TENS/IF 14

K071869 · Emsi · GZJ · Mar 7, 2008 · Neurology

Device Facts

Record IDK071869
Device NameTENS/IF 14
ApplicantEmsi
Product CodeGZJ · Neurology
Decision DateMar 7, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TENS/IF 14 is intended for use by or on the order of a physician for the symptomatic relief of chronic intractable pain and as an adjunctive treatment for the management of post-traumatic or post-surgical pain.

Device Story

TENS/IF 14 is a combination Transcutaneous Electrical Nerve Stimulator (TENS) and Interferential (IF) therapy device. It delivers electrical current to patient skin via electrodes to provide pain relief. Device features 2 output channels supporting 2 or 4 electrodes; output and waveforms are adjustable by the user. Powered by 4 AA batteries or AC adapter (6 VDC). Includes a patient compliance timer capable of storing 60 operation records up to 999 hours. Used in clinical or home settings under physician order. Healthcare providers use the device to manage chronic, post-traumatic, or post-surgical pain, potentially reducing patient pain levels through nerve stimulation.

Clinical Evidence

Bench testing only. The device underwent performance testing in accordance with IEC 60601-1-2 for electromagnetic compatibility and safety. Results confirmed compliance with electrical leakage current, electrode/lead wire safety, and output current/power density requirements.

Technological Characteristics

Combination TENS/IF stimulator; 2 output channels; 4 AA battery or AC adapter (6 VDC) power; adjustable waveforms/output; patient compliance timer (60 records, 999 hours). Safety testing per IEC 60601-1-2.

Indications for Use

Indicated for symptomatic relief of chronic intractable pain and as adjunctive treatment for post-traumatic or post-surgical pain in patients requiring nerve stimulation therapy.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K071869 510(k) Premarket Notification #### MAR - 7 2008 Section 5: 510(k) Summary The following information is provided as required by 21 CFR § 807.87 for the Mini Combo 510(k) premarket notification for and in accordance with FDA's "Guidance Document for Powered Muscle Stimulator 510(k)s", June 9, 1999 and "Guidance for Tens 510(k) Content", August 1994. ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based. The safety and effectiveness of the TENS/IF 14 is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate devices . | Sponsor: | EMSI<br>3504 Cragmont Dr. Suite#100<br>Tampa, Florida 33619<br>Ph: 813-471-0129<br>Fax: 813-471-0130<br>Registration Number: 3003573572 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Cherita James<br>M Squared Associates, Inc<br>719 A Street, NE<br>Washington, DC 20002<br>Ph: 202-546-1262 Ext 257<br>Fax: 202-546-3848 | | Date of submission: | July 3, 2007 | | Proprietary Name: | TENS/IF 14 | | Common Name: | Transcutaneous Nerve Stimulator for pain relief, Interferential<br>Current Therapy | | Classification Status: | 21 CFR 882.5890 | | Class: | II | | Product Code: | GZJ, LIH | {1}------------------------------------------------ Panel: Neurology K060246 INFREX by Johari Digital Healthcare, K021755 TENS Predicate Device: TS1211 by Apex Medical, K952683 IF-4000 by Apex Medical Device Description: The TENS/IF 14 is a combination TENS and Interferential device which delivers nerve stimulation by applying an electrical current to electrodes, which are attached on the patient's skin. The output and waveform is adjustable according to the intended treatment of patient. The stimulator has 2 output channels, accessed through jacks at the top of the housing, so that it may stimulate either 2 or 4 patient electrodes. The device may be powered by either 4 AA batteries or by an AC adapter, that when plugged into a wall outlet provides 6 VDC to the unit. A patient compliance timer can memorize 60 sets of operation records; the total record time is 999 hours. Intended Use: The TENS/IF 14 is intended for use by or on the order of a physician for the symptomatic relief of chronic intractable pain and as an adjunctive treatment for the management of post-traumatic or post-surgical pain. Discussion of performance testing: Testing performed in accordance with the accepted FDA requirements of IEC 60601-1-2, Medical Electrical Equipment, Part 1: General Requirements for Safety. Collateral Standard: Electromagnetic Compatibility-Requirements and Test, found the TENS/IF 14 passed all of the applicable tests. Based on the output measurements, calculations, and safety testing/inspection; the TENS/IF 14 meets standard requirements with respect to electrical leakage current, electrode and lead wire safety, as well as output current and power density. #### Technological Characteristics and Substantial Equivalence Output specifications, device design, waveforms and programmability demonstrated the TENS/IF 14 to be substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with its wings spread, overlaid with three flowing ribbons. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR - 7 2008 EMSI % M Squared Associates, Inc. Ms. Cherita James 719 A Street, NE Washington, DC 20002 Re: K071869 > Trade/Device Name: TENS/IF 14 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: GZJ, LIH Dated: January 22, 2008 Received: January 24, 2008 Dear Ms. James: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Cherita James This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Section 4: Indications for Use Statement To be assigned 510(k) Number: TENS/IF 14 Device Name: Indications for Use: Symptomatic relief of chronic intractable pain Adjunctive treatment for the management of post-traumatic or post-surgical pain AND/OR Over-The-Counter Use Prescription Use (21 CFR 801 Subpart D) (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nishant fn mxm Division of General, Restorative, and Neurological Devices 510(k) Number K071869 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...