K965202 · Bd Becton Dickinson Vacutainer Systems Preanalytic · JKA · Feb 10, 1997 · Clinical Chemistry
Device Facts
Record ID
K965202
Device Name
VACUTAINER BRAND SAFETY-LOK BLOOD COLLECTION SET
Applicant
Bd Becton Dickinson Vacutainer Systems Preanalytic
Product Code
JKA · Clinical Chemistry
Decision Date
Feb 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Intended Use
A needle and attached tubing for venipuncture and the collection of blood specimens from patients. Includes a safety shield to minimize the possibility of needlesticks.
Device Story
VACUTAINER® Brand Safety-Lok™ Blood Collection Set; needle with attached tubing for venipuncture; integrated safety shield to prevent needlestick injuries; used by clinicians for blood specimen collection; device modification involves alternate manufacturing site and material suppliers; functional performance verified via bench testing and simulated use studies.
Clinical Evidence
Bench testing only. Comparative study between current and modified devices evaluated flow rates, shield activation/sustaining/override forces, heat stake break force, pressure, and I.V. penetration/drag forces. Simulated use study evaluated package opening, shield activation, and disposal; results showed equivalent or improved performance for the modified device.
Technological Characteristics
Needle and tubing assembly with integrated safety shield mechanism. Materials sourced from alternate suppliers; functional performance equivalent to predicate. No electronic, software, or energy-based components.
Indications for Use
Indicated for patients requiring venipuncture for blood specimen collection.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
VACUTAINER® Brand Safety-Lok™ Blood Collection Set (K965202)
K113505 — VACUETTE QUICKSHIELD WITH SNAPPY TUBE HOLDER · Greiner Bio-One North America, Inc. · Feb 17, 2012
K202188 — Safety Blood Collection Sets for Single Use · Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. · Feb 19, 2021
K191832 — Kurin Blood Culture Collection Set with Kurin Lock Technology, Push-Button Needle (Product models M-21221, M-21223, D-21221, D-21223, T-21221, T-21223) · Kurin, Inc. · Feb 4, 2020
K213690 — Eclipse Blood Collection Set · Eclipse Medcorp, LLC · Oct 26, 2022
Submission Summary (Full Text)
{0}
Becton Dickinson VACUTAINER Systems
1 Becton Drive
Franklin Lakes, New Jersey 07417-1885
(201) 847-4500
K965202
# BECTON DICKINSON
## G. SUMMARY OF SAFETY AND EFFECTIVENESS
FEB 10 1997
### G.1. Summary
Submitter: Eileen Schweighardt
Becton Dickinson VACUTAINER Systems
1 Becton Drive
Franklin Lakes, NJ 07417-1885
(201) 847-4570
Date: December 19, 1996
Device: VACUTAINER® Brand Safety-Lok™ Blood Collection Set
Classification: Accessory To: Tubes, Vials, Systems, Serum Separators, Blood Collection (75JKA)
Predicate Device: VACUTAINER® Brand Safety-Lok™ Blood Collection Set
Intended Use and Product Description: A needle and attached tubing for venipuncture and the collection of blood specimens from patients. Includes a safety shield to minimize the possibility of needlesticks.
Proposed Modification: Alternate manufacturing site and alternate material suppliers.
Becton Dickinson VACUTAINER Systems has determined that the establishment of an additional manufacturing site and alternate material suppliers does not affect the safety or effectiveness of the product for its intended use. Testing has shown that functional performance of the product is equivalent to the current product.
There are no new product claims associated with this 510(k).
{1}
35
# Synopsis of Test Methods and Results
Equivalent functional performance is demonstrated between the current Safety-Lok™ Blood Collection Set and the modified Safety-Lok™ Blood Collection Set in a study comparing flow rates; shield activation force; shield sustaining force; shield override force; heat stake break force; pressure testing; I.V. penetration and drag forces. The studies identified some differences; however these differences will not affect the overall performance of the modified product.
A simulated use study was conducted to evaluate the ease of package opening, ease of safety shield activation, ease of disposal and successful activation of the safety shield of the modified Safety-Lok™ Blood Collection Set manufactured by BDVS in comparison to the current Safety-Lok™ Blood Collection Set manufactured by Nissho. The principal device manufactured by the BDVS demonstrated equivalent or improved ease of package opening/product accessibility, shield activation and disposal. The Simulated Use and Actual Use studies for successful safety shield activation exhibited successful activation for both the principal and predicate devices.
Eileen Schweighardt
Manager, Regulatory Affairs
12/23/96
Date
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.