | OTC | | OTC | OTC |
| FDA product code | | NUH, NGX | NUH | NGX | NUH, NGX | NGX |
| Power source | | Battery powered, d.c. 3.0V, 2 X AAA<br>batteries | | | Battery powered, d.c. 4.5V, 3 X AAA batteries | |
| User interface | | By LED light and silk-screen indication | | | By LCD display | |
| Output channel | | | Single channel | | Two channels | |
| Number of output models | | TENS and EMS | TENS | EMS | TENS and<br>PMS(EMS) | EMS |
| Number of treatment programs | | 5 | 5 | 10 | 10 | 7 |
| Number of<br>output channels | | Alternating | | | Alternating | Symmetrical<br>Biphasic |
| Number of output channels | | | 1 | | 1 | Dual |
| Method of channel isolation | | | By electrical circuit and software | | | By electrical circuit and software |
| Constant Current or Constant Voltage? | | | Constant Voltage | | | Constant Voltage |
| Waveform | | | Biphasic square | | | Biphasic |
| Software/Firmware/ Microprocessor<br>Control? | | | Yes | | | Yes |
Section 07-Page 6 of
Section 07-page 69
{6}------------------------------------------------
nsor: Famidoc Technology
No: XW-Stimulator A-FDAersion: 1.2
ata: Feb.7, 201
amidoc
Slenderton FLE)
bdominal Trainin output no pul
applied system type 51 lonadjustable Inknowabl Inknowabl nknowab nknowabl nknowab EMS: 0
EMS: MA 25 & 30 Yes Yes Yes Yes Yes astic (ABS) enclosu Predicate device lo output no
ulse applied onadjustab nknowabl nknowabl nknowabl nknowab EMS: 0
LEMS: 0 5 minute 1040™ Yes Yes Yes Yes Yes Yes 25 and 30m odominal Fitne
Belt FDES107 위 대한 함께 이상
onadjustab 1.26 0.8 Yes Pain Relie lastic (ABS) enclosu FDES10 Plaster EMS: NA o output applied pulse New devic (es (es Yes (es res res (es nadjustable 30 minute lulti-function EMS Devic Mini TENS FDES106 0.9 0.8 No TENS: 0
appliec lini TENS&
EMS Device FDES10 oltage/ current lev ngle fault conditi rmal condition age DC current through elect thereige D'O our on through olds.
When device is on but no pulses being applied (uA) On/off status w battery sing materials construct mparison ite omatic Over Current T tient Override Control? omatic No Load Trip omatic overload tri (UA) omatic shut off me range (mi ndicatior eakag function Current Patient
ction 07-Page 7
Section 07-Page 2
{7}------------------------------------------------
nsor: Famidoc Technology Co.,
No: XW-Stimulator A-FDA-07 Version: 1.2
Data: Feb.7, 201
Transparent silic Slenderton FLEX Enclosure - ABS ystem type 51 Abdome lectrode be Yes Predicate device (leg);
For EMS: Any are
which been
escribed in the
ser manual ca Electrode —
Transparent silic For TENS mode
Low back, upper
extremities(arm ower extremiti eatment are Enclosure — AB T1040T" not use Yes 2015 September 1998
Research
Requires (1988) Marquest (1988) Marques (1998) Marquin (1998) North (1998) North (1990)
Buyung (1990) Next (1990) Next (1990) Next (1 המטר המקום של המקום המקום המקום המקום המקום המקום המקום המקום המקום המוקדות המקום המוקדות המוקדות המוקדות המוקדות המוקדות המוקדות המוקדות המוקדות המוקדות המוקדות המונחים של Abdomer ្ម (Gal)
សូមមនុស្សនា
សេចក (Jul tremities Pain Relie ow bacl ranspare FDES10 silica gel
Enclosure upper Plaste ectrode ABS New devic Yes ב-1998 הגדולית
שניירות חיצוניים
Arm, Wais Abdomer Thigh an ow bacl Buttock, ni TENS& EMS lectrode –
ansparent silic or TENS mode
ow back, uppe
xtremities(arm xtremities (leg)
or EMS: Any nclosure - ABS ser manual ca rea which bee escribed in th ose treatmer Device
FDES106 rea (Excep ot use ower mparison ite mpliance with 21 Cl t of patient contact atment are material(s) 898 ?
tion 07-Page 8
{8}------------------------------------------------
Sponsor: Famidoc Technology Co., Ltd.
