K230636 · Merit Medical System, Inc. · KRA · Mar 30, 2023 · Cardiovascular
Device Facts
Record ID
K230636
Device Name
Maestro Microcatheter
Applicant
Merit Medical System, Inc.
Product Code
KRA · Cardiovascular
Decision Date
Mar 30, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1210
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the microcatheter can be used for the controlled and selective infusion of diagnostic, embolic or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.
Device Story
Maestro Microcatheter is a sterile, single-use intravascular catheter; used by physicians in clinical settings for subselective vessel access. Device features include hydrophilic coating on distal 80cm, radiopaque distal tip marker for fluoroscopic visualization, and molded winged hub. Available in lengths 110cm-175cm with straight, 45-degree, or swan neck tip configurations. Physician navigates catheter through vasculature to target site; performs controlled infusion of diagnostic, embolic, or therapeutic agents. Radiopaque marker allows real-time visualization under fluoroscopy to ensure accurate placement. Facilitates targeted delivery of materials, potentially improving therapeutic outcomes while minimizing systemic exposure.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included effective length analysis, proximal shaft/hub dimensions, coating length, leakage (liquid/air), peak tensile force, torque strength, hoop removal, guidewire compatibility, pushability, trackability, torsion, and coil delivery. Results demonstrated that the subject device met all predetermined acceptance criteria.
Technological Characteristics
Intravascular catheter; hydrophilic coating on distal 80cm; radiopaque distal tip marker. Lengths: 110cm, 130cm, 150cm, 165cm, 175cm. Configurations: straight, 45-degree, swan neck. Materials meet ISO 10993 biocompatibility requirements. Complies with ISO 10555-1:2013 for general requirements. Sterile, single-use.
Indications for Use
Indicated for general intravascular use in peripheral and coronary vasculature for controlled and selective infusion of diagnostic, embolic, or therapeutic materials. Contraindicated for use in cerebral vessels.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
K172081 — Maestro Microcatheter · Merit Medical Systems, Inc. · Aug 4, 2017
K173548 — Merit Pursue Microcatheter · Merit Medical Systems, Inc. · Mar 30, 2018
K173430 — SeQure® NF and SeQure® Microcatheters · Accurate Medical Therapeutics, Ltd. · Jan 26, 2018
K050599 — TRACKER EXCEL -14 PRE-SHAPED MICROCATHETER · Boston Scientific Corp · Jun 3, 2005
K083922 — MICRUS MICROCATHETER, COURIER FLEX, MODEL MSF 170000-00 · Micrus Endovascular Corporation · May 6, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
March 30, 2023
Merit Medical System, Inc. Garry Courtney Principal Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095
### Re: K230636
Trade/Device Name: Maestro Microcatheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: February 6, 2023 Received: March 7, 2023
## Dear Garry Courtney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/5 description: The image shows a digital signature. The signature includes the name "Lydia S. Glaw -S". The date of the signature is March 30, 2023, and the time is 15:22:51 -04'00'.
Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230636
Device Name Maestro Microcatheter
#### Indications for Use (Describe)
The Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the microcatheter can be used for the controlled and selective infusion of diagnostic, embolic or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K230636
# 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4583<br>Garry A. Courtney<br>03/03/2023<br>1721504 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Maestro Microcatheter<br>Microcatheter<br>Continuous Flush Catheter<br>2<br>KRA<br>870.1210<br>Cardiovascular |
| Predicate Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Maestro Microcatheter<br>Continuous Flush Catheter<br>K172081<br>Merit Medical Systems, Inc.<br>This predicate has not been subject to a design-related recall |
| Reference Device | | Merit Medical Maestro Microcatheter – K082613 |
{4}------------------------------------------------
| Device<br>Description | The Maestro Microcatheter is available in working length sizes 110cm,<br>130cm, 150cm, 165cm and 175cm lengths. The distal tip of the<br>microcatheter is offered in straight or pre-shaped 45 degree and swan<br>neck configurations. The proximal end of the catheter consists of a<br>molded winged hub with a tapered strain relief. The outer surface of<br>the distal 80cm of the microcatheter shaft is coated with a hydrophilic<br>coating designed to facilitate the introduction of the catheter into the<br>vasculature. The microcatheter incorporates a radiopaque marker at<br>the distal tip to facilitate fluoroscopic visualization.<br>The Maestro Microcatheter is offered with two 3ml syringes. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | There is no change in the Indications for Use Statement from the<br>predicate to the subject device.<br><br>The Microcatheter is intended for general intravascular use, including<br>peripheral and coronary vasculature. Once the subselective region has<br>been accessed, the microcatheter can be used for the controlled and<br>selective infusion of diagnostic, embolic or therapeutic materials into<br>vessels. The catheter should not be used in the cerebral vessels. |
{5}------------------------------------------------
The design and technological characteristics of the subject "longer length" Maestro Microcatheters are substantially equivalent to those of the predicate Maestro Microcatheter. The subject device has the same basic design as the predicate device. The main difference between the subject and the predicate devices is in the working length of the microcatheter – where longer length catheters (165cm and 175cm) are made available to address clinical needs. The comparison between the subject and the predicate devices is based on the following:
- Same intended use ●
- Same indications for use
- Similar material types that meet ISO 10993 biocompatibility requirements
- Same design ●
- Same sterilization methods ●
- Same fundamental technology/principle of operation
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. Performance testing of the subject Maestro Microcatheters was conducted based on the risk analysis and based on the requirements of the following international standard:
- ISO 10555-1:2013, Intravascular Catheters Sterile and single-● use catheters - Part 1: General requirements
- ISO 10993-1:2009, Biological Evaluation of Medical Devices . Part 1: Evaluation and Testing within a risk management process, and FDA quidance Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices, May 1. 1995
Comparison to Predicate Device
{6}------------------------------------------------
The following tests were performed to demonstrate there were no unacceptable risks associated with the changes made to the device:
### Performance Testing-Bench
- Effective Length Analysis •
- Soft Flexible Transition Length .
- Proximal Shaft OD .
- Proximal Hub ID .
- . Coating Length
- Freedom from Liquid Leakage (Pressure Test) .
- Freedom from Air Leakage .
- Peak Tensile Force .
-
| Safety &<br>Performance<br>Tests cont. | Design Validation | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | • Torque Strength | |
| | • Hoop Removal | |
| | • Soft Distal Section | |
| | • Distal Tip | |
| | • Straightener | |
| | • Guidewire Compatibility | |
| | • Pushability | |
| | • Catheter Compatibility | |
| | • Trackability | |
| | • Torsion | |
| | • Reposition Guidewire | |
| | • Negative Pressure | |
| | • Delivery of Coil | |
| | The results of the testing demonstrated that the subject "longer length" Maestro Microcatheters met the predetermined acceptance criteria applicable to the safety and efficacy of the device. | |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, safety and performance testing, the subject "longer length" Maestro Microcatheters meet the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device, the Maestro Microcatheter, K172081, manufactured by Merit Medical Systems, Inc.. | |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.