K965244 · Z Technologies, LLC · GZJ · Aug 27, 1997 · Neurology
Device Facts
Record ID
K965244
Device Name
Z TECH TRANSCUTANEOUS NEURAL STIMULATION
Applicant
Z Technologies, LLC
Product Code
GZJ · Neurology
Decision Date
Aug 27, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended to be used by clinicians in symptomatic relief of chronic intractable pain. Also, for post traumatic and post surgical pain relief.
Device Story
Z Tech Model F7 is a Transcutaneous Electrical Nerve Stimulator (TENS) used by clinicians for pain management. Device generates bi-polar asymmetrical exponentially decaying impulse waveforms (0-60μc) with 1.5ms pulse width and 55Hz maximum frequency. Operates via astable pulse generator; peak voltage, frequency, and pulse width are user-controllable. Output is delivered through isolation transformer to electrodes, eliminating DC components for safety. Powered by internal Ni-Cad batteries; includes recharging circuit that inhibits device use during charging to prevent 110V AC exposure. Housed in Lexan plastic case with manual controls. Clinicians apply electrodes to patient skin; electrical stimulation provides symptomatic relief for chronic intractable, post-traumatic, or post-surgical pain, particularly where other modalities or drug therapies have failed.
Clinical Evidence
Bench testing only. Device performance was validated by recording waveform parameters under various loading conditions (500Ω to 10kΩ) and comparing them to the predicate device and established safety standards (American National Standard for Transcutaneous Electrical Nerve Stimulators). Clinical efficacy is supported by literature references regarding TENS for pain control.
Technological Characteristics
TENS device; bi-polar asymmetrical exponentially decaying impulse waveform; 0-60μc charge; 1.5ms pulse width; 55Hz max frequency. Lexan plastic housing. Isolation transformer output. Internal Ni-Cad battery power with recharging lockout circuit. Complies with American National Standard for Transcutaneous Electrical Nerve Stimulators.
Indications for Use
Indicated for symptomatic relief of chronic intractable pain, post-traumatic pain, and post-surgical pain in patients experiencing acute or chronic pain.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
H-WAVE Model P-4 (TENS DEVICE)
Related Devices
K012643 — APEX MEDICAL LCD TENS-VII · Apex Medical Corp. · Oct 29, 2001
K991397 — TRANSCUTANEOUS ELECTRICAL NERVE STIMULATOR SE-30 TENS · Sein Electronics Co., Ltd. · Feb 11, 2000
K021436 — TTECH MODEL 200E + TENS DEVICE · Theratech, Inc. · Jul 30, 2002
K050435 — E-STIM II · Lhasa Oms, Inc. · Sep 6, 2005
Submission Summary (Full Text)
{0}
K965244
AUG 27
Summary of Safety and Effectiveness
In reference to: 21CFR807.92
Z TECHNOLOGIES, INC.
2615 Woodacres Rd.
Atlanta, Ga. 30345-1644
404-248-0159
Les Horn
May 23, 1997
Z TECH Model F7 (TENS DEVICE)- Z Technologies, Inc.
The Predicate device is the H-WAVE Model P-4 (TENS DEVICE)-
- Electronic Waveform Labs, Inc.
{1}
Safety and effectiveness in accordance with CFR 807.92.
The "Z Tech" TENS Device is technically identical to the "H-Wave" TENS Device, as indicated in the comparison table for the two devices. (SEE EXHIBIT I) The technical characteristics indicated represent the basic output for optimum performance as indicated by "Transcutaneous Neural Stimulation for Relief of Pain"; by Mark Linzer and Doulin M. Long. This article along with the N-S4 article articulates safety standards as well as minimum and maximum levels of output for TENS devices for control of pain. (7μc to 75μc)
The Z Tech Tens Device is designed to produce 0-60μc H-WAVE type pulse which is described as a bi-polar asymmetrical exponentially decaying impulse type waveform. It has a pulse width of approximately 1.5ms and delivers a maximum frequency of 55Hz. These performance characteristics have been shown through clinical experimentation to relieve pain in a significant number of patients in a wide variety of pain states.
Essentially the device originates an astable type pulse of controllable peak voltage, frequency and pulse width, which is amplified and passed to the output electrodes through an isolation transformer, which also eliminates any D.C. component in the output. (Eliminating any D.C. component is a necessary safety feature).
The technological values of charge per phase, charge density, current, voltage, frequency and pulse width, are all controlled and limited to safe levels. These levels are consistent and limited too specifications given and prescribed by clinical testing which is designed to give optimum values for the reduction of pain in human patients.
{2}
The device also employs a recharging type circuit which effectively inhibits any patient use during the recharging of permanently installed ni-cad type batteries. Essentially no 110 Volt A.C. component is ever allowed to be present or available to the output of this TENS DEVICE. (Z TECH) The case which is of a lexan plastic type material. Front and rear controls are clearly marked as to their function for ease of patient use.
The device is intended to be used by clinicians in symptomatic relief of chronic intractable pain. Also for post traumatic and post surgical pain relief.
It is most important to note that a large number of patients receiving benefit from TENS type devices are those incapacitated by chronic pain for whom other treatment modalities have failed.
If care is taken by the placement of electrodes, when this device is in use, then an even greater beneficial result will be forthcoming. The device may be used by all type of patients that experience acute and or chronic pain. (especially in areas where other treatments have failed and where drug type therapy is ill advised)
Waveform parameters were recorded using different loading conditions (500Ω to 10kΩ) for comparison to present and past technology and to the predicate device. These output characteristics and data conform to all safety and effectiveness standards as prescribed by "American National Standard for Transcutaneous Electrical Nerve Stimulators."
The conclusions drawn from clinical and nonclinical tests demonstrate that this TENS type device will effectively relieve patients of chronic and acute pain that otherwise might go untreated. Thus the device may succeed where other types of treatment have been unsuccessful.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
AUG 27 1997
Mr. Les W. Horn
President
Z Technologies, Inc.
2615 Woodacres Road
Atlanta, Georgia 30345-1644
Re: K965244
Trade Name: Z Tech Transcutaneous Neural Stimulation Regulatory Class: II (two)
Product Code: 84GZJ
Dated: August 13, 1997
Received: August 14, 1997
Dear Mr. Horn:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
2 - Mr. Les W. Horn
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Page 1 of 1
510(k) Number (if known): K965244
Device Name: Z Tech Trancutaneous Neural Stimulation
Indications For Use: The device is intended to be used by clinicians in symptomatic relief of chronic intractable pain. Also, for post traumatic and post surgical pain relief.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☐ (Per 21 CFR 801.109)
OR Over-The-Counter Use ☐
Optional Format 1-2-
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.