The CONTOUR® NEXT LINK Wireless Blood Glucose Monitoring System is an over the counter (OTC) device utilized by persons with diabetes in home settings for the measurement of glucose in whole blood, and is for single-patient use only and should not be shared. The CONTOUR® NEXT LINK Wireless Blood Glucose Monitoring System is indicated for use with fresh capillary whole blood samples drawn from the fingertip and palm only. The system consists of a Contour NEXT LINK wireless blood glucose meter, CONTOUR® NEXT test strips and CONTOUR® NEXT control solutions. CONTOUR® NEXT test strips are intended for self-testing by persons with diabetes for the quantitative measurement of glucose in whole blood samples from 20 to 600 mg/dL. The CONTOUR® NEXT control solutions are aqueous glucose solutions intended for use in self-testing by people with diabetes as a quality control check. The CONTOUR® NEXT LINK Wireless Blood Glucose Monitoring System is intended to be used to transmit glucose values to Medtronic MiniMed Paradigm Insulin pumps or Medtronic MiniMed Paradigm REAL-Time Revel Insulin Pumps or Medtronic MiniMed Paradigm® REAL-TIME Insulin Pumps or Guardian REAL-TIME Monitor and facilitate transfer of information to Medtronic MiniMed Carelink Therapy Management Software through use of radio frequency communication. The CONTOUR® NEXT LINK Wireless Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.
Device Story
Handheld electronic blood glucose meter; uses dry reagent strips; measures capillary whole blood glucose via amperometry (FAD-GDH enzyme). Input: 0.6 μL blood sample applied to strip. Operation: Multi-pulse algorithm measures current at multiple time points to compensate for hematocrit/temperature; automatically distinguishes control solution from blood. Output: Glucose concentration displayed on meter; RF transmission of results to compatible Medtronic MiniMed insulin pumps/monitors and Carelink software. Used in home settings by patients. Benefits: Facilitates automated data transfer to insulin pumps/management software; includes 'Double Dip' feature for insufficient sample volume and moisture damage detection. Healthcare providers use transmitted data for therapy management.
Clinical Evidence
Clinical user performance study (N=110) with type 1/2 diabetics (age 19-86). Compared lay-user fingerstick/palm results to YSI reference method. Results: 100% of fingerstick and 97.3% of palm results met ISO 15197 accuracy criteria (±15 mg/dL for <75 mg/dL; ±20% for ≥75 mg/dL). Bench testing confirmed linearity (20-600 mg/dL), precision (CV <2.1%), and interference resistance (acetaminophen, ascorbic acid, etc.).
Technological Characteristics
Glucose dehydrogenase-based electrochemical sensing. Rechargeable battery (3.4-4.2V). OLED display. USB connectivity for PC; 916.5 MHz RF for pump communication. Dimensions/materials similar to Contour USB. Single-patient use. Software-controlled data transmission and pass-through.
Indications for Use
Indicated for persons with diabetes (adults/pediatrics) for quantitative measurement of glucose in fresh capillary whole blood (fingertip/palm). Not for neonates, screening/diagnosis of diabetes, or critically ill patients (e.g., severe hypotension, shock, DKA).
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K122370 — CONTOUR NEXT LINK WIRELESS BLOOD GLUCOSE MONITOR · Bayer Healthcare · Sep 26, 2013
K073231 — ONETOUCH ULTRALINK BLOOD GLUCOSE MONITORING SYSTEM · Lifescan, Inc. · Apr 18, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K110894
MAR 2 8 2012
# 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements
of 21
| Submitter Information | |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Bayer Healthcare, Diabetes Care |
| Address | 430 S Beiger St<br>Mishawaka IN 46544 |
| Phone number | 574-256-3441 |
| Fax number | 547-256-3519 |
| Establishment Registration Number | 1826988 |
| Name of contact person | Roger Sonnenburg |
| Date prepared | 3/26/2012 |
| Trade or proprietary name | Contour® NEXT LINK Wireless Blood Glucose Meter |
| Common or usual name | Blood Glucose Meter |
| Classification name | 75 LFR Glucose Dehydrogenase, Glucose |
| Classification panel | Clinical Chemistry and Toxicology |
| Regulation | 21 CFR 862.1345 |
| Product code(s) | LFR (Glucose Dehyrogenase, Glucose), NBW (System, Test, Blood Glucose, Over The Counter) |
| Legally marketed device(s) to which equivalence is claimed | Predicate 1 Contour USB (K091820), Predicate 2 One Touch Ultralink (K073231), Predicate 3 Care Link USB (K070438) |
| Reason for 510(k) submission | This submission reports a modification of the Contour USB meter (K091820). This modification includes the ability to read a new reagent strip and wireless transmission capabilities with compatible Medtronic Minimed devices |
