GOODLIFE CS-200 (SINGLE PATIENT USE), CS-200 PROFESSIONAL (MULTIPLE PATIENT USE) BLOOD GLUCOSE MONITORING SYSTEMS

K121433 · Hmd Biomedical, Inc. · LFR · Feb 14, 2013 · Clinical Chemistry

Device Facts

Record IDK121433
Device NameGOODLIFE CS-200 (SINGLE PATIENT USE), CS-200 PROFESSIONAL (MULTIPLE PATIENT USE) BLOOD GLUCOSE MONITORING SYSTEMS
ApplicantHmd Biomedical, Inc.
Product CodeLFR · Clinical Chemistry
Decision DateFeb 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The GoodLife CS-200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife CS-200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The GoodLife CS-200 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife CS-200 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The GoodLife KP Blood Glucose Test Strips are for use with the GoodLife CS-200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife CS-200 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in venous or arterial whole blood or fresh capillary drawn from the fingertips. It is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program. This system should only be used with single-use, auto-disabling lancing devices. The GoodLife CS-200 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The GoodLife KP Professional Blood Glucose Test Strip is for use with the GoodLife CS-200 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in venous or arterial whole blood or fresh capillary drawn from the fingertips.

Device Story

System measures glucose in whole blood via amperometric reaction on test strips using glucose dehydrogenase (GDH-FAD). Input: 0.5 μL blood sample applied to strip via capillary action. Meter processes reaction current proportional to glucose concentration; displays result on LCD; stores data. Used by patients (home) or healthcare professionals (clinical). Professional version requires disinfection between patients. Output informs diabetes control monitoring. Benefits: rapid, quantitative glucose monitoring.

Clinical Evidence

Bench testing only. Compliance with ISO 15197:2003 requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus was verified. Results demonstrated that all predetermined acceptance criteria were met.

