HIVOX ELECTRIC STIMULATOR TENS & EMS, HD2

K092448 · Hivox Biotek, Inc. · GZJ · Mar 30, 2010 · Neurology

Device Facts

Record IDK092448
Device NameHIVOX ELECTRIC STIMULATOR TENS & EMS, HD2
ApplicantHivox Biotek, Inc.
Product CodeGZJ · Neurology
Decision DateMar 30, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

For TENS programs: The device is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for (1) Symptomatic relief of chronic (long term) intractable pain, and (2) adjunctive treatment in the management of post-surgical and post-traumatic pain problems. For EMS programs: The device is an electrical powered muscle stimulator indicated for medical purpose to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is indicated for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.

Device Story

HIVOX HD2 is a portable TENS/EMS stimulator; generates small electrical pulses delivered via lead cables to skin-surface electrodes. Operates by passing current through skin to activate underlying nerves (TENS) or induce muscle contraction (EMS). Used by patients or clinicians in clinical/home settings. Device features adjustable frequency (2-120.2 Hz) and fixed pulse width (250 µs). Powered by 3 AAA batteries. Provides symptomatic pain relief and muscle therapy. Healthcare providers use device output to manage pain and physical rehabilitation. Benefits include non-invasive pain management and muscle function maintenance.

Clinical Evidence

No clinical data. Bench testing only, including electrical safety, electromagnetic compatibility, and waveform testing.

Technological Characteristics

Portable TENS/EMS stimulator; DC 4.5V (3x AAA batteries). Constant voltage output. Max output 24.0V/48.0mA (@ 500 Ω). Fixed pulse width 250 µs; frequency 2-120.2 Hz. Plastic housing. Connectivity via lead cables to carbon silicon electrodes (45x45 mm). Standards: IEC/EN 60601-1, IEC/EN 60601-1-2, IEC/EN 60601-2-10.

Indications for Use

Indicated for patients requiring pain relief (chronic intractable, post-surgical, post-traumatic) or muscle stimulation (spasm relaxation, circulation improvement, venous thrombosis prevention, muscle re-education, range of motion maintenance, disuse atrophy prevention).