| Device description | The Contour® NEXT LINK Wireless Blood Glucose Monitoring System consists of a small handheld electronic device is substantially equivalent in look and feel to the Contour® USB predicate system (K091820). The System also contains dry reagent strips and liquid controls to be used for the measurement of glucose in capillary whole blood by persons with diabetes. The System has the same automatic calibration as the predicate device. Blood glucose results are displayed on the meter window and stored in the meter's memory. The |
{1}------------------------------------------------
| | software. | results to compatible Medtronic Minimed insulin pumps. The RF function can<br>also serve as a pass through for data being transmitted from Medtronic<br>Minimed insulin pumps to Medtronic's Minimed PC based data management | |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Intended use of the<br>device | See indications for use below | | |
| Indications for use | | The CONTOUR® NEXT LINK wireless blood glucose monitoring system is an over the<br>counter (OTC) device utilized by persons with diabetes in home settings for the<br>measurement of glucose in whole blood, and is for single-patient use only and should<br>not be shared. The CONTOUR® NEXT LINK wireless blood glucose monitoring system<br>is indicated for use with fresh capillary whole blood samples drawn from the fingertip<br>and palm only. The system consists of a Contour NEXT LINK wireless blood glucose<br>meter, CONTOUR® NEXT test strips and CONTOUR® NEXT control solutions. | |
| | CONTOUR® NEXT test strips are intended for self-testing by persons with diabetes for<br>the quantitative measurement of glucose in whole blood samples from 20 to 600<br>mg/dl | | |
| | The CONTOUR® NEXT control solutions are aqueous glucose solutions intended for<br>use in self-testing by people with diabetes as a quality control check. | | |
| | use of radio frequency communication. | The CONTOUR® NEXT LINK wireless blood glucose monitoring system is intended to<br>be used to transmit glucose values to Medtronic MiniMed Paradigm Insulin pumps or<br>Medtronic MiniMed Paradigm REAL-Time Revel Insulin Pumps or Medtronic MiniMed<br>Paradigm® REAL-TIME Insulin Pumps or Guardian REAL-TIME and facilitate transfer<br>of information to Medtronic MiniMed Carelink Therapy Management Software through | |
| | neonates. | The CONTOUR® NEXT LINK wireless blood glucose monitoring system is not intended<br>for the diagnosis of or screening for diabetes mellitus and is not intended for use on | |
| | 1. 13. 2. 13. 1 | nological Characteristics of the New Device Compared to Pre | |
| | Similarities to Predicate 1 | | |
| Characteristic | Predicate 1 | New Device | |
| | Contour USB K091820 | Contour NEXT LINK Wireless | |
| Blood sample volume<br>Meal markers | 0.6ul | Same as predicate | |
| Automatic calibration | Yes<br>Yes | Same as predicate<br>Same as predicate | |
| Communication port | USB | Same as predicate | |
| User interface | Alphanumeric, Iconic, Native<br>Language | Same as predicate | |
| Display (technology) | Graphical (OLED) | Same as predicate | |
| Display visibility | Day & Night | Same as predicate | |
| Illuminated strip port | Yes | Same as predicate | |
| Operational buttons | 4 | Same as predicate | |
| Battery type | Rechargeable (3.4-4.2V) | Same as predicate | |
| | | | |
| Communication Link to<br>Computer<br>Displayed countdown | Direct USB Connection or<br>optional USB Cable<br>5 seconds | Same as predicate<br>Same as predicate | |
| Characteristic | Predicate 1<br>Contour USB K091820 | New Device<br>Contour NEXT LINK Wireless | |
| Control solution ranges | Low/Normal/High | Level 1/Level 2 | |
| Control solution buffer<br>concentration | 50-100 mM | 22 mM | |
| Reagent strip<br>compatibility | Only compatible with Contour<br>reagent strip | Only compatible with Contour NEXT<br>reagent strip | |
| Total test time Contour<br>NEXT reagent strip | N/A | 7 Seconds | |
| Software system<br>requirements | Windows:XP,SP3,Vista SP2, 7;<br>MAC OS 10.6.3,10.5.8;<br>Java 1.6.0_07 or higher, High<br>power USB port | High power USB port | |
| Double dip function<br>Contour NEXT reagent<br>strip | N/A | Yes | |
| Indications for Use | Cleared for home and<br>professional use and fingertip,<br>palm and forearm sampling. | Intended for home use, single patient use<br>only fingertip and palm sampling only.<br>RF Capability to communicate with<br>compatible insulin pumps and act as a<br>data pass through between insulin pumps<br>and data management software (see<br>Indications for Use section). | |
| Test memory | 2000 results | 1000 results | |
| User accessible data<br>storage | Yes (500MB) | No | |
| Data management<br>software application on- | Yes | No | |
.