Technological Characteristics

Electrochemical biosensor (amperometry) using Glucose Dehydrogenase (FAD). Meter includes LCD, coding slot for digital code strip, and PCB-based check strip. Operating range 10-40°C. Sample volume 0.5µL. Calibration via digital code strip.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary (single-patient) or venous/arterial/capillary (professional) whole blood. Intended for diabetes management monitoring. Contraindicated for neonatal use, diabetes screening/diagnosis, and use on critically ill, shock, dehydrated, hypotensive, or hyperosmolar patients.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121433 ## HMD BioMedical Inc. ## 510(k) Summary # FEB 1 4 2013 ## 1. Submitter Information | Company name: | HMD BioMedical Inc. | |-----------------|------------------------------------------------------------------| | Contact person: | Jessica Tung | | Address: | No. 181, Minsheng St., Xinpu Township, Hsinchu<br>County, Taiwan | | Phone: | 886-3-6017000 | | FAX: | 886-3-5885500 | | E-mail: | jessica-tung@hmdbio.com | | Date Prepared: | February 14, 2013 | # 2. Name of Device Trade/Proprietary Name: For single patient use GoodLife CS-200 Blood Glucose Monitoring System - For multiple patient use GoodLife CS-200 Professional Blood Glucose Monitoring System | Common name: | Blood Glucose Test System | |-----------------------|---------------------------| | Classification name: | Glucose Test System | | Classification Panel: | Clinical Chemistry (75) | | Regulation no .: | 862.1345 (Class II) | | Product code: | NBW, LFR | - 3. Predicate Device | Trade/Proprietary name: | U-RIGHT TD-4279A Blood Glucose<br>Monitoring System | |-------------------------|-----------------------------------------------------| | Common name: | Blood Glucose Test System | | Submitter: | TaiDoc Technology Corporation | | 510(k) no.: | K101509 | | Product code: | NBW, LFR | 4. Device Description {1}------------------------------------------------ GoodLife CS-200 Blood Glucose Monitoring System and GoodLife CS-200 Professional Blood Glucose Monitoring System consist of: - (1) Glucose Meter - (2) Glucose Test Strips - (3) Two levels of glucose control solutions (Level I and Level II) may be purchased separately. Glucose control solutions were - previously cleared under K032985. - (4) Check Strip - (5) Instruction for use #### [Test Principle] The GoodLife CS-200 & GoodLife CS-200 Professional Blood Glucose Monitoring System are electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor. When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use. #### [Control Solution] The GoodLife Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) assessing the performance of the GoodLife CS-200 & GoodLife CS-200 Professional Blood Glucose Monitoring System and GoodLife KP & GoodLife KP Professional Blood Glucose Test strips. There are two levels of controls (Levels 1, 2). #### [Check Strip] The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover. [Device Calibration] {2}------------------------------------------------ HMD BioMedical Inc. The device is calibrated by Digital Code Strip. While inserting the Digital Code Strip into the coding slot, the meter will turn on automatically and complete the coding. The meter will apply formula including this parameter of code to calculate the glucose value. #### 5. Intended Use The GoodLife CS-200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife CS-200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The GoodLife CS-200 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife CS-200 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes, nor for neonatal use. The GoodLife KP Blood Glucose Test Strips are for use with the GoodLife CS-200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife CS-200 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in venous or arterial whole blood or fresh capillary whole blood drawn from the fingertips. It is intended for testing outside the body (in vitro diagnostic use) and intended for multiple-patient use in professional healthcare setting as an aid to monitor the effectiveness of a diabetes control program. This system should only be used with single-use, auto-disabling lancing devices. The GoodLife CS-200 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes, nor for neonatal use. The GoodLife KP Professional Blood Glucose Test Strip is for use with {3}------------------------------------------------ HMD BloMedical Inc. the GoodLife CS-200 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in venous or arterial whole blood or fresh capillary whole blood drawn from the fingertips ## 6. Comparison of subject devices and predicate device Technological Characteristics Comparison Table of GoodLife CS-200 Blood Glucose Monitoring System for single patient use & GoodLife CS-200 Professional Blood Glucose Monitoring System for multiple patient use and U-RIGHT TD-4279A Blood Glucose Monitoring System (K101509) | Item | Subject Device<br>GoodLife CS-200 & CS-200 Professional<br>BGMS | | Predicate Device<br>U-RIGHT TD-4279A<br>BGMS (k101509) | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------------------------------------------------------------------------------------------------------------| | Similarities | | | | | Intended use | For single patient use (CS-200) -<br>It is designed to quantitatively measure<br>the concentration of glucose in fresh<br>capillary whole blood | | Same | | Detection method | Amperometry:<br>Current produced by chemical reaction | | Same | | Coding function | By code strip | | Same | | Test range | 20~600mg/dL | | Same | | Operating conditions | 50~104°F (10~40°C) | | Same | | Test time | 5 sec | | Same | | Capillary testing site | For single patient use (CS-200) -<br>Fingertips only | | Same | | Enzyme | Glucose Dehydrogenase (FAD) | | Same | | Differences | | | | | Intended use | For multiple patient use (CS-200<br>Professional) -<br>It is designed to<br>quantitatively measure the<br>concentration of glucose in<br>fresh capillary drawn<br>from the fingertips, venous or | | It is designed to<br>quantitatively measure the<br>concentration of glucose in<br>fresh capillary whole blood. | {4}------------------------------------------------ HMD BioMedical Inc. | | arterial whole blood samples. | | |----------------------|-------------------------------|---------------------| | Sample volume | 0.5ul | 1.1ul | | Storage condition | 50~104°F (10~40°C) | 35.6~89.6°F(2~32°C) | | PC data transmission | N/A | USB cable | ## 7. Discussion of Clinical Tests Performed GoodLife CS-200 Blood Glucose Monitoring System and GoodLife CS-200 Professional Blood Glucose Monitoring System (Subject Device) is compliant to the standard of ISO 15197:2003 In vitro diagnostic test systems- Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. All the relevant activities were performed by professionals and the results demonstrated that the predetermined acceptance criteria were fully met. #### 8. Conclusion The subject device was tested and fulfilled the requirements from those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device. {5}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 25, 2013 HMD BioMedical Inc. c/o Jessica Tung No. 181. Minsheng St., Xinpu Township, Hsinchu County China (Taiwan) 305 Re: k121433 Trade/Device Name: GoodLife CS-200 Blood Glucose Monitoring System GoodLife CS-200 Professional Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: LFR, NBW Dated: February 07, 2013 Received: February 8, 2013 Dear Ms. Tung: This letter corrects our substantially equivalent letter of February 14, 2013. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {6}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): k121433 Device Name: GoodLife CS-200 Blood Glucose Monitoring System Indications for Use: The GoodLife CS-200 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife CS-200 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The GoodLife CS-200 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife CS-200 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The GoodLife KP Blood Glucose Test Strips are for use with the GoodLife CS-200 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. AND/OR Over-The-Counter Use _ Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) The States of the Comments of the Children Comments on Child African Comments of Child Annual Children Comments of Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR) # Katherine Serrano Division Sign-Off Office of In Vitro Devices and Radiologic Health 510(k) k121433 {8}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): k121433 Device Name: GoodLife CS-200 Professional Blood Glucose Monitoring System Indications for Use: The GoodLife CS-200 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in venous or arterial whole blood or fresh capillary whole blood drawn from the fingertips. It is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of a diabetes control program. This system should only be used with single-use, auto-disabling lancing devices. The GoodLife CS-200 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The GoodLife KP Professional Blood Glucose Test Strip is for use with the GoodLife CS-200 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in venous or arterial whole blood or fresh capillary drawn from the fingertips. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use _ × (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR) # Katherine Serrano Division Sign-Off Office of In Vitro Devices and Radiologic Health 510(k) k121433
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...