PRECICHEK NS-101 POCT PROFESSIONAL BLOOD GLUCOSE MONITORING SYSTEM
K113314 · Hmd Biomedical, Inc. · LFR · Nov 9, 2012 · Clinical Chemistry
Device Facts
Record ID
K113314
Device Name
PRECICHEK NS-101 POCT PROFESSIONAL BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Hmd Biomedical, Inc.
Product Code
LFR · Clinical Chemistry
Decision Date
Nov 9, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by health care professionals in a clinical setting: The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices. The PRECICHEK KP Blood Glucose Test Strip is for use with the PRECICHEK NS-101 POCT Professional Blood Glucose Meter to quantitatively measure glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples. The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to assess the performance of the PRECICHEK NS-101 POCT Professional Blood Glucose monitoring system and the PRECICHECK KP Blood Glucose Test strips to check that the meter and test strips are working together properly and that the test is performing correctly.
Device Story
System measures glucose in capillary, venous, or arterial whole blood; utilizes electrochemical biosensor technology (GDH-FAD) and capillary action; requires 0.5 μL sample volume. Meter processes electrical current generated by glucose-reagent reaction; displays quantitative glucose concentration. Used in clinical settings by healthcare professionals for multi-patient monitoring; requires disinfection between uses. Includes meter, test strips, control solutions, and check strip. Meter features barcode scanner for lot identification; calibration via code entry. Provides rapid results (5 seconds) to aid diabetes management; supports clinical decision-making regarding glycemic control.
Clinical Evidence
Bench testing only. Compliance with ISO 15197:2003, IEC 60601-1, IEC 60601-1-2, and CLSI standards (EP05-A2, EP06-A, EP07-A2, EP09-A2) confirmed performance criteria met.
Indicated for quantitative glucose measurement in fresh capillary (fingertip), venous, and arterial whole blood by healthcare professionals in clinical settings to monitor diabetes control. Contraindicated for neonatal use, diabetes diagnosis/screening, and patients who are dehydrated, hypotensive, in shock, critically ill, or in a hyperosmolar state.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
SureStepFlexx Professional Blood Glucose Management System (K023194)
Related Devices
K123010 — ON CALL VIVID BLOOD GLUCOSE MONITORING SYSTEM · ACON Laboratories, Inc. · Dec 14, 2012
K123007 — VERI-Q MGD-1001 BLOOD GLUCOSE MONITORING SYSTEM, VERI-Q PLUS MGD-1001 BLOOD GLUCOSE MONITORING SYSTEM · Mico Biomed Co., Ltd. · Jun 19, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
K113314
.
# 510(k) Summary
ﺍﻟﺘﻲ ﺗﺘﺮﺗﺒﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﺍﺻﻞ
ﻬﺎ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ
| | 1. Submitter Information | | NOV | 9 2012 | |
|--|--------------------------|------------------------------------------------|-----|--------|--|
| | Company name: | HMD BioMedical Inc. | | | |
| | Contact person: | Jessica Tung | | | |
| | Address: | No. 181, Minsheng St., Xinpu Township, Hsinchu | | | |
| | | County, Taiwan | | | |
| | Phone: | 886-3-6017000 | | | |
| | FAX: | 886-3-5885500 | | | |
| | E-mail: | jessica-tung@hmdbio.com | | | |
| | Date Prepared: | Apr. 30, 2012 | | | |
- 2. Name of Device
:
:
.
Trade/Proprietary Name:
| | PRECICHEK NS-101 POCT Professional Blood |
|-----------------------|------------------------------------------|
| | Glucose Monitoring System |
| Common name: | Blood Glucose Test System |
| Classification name: | Glucose Test System |
| Classification Panel: | Clinical Chemistry (75) |
| Regulation no.: | 862.1345 (Class II), 862.1660 (Class I) |
| Product code: | LFR, JJX |
| Panel: | Clinical Chemistry (75) |
- 3. Predicate Device
| Trade/Proprietary name: | SureStepFlexx Professional Blood Glucose<br>Management System |
|-------------------------|---------------------------------------------------------------|
| Common name: | Blood Glucose Test System |
| Submitter: | Life Scan, Inc. |
| 510(k) no.: | K023194 |
| Product code: | NBW, CGA |
:
{1}------------------------------------------------
Common name: Glucose Control Solution Submitter: HMD Biomedical, Inc. K032985 510(k) no.: אננ Product code:
- 4. Device Description
PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System consists of:
- (1) Glucose Meter
- (2) Glucose Test Strips
- (3) Two levels of glucose control solutions (Level I and Level II) may be purchased separately. Glucose control solutions were previously cleared under K032985.
- (4) Check Strip
- (5) Instruction for use
[Test Principle]
The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor.
When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.
# [Control Solution]
The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to
2
{2}------------------------------------------------
assess the performance of the PRECICHEK NS-101 POCT Professional Blood Glucose monitoring system and the PRECICHECK KP Blood Glucose Test strips. There are two levels of controls (Levels 1,2).
# [Check Strip]
The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.
### [Device Calibration]
The device is calibrated by implicit coding process. The code number is the last two digits of the strip lot number. The user should input the strip lot number into the memory of the meter before use.
