EGV1.1 SELF MONITORING BLOOD GLUCOSE SYSTEM AND EGV1.1 PRO MONITORING BLOOD GLUCOSE SYSTEM

K133537 · Eps Bio Technology Corp. · LFR · Apr 28, 2014 · Clinical Chemistry

Device Facts

Record IDK133537
Device NameEGV1.1 SELF MONITORING BLOOD GLUCOSE SYSTEM AND EGV1.1 PRO MONITORING BLOOD GLUCOSE SYSTEM
ApplicantEps Bio Technology Corp.
Product CodeLFR · Clinical Chemistry
Decision DateApr 28, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The EGV1.1 Self Monitoring Blood Glucose System and EGV1.1 Pro Monitoring Blood Glucose System are intended for single- and multiple-patient use, respectively.

Device Story

System measures glucose in venous or capillary whole blood samples via test strips. EGV1.1 Pro intended for professional clinical settings; EGV1.1 intended for home use. Devices utilize electrochemical sensing to quantify glucose levels. Results displayed on meter screen to assist healthcare providers or patients in monitoring diabetes control. Professional version requires single-use lancing devices to prevent cross-contamination. Home version restricted to single-patient use to avoid blood-borne pathogen transmission. Output informs diabetes management decisions. Benefits include rapid, point-of-care glucose monitoring.

Clinical Evidence

No clinical data provided in the document.

Technological Characteristics

Handheld blood glucose meter. Powered by Alkaline batteries. Features three-button interface. Validated for durability against 10,950 cleaning/disinfection cycles using PDI Super Sani-Cloth Germicidal Disposable Wipes (EPA #9480-4).

Indications for Use

Indicated for single-patient (EGV1.1) or multiple-patient (EGV1.1 Pro) use for the self-monitoring of blood glucose levels.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 EPS BIO TECHNOLOGY CORP. C/O CYNTHIA HUNG REGULATORY SPECIALIST NO.8 R&D RD III. HSINCHU SCIENCE PARK HSINCHU CITY. TAIWAN 30077 April 28, 2014 Re: K133537 Trade/Device Name: EGV . I Self Monitoring Blood Glucose System EGVI.I Pro Monitoring Blood Glucose System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. LFR Dated: March 27, 2013 Received: March 31, 2013 Dear Ms. Cynthia Hung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however. that device labeling must be truthful and not misleading. If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the {1}------------------------------------------------ Page 2-Ms. Cynthia Hung electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. # Katherine Serrano -S For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K133537 Device Name EGV1.1 Pro Monitoring Blood Glucose system #### Indications for Use (Describe) The EGV1.1 Pro Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in venous whole blood or fresh capillary whole blood from fingertip. Testing is done outside the body (In Vitro diagnostic use). It is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. The system is only used with single-use lancing devices. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. The system consists of the EGV1.1 Pro meter and the EGV1.1 Pro meter only is used with the EGV1.1 Pro test strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from fingertip. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) ## Stayce Beck -S This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K133537 Device Name EGVI.I Self Monitoring Blood Glucose system #### Indications for Use (Describe) The EGV . I Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose in fresh capillary whole blood from fingertip, palm, or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by a single patient with diabetes and should not be shared, as an aid to monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. Alternative site testing can be only used during steady-state blood glucose conditions. The system consists of the EGV I.I meter and the EGV I.I test strips. The EGV I.I meter only is used with the EGV I.I test strips to quantitatively measure glucose in fresh capillary whole blood from fingertip, palm, or forearm. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # Stayce Beck -S This section applies only to requirements of the Paperwork Reduction Act of 1995. ### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@Ida.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." : Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...