Reliefband® is a digital therapeutic device indicated for use in the treatment of nausea, retching and vomiting due to motion sickness, morning sickness associated with pregnancy, and chemotherapy. Reliefband® is also indicated as an adjunct to antiemetic therapy in reducing postoperative nausea.
Device Story
Non-invasive wrist-worn nerve stimulation device; provides relief from nausea, retching, and vomiting. Patient-operated; worn on ventral/palmar surface of wrist, two fingers breadth above distal skin crease between palmaris longus and flexor carpi radialis tendons. Device delivers electrical stimulation via two captive electrodes. User interface includes pushbutton for intensity selection and five blinking LEDs to indicate intensity level; sixth LED indicates low battery. Stimulation uses asymmetric biphasic pulses with ramp-up, constant, and ramp-down phases. Output range 10-40 mA across five discrete levels. Device regulates stimulus output relative to battery voltage; shuts off if regulation cannot be maintained. Benefits patient by providing non-pharmacological nausea relief.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological identity to previously cleared predicate devices.
Technological Characteristics
Wrist-worn transcutaneous electrical nerve stimulator. Two captive electrodes. Output: asymmetric biphasic pulses, 10-40 mA (500 ohm load). Power: two 3V CR2025 batteries. User interface: membrane keypad, 5 intensity LEDs, 1 low battery LED. Operating temp: 10°C to 45°C. Regulation: ±5% of nominal output. Dimensions: ~1.5" x 2" x 0.45".
Indications for Use
Indicated for patients experiencing nausea, retching, and vomiting due to motion sickness, pregnancy-related morning sickness, or chemotherapy. Also indicated as an adjunct to antiemetic therapy for postoperative nausea.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
ReliefBand® Device Models RB-DL, RB-EL, and RB-RL (OTC) (K020180)
ReliefBand® Device Models WB-2L, WB-6L, and WB-RL (Rx) (K994387)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary K173233
| Contact: | Barbara J. Whitman<br>Regulatory Affairs Consultant<br><br>ReliefBand Technologies<br>220 Gibraltar Road<br>Suite 270<br>Horsham, PA 19044 USA<br>Phone: 908-763-5225<br>Email: whitmanregulatory@gmail.com<br>Web: www.reliefband.com | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date Prepared: | September 27, 2017 | |
| Trade name: | ReliefBand® Device | |
| Common Name: | Nerve Stimulation Therapy Device | |
| Classification Name: | Transcutaneous electrical nerve stimulator for pain relief<br>(21 CFR 882.5890) | |
| Product Code: | GZJ | |
| Device Classification: | Class II | |
| Manufacturer and<br>Distributor: | ReliefBand Technologies<br>220 Gibraltar Road<br>Suite 270<br>Horsham, PA 19044 USA | |
### Predicate Devices:
| Predicate Device | Manufacturer | 510(k) No. | Date Cleared |
|--------------------------------------------------------------------------------------|------------------------------|------------|----------------|
| (PRIMARY PREDICATE)<br>ReliefBand® Device<br>Models RB-DL, RB-EL,<br>and RB-RL (OTC) | Woodside Biomedical,<br>Inc. | K020180 | March 21, 2002 |
| ReliefBand® Device<br>Models WB-2L, WB-6L,<br>and WB-RL (Rx) | Woodside Biomedical,<br>Inc. | K994387 | March 16, 2000 |
{1}------------------------------------------------
#### Description of the Subject Device
The ReliefBand® 1.5 is a non-invasive device that is indicated for the counter use in the relief of mild to moderate nausea, retching and vomiting associated with chemotherapy, motion sickness, mild to moderate nausea and vomiting associated with pregnancy, and as an adjunct to antiemetics in reducing mild to moderate postoperative nausea. This device was originally marketed under its predicate name of Reliefband RB-EL (K020180) when owned by Woodside Biomedical, Inc. The devices are technologically identical. The same contract manufacturer has made this device for decades.
Exactly like its predicate, the Reliefband 1.5 is contained within a wristband, and provides relief of nausea, retching and vomiting through electrical stimulation of the nerves on the ventral side of the patient's wrist. The device can be worn on the ventral or palmar (i.e., inside) surface of either or both wrists, approximately two fingers breadth above the distal skin crease of the wrist joint, between the tendons of the palmaris longus and flexor carpi radialis muscles.
The ReliefBand® 1.5 has a user display that incorporates five blinking LEDs which are used to identify the intensity level (5 discrete LEDs, one for each intensity level), so that the patient can easily select the desired stimulation. Selection of the intensity level is performed via a pushbutton located on the user display, which controls the peak pulse amplitude of the electrical impulse and thereby determines the intensity of the stimulation. A sixth blinking LED is used to display the low battery indicator.