While performing the blood glucose test, the strip lot number printed on the vial label is scanned to match the one in the memory. The meter will apply formula including this parameter to calculate the glucose value.
#### 5. Intended Use
PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System
The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from the fingertips, venous whole blood and arterial whole blood. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This
3
{3}------------------------------------------------
system should only be used with single-use, auto-disabling lancing devices.
The PRECICHEK KP Blood Glucose Test Strip is for use with the PRECICHEK NS-101 POCT Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood from the fingertips, venous whole blood and arterial whole blood.
The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to assess the performance of the PRECICHEK NS-101 POCT Professional Blood Glucose monitoring system and the PRECICHECK KP Blood Glucose Test strips. There are two levels of controls (Levels 1, 2).
- 6. Comparison of Subject Devices and predicate device
Technological Characteristics Comparison Table of PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System and LifeScan SureStep Flexx Professional Blood Glucose Management System (K023194)
| Item | Subject Device | Predicate Device |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Similarities | | |
| Indication of Use | It is designed to quantitatively<br>measure the concentration of<br>glucose (sugar) in capillary,<br>venous, and arterial whole<br>blood by healthcare<br>professionals to monitor<br>patient's blood glucose in<br>clinical settings. The system is<br>used outside the body (in vitro<br>diagnostic). | Same |
| Code system | Bar code | Same |
| | PRECICHEK NS-101 POCT<br>Professional BGMS | LifeScan SureStep Flexx<br>Professional BGMS |
| Blood sample<br>source | Fresh whole blood (capillary,<br>venous, arterial) | Same |
| Warming &<br>Precaution | For <i>in-vitro</i> diagnostic use only | Same |
| Calibration | Plasma Equivalent | Same |
| Operating<br>Procedure | Menu control<br>Operational ID, Patient ID,<br>Strip Lot | Same |
| Test Mode | Glucose Control Solution test,<br>Capillary test, Venous test,<br>Arterial test. | Same |
| Characteristics | For professional use only | Same |
| Differences | | |
| Dimensions | 143 x 65 x 33mm | 161 x 90 x 41mm |
| Biological Source | Glucose Dehydrogenase | Glucose Oxidase |
| Principle | Electrical chemistry reaction | Photochemical Reaction |
| Blood sample<br>volume | 0.5uL | 5-30uL |
| Buttons | Power, Scan, numbers | Power |
| Weight | 200g (with battery) | 375g (with battery) |
| Reaction time | 5 seconds | 30 seconds |
| Blood Glucose<br>Range | 20-600mg/dl | 0-500 mg/dl |
| Operational<br>Environment | 10-40℃, 50-104℉ | 18-30℃, 64-86℉ |
| Environment | 20-80% RH | 30-70% |
| Power Source | 3.7V Li rechargeable via USB<br>battery | 3 AA 1.5v alkaline batteries |
{4}------------------------------------------------
# 7. Discussion of Clinical Tests Performed
PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System (Subject Device) is compliant to the standard of ISO 15197:2003 In vitro diagnostic test systems- Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. All the relevant activities were performed by professionals and
{5}------------------------------------------------
the results demonstrated that the predetermined acceptance criteria were fully met.
- 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The subject device was tested to evaluate its safety and effectiveness, including the followings:
- ISO 15197:2003 In vitro diagnostic test systems Requirements (1) for blood-glucose monitoring systems for self-testing in managing diabetes mellitus
- IEC 60601-1:1988 + A1:1991 + A2:1995 Medical electrical (2) equipment - Part 1: General requirements for safety
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: (3) General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility-Requirements and tests
- (4) . CLSI/NCCLS EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods
- (5) CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement.
- CLSI EP07-A2: Interference Testing in Clinical Chemistry (6)
- CLSI/NCCLS EP09-A2: Method Comparison and Bias Estimation (7) Using Patient Samples
- FDA Guidance: Guidance for the Content of Premarket (8) Submissions for Software Contained in Medical Devices, May 11, 2005
- 9. Conclusion
The subject device was tested and fulfilled the requirements from those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure. The text is in all caps and appears to be in a sans-serif font.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 9, 2012
HMD Biomedical Inc. c/o Wang Yunhan No. 181, Minsheng St Xinpu Township, Hsinchu County Taiwan
Re: k113314
Trade/Device Name: PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: LFR, JJX Dated: October 26, 2012 Received: October 31, 2012
Dear Wang Yunhan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
#### Page 2 - Wang Yunhan
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
# Carol C. Benson
for
Courtney H. Lias, Ph.D Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): | ||3314
Device Name [Trade Name]:
PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System
#### Indications for Use:
The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by health care professionals in a clinical setting:
The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The PRECICHEK KP Blood Glucose Test Strip is for use with the PRECICHEK NS-101 POCT Professional Blood Glucose Meter to quantitatively measure glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples.
The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to assess the performance of the PRECICHEK NS-101 POCT Professional Blood Glucose monitoring system and the PRECICHECK KP Blood Glucose Test strips to check that the meter and test strips are working together properly and that the test is performing correctly.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use _ AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Lallum Lumo
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) KII 33 14
Page 1 of __
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.