MEDISENSE PRECISION PCX POINT OF CARE MANAGEMENT SYSTEM FOR BLOOD GLUCOSE TESTING

K022941 · Abbott Laboratories · LFR · Oct 28, 2002 · Clinical Chemistry

Device Facts

Record IDK022941
Device NameMEDISENSE PRECISION PCX POINT OF CARE MANAGEMENT SYSTEM FOR BLOOD GLUCOSE TESTING
ApplicantAbbott Laboratories
Product CodeLFR · Clinical Chemistry
Decision DateOct 28, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The MediSense Precision PCx Point of Care Management System for Blood Glucose is intended for the in vitro diagnostic use (i.e. for external use only) for the quantitative measurement of glucose in fresh capillary whole blood. The MediSense Precision PCx Point of Care Management System is for home (lay user) or professional use. The PCx Point of Care Management System for Blood Glucose Testing is for monitoring blood glucose concentration in patients with diabetes mellitus or other conditions. The product may also be used by healthcare professionals for the quantitative measurement of glucose in venous, arterial, or neonatal whole blood, provided that the sample is used within 30 minutes.

Device Story

System uses amperometric biosensor technology to measure blood glucose. Input: fresh capillary, venous, arterial, or neonatal whole blood applied to test strips. Operation: biosensor generates electrical current proportional to glucose concentration in sample. Output: quantitative glucose measurement displayed to user. Used in home settings by patients or in clinical settings by healthcare professionals. Output informs glucose monitoring and clinical decision-making for diabetes management. Benefits patient by providing rapid, point-of-care glucose data.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use comparison to predicate.

Technological Characteristics

Amperometric biosensor technology; quantitative measurement of glucose via current generation proportional to sample glucose concentration. System designed for point-of-care use.

Indications for Use

Indicated for quantitative blood glucose monitoring in patients with diabetes mellitus or other conditions. Suitable for home (lay user) and professional use. Professional use includes testing of capillary, venous, arterial, or neonatal whole blood samples.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ OCT 2 8 2002 #### Premarket Notification [510(k)] Summary (21 CFR 807.92) | | Submitted by: Abbott Laboratories, MediSense Products<br>4A Crosby Drive<br>Bedford, Massachusetts 01730 | |-----------------|----------------------------------------------------------------------------------------------------------| | Device Name: | MediSense Precision PCx Point of Care Management System for Blood<br>Glucose Testing | | Common Name: | Blood Glucose Testing System | | Classification: | Glucose Test System, Class II per 21 CFR 862.1345 | Predicate Devices: Precision PCx Point of Care Management System for Blood Glucose Testing (K982303) that uses the MediSense blood glucose test strip cleared under K971812. Description: The Precision PCx Point of Care Management System for Blood Glucose Testing utilizes amperometric biosensor technology to generate a current. The size of the current is proportional to the amount of glucose present in the sample providing a quantitative measure of glucose in whole blood and control solutions. Intended Use: The Precision PCx Point of Care Management System for Blood Glucose Testing is intended for in vitro diagnostic use (i.e. for external use only) for the quantitative measurement of glucose in fresh capillary whole blood. The Precision PCx Point of Care Management System for Blood Glucose Testing is for home (lay user) or professional use. The PCx Point of Care Management System for Blood Glucose Testing is for monitoring blood glucose concentration in patients with diabetes mellitus or other conditions. Healthcare professionals may also use the product for the quantitative measurement of glucose in venous, arterial or neonatal whole blood, provided the sample is used within 30 minutes. Comparison to Predicate Device: The changes to the Precision PCx Point of Care Management System for Blood Glucose Testing do not affect the technological characteristics or the intended use of the predicate Precision PCx Point of Care Management System (K982303) when using the test strips described in K971812. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an abstract design featuring three stylized human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 NOV - 6 2002 Ms. Iris R. Gibbs Senior Regulatory Affairs Specialist Abbott Laboratories Inc. MediSense Products 4A Crosby Drive Bedford, MA 01730-1402 Re: k022941 > Trade/Device Name: MediSense Precision PCx Point of Care Management System for Blood Glucose Testing Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: LFR Dated: August 30, 2002 Received: September 3, 2002 Dear Ms. Gibbs: This SE Letter corrects SE Letter K022941 dated October 28, 2002. It corrects the Product Code to read LFR. The letter dated October 28, 2002 had the Product Code as NBW which is incorrect. The letter is now correct. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ ### Page 2 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ # INDICATIONS FOR USE ## 510(k) Number (if known): : K023941 Device Name: MediSense Precision PCx Point of Care Management System for Blood Glucose Testing ### Indications For Use: The MediSense Precision PCx Point of Care Management System for Blood Glucose is intended for the in vitro diagnostic use (i.e. for external use only) for the quantitative measurement of glucose in fresh capillary whole blood. The MediSense Precision PCx Point of Care Management System is for home (lay user) or professional use. The PCx Point of Care Management System for Blood Glucose Testing is for monitoring blood glucose concentration in patients with diabetes mellitus or other conditions. The product may also be used by healthcare professionals for the quantitative measurement of glucose in venous, arterial, or neonatal whole blood, provided that the sample is used within 30 minutes. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) 第二次第二十六年(令)) 11月1日:11時10分 11時10分 11時10分 11時間 17時間 17時間 1970年 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1970 1 Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use or Over-The-Counter Use (Per 21 CFR 801.108) Jean Copin (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ Abbott Laboratories, Inc., MediSense Products
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...