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the word "HIVOX" in a bold, sans-serif font, followed by the registered trademark symbol. To the right of the word is a stylized graphic that resembles a plus sign or cross, composed of interconnected squares or blocks. The graphic has a three-dimensional appearance due to the shading on some of the blocks. ### HIVOX BIOTEK INC 8/F, No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC Fax: +886 2 8511 2669 Phone: +886 2 8511 2668 # SUMMARY OF SAFETY AND EFFECTIVENESS for TENS & EMS #### 510(k): K092448 DATE OF SUBMISSION i MAR 3 0 2010 | SUBMISSION: | July 27, 2009 | |-------------|----------------------------------------| | SUBMITTER: | HIVOX BIOTEK INC. | | | 8F, No.98, Shinde Road, | | | Sanchong City, Taipei, 24158, TAIWAN | | | TEL: 886-2-85112668 FAX:886-2-85112669 | | | Email: ceirs.jen@msa.hinet.net | #### ESTABLISHMENT REGISTRATION NO: 9611558 OFFICIAL CONTACT: Dr. JEN, KE-MIN NO 58, Fu-Chiun Street, Hsin-Chu City, 30067, TAIWAN TEL: 886-3-5208829 FAX:886-3-5209783 Email: ceirs.jen@msa.hinet.net HIVOX Electric Stimulator TENS & EMS, HD2 TRADE NAME: COMMON/USUAL Transcutaneous Nerve Stimulator & NAME: Electric Muscle Stimulator Transcutaneous Nerve Stimulator CLASSIFICATION NAME: GZJ, Class II, 882.5890 REGULATION NUMBER: ID# Class II 800 59850 SECONDARY PRODUCT CODE: IPF, Class II, 890.5850 PREDICATED DEVICE: MEDIHIGHTEC Medical Co., Ltd. TENS/EMS Combo, EMS, TENS K082514 {1}------------------------------------------------ | HIVOX<br>Image: [logo] | HIVOX BIOTEK INC | | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | | 8/F, No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC<br>Phone: +886 2 8511 2668 Fax: +886 2 8511 2669 | | | Intended Use: | For TENS programs: The device is an electrical nerve stimulator<br>indicated for use for pain relief by applying an electrical current to<br>electrodes on a patient's skin to treat pain. In particular, this device<br>is indicated for use for (1) Symptomatic relief of chronic (long<br>term) intractable pain, and (2) adjunctive treatment in the<br>management of post-surgical and post-traumatic pain problems.<br>For EMS programs: The device is an electrical powered muscle<br>stimulator indicated for medical purpose to repeatedly contract<br>muscles by passing electrical currents through electrodes<br>contacting the affected body area. In particular, this device is<br>indicated for use for (1) Relaxing muscle spasms, (2) Increasing<br>local blood circulation, (3) Immediate post-surgical stimulation of<br>calf muscles to prevent venous thrombosis, (4) Muscle<br>re-education, (5) Maintaining or increasing range of motion, and<br>(6) Preventing or retarding disuse atrophy. | | | Description of Device: | The HIVOX Electric Stimulator TENS & EMS, HD2 generates<br>small pulses of electrical current. Delivered along lead cables to<br>electrodes placed on your skin, these pulses pass through the skin<br>and activated underlying nerves. The relief from chronic and<br>acute pain that the HD2 can provide results from this electrical<br>stimulation. | | | Performance Tests<br>Submitted: | The relevant standards including:<br>1. IEC/EN 60601-1 : Medical electrical equipment Part 1.<br>General requirements for safety, 1996.<br>2. IEC/EN 60601-1-2 : Medical electrical equipment, Part 2.<br>Electromagnetic compatibility - Requirements and tests, 2004.<br>3. IEC/EN 60601-2-10 : Medical electrical equipment Part 2.Safety<br>of Nerve and Muscle Stimulators, 1987. | | | Non-Clinical Tests<br>Submitted: | The HIVOX Electric Stimulator TENS & EMS, HD2 has been<br>tested in accordance with applicable standards for medical device<br>electrical safety, electromagnetic compatibility, and the particular<br>requirements for safety of nerve stimulators.<br>Accessories also meet safety requirements: 510(k) electrodes are<br>specified, and the patient cable utilizes shrouded connectors to<br>meet lead wire safety requirements.<br>System level testing including waveform testing was performed in<br>combination the HIVOX Electric Stimulator TENS & EMS, HD2. | | | Clinical Tests<br>Submitted: | None | | | Conclusion: | As the product description and tests as above, the new device:<br>HIVOX Electric Stimulator TENS & EMS, HD2 is as safe and<br>effective as, and the function in a manner equivalent to the<br>predicate device: MEDIHIGHTEC TENS/EMS Combo, EMS,<br>TENS, K082514. | | ﺗ . . devices in this aspect. {2}------------------------------------------------ # HIVOXiff HIVOX BIOTEK INC 8/F . No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC Fax: +886 2 8511 2669 Phone: +886 2 8511 2668 # Comparison for Predicate Device & Subject Device | Comparison Areas | NEW DEVICE | PREDICATE DEVICE | |-----------------------------------------------------|-------------------------------|--------------------------------------------| | 510(k) number | K092448 | K082514 | | Manufacturer | HIVOX Biotek Inc | Medihightec Medical Co. Ltd. | | Device name | HD2<br>TENS & EMS | MH8000/6000<br>TENS/EMS Combo | | Power source | DC 4.5V<br>(AAA Battery x 3 ) | One DC 9V battery | | Channels<br>Synchronous<br>Reciprocal | SAME | 2<br>YES<br>YES | | Computerized | SAME | No | | Software provided | SAME | No | | Constant current | No | YES | | Constant voltage | YES | No | | Max output voltage<br>(Vp ±20%) | 24.0V ±20% @ 500 Ω load | 20.4V±20% @500 Ω load<br>(40.8 Vp-p) | | Max output current<br>(mA ±20%) | 48.0mA ±20% ( 500 Ω load ) | 40.8 mA±20% ( 500Ω load)<br>(81.6 mA, p-p) | | Pulse Width ( µs ) | 250 µs (fixed) | 50~300 µs | | Maximum Frequency | 2~120.2 Hz (adjustable) | 2~150 Hz (adjustable) | | Max phase charge | □12.0 µC @ 500 ohms load | 24.0 µC @ 500 ohm | | Maximum Current<br>Density | 0.59 mA/cm² @500Ω | 1.06 mA/cm² @500Ω | | Maximum Power<br>Density | 0.014 W/cm² @500Ω | 2.4 m W/cm² @500Ω | | Comparison Areas | NEW DEVICE | PREDICATE DEVICE | | Channel isolation | SAME | YES | | Line current isolation | SAME | N/A | | Patient override<br>control method | SAME | YES | | Burst mode | SAME | YES | | Housing<br>material<br>and construction: | SAME | Plastic material | | Standards: | SAME | IEC 60601-1:1988<br>EN 60601-1-2:1990 | | Timer setting<br>Automatic shut off | SAME<br>2 minutes | YES (05~90Min)<br>30 Sec | | Automatic overload<br>trip current/voltage<br>level | N/A | YES | | Automatic no load trip | N/A | YES | | Indicator display<br>Unit functioning: | Power ON | Power ON, Vpp | | Low battery | Yes | NO | | indicator:<br>Voltage level: | NO | YES | | Weight ( unit ) | 105 g w/o Battery | 162 grams w/ battery | | Dimensions ( unit ) | 120 x 52 x 17 mm | 136 x 70 x 27 mm | | Size of electrode | Carbon silicon of 45x45 mm | Irregular: 10x15cm | {3}------------------------------------------------ K092448 Image /page/3/Picture/1 description: The image shows the word "HIVOX" in all caps, followed by a small superscript symbol and a stylized plus sign. The plus sign is made up of several small squares connected in a winding pattern. The text and symbol are all in black and are set against a white background. # HIVOX BIOTEK INC 8/F, No.98 Shingde Road Sanchong City Taipei 241 Taiwan ROC Phone: +886 2 8511 2668 Fax: +886 2 8511 2669 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002 Hivox Biotek Inc. c/o Dr. Ke-Min Jen 8F, No. 98, Shinde Rd., Sanchong City Taipei Hsien China (Taiwan) 24158 MAR 3 0 2010 Re: K092448 Trade/Device Name: Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ, IPF Dated: March 12, 2010 Received: March 19, 2010 Dear Dr. Jen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {5}------------------------------------------------ Page 2 - Dr. Ke-Min Jen forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Kevin Alexander Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use #### 510(k) Number: K092448 Device Name: #### HIVOX Electric Stimulator TENS & EMS, HD2 #### Indications for Use : The TENS programs of HIVOX Electric Stimulator TENS & EMS, HD2: The device is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for (1) Symptomatic relief of chronic (long term) intractable pain, and (2) adjunctive treatment in the management of post-surgical and post-traumatic pain problems. The EMS programs of HIVOX Electric Stimulator TENS & EMS, HD2: The device is an electrical powered muscle stimulator indicated for medical purpose to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is indicated for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy. Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Page 1 of 1 510(k) Number K092448
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...