File No: XVV-Stimulator A-FDA-07
Version: 1.2
Data: Feb.7, 2014
| Comparison item | New device | | | Predicate device | | |
|-------------------------|------------------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------|--------------------------------------|------------------------------|----------------------------------------------------------|
| | Mini TENS&<br>EMS Device<br>FDES106 | Multi-function<br>Mini TENS<br>&EMS Device<br>FDES106A | Pain Relief<br>Plaster<br>FDES105 | Abdominal<br>Fitness Belt<br>FDES107 | T1040™ | Slenderton FLEX<br>Abdominal Training<br>system type 515 |
| Classification | Type of protection<br>against electric shock | Internally powered equipment | | | Internally powered equipment | |
| | Degree of protection<br>against electric shock | Type BF applied part | | | Type BF applied part | |
| | Device Class | Class II | | | Class II | |
| | Biocompatibility | Compliant with requirements of ISO 10993-5 and ISO 10993-10 | | | Same | Same |
| Compliance<br>with | Mechanical Safety | Compliant with requirements of IEC 60601-1, IEC 60601-2-10 | | | Same | Same |
| Voluntary<br>Standards? | Electrical Safety | Compliant with requirements of IEC 60601-1, IEC 60601-2-10,<br>IEC 60601-1-2 safety standards | | | Same | Same |
| | Energy delivered | The delivered energy is limited according to requirements of<br>collateral IEC 60601-2-10 safety standards | | | Same | Same |
| | Other | Compliant with requirements of IEC 60601-11 safety standard | | | Unknowable | Unknowable |
| | Applied part | Electrode pad | Electrode<br>belt | Electrode<br>pad | Abdominal<br>electrode belt | Abdominal<br>electrode belt |
| | Weight (lbs., oz.) | 0.093 | 0.093 | 0.093 | 0.18 | 0.19 |
| | Dimensions(in.) [W × H × D]<br>For unit | 2.25"×1.77"×0.36" | | | 3.62"×3.19"×0.75" | 5.91"x2.68"x<br>1.02" |
| | | | | | | Unknowable |
| | | | | | | Unknowable |
Section 07-Page 9 of
{9}------------------------------------------------
Sponsor: Famidoc Technology Co., Ltd.
File No: XVV-Stimulator A-FDA-07
Version: 1.2
Data: Feb.7, 2014
| Comparison item | Mini TENS&<br>EMS Device<br>FDES106 | Multi-function<br>Mini TENS<br>&EMS Device<br>FDES106A | Pain Relief<br>Plaster<br>FDES105 | Abdominal<br>Fitness Belt<br>FDES107 | New device | T1040™ | Abdominal Training<br>Slenderton FLEX<br>system type 515 | Predicate device |
|---------------------------------------|-------------------------------------|--------------------------------------------------------|-----------------------------------|--------------------------------------|------------|------------|----------------------------------------------------------|------------------|
| Operating temperature and<br>humidity | | 5-40°C, 30%-85% | | | | Unknowable | 0-35°C, 20%-65% | |
| Storage temperature and humidity | | -10-50°C, 10%-90% | | | | Unknowable | 0-52°C, 10%-90% | |
8.2 Comparison of significant output specifications
| Comparison item | | New device | | | Predicate device | | |
|------------------------|-----------|-------------------------------------|--------------------------------------------------------|-----------------------------------|--------------------------------------|----------|----------------------------------------------------------|
| | | Mini TENS&<br>EMS Device<br>FDES106 | Multi-function<br>Mini TENS<br>&EMS Device<br>FDES106A | Pain Relief<br>Plaster<br>FDES105 | Abdominal<br>Fitness Belt<br>FDES107 | T1040™ | Slenderton FLEX<br>Abdominal Training<br>system type 515 |
| Waveform | TENS mode | Biphasic | Biphasic | Biphasic | N/A | Biphasic | N/A |
| | EMS mode | Biphasic | Biphasic | N/A | Biphasic | Biphasic | Biphasic |
| Shape | TENS mode | Square | Square | Square | N/A | Square | N/A |
| | EMS mode | Square | Square | N/A | Square | Square | Square |