:
:
:
.
.
:
{2}------------------------------------------------
. . . . . . .
:
.
:
:
.
:
.
.
:
:
:
the contraction of the country of
.
:
{3}------------------------------------------------
Summary of the Technological Characteristics of the New Device Compared to Predicate 2 (One Touch Ultralink K073231) - - - - - - -Similarities to Predicate 2 Predicate 2 New Device
| Characteristic | One Touch Ultralink | Contour NEXT LINK Wireless |
|----------------------------------------------------------------------|---------------------------------------------------------------|------------------------------------------|
| Communication | Ability to communicate with<br>compatible insulin pump via RF | Same as predicate |
| Frequency | 916.5 MHz | Same as predicate |
| Differences from Predicate 2 | | |
| Characteristic | Predicate 2<br>One Touch Ultralink | New Device<br>Contour NEXT LINK Wireless |
| User<br>confirmation<br>regarding<br>sending via RF<br>of BGM result | RF on, RF off | RF on, RF off and RF prompt |
Summary of the Technological Characteristics of the New Device Compared to Predicate 3 (Care Link USB K070438) - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1 - 3 - 1
| Characteristic | Predicate 3<br>Care Link USB | New Device<br>Contour NEXT LINK Wireless |
|--------------------------|------------------------------|------------------------------------------|
| Pass through<br>function | Yes | Same as predicate |
### SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
#### Performance Test Summary-New Device
| Performance Test Summary-New Device | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Results Summary |
| Precision | Repeatability (ISO 15197 Section 7.2.2)<br>Reference: 510(k) submission, Section Labeled Performance Testing Bench<br>Protocol: The repeatability of the CONTOUR NEXT LINK wireless blood glucose monitoring<br>system was tested at glucose levels of 38, 78, 122, 204 and 326 mg/dL. Three lots of Contour<br>Next reagent strips were tested across 10 Contour NEXT LINK wireless meters.<br>Acceptance Criteria: No ISO criteria stated. Internal acceptance criteria are $Cpk > 0.65$ at all<br>levels. (Cpk is difference between the mean result and the nearest limit, divided by 3 standard<br>deviations.)<br>Results: Pass, All Cpk's greater than 0.65. |
| | Intermediate Precision (ISO 15197 Section 7.2.3)<br>Reference: 510(k) submission, Section Labeled Performance Testing Bench<br>Protocol: The intermediate precision of the CONTOUR® NEXT LINK wireless blood glucose<br>monitoring system was determined using three levels of Contour NEXT control on three<br>Contour NEXT strip lots on 10 Contour NEXT LINK wireless meters over the course of 10<br>days.<br>Acceptance Criteria for Contour reagent strips: No ISO criteria stated. Internal acceptance<br>criteria are $Cp > 1.0$ at all levels. (Cp is the tolerance range divided by 6 standard deviations.)<br>Acceptance Criteria for Contour NEXT reagent strips: Internal acceptance criteria are $Cp > $<br>1.0 at all levels<br>Results: Pass. All Cp’s above 1.0 |
{4}------------------------------------------------
| Accuracy | System Accuracy Evaluation (ISO 15197 Section 7.3)<br>Reference: 510(k) submission, Section Labeled Performance Testing Bench<br>Protocol: The accuracy of the CONTOUR®NEXT LINK wireless blood glucose monitoring<br>system was assessed using 100 fresh capillary blood specimens that were collected during a<br>fingerstick study at the Mishawaka site. The blood samples were tested using three<br>CONTOUR® NEXT reagent lots, 10 vials per lot to yield 300 results. The reference<br>measurement was done on a YSI™ glucose analyzer.<br>Acceptance Criteria: 95% ± 15 mg/dL < 75 mg/dL, 95% ± 20% ≥ 75 mg/dL<br>Results: Pass for Contour NEXT reagent strips. | | | | |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|--------------------|-------------------|-------------------|
| | | | | | |
| | Summary of Results for Contour Next reagent strips on Contour Next Link Wireless Meter: | | | | |
| | Results within: | ±5mg/dL | ±10mg/dL | ±15mg/dL | |
| | YSI Glucose <75 mg/dL | 48 of 51 (94.1%) | 51 of 51 (100%) | 51 of 51 (100%) | |
| | Results within: | ±5% | ±10% | ±15% | ±20% |
