ECHELON MICRO CATHETER, MODELS 105-5092-150SA, 145-5092-150SA AND 190-5092-150SA AND SYRINGE ADAPTER ACCESSORY

K090870 · Micro Therapeutics, Inc. · KRA · May 1, 2009 · Cardiovascular

Device Facts

Record IDK090870
Device NameECHELON MICRO CATHETER, MODELS 105-5092-150SA, 145-5092-150SA AND 190-5092-150SA AND SYRINGE ADAPTER ACCESSORY
ApplicantMicro Therapeutics, Inc.
Product CodeKRA · Cardiovascular
Decision DateMay 1, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1210
Device ClassClass 2

Intended Use

The Echelon Micro Catheter is intended to access the peripheral and neuro vasculature for the controlled selective infusion of physician-specified therapeutic agents such as embolization materials and or diagnostic materials such as contrast media.

Device Story

End-hole, single-lumen microcatheter designed for navigation through peripheral and neuro vasculature over a steerable guidewire; features semi-rigid proximal shaft transitioning to flexible distal shaft; includes dual radiopaque markers for fluoroscopic visualization and lubricious outer coating. Packaged with a syringe adapter accessory to reduce dead space within the luer hub, minimizing potential mixing of Onyx and DMSO during connection and injection. Used by physicians in clinical settings for selective infusion of therapeutic or diagnostic agents. Device facilitates precise delivery of embolization materials, potentially benefiting patients requiring targeted vascular treatment.

Clinical Evidence

Bench testing only. Verification studies included dimensional luer fitting integrity, Onyx infusion pressure, catheter dead space volume, radiopacity of Onyx injection, and static burst pressure.

Technological Characteristics

Single-lumen, end-hole catheter with semi-rigid to flexible shaft transition. Features dual radiopaque markers and lubricious coating. Includes a syringe adapter accessory for luer hub dead space reduction. Mechanical device; no energy source or software.

Indications for Use

Indicated for accessing peripheral and neuro vasculature for controlled selective infusion of therapeutic agents (e.g., embolization materials) or diagnostic materials (e.g., contrast media).

Regulatory Classification

Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Special 510(k) Premarket Notification K090870 # 510(k) Summarv #### Applicant - Manufacturer Name and Address Micro Therapeutics dba ev3 Neurovascular 9775 Toledo Way Irvine. CA 92618 Establishment Registration No. 2029214 MAY - 1 2009 March20, 2009 Date #### Contact Information Tom Daughters Director, Regulatory Affairs #### Proprietary Device Model & Trade Name Model: 105-5092-150SA, Echelon Micro Catheter (.014) Model: 145-5092-150SA, Echelon Micro Catheter (.014 Pre-Shape) Model: 190-5092-150SA, Echelon Micro Catheter (.014 Pre-Shape) Model: 103-(5095) 1207 Syringe Adapter Accessory #### Device Classification & Common Name Classification Code: KRA, 21CFR Part 870.1210 Common Name: Catheter, Continuous Flush Classification: Class II #### Predicate Devices - 510(k) References K030688 Model: Echelon Micro Catheter (.014) K031992 Model: Echelon Micro Catheter (.010) K042187 Models: Echelon Micro Catheter (.010; .014) - Pre-Shaped Tips K051990 Models: Echelon Catheters (.010, .014) - Revised Contraindication #### Description of the Device Subject to Premarket Notification The MTI Echelon™ Micro Catheter is an end-hole, single-lumen catheter designed to be introduced over a steerable guidewire into the vasculature. The catheter has a semi-rigid proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated to increase lubricity. The Echelon Micro Catheter is packaged with an ev3 Syringe Adapter. This device, attached to an ev3 1ml syringe filled with Onyx®, will reduce the dead space within the micro catheter luer hub. Reducing the dead space within the hub {1}------------------------------------------------ is intended to minimize the potential mixing of Onyx® and DMSO in the hub of the catheter during connection and injection. #### Indications for Use The Echelon Micro Catheter is intended to access the peripheral and neuro vasculature for the controlled selective infusion of physician-specified therapeutic agents such as embolization materials and or diagnostic materials such as contrast media. #### Performance Data Because the Echelon catheter itself has not changed, the scope of design verification focused on performance and compatibility of the Syringe Adapter accessory device. Verification studies included: Dimensional Luer Fitting Integrity Onyx Infusion Pressure Catheter Dead Space Volume Radiopacity of Onyx Injection Static Burst Pressure #### Substantial Equivalence The risk assessment for the Echelon catheter and the Syringe Adapter accessory was reviewed to identify any new or unique risks. No new risks were identified. The indications for use, the scientific technology and the performance data demonstrate the Echelon Microcatheter with the Syringe Adapter is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. In the center of the seal is an abstract image of an eagle. MAY - 1 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Micro Therapeutics dba ev3 Neurovascular c/o Mr. Tom Daughters Director, Regulatory Affairs 9775 Toledo Way Irvine. CA 92618 Re: K090870 Echelon Micro Catheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: March 20, 2009 Received: March 30, 2009 #### Dear Mr. Daughters: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ### Page 2 - Mr. Tom Daughters Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Statement of Indications for Use ## Indications for Use 510(k) Number (if known): _ K090870 Device Name: Echelon Micro Catheter Indications for Use: The Echelon Micro Catheter is intended to access the peripheral and neuro vasculature for the controlled selective infusion of physician-specified therapeutic agents such as embolization materials and or diagnostic materials such as contrast media. Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K090870 Page Page 10 of 58 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...