| Max Output Voltage (V) | TENS mode | 30.4 | 30.4 | 28 | N/A | 40.7 | N/A |
| 500Ω load | EMS mode | 28 | 28 | N/A | 46.4 | 40.7 | 47.3 |
| 2kΩload | TENS mode | 44.8 | 44.8 | 36.8 | N/A | 105.1 | N/A |
| | EMS mode | 44.8 | 44.8 | N/A | 72 | 105.1 | 89.3 |
# Section 07-Page 10 of
{10}------------------------------------------------
nsor: Famidoc Technology No: XW-Stimulator A-Fl ersion: 1.2
ata: Feb.7, 201
Slenderton FLE)
bdominal Traini ystem type 51 Inknowable 200-300 45-75 28.41 91.3 94.7 89.3 60.7 n/A N/A n/A N/A N/A N/A N/A N/A redicate device 1040 T 4.1-50 4.1-50 154.1 47.8 47.8 15.4 15.4 154. 81.4 81.4 245 245 40.7 40.7 2.71 2.71 Fitness Bel Abdomina FDES10 92.8 3-80 23.2 23.2 120 N/A N/A 250 N/A N/A N/A N/A N/A N/A ਰੇ ਟ ਮ Pain Relie FDES10: 00-25 Plaste 41.6 15.2 15.2 18.4 2-80 N/A N/A N/A n/A 4.2 N/A N/A N/A N/A ്ക് New devic ulti-functior EMS Devic Mini TENS FDES106 00-250 60.8 22.4 15.2 15.2 54.4 54.4 22.4 2-80 2-50 5.4 5.4 250 す す റ്റ് ni TENS& MS Device
FDES106 00-25 15.2 54.4 54.4 60.8 22.4 22.4 2-80 2-50 5.2 5.4 5.4 250 56 ম す ENS mode ENS mode ENS mode ENS mod ENS mod ENS mod ENS mod ENS mod MS mode MS mode MS mode MS mode MS mode MS mode MS mode EMS mode omparison ite x Output Current (mA) Maximum Phas ulse cycle (ut let Charge pe equency(H: Charge (uC Pulse Widt Range(uS 00Ω load 0kΩ load 0kΩ load kQload 2000) 500Ω
tion 07-Page 1
amido
{11}------------------------------------------------
Sponsor: Famidoc Technology Co., Ltd.
File No: XVV-Stimulator A-FDA-07
Version: 1.2
Data: Feb.7, 2014
| Comparison item | | Mini TENS&<br>EMS Device<br>FDES106 | Multi-function<br>Mini TENS<br>&EMS Device<br>FDES106A | Pain Relief<br>Plaster<br>FDES105 | Abdominal<br>Fitness Belt<br>FDES107 | T1040™ | Slenderton FLEX<br>Abdominal Training<br>system type 515 |
|----------------------------------------------------------------------------------------------------|------------------------------------|--------------------------------------------------------|--------------------------------------------------------|--------------------------------------|--------------------------------------|----------------------------------------------------------|----------------------------------------------------------|
| Maximum Current Density<br>(mA/cm², 500Ω, r.m.s) | TENS mode | 4.1 | 3.4 | 4.1 | N/A | 2.71 | N/A |
| | EMS mode | 3.77 | 3.13 | N/A | 1.9 | 2.71 | Unknowable |
| Maximum Average<br>Current (average absolute<br>value) [mA, 500Ω] | TENS mode | 0.51 | 0.42 | 0.48 | N/A | Unknowable | N/A |
| | EMS mode | 0.41 | 0.34 | N/A | 0.3 | Unknowable | Unknowable |
| Maximum Power Density<br>(W/cm², 500Ω, r.m.s) | TENS mode | 0.124 | 0.103 | 0.105 | N/A | 0.099 | N/A |
| | EMS mode | 0.105 | 0.087 | N/A | 0.039 | 0.099 | Unknown |
| Maximum Average Power<br>Density (mW/cm²) [using<br>smallest electrode<br>conductive surface area] | TENS mode | 2 | 1.4 | 1.7 | N/A | 5.35 | N/A |
| | EMS mode | 1.2 | 1 | N/A | 1.8 | 5.35 | Unknowable |
| Duration of primary<br>(depolarizing) phase (ms) | TENS mode | 0.25 | 0.25 | 0.25 | N/A | 0.5 | N/A |
| | EMS mode | 0.25 | 0.25 | N/A | 0.25 | 0.5 | 0.3 |
| Burst Mode (For TENS mode only) | | | | | | | |
| a. Pulse per burst | | 7 | 7 | 7 | N/A | N/A | N/A |
| b. Bursts per second | | 2 | 2 | 2 | N/A | N/A | N/A |
| c. Burst duration (ms) | | 0.2 | 0.2 | 0.2 | N/A | N/A | N/A |
| d. Duty Cycle (ms) | | 1.4 | 1.4 | 1.4 | N/A | N/A | N/A |
| Comparison item | New device | | | | Predicate device | | |
| | Mini TENS<br>EMS Device<br>FDES106 | Multi-function<br>Mini TENS<br>&EMS Device<br>FDES106A | Pain Relief<br>Plaster<br>FDES105 | Abdominal<br>Fitness Belt<br>FDES107 | T1040™ | Slenderton FLEX<br>Abdominal Training<br>system type 515 | |
| For EMS mode only | | | | | | | |
| On Time (S) | 5 | 5 | N/A | 4-8 | N/A | Unknowable | |
