EME SELF MONITORING BLOOD GLUCOSE SYSTEM, EME PRO SELF MONITORING BLOOD GLUCOSE SYSTEM
K112901 · Eps Bio Technology Corp. · LFR · Dec 23, 2011 · Clinical Chemistry
Device Facts
Record ID
K112901
Device Name
EME SELF MONITORING BLOOD GLUCOSE SYSTEM, EME PRO SELF MONITORING BLOOD GLUCOSE SYSTEM
Applicant
Eps Bio Technology Corp.
Product Code
LFR · Clinical Chemistry
Decision Date
Dec 23, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The EME Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in fresh capillary whole blood from fingertip, palm, or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by a single patient with diabetes and should not be shared, as an aid to monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. Alternative site testing can be only used during steady-state blood glucose conditions. The system consists of the EME meter and the EME test strips. The EME meter only is used with the EME test strips to quantitatively measure glucose in fresh capillary whole blood from fingertip, palm, or forearm. The EME Glucose Control Solution For use with EME Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results. The EME Pro Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in venous whole blood or fresh capillary whole blood from fingertip. Testing is done outside the body (In Vitro diagnostic use). It is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. The system is only used with single-use lancing devices. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. The system consists of the EME Pro meter and the EME Pro test strips. The EME Pro meter only is used with the EME Pro test strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from fingertip. The EME Glucose Control Solution For use with the EME Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.
Device Story
The EME and EME Pro systems are blood glucose monitoring devices. They utilize test strips to measure glucose concentrations in fresh capillary (fingertip, palm, forearm) or venous whole blood. The system consists of a meter and test strips, supported by a control solution for accuracy verification. The EME system is designed for single-patient home use, while the EME Pro is intended for multiple-patient use in professional clinical settings. Users apply a blood sample to the test strip, which the meter analyzes to provide a quantitative glucose reading. This output assists healthcare providers and patients in monitoring diabetes control. The devices are not intended for neonatal use or diabetes diagnosis.
Clinical Evidence
No clinical data. Bench testing only. Disinfection efficacy studies demonstrated complete inactivation of Hepatitis B Virus (HBV) using PDI Super SANI-CLOTH Germicidal wipes. Robustness testing confirmed no performance degradation after 20,000 disinfection cycles for the meter and 14,600 cycles for the lancing device.
Technological Characteristics
Glucose test system (21 CFR 862.1345). Consists of a meter, test strips, and control solution. Measures glucose in capillary or venous whole blood. Intended for in vitro diagnostic use.
Indications for Use
Indicated for single patient home use (EME Self Monitoring Blood Glucose System) and multiple patient use in a professional healthcare setting (EME Pro Self Monitoring Blood Glucose System) for blood glucose monitoring.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K111728 — EMV3 SELF MONITORING BLOOD GLUCOSE SYSTEM, EMV3 PRO SELF MONITORING BLOOD GLUCOSE SYSTEM · Eps Bio Technology Corp. · Sep 16, 2011
K112272 — EM44 SELF MONITORING BLOOD GLUCOSE SYSTEM, EM44 PRO SELF MONITORING BLOOD GLUCOSE SYSTEM · Eps Bio Technology Corp. · Sep 7, 2011
K161299 — BGM014 Blood Glucose Monitoring System, BGM014 Pro Blood Glucose Monitoring System · Apex BioTechnology Corp. · Dec 23, 2016
K092887 — EASYPLUS MINI MASTERDRIVER SELF MONITORING BLOOD GLUCOSE SYSTEM, MODEL EPS09009 · Eps Bio Technology Corp. · Oct 21, 2009
K113243 — EM50 SELF MONITORING BLOOD GLUCOSE SYSTEM · Eps Bio Technology Corp. · Mar 28, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure or a caduceus, which is a symbol often associated with healthcare.
10903 New Hampshire Avenue Silver Spring, MD 20993
EPS Bio Technology Corp. c/o Cynthia Hung No. 8, R&D RD. III Hsinchu Science Park Hsinchu City, 30077 Taiwan
DEC 2 3 2011
Re: k112901
Trade/Device Name: EME Self Monitoring Blood Glucose System, EME Pro Self Monitoring Blood Glucose System
Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: LFR, NBW Dated: November 29, 2011 Received: December 1, 2011
Dear Ms. Hung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{1}------------------------------------------------
## Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known):
### Device Name: EME Self Monitoring Blood Glucose system
Indications for Use:
The EME Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in fresh capillary whole blood from fingertip, palm, or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by a single patient with diabetes and should not be shared, as an aid to monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. Alternative site testing can be only used during steady-state blood glucose conditions.
The system consists of the EME meter and the EME test strips. The EME meter only is used with the EME test strips to quantitatively measure glucose in fresh capillary whole blood from fingertip, palm, or forearm.
## The EME Glucose Control Solution
For use with EME Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
signature
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k 112901
Page 1 of 2
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: EME Pro Self Monitoring Blood Glucose system
Indications for Use:
The EME Pro Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in venous whole blood or fresh capillary whole blood from fingertip. Testing is done outside the body (In Vitro diagnostic use). It is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. The system is only used with single-use lancing devices. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus.
The system consists of the EME Pro meter and the EME Pro test strips. The EME Pro meter only is used with the EME Pro test strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from fingertip.
#### The EME Glucose Control Solution
For use with the EME Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Signature
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K112901
Page Dof 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.