#### Indications For Use
Reliefband® is a digital therapeutic device indicated for use in the treatment of nausea, retching and vomiting due to motion sickness, morning sickness associated with pregnancy, and chemotherapy. Reliefband® is also indicated as an adjunct to antiemetic therapy in reducing postoperative nausea.
#### Technical Characteristics
A comparison of the technological characteristics of the ReliefBand® 1.5 Device family and the predicate devices has been performed. The results of the comparison demonstrate that the ReliefBand® 1.5 is substantially equivalent to the marketed predicate devices. The differences between the ReliefBand® 1.5 and the predicate models are based solely on marketing changes and do not affect the safety or effectiveness of the device. The Reliefband® 1.5 is technologically identical to its previously cleared predicate device Reliefband® RB-EL. (Please see table below).
{2}------------------------------------------------
# Technical Characteristics - Substantial Equivalency Table
| Performance<br>Characteristic | Rx Predicate Device (K994387) | OTC Predicate Device RB-xL (K020180) | K173233<br>OTC ReliefBand® 1.5 |
|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The ReliefBand NST Device is indicated<br>for use in the treatment<br>of nausea and vomiting (NV) due<br>to chemotherapy, motion sickness, and<br>pregnancy. The ReliefBand NST Device is<br>indicated as an adjunct to antiemetics in<br>reducing postoperative nausea (PON) | The ReliefBand Devices are indicated for<br>over the counter use in the relief of<br>nausea and vomiting (NV) due to motion<br>sickness and for relief of mild to<br>moderate nausea and vomiting<br>associated with pregnancy | Reliefband® is a digital therapeutic<br>device indicated for use in the treatment<br>of nausea, retching and vomiting due to<br>motion sickness, morning sickness<br>associated with pregnancy, and<br>chemotherapy. Reliefband® is also<br>indicated as an adjunct to antiemetic<br>therapy in reducing postoperative<br>nausea. |
| Size | Teardrop shape (approximately 1.5" by 2" by 0.45" thick enclosure) | | |
| Weight | Approximately 1.2 ounce (34 grams) | | |
| Operating<br>Temperature | 10°C to 45°C (50°F to 113°F) (±18% output tolerance at operating extremes) | | |
| User Controls | Single push button (membrane keypad) | | |
| User Displays | Five flashing LED setting indicators (green), 2 second flash period during normal operation | | |
| | One flashing low battery LED indicator (red), 4 second flash period alternating with setting LED when in Low Battery operation | | |
| Operating Modes | | OFF | |
| Performance<br>Characteristic | Rx Predicate Device (K994387) | OTC Predicate Device RB-xL (K020180) | K173233<br>OTC ReliefBand® 1.5 |
| Output Range<br>(Peak pulse current at<br>nominal, 500 ohm<br>load) | Five discrete levels between 10 mA (± 15%) – 40 mA (± 15%). | | |
| Output Waveform<br>Characteristics<br>(At nominal, 500 Ohm<br>load) | Asymmetric biphasic pulses.<br>Ramp up, constant and ramp down phases. | | |
| Maximum Charge<br>Delivered<br>(Setting 5, charge per<br>pulse at nominal, 500<br>ohm load) | Approximately 3.1 microcoulombs (max. 3.56 microcoulombs including 15% output tolerance) | | |
| Maximum Output<br>Voltage<br>(open circuit, no load) | 110 Volts | | |
| Output Channels | Two captive electrodes. | | |
| Regulation of<br>Stimulus Output with<br>Decreasing Battery<br>Voltage | Regulation is ± 5% of nominal output. Device stops stimulation when regulation cannot be maintained (device turns itself off). | | |
| Battery Life:<br>(Setting 3 at nominal) | 150 hours continuous or intermittent use when two 3V CR2025 used.<br>Low battery indicator flashes when approximately 10% of battery life remains. | | |
{3}------------------------------------------------
# Performance Data
The performance data indicate that the Relief anily is substantially equivalent to the predicate devices cleared to market via 510(k) #s: K994387 and K020180.
{4}------------------------------------------------
### Basis for Determination of Substantial Equivalence
Substantial equivalence was determined on the basis that there have been no changes in the mechanism of action, intended use and safety and effectiveness. The technology of the subject device has remained identical to that of its predicates and is therefore substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 30, 2017
ReliefBand Technologies % Barbara Whitman President Whitman Regulatory Consulting 616 Revere Court Flemington, New Jersey 08822
Re: K173233
Trade/Device Name: ReliefBand 1.5 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: September 27, 2017 Received: October 5, 2017
Dear Barbara Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{6}------------------------------------------------
#### Page 2 - Barbara Whitman
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.