| | YSI Glucose ≥75 mg/dL | 212 of 249 (85.1%) | 247 of 249 (99.2%) | 249 of 249 (100%) | 249 of 249 (100%) |
| Linearity/assay reportable range | For Contour reagent strip, established in current Predicate (K091820) and previous Contour systems<br>Reference: 510(k) submission, Section Labeled Performance Testing Bench<br>Protocol: A linearity study was performed with three lots of Contour Next reagent strips. A<br>fresh venous blood pool was divided into eight aliquots and glycolyzed at room temperature or<br>supplemented with a glucose stock solution to produce whole blood samples with plasma<br>across the intended glucose measuring range. The evaluation included 24 replicates on eight<br>meters. In addition, all three reagent strip lots were tested with a blood pool adjusted to 15<br>mg/dL per the YSI reference, using meters with the low-end cutoff disabled.<br>Acceptance Criteria: Cpk ≥ 0.65 at all levels<br>Results: Pass, Values measured across the intended glucose measuring range | | | | |
| Traceability | The Yellow Springs Instruments Stat Plus 2300 analyzer (YSI) is traceable to the hexokinase<br>method developed collaboratively by the FDA, CDC, NIST and AACC. The hexokinase<br>method utilizes NIST Standard Reference Material 917b, dry D-glucose. For the YSI<br>qualification, six-level serum controls were assayed on six different YSI analyzers over a one-<br>month period. Each day that YSI instruments were used as the reference method, the serum<br>controls were analyzed to ensure that the instruments were in control.<br>Summary of Results: Pass, Results within limits | | | | |
.
・
·
{5}------------------------------------------------
| Detection limit | Established in current Predicate (K091820) and previous Contour systems<br>Reference: 510(k) submission, Section Labeled Performance Testing Bench |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Protocol: To ensure that the performance of Contour NEXT LINK wireless blood glucose<br>system is acceptable at glucose concentrations near the low end of the analytical range, the<br>three clinical trial lots were evaluated with glycolyzed 42% Hct whole blood samples at<br>approximately 10, 15, and 20mg/dL glucose. The study was done by collecting 24 replicates<br>on eight breadboard meters which did not have the low limit set. |
| | In addition, the system was tested with blood adjusted to extremely low and extremely<br>high glucose levels of 5, 900, 1200, 1500 and 1800 mg/dL. Eight meters were tested with a<br>total of 72 reagent strips at 5 mg/dL, and 288 reagent strips at 900 mg/dL and above.<br>Acceptance Criteria: |
| | For blood with extreme glucose levels, the acceptance criterion is for the meter to display<br>"Low" or "High" glucose error messages. |
| | Results: Pass. All extremely low and extremely high samples generated "Low" or "High"<br>error messages as appropriate. |
| Analytical<br>specificity | The effect of changing hematocrit was evaluated at 15, 20, 30, 42, 55 and 65% Hct at blood<br>glucose concentrations of 40, 90, 127 and 329 and 450 mg/dL for three reagent strip lots on 5<br>Contour NEXT LINK wireless meters in triplicate. |
| | A high altitude simulation study was also performed which evaluated 15, 41 and 65 % Hct at<br>blood glucose concentrations of 40, 90 and 450 mg/dL for two reagent strip lots on 5 Contour<br>NEXT LINK wireless meters in duplicate. |
| | The acceptance criterion was <10 mg/dL or 10% mean difference from the YSI reference<br>values. |
| | The following interfering substances were tested: hematocrit, ascorbic acid, bilirubin, uric acid,<br>maltose, galactose |
| | Other exogenous interfering substances were also tested, such as cholesterol, creatinine,<br>dopamine, etc. |
| | Summary of Results for hematocrit (Contour NEXT reagent) : |
| | At all Hct levels, the mean difference from the reference values met the acceptance criterion. |
| | Summary of Results for interfering substances (Contour NEXT reagent) : |
| | Interfering substances, except for ascorbic acid, did not have a significant effect on<br>performance. |
| | Ascorbic acid above 10 mg/dL can cause a +10% assay bias. The high end of the therapeutic<br>range for ascorbic acid is 2 mg/dL. |
| Assay cut-off | Not applicable |
:
:
.