| Off Time (S) | 10 | 10 | N/A | 2-10 | N/A | Unknowable | |
| Ramp up time(S) | 2 | 2 | N/A | 2-10 | N/A | Unknowable | |
| Ramp down time(S) | 2 | 2 | N/A | 0-12 | N/A | Unknowable | |
Section 07-Page 12 of 13
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the word "Famidoo" written vertically. The letters are outlined in black. The word is oriented so that the first letter, "F", is at the bottom and the last letter, "o", is at the top.
sor: Famidoc Technolog No: XW-Stimulator A-Fl Data: Feb.7, 201 ersion: 1.2
# Conclusion कं
ndamental scientific technology of the device. Thus, the FDES106(ED406) Series Electro-Stimulator is substantially equivalent t ES106(ED406) Series Stimulator, which includes models FDES106A(ED406A) and FDES107(ED407 tended use and technological characteristics and the similar technological characteristics as the predicate device T1040 lendertion FLEX Abdominal Training system type 515 (K030708). Moreover, bench testing, safety report and Risk Analysi umentation supplied in this submission demonstrates that the difference in the submitted models could maintain the same ectiveness as that of predicate device. In the other words, those engineering difference do not affect the intended use oredicate device
# ction 07-Page 13
{13}------------------------------------------------
EPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### February 7, 2014
Famidoc Technology Co., Ltd. c/o Reanny Wang Vice-General Manager No. 212 Yilong Road, Hexi Industrial Zone, Jingxia, Changan Town, Dongguan 523853, Guangdong Province CHINA
Re: K130723
> Trade Name: FDES106 (ED406) Series OTC Stimulator Models: Mini TENS&EMS Device Model FDES106 (ED406), Multi-function Mini TENS&EMS Device Model FDES106A (ED406A), Pain Relief Plaster Model FDES105 (ED405), Abdominal Fitness Belt Model FDES107 (ED407) Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II
Product Code: NGX, NUH
Dated: January 7, 2014
Received: January 13, 2014
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{14}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21
CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pemal-S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
### Indications for Use
510(k) Number: K130723
FDES106 Series OTC Stimulator Models: Device Name: Mini TENS&EMS Device Model FDES106 (ED406), Multi-function Mini TENS&EMS Device Model FDES106A (ED406A), Pain Relief Plaster Model FDES105 (ED405), Abdominal Fitness Belt Model FDES107 (ED407),
Indications For Use:
FDES106 (ED406) Mini TENS &EMS Device
For program N1, B and H of TENS mode
To be used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household work activities.
#### For program E1 and E2 of EMS mode
Used to stimulate healthy muscles in order to improve and facilitate muscle performance.
#### FDES106A (ED406A) Multi-function Mini TENS &EMS Device
#### For program N1, B and H of TENS mode
To be used for temporary relief of pain associated with sore and aching muscles in the lower back, abdomen, thigh, and arm due to strain from exercise or normal household work activities.
#### For program E1 and E2 of EMS mode
Used to stimulate healthy muscles in the lower back, abdomen, thigh, and arm in order to improve and facilitate muscle performance.
#### FDES105 (ED405) Pain Relief Plaster
To be used for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household work activities.
Page 1 of 2
{16}------------------------------------------------
#### FDES107 (ED407) Abdominal Fitness Belt
Used to stimulate healthy muscles in order to improve and facilitate muscle performance for abdominal.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR .
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
.
Carlos L.JRenia -S
Page 2 of 2
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.