K961668 · Cook, Inc. · KRA · Nov 4, 1996 · Cardiovascular
Device Facts
Record ID
K961668
Device Name
MICROFERRET CATHETER
Applicant
Cook, Inc.
Product Code
KRA · Cardiovascular
Decision Date
Nov 4, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1210
Device Class
Class 2
Intended Use
The Microferret™ catheter is used for percutaneous access to small vessel, distal anatomy.
Device Story
Single-lumen catheter for percutaneous access to small vessel, distal anatomy; graduated stiffness shaft (stiff proximal, soft distal) constructed of polyethylene; proximal hub conforms to ANSI Luer lock standards for syringe compatibility. Used by clinicians in vascular procedures to navigate distal anatomy. Device facilitates infusion or access; performance verified via design validation studies. Marketed internationally since 1992 with established safety profile.
Clinical Evidence
No clinical trials presented. Evidence based on worldwide marketing history since 1992, design validation studies, and established biocompatibility of materials. No adverse health effects reported in marketing history.
Technological Characteristics
Single-lumen catheter; polyethylene construction with graduated flexibility; proximal hub compliant with ANSI Luer lock standards. Sterilization via standard methods used for existing COOK catheters.
Indications for Use
Indicated for percutaneous access to small vessel, distal anatomy in patients requiring vascular catheterization.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
K250972 — Primum Hydrophilic Guiding Catheter · Pendracare · Jun 29, 2025
K954457 — MEDI-TECH EDDY INTRAVASCULAR INFUSION CATHETER · Medi-Tech, Inc. · May 15, 1996
K963988 — INTRAVASCULAR INFUSION DEVICE · Boston Scientific Corp · Dec 12, 1996
K233205 — Distal Access Catheter · Shanghai Heartcare Medical Technology Co., Ltd. · Dec 22, 2023
Submission Summary (Full Text)
{0}
K96/1668
510(k) Premarket Notification
Microferret™ Catheter
COOK INCORPORATED
NOV A 1996
## J. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
**Submitted By:**
April Lavender, RAC
Vice President Regulatory Affairs
COOK INCORPORATED
925 South Curry Pike
P.O. Box 489
Bloomington, IN 47402
24 April 1996
**Device:**
Trade Name: Microferret™ Catheter
Proposed Classification Name: Class II 74 DQY
## Predicate Devices:
The Microferret™ catheter is substantially equivalent to other devices intended for percutaneous vascular use in terms of indications for use, design, construction and materials equivalence. Specifically, this device is similar to numerous pre-Amendment COOK catheters, Tracker™ catheters (K#853997, K#862117, K#874751), manufactured by Target Therapeutics®, San Jose, California, Evolution™ Microcatheters (K#953114) manufactured by Boston Scientific, Watertown, Massachusetts, Venture™ Infusion Catheters (K#931335) manufactured by Scimed Life Systems and distributed by Boston Scientific, Watertown, Massachusetts and others.
## Device Description:
The Microferret™ catheter is used for percutaneous access to small vessel, distal anatomy. Its single lumen construction provides graduated stiffness, from proximal being stiff to distal being very soft. The material used to construct this device is polyethylene with a graduated transition of flexibility along the shaft. The proximal hub of the catheter is designed to conform with Luer lock standards (ANSI) which are compatible with commercially available syringes. These materials are widely used in catheter device manufacturing and their biocompatibility has been verified. In addition, design validation studies have been performed to assure the device can be expected to perform its intended function when used according to the recommendations in the suggested instructions for use provided with the device.
## Summary of Clinical Use
This device has been marketed in Europe since 1992 by a European manufacturing firm owned by COOK Group Incorporated. There has been good clinical success with the device in its worldwide use to date. The firm has identified the primary risks of devices such as the Microferret™ catheter and have included these potential risks in the device labeling. These risks are not specific to the Microferret™ catheter, but apply to vascular catheters in general. There have been no adverse health effects associated with the use of this catheter during its marketing history.
## Substantial Equivalence
This device will be manufactured according to specified process controls and a Quality Assurance Program. This device will undergo packaging similar to the devices currently marketed and distributed by COOK INCORPORATED. This device will undergo sterilization similar to the devices currently marketed and distributed. Being similar with respect to indications for use, materials and physical construction to predicate devices, this device meets the requirements for section 510(k) substantial equivalence.
12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.