.
:
{6}------------------------------------------------
| RF of BGM<br>result to<br>insulin pump | Protocol: Electromagnetic compatibility testing was performed that included interferences; cell<br>phone, metal detectors, EAS immunity and household emitter immunity. (N of 30 meter and<br>pump pairs)<br>Acceptance criteria: Medtronic pumps were required to successfully receive the values<br>transmitted by the Contour NEXT LINK wireless system at a distance of 4 feet with the pump in<br>eight different orientations. In the event there is interruption of RF Communication, any<br>corrupted data shall be recognized and not processed. Testing was performed against RF<br>specifications(ES9411) and product specifications (ES10255). This included the following:<br>Assuring the meter transmitted blood glucose values in the mg/dL to the receiving<br>device at 4 feet at 916.5 MHz every 5 seconds for up to 12 attemptsMeter stopped transmitting after it successfully received ACK from the receiving<br>device, and the meter stopped transmitting after 60 seconds if an ACK was not<br>received.Meter is capable to turn RF transmit feature On or OFF and if RF transmit feature<br>OFF, meter will still display BG value but does not transmit BG value.Meter has a 6 digit ID numberResults: EMC testing with various interferences confirmed that in no case was an incorrect<br>value accepted or displayed by the pump per RF and product specifications. |
|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pass through<br>function (RF of<br>pump data<br>through BGM<br>to PC based<br>data<br>management<br>software when<br>connected to<br>PC USB port) | Protocol: Testing was performed that assured the Contour NEXT LINK wireless meter, when<br>connected to a PC USB port, enabled communications between the compatible Medtronic<br>MiniMed device and a PC for the purpose of allowing meter data and settings to be<br>downloaded to Medtronic data management software.<br>Acceptance criteria: Testing was performed against RF specifications (ES9411) and product<br>specifications (ES10255).<br>Results: Data transfer and pass-through mode testing demonstrated successful data transfer<br>operation for the pass through mode. |
.
·
:
> . . . . . . . . . -
.
. . . . . .
:
:
. .
{7}------------------------------------------------
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL<br>EQUIVALENCE AND/OR OF CLINICAL INFORMATION | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------------------|-----------------------|-----------------------|
| Clinical<br>Performance<br>Data/Information | Reference: 510(k) submission, Section Labeled Clinical Trials<br>Protocol for Contour NEXT LINK wireless meters using Contour NEXT reagent strips:<br>The Contour NEXT wireless blood glucose monitoring system clinical trial included 110 adults<br>aged 19-86. 88% of the subjects managed their diabetes at home with insulin.<br>Three lots of reagent were tested in the study with each subject randomized to one lot. Each<br>subject performed one fingerstick self test, followed by testing of the subject's blood by the study<br>staff. Subjects also performed AST self tests on the palm. All meter results were compared to<br>the YSI reference results.<br>Acceptance Criteria for Contour NEXT LINK wireless meter using Contour NEXT reagent<br>strips:<br>95% of results within ±20% (≥ 75 mg/dL) or within ±15 mg/dL (<75 mg/dL) of YSI results | | | | |
| Clinical trial results for Contour®Next Link Wireless System | | | | | |
| Clinical Trial Results for Contour®Next Link Wireless - Contour®Next Strips:<br>Fingerstick 110 results: 110 subjects, 1 lot per subject, 3 lots overall | | | | | |
| | Results<br>within: | ±5 mg/dL | ±10 mg/dL | ±15 mg/dL | ±20 mg/dL |
| | YSI Glucose<br><75 mg/dL | 8 of 8 (100%) | 8 of 8 (100%) | 8 of 8<br>(100%) | 8 of 8 (100%) |
| | Results<br>within: | ±5% | ±10% | ±15% | ±20% |
| | YSI Glucose<br>≥75 mg/dL | 82 of 102<br>(80.4%) | 101 of 102<br>(99.0%) | 101 of 102<br>(99.0%) | 102 of 102 (100%) |
| | Results<br>within: | ±5 mg/dL or<br>5% | ±10 mg/dL or<br>10% | ±15 mg/dL<br>or 15% | ±15 mg/dL or 20% |
| | Total | 90 of 110<br>(81.8%) | 109 of 110<br>(99.1%) | 109 of 110<br>(99.1%) | 110 of 110<br>(100%) |
| Clinical Trial Results for Contour®Next Link Wireless - Contour®Next Strips: AST Palm<br>109 results: 109 subjects, 1 lot per subject, 3 lots overall |…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.